Agilent 1146A
Agilent 1146A
User’s Guide
Contents
Receiving your Shipment. . . . . . . . . . . . . . . . . .              2
Packaging . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   2
  Warning. . . . . . . . . . . . . . . . . . . . . . . . . . . . . .    2
Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   3
Compatibility . . . . . . . . . . . . . . . . . . . . . . . . . . .     3
Control and Connector Identification . . . . . . . .                    3
Specifications . . . . . . . . . . . . . . . . . . . . . . . . . . .    4
   Electrical Specifications . . . . . . . . . . . . . . . .            4
   Environmental Specifications . . . . . . . . . . .                   5
   Mechanical Specifications . . . . . . . . . . . . . .                5
   Safety Specification . . . . . . . . . . . . . . . . . . .           5
   Electromagnetic compatibility . . . . . . . . . . .                  5
   Typical Response Curves . . . . . . . . . . . . . . .                6
   Linearity for DC signal Range 10 mV/A . . . .                        7
Operation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .   7
   Zero Adjustment . . . . . . . . . . . . . . . . . . . . . .          7
   Current Measurement . . . . . . . . . . . . . . . . .                7
   Battery Indication . . . . . . . . . . . . . . . . . . . . .         8
   Overload Indication . . . . . . . . . . . . . . . . . . .            8
Maintenance . . . . . . . . . . . . . . . . . . . . . . . . . . . .     8
   Battery Replacement . . . . . . . . . . . . . . . . . .              8
   Calibration. . . . . . . . . . . . . . . . . . . . . . . . . . .     8
Technical Assistance . . . . . . . . . . . . . . . . . . . . .          8
          Packaging
          The 1146A AC/DC Current Oscilloscope Probe is shipped with a separate battery
          (not installed) and a User’s Guide.
WARNING   • Connect the probe to the oscilloscope or voltage measuring instrument before
            clamping the probe around a conductor.
          • Never use the probe on circuits rated higher than 600 V or with float voltage
            greater than 600 V.
          • Never leave the probe clamped around a conductor while it is not connected
            to an oscilloscope or voltage measuring instrument.
          • Carefully center the conductor inside the probe jaws and ascertain that the
            probe is perpendicular to the conductor before closing the jaws.
          • Avoid, if possible, the proximity of other conductors which may create noise.
          • Check the magnetic mating surfaces of the probe jaws; these should be free
            of dirt, rust, or other foreign matter
          • Do not use a probe which is cracked, damaged or has defective leads.
          This symbol signifies that the 1146A AC/DC Oscilloscope Current Probe is
          protected by double or reinforced insulation. Only use specified replacement
          parts when servicing the instrument.
          This symbol signifies CAUTION! and requests that the user refer to the user
          manual before using the instrument.
          2
                                                   Agilent 1146A AC/DC Oscilloscope Current Probe
                                                                                      Description
                Description
                The 1146A AC/DC Current Oscilloscope Probe expands oscilloscope applications
                in industrial, automotive or power environments, and is ideal for analysis and
                measurement of distorted current waveforms and harmonics. The probe permits
                accurate display and measurement of currents from 100 mA to 100A rms, DC to
                100 kHz without breaking into the circuit. The probe uses Hall effect technology
                to measure AC and DC signals. The probe connects directly to an oscilloscope
                through a 2 meter coaxial cable with an insulated BNC.
                Compatibility
                The 1146A AC/DC Current Oscilloscope Probe is compatible with any analog or
                digital oscilloscope or other voltage measuring instrument which has the
                following features:
                • BNC input connector.
                • Range capable of displaying 0.2 to 0.5 V per division.
                • Minimum input impedance of 1 MΩ.
                                                                                 φ = 0.46”
                                                                                     (11.8 mm)
                                                                                                 3
Agilent 1146A AC/DC Oscilloscope Current Probe
Specifications
Specifications
Electrical Specifications
   Current Range 100 mV/A:100 mA to 10 A peak
                    10 mV/A:1 to 100 A peak
   Output Signal 1000 mV peak max
   AC current accuracy (After calibration and for one year)
      (zero probe before making measurement)
          Range                               Accuracy
          100 mV/A (50 mA to 10 A peak)       3% of reading ±50 mA
          10 mV/A (500 mA to 40 A peak)       4% of reading ±50 mA
          10 mV/A (40 A to 100 A peak)        15% max at ±100A
    Phase Shift1 < 1º from dc to 65 Hz on 10 mV/A
                   < 1.5º from dc to 65 Hz on 100 mV/A
    Frequency Range DC to 100 kHz (-3dB with current derating)
    Noise Range 10 mV/A: 480 µV
            Range 100 mV/A: 3 mV
    Slew Rate Range 10 mV/A: 20 mV/µs
                Range 100 mV/A: 0.3 V/µs
    Load Impedance >1 MΩ / 100 pF
    Insertion Impedance (50/60 Hz) 100 mV/A: 0.01Ω
                                            10 mV/A: 0.01Ω
    Rise or Fall time Range 100 mV/A: 3 µs
                         Range 10 mV/A: 4 µs
    Working Voltage 600 Vrms max.
