URIT-8031 Operation Manual (2010)
URIT-8031 Operation Manual (2010)
Catalogue
COPYRIGHT AND DECLARATION ....................................................................................... 1
PREFACE.............................................................................................................................. 3
CHAPTER 1 SAFETY GUIDELINES ..................................................................................... 4
   1.1 Safety Cautions ............................................................................................................ 4
   1.2 Using Precaution .......................................................................................................... 7
   1.3 Installation .................................................................................................................... 9
     1.3.1 Instrument Inspection ............................................................................................. 9
     1.3.2 Installation.............................................................................................................. 9
     1.3.3 Environmental Requirements ................................................................................. 9
     1.3.4 Location Requirements ........................................................................................ 10
     1.3.5 Power Requirements ............................................................................................ 10
     1.3.6 Instrument Connection ......................................................................................... 10
CHAPTER 2 INSTRUMENT INTRODUCTION .................................................................... 12
   2.1 Brief Introduction ........................................................................................................ 12
   2.2 Intended Use .............................................................................................................. 12
   2.3 Main Structure ............................................................................................................ 13
     2.3.1 Performance and Composition ............................................................................. 13
     2.3.2 Top View .............................................................................................................. 14
     2.3.3 Rear View ............................................................................................................ 14
     2.3.4 Function of Main Part ........................................................................................... 14
   2.4 Function and Specification .......................................................................................... 15
   2.5 Technical Parameter ................................................................................................... 17
CHAPTER 3 MEASURING PRINCIPLE .............................................................................. 19
   3.1 Principle ..................................................................................................................... 19
   3.2 Assay Methods ........................................................................................................... 19
     3.2.1 Endpoint Assay .................................................................................................... 19
     3.2.2 Two-point Assay ................................................................................................... 20
     3.2.3 Rate Assay (Kinetic Method) ................................................................................ 21
CHAPTER 4 OPERATION ................................................................................................... 22
   4.1 Operating Step ........................................................................................................... 22
     4.1.1 Operating Step ..................................................................................................... 22
     4.1.2 Start up ................................................................................................................ 23
     4.1.3 Create Standard Curve ........................................................................................ 23
     4.1.4 QC Test ................................................................................................................ 23
     4.1.5 Test Patient’s Sample ........................................................................................... 23
     4.1.6 Finish ................................................................................................................... 24
   4.2 Basic Operation .......................................................................................................... 24
     4.2.1 Start up ................................................................................................................ 24
     4.2.2 Check Instrument State........................................................................................ 25
     4.2.3 Preparation of Reagent ........................................................................................ 25
     4.2.4 Preparation of Sample, Standard Solution and Detergent..................................... 25
     4.2.5 Registration of Standard and QC.......................................................................... 26
     4.2.6 Routine Sample Registration ................................................................................ 27
     4.2.7 Analyses .............................................................................................................. 30
                                                                    I
                                                URIT-8031 Operating Manual
                                                                 III
                                   URIT-8031 Operating Manual
All contents of this manual are complied with the related laws and regulations of the
People’s Republic of China, as well as the specific conditions of the URIT-8031 Automatic
Chemistry Analyzer. All the updated information is included in this manual before printing.
URIT is fully responsible for the revision and explanation of this manual, and reserves the
right to renovate the relevant contents without separate notice. Some of the schematic
pictures in this manual are for reference, if there is any difference, please according to the
real object.
All the information of this manual is protected by the Copyright Law. No part of this manual
may be reproduced, stored or transmitted in any form, or by any means without the
express written permission of URIT.
All instructions must be followed strictly during operation. In no event should URIT be
responsible for failures, errors and other liabilities resulting from user's noncompliance
with the procedures and precautions described in this manual.
If a malfunction occurred under normal use because of the material and workmanship,
URIT will provide one year’s warranty service from the date of installation to this
instrument which sold by URIT and agents. The using period of this instrument is 10 years.
Once the following situations occurred, URIT assumes no liability to the safety, reliability
and operation condition of the instrument, and all agreed right of free service is deem to be
waived permanently and unconditionally.
                                               1
                                    URIT-8031 Operating Manual
Add: No.4 East Alley, Jiuhua Road, Guilin, Guangxi 541001, PR China
Tel: +86(773)2288586
Fax: +86(773)2288560
Web: www.urit.com
E-mail: service@uritest.com
Supplied by:
Version: 03/2010-3
                                                2
                                   URIT-8031 Operating Manual
PREFACE
This document is the operating manual for URIT-8031 Automatic Chemistry Analyzer. It
describes the structure, operation, maintenance and troubleshooting concerning the
instrument in details. Users should read carefully the manual and get special training
before operating to guarantee instrument precision, normal operation and personal safety.
Sign Illustration
                                                                    In vitro diagnostic
                           Power off
                                                                      Medical device
                    Environmental Protection                    Keep away from heat and
                            lifetime                               radioactive source
                                                3
                                   URIT-8031 Operating Manual
Please comply with the following rules for safety and effective use.
         CAUTION
         1)   Installation should be complied with installed instruction of the manual.
         2)   If relocation is necessary, contact your local distributor or URIT firstly.
         CAUTION
         1)   DO NOT move the screws on the cover, such as the screws on the board at
              back and side.
         2)   If a spill occurs or liquid gets into the instrument, please contact URIT.
              Neglecting the liquid may cause electric shock.
         CAUTION
         1)   While the instrument is in motion, DO NOT touch the moving parts, such as
              sample probe, reagent probe and stirrer, etc. And DO NOT put your finger or
              hand into the open part of instrument.
         2)   Before replacing lamp, turn off the refrigeration switch and wait until the lamp
              is cooled down. Otherwise, you may get burned.
Eyes Protection
Please comply with the following precaution for eyes protection.
         CAUTION
         DO NOT directly look at the light source when the instrument is in motion.
                                               4
                                  URIT-8031 Operating Manual
         CAUTION
         1)   DO NOT open the top cover, rear cover and reagent tray when the
              instrument is under analyzing condition.
         2)   Please check the accuracy of instrument by quality-control before using.
         3)   Please comply with the manual to maintain, check and replace the assembly
              unit.
         4)   Please comply with the corresponding explanation to handle the reagent,
              quality-control materials and reference materials.
         5)   Please handle the sample according to the requirements in the manual.
         Biological Hazard
         If chemical adheres to the human body, contagion may occur. DO NOT touch the
         sample, mixed solution and waste solution directly. Be sure to put on protective
         gloves, clothes, or even glasses if necessary. If the sample splashes to the skin
         accidentally, please treat immediately according to the working standards and
         consult a doctor.
         CAUTION
         Some reagents are strong acid or alkaline. Please use them carefully avoiding
         direct contact. If the reagent spill to the human body, immediately wash it off with
         water and soap. If the reagent splashes into eyes accidentally, wash it off with
         water and consult an oculist.
         Biological Hazard
         Some substances contained in QC solution, standard solution and waste
         solution are regulated by discharge standards and pollution control regulations,
         waste must be disposed according to the relevant environmental protection
         regulations.
                                              5
                                 URIT-8031 Operating Manual
Systematic Usage
        CAUTION
        1)   URIT-8031 Automatic Chemistry Analyzer is intended use for medical
             institution and laboratory to analyze some specific chemical composition of
             human body fluid. If the instrument to be used beyond this scope, consult
             URIT firstly.
        2)   Please consider together with the clinical symptom or other analyzing result
             when make the clinical judgment.
Operational Environment
        CAUTION
        1)   Please install the instrument according to the specified installed instruction
             in the manual. Otherwise, the results may not reliable even may cause
             system damage.
        2)   Please contact URIT if system state is changed.
        CAUTION
        1) The operator must be trained before operating the instrument. Please follow
           the instruction of manual to operate. Improper operation may cause personal
           injury, system damage and improper result.
        2) Please make a calibration and quality-control test before use the system for
           the first time to ensure it can be used normally.
        3) A quality-control test must be done when use the system. Otherwise, the
           reliability of the result could not be guaranteed.
        4) The communication joint of analytical part is set to connect with the
           communication joint of operational part. Please use the cables of URIT for
           connecting.
        5) The operation part is an external computer which is installed the specified
           operational software. The computer should be for the instrument exclusive
           use. DO NOT run any other software when it is connected with the
           instrument. Inappropriate manner may result in computer virus infection
        6) Do not touch the keyboard, display or mouse with wet hands or hands with
           chemical.
                                             6
                                 URIT-8031 Operating Manual
        CAUTION
        1)   URIT-8031 could be used for processing serum, urine and cerebrospinal
             fluid sample. Some samples or reagents may not be applicable to
             measurements in this instrument. In such case, please consult the reagent
             suppliers.
        2)   Sample must not contain insoluble substances such as fibrin and dust.
             Coagulation and impurity may block the sample probe thus causing bad
             effect on tests.
        3)   Anticoagulant and preservative contained in sample may interfere the
             accuracy of test results.
        4)   Take note of the storage condition of sample. Improper storage may
             deteriorate sample. For example, the serum sample could be stored for 2
             weeks at the temperature of -20℃ , while only 2 hours at 22~24℃. Beyond
             expiration, concentration of substances contained in serum will increase or
             decrease significantly, especially the enzyme such as ALT and AST.
        5)   If visible suspended substances exist in urine or cerebrospinal fluid samples,
             centrifuge the samples to remove suspended substances before analyses.
        6)   If pretreatment is needed for samples, or if you want to centrifuge serum
             samples, please consult your reagent provider firstly
        7)   DO NOT expose samples in the air for long because they may be
             contaminated or boiled off and thus erroneous test result may occur.
        CAUTION
        1)   For storage, handling and usage of reagent, or selection of reference
             materials and QC material, refer to the Instruction for Use provided by their
             manufacturers. Improper storage may not guarantee the accuracy of test
             result even though they are not expired.
        2)   Please consult the reagent supplier about the using instruction of reagent,
             standard solution and control serum.
        3)   The instruction of maintenance of reagent, standard solution and control
             materials is signed on the package or enclosure in the manual. Maintenance
             by improper way may cause incorrect result even if within its validity.
        4)   For the stability of reagent in open environments, contact your reagent
             supplier.
        5)   Be sure to perform calibration when replacing reagent. Otherwise, inexact
             test result may occur.
        6)   Set up the reagent parameters provided by reagent manufacturer before
             analyses. Only proper reagent setup can guarantee the correct test results.
                                             7
                                  URIT-8031 Operating Manual
        CAUTION
        Cross-contamination among reagents may influence test results. Contact your
        reagent supplier for details. As for how to check cross-contamination among
        reagents, contact URIT please.
        CAUTION
        1)   If serum is hemolytic, chylific or jaundiced, test results may be influenced.
             Please contact your reagent supplier for details.
        CAUTION
        Keep the instrument away from strong noise source and electromagnetic wave.
        Turn off mobile phones and transmitter-receiver when operating the instrument
        since the electromagnetic wave may cause an adverse effect on instrument.
Other Cautions
        CAUTION
        1)   DO NOT touch the keyboard, display or mouse with wet hands or hands with
             chemical.
        2)   Check samples for contamination (dust, or fibrinogen) and air bubble before
             analyses.
        3)   To make periodic maintenance, test and replacing according to the manual
             for getting the exact result.
        4)   For replacements of major parts, such as light source lamp, sample probe,
             reaction cuvette, etc., Please contact URIT.
        5)   For settings of sample volume, reagent volume, wavelength, standard
             values, etc., please refer to the instruction in reagent kit as well as this
             operating manual. Checking the quality of distilled water and detergent,
             check calibration results, control results, and sample results after analyses.
             Make sure there is no air bubble in the flow paths.
                                              8
                                    URIT-8031 Operating Manual
1.3 Installation
            CAUTION
            Only the URIT technician can perform the installation of URIT-8031.
Users shall provide a place satisfying the installation requirements in accordance with this
manual before installation. If relocation is necessary, please contact your local distributor
or URIT.
1) Carefully unpack the package and take out the URIT-8031 Automatic Chemistry
   Analyzer and the accessories.
2) Inspect the instrument and accessories for quantity and visible signs of damage
   according to the accompanying Packing List.
1.3.2 Installation
The instrument is high sophisticated thus proper installation is very important to its
performance. User should guarantee the environment and electrical condition are comply
with the recommended conditions. Provide a distance of 50cm at least for each side for
operating and maintaining.
                                                9
                                   URIT-8031 Operating Manual
Measurement: cm
The following power must be prepared; switchboard should be located within 10m.
1) Power
2) Grounding
3) Plug board
     A 15A output plug board with more than three 5A sockets. Heavy-duty devices should
     not share the plug board with the instrument, such as refrigerator, air conditioner etc.
4) 3 core power cable cat is using; the type of wire and plug is depended on voltage.
          CAUTION
          Make sure the instrument is grounded properly. Poor grounding may cause bad
          effects on test result and even damage to the instrument.
1. Take out the power line, one end insert into the power interface of instrument, the
                                              10
                                   URIT-8031 Operating Manual
2.   Using the communication line which provided by URIT for connection. One end
     connects to the COM serial port of computer; the other end connects to the COM
     serial port of instrument. And please tighten by the screws.
3.   Distilled water and detergent are needed for analyses; meanwhile, the waste solution
     will discharge. There are liquid-level inductive interface on the rear of instrument for
     connecting to distilled water barrel, detergent barrel and waste barrel. Connect one
     end of tube and liquid-level inductive line to the instrument, the other end connect to
     the corresponding barrel. The connection of instrument please see figure 1.3.6.
          CAUTION
          Please connecting the flow path tubes under the instruction of URIT service
          engineer or the personnel authorized by URIT.
          Note
          Do not locate the instrument on a place where difficult to operate cutoff device.
                                              11
                                    URIT-8031 Operating Manual
Work Unit of URIT-8031 consists of optical unit, mechanical operation unit, liquid path
control unit, hardware circuit unit and operating unit.
1)   The optical unit consists of 120 cuvettes, high-resolution filter and halogen light. It’s a
     whole sealing, statics, array and rear spectrophotometry optical system.
2)   The mechanical operation unit consists of sample processing system and reagent
     processing system. The sample processing system includes one sample tray, one
     sample probe arm, one sample syringe (the materials of syringe is ceramic inner core,
     sampling accurately and free maintenance), and a sample probe washing pool.
     Reagent processing system includes one reagent tray, one reagent probe arm, one
     reagent syringe and a reagent probe washing pool. Furthermore, a stirring
     mechanism and an eight-step washing system are included.
3)   The liquid path control unit consists of vacuum pump, solenoid valve, rinse system
     and pipeline system, etc.
4)   The hardware circuit unit consists of power board, main board, terminal board and
     circuit transfer board.
5)   The operating unit is an external computer (CPU: Intel Pentium4, 2.4GHz or above;
     Hard Disk: 40G or above; Memory: 512M or above; Display Resolution: 1024*768 or
     above; Peripherals serial port, net mouth, parallel port and USB port), the application
     software should be setup under the Windows XP (Home/Professional SP1 ) or the
     above operation system.
6)   The instrument is easy to operate. The layout of the screen menu is reasonable,
     name is simple. Such as testing parameter setup, patient’s information input,
     quality-control, reagent, data query, standard, running test and hardware parameter.
     After setting, put the sample and reagent to the instrument and begin to analyzing.
     Print out the result by the external printer at last.
The instrument is only for professional, in vitro use in hospitals, clinics and laboratories.
                                               12
                                 URIT-8031 Operating Manual
         CAUTION
         Some samples may not be analyzed according to the tested parameter and
         reagent. For the case of these samples, please contact reagent manufacturer or
         distributor.
The analyzer is mainly consists of analytical unit, computer and printer. Computer and
printer are optional accessories.
The analytical unit is composed of reagent tray, sample tray (reagent tray and sample tray
are combined to one tray), reaction tray, aspirating mechanism, stirring mechanism,
optical system, liquid path system, hardware circuit and cooling device.
          NOTE
          Computer is a part of analysis system; customer could prepare by themselves
          or purchase it from manufacturer.
                                            13
                                   URIT-8031 Operating Manual
1)   Sample Tray: Holds sample cups, and convey the cup to sample aspirating position
     for aspirating.
2)   Reagent Tray: Holds reagent bottles, and convey the bottle to reagent aspirating
     position for aspirating. The reagent tray with 24 hours non-stop refrigeration function,
                                              14
                                   URIT-8031 Operating Manual
3)   Reaction Tray: Fixed the cuvette. Sample and reagent reacted in the fixed cuvette in
     37℃ therm ostat m eanwhile colorim etric directly.
7)   Rinsing Mechanism: Drains out reaction solution and washes the cuvettes, add and
     drain out the pure water. The design of eight-step washing can clean the cuvettes
     adequately.
9)   Sample Probe Washing Pool: Omni directionally wash the inside and outside of
     sample probe by distilled water.
10) Reagent Probe Washing Pool: Omni directionally wash the inside and outside of
    reagent probe by distilled water.
3)   Rinse: Rinsing the inside and outside of aspirating probe (sample probe and reagent
     probe) with distilled water, and omni directionally clean the stirrer; washing station of
     eight-step rinse the cuvettes with detergent; separate rinse available among items.
4)   Sample probe: With liquid level sensing, volume tracking function, auto-protective
     function to prevent from collision. The function of probe-clog detected is optional.
5)   Reagent probe: With liquid level sensing, volume tracking function, auto-protective
     function to prevent from collision. The reagent preheating function is optional.
7) Waste disposal: Two-stage distributaries for high and low concentration waste.
8)   Backup reagent position: Three reagent positions available for the same item. When
     the first alarm to lack of reagent occurs, aspirating probe will turn to the second
     reagent position to aspirate reagent automatically, and the second alarm occurs,
     aspirating probe will turn to the third reagent position to aspirate reagent.
                                              15
                                   URIT-8031 Operating Manual
9)   Reagent capacity expansion: It’s provides the function that a reagent position can be
     set to test several items but once only test one item, and four reagents item test.
10) Test method: End point, rate assay(kinetic method), 2-point end point(2-point kinetic
    method), dual-wavelength, blank method(reagent blank, sample blank and water
    blank), immune turbidimetry, double reagent and electrode, colorimetric, sample
    inspection(serum index, such as jaundice, hemolysis and so on ), non-linearity
    detection etc.
11) Calibration method: At least linearity (single point, two points or multi points) and
    non-linearity calibration. Multiple calibration formula including Logit-Log4P,
    Logit-Log5P and exponential function, spline, exponential 5P, parabola, Wei Bull,
    K-factor etc.
12) Calibration system: Selecting best test point according to reaction curve, do not need
    to calibrate for second times; Calibration result is calculated automatically. Tracking
    calibration function, the change of K value is presented on a drawing. 8 calibrations in
    different concentration can be used for each item.
15) QC Predefine different controls. More than 4 controls can be tested simultaneously
    and QC could be insert randomly in the course of testing. QC diagram could be
    storage, displayed and printed.
16) Colorimetric method: Colorimetric in reaction cup directly, and single-hole detection.
17) Monitor: Monitoring cuvette online, display whole reaction process in real time, skips
    and marks the unqualified cuvette automatically.
18) Pre-diluents/Retest: The software could identify the sample which substrate is use up
    and linearity is over range, for these samples, system could select pre-diluents test
    and retest manually or automatically. The diluents time could be programmed. Max
    dilute multiple is up to 250.
19) Data reset: Reselect measure point for abnormal sample (Substrate use up, over
    range of linearity) and recount without retest.
20) Enzyme linearity check and expansion function: Search and check the linearity
    reaction interval of enzyme automatically to get the real result.
21) User mode: Hospital mode, blood station mode, physical examination center mode
    etc.
22) Item sequence: Item print and measuring sequence could be programmed.
23) Patient result and data storage: Store and backup automatically and permanently in
    infinite quantity.
                                              16
                                  URIT-8031 Operating Manual
26) Barcode reader: The instrument supports the barcode scanning function. Supporting
    codabar, interleaved 2of 5, code128, code39, code93, UPC/EAN and any other
    barcode rules.
27) Printing function: Various printing format, support Chinese/English printing. User
    could edit the format of report.
28) Light source: Long-life halogen lamp, auto-sleep. With liquid cooling system to cool
    down the light.
1) Testing Speed: 360 tests/h (pure biochemistry), 600 tests/h (with ISE)
2)   Sample Tray: 71 sample positions, they are consist of routine sample positions,
     standard positions, quality control positions, probe washing position and STAT
     position; Various samples can be mixed placed. Neonate ultramicro quantity cup,
     primitive tube and plastic tube are appropriate for those positions. Offer up to 20
     virtual sample trays, 1080 samples could be edited simultaneously in single time.
5)   Optical System: A whole sealing, static, array and rear spectrophotometry system
     composed of high resolution and halogen lamp, with 10 wavelengths for selecting
     (340nm, 405nm, 450nm, 492nm, 510nm, 546nm, 578nm, 630nm, 700nm, 800nm)
     and other two positions for option.
10) Cuvette Light Path: It’s provides 5mm, 6mm and 7mm light paths for selecting.
11) Reaction Mixture Volume: 110μL~600μL (light path is 5mm), 130μL~750μL (light path
    is 6mm), 150μL~900μL (light path is 7mm).
                                             17
                                  URIT-8031 Operating Manual
                                             18
                                   URIT-8031 Operating Manual
3.1 Principle
When a parallel monochrome light beam goes through a light-absorbing object (gas, liquid
or solid), some photons are absorbed and the light intensity decreases from I0 to I, the
formula is:
                                              I
                                      − lg       = εLC
                                              I0
ε: Proportionality factor
I: Intensity of shoot-out light
I0: Intensity of shoot-in light
Measure I, I0, and L, and then calculate the concentration C according to the formula
above.
Endpoint assay means the concentration of measured item is calculated according to the
absorbance at the end of the reaction process.
Endpoint assay in which absorbance is measured at a time point when absorbance stops
changing at the end of the reaction process. The reaction curve below explains the assay.
                                              19
                                   URIT-8031 Operating Manual
End-point assay in which measurement is made twice at different points(at the beginning
and at the end of the reaction time) to obtain the difference of absorbance. Observe the
reaction curve below.
Measurement is made twice at two different optional points (except the start point and end
point of the reaction time) to obtain the difference of absorbance. Observe the reaction
curve below.
This assay is helpful to solute some non-specificity problem in some of the reaction.
                                              20
                                URIT-8031 Operating Manual
The concentration is obtained from the change in absorbance (ΔA/min) between the
specified measurement points. Observe the reaction curve below.
                                           21
                                    URIT-8031 Operating Manual
CHAPTER 4 OPERATION
Following show the operation step of instrument. Suggesting perform standard analyses
everyday after start-up and do control analyses.
          Power on
             4.2.1 Startup URIT-8031
             4.2.2 Check instrument state
          Routine operation
             4.2.5 Standard order
             4.2.5 QC analyses order
                                                                 Preparation of routine sample
             4.2.6 Routine order
                                                                 Place reagent on instrument
                                                                 Place sample on instrument
                                                                 4.2.6 Routine
          Start analysis                                         4.2.7 Start analysis
             4.2.7 Start analysis                                4.2.9 Finish
Analysis
          Finish
                                                            4.2.8 STAT analyses
             4.2.9 Finish
                                               22
                                    URIT-8031 Operating Manual
4.1.2 Start up
Firstly turn on power switch of analyzer and computer. Then check whether the distilled
water and detergent is sufficient, and be sure the waste solution barrel is empty. Wait
about 30 minutes for instrument preheating.
           CAUTION
           1)   There are two switches, testing switch and power switch. Before
                analyzing, please turn on the power switch firstly and then turn on the
                testing switch.
           2)   Please put on the cover of reagent tray to keep the cooling effect.
It is suggested to perform standard analyses every day to obtain preferable result. The
following are the steps for standard analyses.
1)   Place reagent on reagent tray and put a known concentration standard solution on
     standard position of sample tray.
2)   Apply to system for standard analyses.
3)   Click “test” to start standard analyses.
4)   Confirm new standard curve is created.
4.1.4 QC Test
There are two kinds of QC analyses: within-day variation QC and between-days variation
QC. Within-day QC is to check the data measured within a day. Between-days QC are to
check the data measured within different days. It is suggested to measure two control
samples in different concentrations every day, and each control sample should be
measured at least twice.
                                                23
                                  URIT-8031 Operating Manual
4.1.6 Finish
Perform rinse and log off the system after finishing analyses and then turn off power
switch.
4.2.1 Start up
                                         Figure 4.2.1
5)   Input the registered doctor’s name and the corresponding password to enter system
     to operate analyzer.
          CAUTION
          User authority management is applied to limit certain function for low permission
          user, which is benefit for user to manage the software reasonably.
NOTE
Please refer to figure 5.5.1.1 to set the name and password.
NOTE
Instrument with the function of 24 hours stand-by, auto sleep and one-key setup.
                                               24
                                    URIT-8031 Operating Manual
1)   Check the distilled water, detergent and waste solution, if distilled water or detergent
     is insufficient, replenish it. Empty the waste barrel before analyses.
        Biological Hazard
        DO NOT touch the detergent or waste solution. If it adheres to hands or clothes,
        wash it off with water immediately. If it splashes into eyes, wash it off with water
        immediately and consult a doctor.
2)   Verify that reagent lids have been removed and the rest of reagent volume is sufficient
     for your analyses.
          CAUTION
          If the rest volume in reagent bottle is insufficient for analyses, replace the whole
          reagent bottle. DO NOT refill the reagent to the bottle. If refilled, data error may
          be caused.
3)   Check aspirating probe and stirrer. Verify that the aspirating probe is not blocked.
     Make sure there are no stains and crystals adhere on the surface of probes and
     stirrer.
1)   Be sure to use the certified reagents. Read through the reagent instructions and set
     up parameters properly before analyses. For item parameter setup, refer to 5.2.1 and
     5.2.2.
2)   If reagent is insufficient, replace it with a whole new bottle. Place the reagent bottle in
     the specified position according to the preset reagent parameters.
3)   Reagent should be stored at temperature of 2-8 ℃. Long-time exposure in the air may
     deteriorate the reagents.
          Biological Hazard
          Be careful not to spill out reagent. If reagent is spilled, wipe the area by dry fabric
          immediately.
Add samples into special sample cups or tubes, and then select corresponding sample
                                               25
                                   URIT-8031 Operating Manual
          CAUTION
          Before transferring samples, observe their appearances and shapes to see if
          icterus, hemolysis or chyle exists.
Add proper volume of standard solution into sample cup and place the cup in the standard
position.
Dilute detergent proportionally, and then add into the detergent barrel. Be sure the solution
level not exceeds the marked highest line on detergent barrel.
           Biological Hazard
              1)   DO NOT mistake the detergent, e.g., acid solution and alkaline solution
                   will produce toxic gas when they are mixed.
              2)   If detergent is spilled, wipe the area by dry fabric or tissue. Leaving the
                   spill will erode the parts of the instrument.
Biological Hazard
             1) When opening the lid of detergent barrel, be careful to avoid touching the
                solution. If the detergent adheres to your skins or clothes, wash it off with
                water.
             2) If the detergent splashes into eyes, wash it off with water immediately and
                consult a doctor.
4.2.5.1 QC Registration
NOTE
                                              26
                                  URIT-8031 Operating Manual
The QC test could be done when the routine test is going on, just set the “sample position”
as QC position, such as the position of C1 or C2. The system will identified it as QC and
process the QC result automatically.
It is necessary to register analytical items and patient information for each routine sample
before analyses according to application form. Figure 4.2.6-1.
Figure 4.2.6-1
1)   Select “Routine Analyses” from “Chemistry Analyses” or click the shortcut icon
     to enter the “Routine Entry Screen.
2)   Setup the basic information of sample ID, sex, age, sample cup, initial cuvette No. and
     analyzing date etc.
3)   Select analytical item by clicking the target item. Click again could switch between
     “select” and “unselect”.
                                             27
                                   URIT-8031 Operating Manual
5)   Click “NEW” to edit next sample. Repeat the step of 2) to 4) could add more sample
     analyses.
 NOTE
 Select combined item directly for choosing analytical item. For editing of combined item,
 see 5.2.3; for editing of chemistry item, see 5.2.1.
2)   Click “COPY”, select the “Same Cup” (by placing “√”mark) and enter the repeating
     times, see figure 4.2.6-2.
Figure 4.2.6-2
           NOTE
           If the “Same Cup” is not selected, the system will recognize the current sample
           position as the initial sample position and orderly test samples at different
           positions for the same analytical items.
The created List will include sample ID, sample cup No. and analytical item. See figure
4.2.6-3.
                                              28
                                   URIT-8031 Operating Manual
                                        Figure 4.2.6-3
The function of “Pre-dilution” could be used in the created list.
    Function of keys:
 NOTE
 System provides more than 20 simulative sample trays, 1080 samples could be input for
 each time.
                                              29
                                   URIT-8031 Operating Manual
4.2.7 Analyses
For sample that needs to be tested emergently, you can use the function of STAT
analyses.
Click “STAT Analyses” or click the icon            to perform emergent sample test. Specific
operation is the same as routine sample. See figure 4.2.8
Figure 4.2.8
CAUTION
           The STAT Entry Screen is similar to the Routine Entry Screen. Pay attention
           not to confuse them when doing analyses. STAT samples must be placed at
           the STAT positions, such as E1,E2 etc.
           CAUTION
           The STAT analyses are performed only after routine test started.
                                              30
                                   URIT-8031 Operating Manual
1) Normal Exit
     Upon completing the test, system will stop and the test data is processed and stored
     automatically.
2) Emergent Exit
     If you have to abort the test, click the “Emergency exit” key. In this case, the test is
     cancelled.
          CAUTION
          If the emergent exit function is executed, all the sample tests which applied will
          stop and data will loss. Re-applied if it is need to test these samples
          continuously.
It’s necessary to check the test result after finishing the analyses.
After finishing the test, enter the corresponding patient’s information to complete the test
report.
Select “Register” or click the icon         to enter the registration interface. The sample
information could be registered in this interface, such as “Report Date”, “Sample ID”,
“Name”, “Sex”, “age”, “Inpatient ID”, “Patient No.”, “Department”, etc.
Click “Calculated Item”, the item will display. Click “SAVE” to edit next sample and repeat
this operation until complete all the sample editing.
Printout: Offer various print formats for different needs. Support Chinese/English printing
mode.
Click “Data Processing      Result Modification” to modify the result when abnormal result
 appear. Select the date and item that need to modify, system will display the data of
 selected item. Suggest calculating coefficient according to the below formula:
Input the calculated coefficient into the blank box of coefficient, click “Modify”, system will
modify all intraday data under this item automatically, and then click “SAVE” to save the
                                              31
                                  URIT-8031 Operating Manual
data.
         CAUTION
         Take objective factors, such as clinical symptoms or sample characters into
         consideration before deciding to modify test results. The modified coefficient
         should be calculated combined with QC results in recent days. To avoid
         misdiagnosis, doctor or lab personnel should supervise the modification of test
         results.
Select “Data Processing         Result Review” or click icon           to enter the query
interface. The analytical result could be queried in this interface. System provided various
index functions, such as query by analytical date, patient’s name, patient No. and doctor
etc.
         NOTE
         Reports can also be printed out in accordance with the laboratory’s requirement
         in this screen.
System provides the interface of “Chemistry Analytical Data Display” for lab to depth
analyze the result. Select “Data Processing   Analytical Curve” to analyze the analytical
data. Furthermore, this function could examine the QC and standard curve, and analyze
the accuracy of the QC and standard.
Backup or restoration various test data to avoid data loss. Database maintenance consists
of two parts. Operating as following steps:
                                             32
                                      URIT-8031 Operating Manual
2)   Select “Database Recover” and then restore data according to the requirements.
3)   Click “Recovery” and exit.
         Special Analyses
                                 Standard and QC Analyses
Cuvette Rinse
                                 Parameter Setup
        System
        Parameter            Setup of Communication Port and Hospital Name
Chart Color
                                    Query of Analytical
                                                  33 Result
Parameter setup of routine item is the first step of chemistry analyses. Setup the
parameter accurately could guarantee the accuracy of analytical result. The basic factors
for each analytical method includes name of chemistry item, testing method, wavelength,
reagent volume, reagent position, incubation time, measuring time, sample volume, blank
low value, blank high value, normal low value, normal high value, unit and decimal etc.
The setup steps are as follows:
1)   Select the “Chemistry Parameter —> Item Parameter” or click icon          to enter the
     parameter setup interface.
2) Input the item name, includes the abbreviation and other name. See figure 5.2.1-1.
Figure 5.2.1-1
3)   Setup the parameter of new adding analytical item according to the instruction or lab
     item file. See figure 5.2.1-2.
                                             34
                                     URIT-8031 Operating Manual
Figure 5.2.1-2
1.   Method: Endpoint, Rate method (Kinetic method), 2-point endpoint, 2-point rate
     method (2-point kinetic method), Multi-standard method and Dual-wavelength method
     etc.
6. Priority: PRI-1with the highest priority level; PRI-9 with the lowest priority level.
8.   Standard Factor and Standard Value: Endpoint, Rate method (Kinetic method),
     Permanent Time method and Dual-wavelength method are all analyzed with standard
     and factor. See figure 5.2.1-3. Click icon       , select standard position and input the
     corresponding standard value. If the Factor method is adopted, the corresponding
     standard factor has to input and set the standard value as “0”. As shown in figure
     5.2.1-3, if the standard number is more than“1” (not include “1”), that is the
     multi-standard method, so the Multi-standard method could be used.
                                                35
                                   URIT-8031 Operating Manual
Figure 5.2.1-3
9.   Sample volume: The sample volume to be aspirated into cuvette. Please setup the
     volume according to the reagent instruction and lab relative file.
10. First Reagent Volume: The reagent volume to be aspirated into cuvette in the Single
    Reagent Method. About the reagent volume in the Dual-Reagent Method, please refer
    to the reagent instruction and lab requirement.
11. Secondary Reagent Volume: The secondary reagent volume in the Dual-Reagent
    Method, please refer to the reagent instruction and lab requirement.
12. Incubation Time: For endpoint method, incubation time is from the mix of sample and
    reagent to the end of reaction process. For Two-point method, incubation time is from
    the first measured point to the second measured point. Two incubation time has to be
    set in the Dual-Reagent method. Please refer to the reagent instruction and lab
    requirement.
13. Test Point: The number of test points when reading absorbance continuously after
    incubation.
14. Normal Value Range Setup: blank low value and blank high value are used for judging
    the reagent quality, please refer to the max absorbance value to input the value. The
    normal low value and normal high value are used for judging the analytical result. The
    linear range could be input by refer to the corresponding reagent information.
           NOTE
           Lab could set the range of item reference value according to local actual test
           situation and lab’s data statistical.
4) Please click “SAVES” to complete setup after setting all the parameters.
           CAUTION
           All the new setting data will lose if not click “SAVE”.
The software system expands the function of “Special Item Parameter” for users. It
provides that the analytical results detected by other instruments can be transmitted to the
software system for a unified printing format. See figure 5.2.2
                                              36
                                   URIT-8031 Operating Manual
Figure 5.2.2
     Char: Test results will be indicated with character system, such as “+”, “_”, “Negative”
     or “Positive”.
     Float: Test results will be indicated with numerical data. The decimal, normal high
     value and normal low value of result could be input.
The analytical item will group together and to be given a separate code which name as
“Combined Item”. Add or delete item according to the lab requirement. The combined item
will appear in the interface of “Combined Item Region” for analyses after the correlative
setup has been done.
 NOTE
 “Item Combined” means multi items combined to a separate code, that is a process; But
 “Combined Item” is a code after the “Item Combined” has been proceeded, which is a
 name.
Some of the analytical items have to be calculated and the corresponding result is for
assisting the doctor to do diagnosis. These results are the calculated results, and the
items named as calculated item which function is completed setting in the selection of
“Calculated Item Setup”.
           NOTE
           Take note of replacing the “-”with “_” contained in item name when entering
           arithmetic expression.
Items can be sequenced for display and printout, including analytical items, external items
and calculated items. Click “Item Sequence” to enter the interface for setup.
                                              37
                                 URIT-8031 Operating Manual
The applied analyses of standard and QC could be done in this interface. Please see 4.2.5
for specific operations.
The A/D value and absorbance value of all filters could be read. This function is only for
the personnel who authorized by URIT to debug.
          CAUTION
          URIT is not responsible for any consequence resulting from changed A/D
          values without permission or authorization by URIT.
System support multilevel QC test. The correlative QC parameter has to be set for QC test.
Click “Chemistry QC Management” for setting. See as the figure 5.4.
Figure 5.4
Some relevant parameter materials will be accompanied with QC materials when delivery.
The general characteristics of the QC material could be confirmed by inputting its basic
data. The laboratory personnel could input the basic parameter into the system when
introduce the new QC materials. The “QC setup” includes the basic setting of target value,
SD value and batch number.
 NOTE
 Chemistry Item list of QC is added or deleted in the selection of “Routine Item
 Parameter”. See 5.2.1.
                                            38
                                  URIT-8031 Operating Manual
          NOTE
          The laboratory should set the target value and SD value for the new QC
          material. The target value must be defined according to the presented
          determination method. The parameter which supplied by manufacturer only for
          reference. See 6.2.3 for specific setup steps of target value and SD value.
          NOTE
          Laboratory could select individual QC standard, see 6.3 for QC diagram
          analytical method.
Statistics analyses of the operator’s job, convenient for laboratory management. The
figure as below:
                                             39
                                 URIT-8031 Operating Manual
Figure 5.5.1-1
Startup URIT-8031 chemistry operating software, figure 5.5.1-2 will appear, suggested to
input name of “Operator” and corresponding “Password”, setup the work cord and name
here. The work cord is used for work registration and printing, utilize the examination
doctor cord to input the corresponding password in the below figure 5.5.1-2.
Figure5.5.1-2
Corresponding password could be setup in the figure 5.5.1-2, if another password existed
in the system, the input filed of “Old Password” in the input status. System will save the
new password and default as operating password.
High priority will allow the operator to modify “Routine Item Parameter”, “Item Combined
Setup”, “Calculated Item Setup” etc; while low priority won’t.
         NOTE
         System software adopts strict multi-authority management to guarantee the
         safety of data. Please remember and keep your password in private.
                                            40
                                   URIT-8031 Operating Manual
Cord can be used to represent doctor, such as number, symbol or letter; doctor’s name
also can be used. Setup according to specific condition.
Cord can be used to represent department, such as number, symbol or letter, department
title also can be used. Setup according to specific condition.
NOTE
“Doctor” and “Department” are setup by laboratory itself.
Laboratory can setup the chart color individually, see figure 5.5.2:
Figure 5.5.2
                                              41
                                  URIT-8031 Operating Manual
Figure 5.5.3
System will test all communication ports automatically and a dialog box will appear to
prompt user to select communication port. User’s name should be input in this function
module, and the name will be the title of report sheet and other diagram.
The reliability of test result is determined by two aspects: One is the precision, which
means test results are stable in repeatability. Precision can be guaranteed by establishing
perfect room quality control system; the other one is accuracy, which means test results
close to target value. Accuracy can be guaranteed by proper assay and calibration. It is,
therefore, necessary to select certified control sample and calibration solution, and use
them strictly according to their instructions.
1)   Freeze-dry control, liquid control and mixed control serum, classified according to
     physical property.
2)   Fixed value and non-fixed value control sample, classified according to the presence
     and absence of fixed value. Different inspection body can choose more than one
     quality control as quality control.
                                             42
                                    URIT-8031 Operating Manual
QC target value and control limit are usually provided by the control samples supplier, also,
you can determine them through the following methods:
     Perform QC analyses at least 20 times with a new batch of control sample. Calculate
     the mean value and standard deviation from the obtained QC data.
     Obtain the accumulated mean value of original 20 QC data as the target value while
     the accumulated mean value of 3~5 month QC data as the standard deviation.
     Control limit is the multiplier of standard deviation. Control limit of analytical item is
     determined according to different QC rules.
6.2.4 QC Procedures
QC procedure consists of two systems; they are “Chemistry QC Management” and “QC
Analyses”. Select “Chemistry QC Management” in the “Chemistry Analyses”, its function is
setup the parameter before QC Analyses and process the QC data, such as setup the QC
target value, QC batch number and SD value etc. Select “QC Analyses” in “Special
Analyses” to begin QC Analyses, QC data will saved in system automatically and a QC
diagram will generate for chemistry analyses; QC Analyses could be add to routine test
interface. See 4.2.5 for application of QC Analyses.
                                               43
                                 URIT-8031 Operating Manual
1-2S rule exceed the average value ±2SD is judged as out of control.
1-3S rule exceed the average value ±3SD is judged as out of control.
1-2S                  Warning
1-3S                  Random Error
2-2S                  Systematic Error
R-4s                  Random Error
4-1S                  Systematic Error
10X                   Systematic Error
1-2s Rule: When one QC result exceeds ±2SD, regard as “warning” into next standard.
                                            44
                                URIT-8031 Operating Manual
1-3S Rule: When one QC result exceeds ±3SD limit, this is judged as random error.
2-2S Rule: The QC results of two successive are both exceed +2SD or -2SD, this is
judged as systematic error.
R-4s rule: When one QC result exceeds mean value +2SD, and the other exceeds -2SD,
this is judged as random error.
                                           45
                               URIT-8031 Operating Manual
4-1s rule: When four consecutive QC results exceed +1SD or -1SD, this is judged as
systematic error.
Figure6.3-6,4-1s QC rule
10X Rule: When 10 consecutive QC result fall on the + or – side, this is judged as
systematic error.
                                          46
                                     URIT-8031 Operating Manual
3) Metamorphic Reagent
     Metamorphic Reagent
5) Inadequate wash
     Stirrer wash is inadequate.
6) Adverse mixture
     The depth of stirrer to cuvette is excursion; stirrer mechanism faulty.
1) Inaccurate standard
     The dissolvent of standard solution is inappropriate
2) Metamorphic reagent
     The reagent is metamorphic, and the batch number is various.
3) Temperature
     The temperature control is inappropriate.
     Retest the out-of-control item by using the same bottle of control sample. Strictly obey
      the operation flow to check if the out-of-control is due to operation incorrectly or
      random error. If retest result falls outside the acceptable range, proceed to the next
      step.
     Retest the out-of-control item by using a new bottle of control sample (same lot). If
      retest result is in control, the previous bottle of control sample may be to blame. If
                                                 47
                                    URIT-8031 Operating Manual
retest result still falls outside the allowable range, proceed to the next step.
    Retest the out-of-control item by using a new lot of control sample. If retest result is in
     control, the previous lot of control sample may be deteriorated. Then check the
     expiration date and storage condition. If retest result still falls outside the allowable
     range, proceed to the next step.
    Perform instrument maintenance; retest the out-of-control item. Check the instrument
     state; check whether the light source or filter are changed or not; and whether the
     cuvette need to wash or replaced. Furthermore, replacing reagent. If the retest result
     still falls outside range, proceed to the next step.
    Obtain technical help. If you cannot get the in-control result after performing the above
     five steps, contact reagent manufacturer or URIT to get more technical support.
Calibration solution contains the known quantity object, which is used for calibrating the
value of this method; the calibration solution is concerned with the method, reagent and
instrument. The function of calibration solution is to reduce or eliminate systematic error
caused by instrument and reagent. It should be better to use human serum matrix to
reduce matrix effect.
It is suggested to perform calibration every six months or under the following situations:
    When changing reagent batch number or type, unless specified by the lab that the
     change will not influence the precision.
    After replacing the major components, such as lamp, sampling mechanism, probe, or
     cuvette etc.
    When control result shows abnormal offset, tendency, or falls out of the acceptable
     range and it cannot be corrected by routine tests.
Select “Special Analyses” to applying standard analyses for each item, analytical data
                                               48
                                   URIT-8031 Operating Manual
Input and save the concentration value of calibration material into the standard column.
Enter “Routine Analyses” interface, if “Concentration of Standard Material” has been input
when setup the parameter of analytical item, the selection of this item will appear in the
“Standard list”.
Select the item in the “Standard” list for standard analyses. Analytical result will be
processed and saved automatically and to be the standard of this item (standard factor).
          NOTE
          See 4.2.5 for standard application and test.
          NOTE
          System default the value of “Standard Factor” as 1 when newly built an item; and
          system will calculate the result according to the standard factor “1” if not
          modified.
          NOTE
          It is suggested that all calibration work should be written down and complied into
          documents. What should be emphasized is that the fixed-value control serum
          should never be used as standard solution.
          NOTE
          The calibration cycle could be set manually or automatically.
CHAPTER 7 MAINTENANCE
In order to keep the analyzer in the best work state, periodic maintenance for specific
components should be done since it can guarantee the analyzer under the no fault work
status and stabilize the test result.
                                              49
                                   URIT-8031 Operating Manual
Figure 7.1-1
Select “Probe and Stirrer Cleaning” in the menu of “Maintenance” or click icon              to
rinse the probe and stirrer. It is aimed at ensuring cleanness and avoiding resides in tubing.
The rinsing time could be setup. See figure 7.1.1.
Figure 7.1.1
Select “Cuvette Rinsing” in the menu of “Maintenance” or click the icon            to rinse the
cuvette. It is aimed at keeping cleanness and avoiding solution resides. The functional
interface as follows:
                                              50
                                 URIT-8031 Operating Manual
Figure 7.1.2
         NOTE
         It is very important to rinse all the cuvettes at least one time when turn on and
         turn off the instrument.
                                       Figure 7.1.3
Functional Button:
                                            51
                                   URIT-8031 Operating Manual
                     : Pump off all the distilled water from cuvette. The function is effective
only there is water in the cuvette
: Reading the absorbance of all cuvettes and result will not be saved
7.2 Maintenance
Please do the following maintenance job to keep the analyzer in the best work state and
use safety:
1)   Before daily work, verify that detergent and distilled water are enough for your
     analyses. If insufficient, replenish the detergent barrel and distilled water barrel.
2)   Empty the waste barrel if waste solution exceeds the warning line marked on the
     waste barrel.
3)   Check the mechanical movement of sample probe, reagent probe and stirrer. Check if
     probes are clogged or not. Inspect the exterior of probes and stirrer for contamination
     and dirty. Clean them if necessary.
        Biological Hazard
        Put on protective gloves to avoid contacting chemical solution. If it spill to human
        body, wash it off with water immediately consult doctor if necessary.
4)   Perform at least 5 times “Probe and Stirrer Cleaning” and one time “Cuvette Rinsing”.
     Consult 7.1.1 and 7.1.2 for details.
5)   Perform “Add water “before power off to guarantee the cuvette full of distilled water.
     See 7.1.3 for specific operation.
                                               52
                                   URIT-8031 Operating Manual
     b)    Wipe the exterior of sample probe, reagent probe and stirrer using a fabric
           moistened with alcohol.
2)   Rinse the needles of eight-step washing station according to the method that rinses
     the sample probe and stirring rod.
          NOTE
          Be careful not to bend or scratch the probe when cleaning them.
3)   Select “Cuvette Signal” to check if cuvette blank value is within acceptable range or
     not. If outside the range take out the relevant reaction cuvettes together with the
     cuvette racks from the reaction tray. Observe the cuvettes carefully and replace them
     if necessary.
          NOTE
          To get optimum data, DO NOT read the cuvette blank until the light source lamp is
          stable. It may take 25 minutes to complete a warm up time.
          NOTE
          Be careful not to scratch the reaction cuvettes when taking them out.
          Biological Hazard
          Put on protective gloves to avoid contacting chemical solution. If it spill to human
          body, wash it off with water immediately.
4)   Turn off the refrigeration switch of the instrument. Using fabric moistened with water
     or detergent, wipe the instrument’s panel board.
          CAUTION
          Operate carefully to avoid scratch to instrument.
1) Do as weekly maintenance.
2) Check flow paths and tube connectors for leakage or clogs.
                                              53
                                  URIT-8031 Operating Manual
If instrument has been disused over two days, perform the following steps before turning
off and after starting up.
1) Add new distilled water to distilled water barrel, meanwhile, add detergent to detergent
   barrel.
2) Enter the maintenance program to wash the flow paths and cuvettes at least twice.
The preventive maintenance includes not only checkup and repair, but also the following.
Some assembly units which used for a long time or easy to damage are needed to replace
regularly.
The light source lamp should be replaced when it was ageing gradually and caused noise
during the test. If the cuvette blank or the attenuation of source intensity has exceeded the
range or light source has been worked exceed 2000 hours, the light source lamp should
be replaced.
Caution
         The operation of replacing the light source lamp must be permitted by URIT or
         under the guide of engineer who authorized by URIT.
This operation only can be done by the personnel who authorized by URIT. Please contact
URIT if you need to replace the tubes.
                                             54
                                     URIT-8031 Operating Manual
Suggest replacing the cuvettes and cuvette bracket every 3 months or replacing by needs.
The operating steps are as follows:
          Caution
          1)   Do not touch the walls of cuvette to avoid stain with fingerprint and affect the
               test result.
          2)   If most of the cuvettes has been stained, suggest to replace the cuvettes
               with the cuvette bracket.
This operation only can be done by the personnel who authorized by URIT. Please contact
URIT if you need to replace the vacuum pump.
This operation only can be done by the personnel who authorized by URIT. Please contact
URIT if you need to replace the solenoid valve.
This operation only can be done by the personnel who authorized by URIT. Please contact
URIT if you need to replace the sampling probe.
                                                55
                                   URIT-8031 Operating Manual
CHAPTER 8 TROUBLESHOOTING
This chapter lists the various malfunctions, along with probable causes and recommended
remedies to correct the problem quickly and easily. If the problem still exists after following
the recommended remedy, contact URIT for Technical support.
        CAUTION
        Handle malfunction with utmost care and confirm if it is necessary to cut off the
        power supply at the first.
        Biological Hazard
        Put on protective gloves to avoid contacting chemical solution. If it spill to human
        body, wash it off with water immediately.
                                              56
                                   URIT-8031 Operating Manual
To eliminate malfunction easily and correctly, users should read through the Operating
Manual and be familiar with the routine operation and maintenance of URIT-8031.
Users should not only confirm the malfunction, but also clearly know what the normal
status should be when the malfunction is eliminated.
If malfunction relates to the hardware or software, contact your local distributor or URIT. If
malfunction relates to the operation and analyses, refer to the troubleshooting table below
for solution.
The maintenance engineer authorized by URIT takes proper measures to correct the
problem.
Users can also eliminate the malfunction under the directions of maintenance engineer.
Our Customer Service Office is available to help if the problem is beyond the scope of this
manual or if you need more technical assistance from URIT. Before calling, please identity
the following information:
The troubleshooting table below presents the various problems and malfunctions that may
occur during operation. If the problem can not be solved through the recommended
methods, contact URIT please.
                                              57
                                      URIT-8031 Operating Manual
           NOTE
           For replacing parts of the instrument, refer to Appendix A.
                                    Table 1 Troubleshooting
SN       SYMPTOM                POSSIBLE CAUSE                                REMEDY
                                                              1) Connect power cord correctly.
                                                              2) Check if the power receptacle is in
                           1) Incorrect connection with
                                                              good condition.
                           power cord.
                                                              3) Replace fuse(8A)
                           2) No electricity with power
     Instrument is not                                        4)Confirm select the correct connector
                           receptacle.
1    active when power                                        at     [System     parameter—instrument
                           3) The safety fuse is fusing
     is on.                                                   parameter—COM set]
                           4) Improper COM interface is
                                                              5) Make sure the RS232 communication
                           selected.
                                                              cable is connected to PC correctly.
                           5) Communication cable error.
                                                              If the problem still persists, contact your
                                                              local distributor or URIT.
                                                              1) Select [Maintenance—Cuvette rinse],
                                                              Check if reaction cuvettes are dirty or
                           Cuvette dirty or damage            damage. Replace them if necessary
2    Cuvette blank error
                           Light source aging                 2) Replace light source.
                                                              3) If the problem persists, contact your
                                                              local distributor or URIT.
                                                              1) Check or replace the lamp holder
                           1) Bad contact of lamp holder.     2) Replace the lamp.
3    Lamp is dark
                           2) Lamp is burned out.             3) If the problem persists, contact your
                                                              local distributor or URIT.
                                                              1) Check flow path tubes. Reconnect or
                                                              replace tubes if necessary.
                       1) Air leaks in the flow paths.        2) Reconnect meter regulator. Eject air
     Inaccurate
                       2) Air bubbles are formed in           bubbles.
     aspirated volume
4                      meter regulator.                       3) Unclog or replace probe.
     of    reagent  or
                       3) Probe is clogged.                   4) Check magnetic valve and replace it if
     sample.
                       4) Magnetic valve problem.             necessary.
                                                              5) If the problem persists, contact your
                                                              local distributor or URIT.
     Water or detergent
                                                              1) Reconnect the tube or replace it.
     does not come out     1) Flow path tube is leaky.
                                                              2) Unclog the tube.
     through the probe     2) Flow path tube is clogged.
5                                                             3) Replenish water or detergent.
     washing pool or       3) Water or detergent is used
                                                              4) If the problem persists, contact your
     stirrer   washing     up.
                                                              local distributor or URIT.
     pool.
                           1) Flow path tube is leaky.        1) Reconnect the tube or replace it.
     Adding or draining
6                          2) Flow path tube is clogged.      2) Unclog the tube.
     water is abnormal
                           3) Vacuum pump error.              3) Check the vacuum pump and replace
                                                 58
                                           URIT-8031 Operating Manual
                                                                    it if necessary.
                                                                    4) If the problem persists, contact your
                                                                    local distributor or URIT.
                                                                    1) Check the wire of light coupling or
     A certain movable
                               Light coupling is short circuited    replace the light coupling.
7    part is out of
                               or broken.                           2) If the problem persists, contact your
     control.
                                                                    local distributor or URIT.
                                                                    1) Check the liquid level sensor board
                               1) Liquid level sensor board is      and replace it if necessary.
     Liquid level sensor       defective.                           2) Reconnect the liquid level sensor
8
     is out of order           2) Bad contact with liquid level     board.
                               sensor board.                        3) If the problem persists, contact your
                                                                    local distributor or URIT.
                                                                 1) Select “maintenance” –-- “cuvette
                                                                 rinse” to check whether cuvette is dirty.
                                                                 Clean or replace the cuvette.
                               1) Cuvette is dirty or breakage.
                                                                 2) Check sample injector and tube
                               2)    Inaccurate        aspirated
                                                                 whether leakage existed.
                               volume of reagent or sample.
                                                                 3) Replace the lamp.
     Inaccurate         test   3) Lamp is deteriorated.
                                                                 4) Set the parameter follows the
9    result   or       poor    4) Parameters of analytical
                                                                 operation manual. Make sure the
     repeatability.            item are set improperly.
                                                                 instrument is well grounded by means of
                               5) Ground wire is absent with
                                                                 the ground pole.
                               power supply.
                                                                 5) Check if the reagent is certified.
                               6) Reagent problem.
                                                                 Perform recalibration.
                                                                 6) If the problem persists, contact your
                                                                 local distributor or URIT.
                                                                1) Turn the refrigeration switch off,
                                                                slowly rotate the problem parts and
                                                                observe if there is abnormal noisy or if it
                                                                is stuck.
                               1) Communication error.
                                                                2) Enter into system parameter”
                               2) Mechanical parts are loose
                                                                interface, select “Engineer set” to adjust
     Abnormal         motor    or stuck.
10                                                              parameter.      (Only     accessible    to
     movement                  3) Light coupling joint of motor
                                                                professional person authorized by
                               is loose.
                                                                URIT).
                               4) Light coupling is defective.
                                                                3) Check the light coupling and replace it
                                                                if necessary.
                                                                4) If the problem persists, contact your
                                                                local distributor or URIT.
                                                                    1) Replace the stirrer motor.
     Stirrer   does     not    1) Stirrer motor is broken.          2) Reinstall the stirrer motor.
11
     work.                     2) Bad contact of stirrer circuit.   3) If the problem persists, contact your
                                                                    local distributor or URIT.
                                                       59
                                        URIT-8031 Operating Manual
                                                   60
                                 URIT-8031 Operating Manual
9.1 Storage
The wrapped instrument should be stored at a ventilated room, with temperature range
from -40℃ to 55℃, ambient humidity not exceeding 95% and atmosphere pressure is
75kPa~106kPa. DO NOT store the instrument along with any poison or corrosive. The
instrument stored for over one year may fall short of the precision of measurement.
Therefore, it is suggested that perform mechanical calibration and alignment procedure
when using the instrument.
         CAUTION
         Please contact URIT to perform calibration for mechanism of the instrument.
9.2 Transportation
The transportation must strictly follow the terms and conditions specified in the order
contract.
CAUTION
                                            61
                                URIT-8031 Operating Manual
        Under the packing sound condition, the transport temperature is -40~55℃, the
        relative Humidity is ≤95% and atmosphere pressure is 75kPa~106kPa.
SN            Name                                           Remark
1    Fuse                     T8AL 250V
2    Reaction Cuvette         Replacing the cuvettes every 3 months or by needs
3    Reaction Cuvette Rack    Replacing the cuvettes every 3 months or by needs
     Light Source Lamp        Replacing when its using time exceeded 2000 hours or
4
                              system prompt to replace
5    Teflon Tube
6    PU Tube
     Probes                   Sample probe, reagent probe; replacing when it is
7
                              damaged or bend
8    Step Motor               Replacing when it is failure
9    Vacuum Pump              Replacing when it is failure
10   Solenoid Valve           Replacing when it is failure
11   Temperature Controller   Replacing when it is failure
12   Temperature Sensor       Replacing when it is failure
13   Cooling fan              Replacing when its working time exceeded 40,000 hours
NOTE
                                           62
                       URIT-8031 Operating Manual
NOTE
This replaceable component list just for reference. URIT reserves the right to
change the list without notice.
63