    Common mode voltage 600 Vrms max.
    Influence of adjacent conductor <0.2 mA/A AC
    Influence of conductor position in jaw 0.5% of reading at 1 kHz
    Battery 9 V alkaline (NEDA 1604A, IEC 6LR61)
    Low battery green LED when ≥6.5 V
    Overload indication red LED indicates input greater than the selected
                             range.
    Typical Consumption 8.6 mA
    Battery Life 55 hours typical
1 Reference conditions 23 °C ±5 °C, 20 to 75% relative humidity, DC to 1kHz, probe zeroed,
1 minute warmup, batteries at 9 V ±0.1 V, external magnetic field <40 A/m, no DC component, no
external current carrying conductor, 1 MΩ / 100 pF load, conductor centered in jaw.
4
                                Agilent 1146A AC/DC Oscilloscope Current Probe
                                                                Specifications
Environmental Specifications
  Operating Temperature 0º to 50 ºC
  Storage Temperature -30º to 80 ºC
  Temperature Influence <0.2% per ºC
  Operating Relative Humidity
               10º to 30 ºC: 85 5% RH (without condensation)
               40º to 50 ºC: 45 5% RH (without condensation)
  Altitude Operating:0 to 2000 m
           Non operating:0 to 12,000 m
Mechanical Specifications
  Zero Adjustment 20 turn potentiometer
  Maximum cable diameter 11.8 mm
  Case Protection IP20 per IEC 529
  Drop Test 1.0 m on 38 mm of oak on concrete; test according to IEC 1010
  Mechanical Shock 100 G; test per IEC 68-2-27
  Vibration Test per IEC 68-2-6, frequency range 10 Hz to 55 Hz, amplitude
              0.15 mm
  Handle Lexan¨ 920A, UL 94 V2
  Dimensions 231 x 36 x 67mm
  Weight 330 g (11.6 oz) with battery
  Color light gray
  Output lead insulated coaxial cable with insulated BNC connector
  Cable length 2 m
Safety Specifications
  • Double insulation or reinforced insulation between primary or secondary
     and outer case of the handle, per IEC 1010.
  • 600 V category III, Pollution: 2.
  • 300 V category IV, Pollution: 2.
  • 5550 V 50/60 Hz between primary or secondary and the outer case of the
     handle.
  • 3250V 50/60 Hz between primary and secondary.
Electromagnetic compatibility
   EN 50081-1 Class B
   EN 50082-2 Electrostatic discharge IEC 1000-4-2
        Radiated Field IEC 1000-4-3
        Fast transients IEC 1000-4-4
        Magnetic Field at 50/60 Hz IEC 1000-4-8
                                                                            5
Agilent 1146A AC/DC Oscilloscope Current Probe
Specifications
Frequency
Phase Shift
6
                                        Agilent 1146A AC/DC Oscilloscope Current Probe
                                                                             Operation
-2%
-4%
              -6%
Uncertainty
-8%
-10%
-12%
-14%
              -16%
                     0A   20A    40A        60A         80A       100A
Accuracy
Operation
Zero Adjustment
The probe has a zero adjustment which should be adjusted before measurement.
Alternatively, you may "zero" with the oscilloscope instead.
Current Measurement
Connect the current probe to the proper input channel on the oscilloscope. Begin
with the least sensitive range on the current probe (10 mV/A). Select the
0.5 V/Division range on your oscilloscope. Clamp the probe on the conductor to
be measured and read the current flowing directly on your oscilloscope.
You may also use your oscilloscope to amplify the signal while using the 100 mV/A
probe range (which offers the best accuracy and least phase shift).
    Important
    It is possible to change the range on the current probe without removing the probe
    from the current carrying conductor, but it is important to remember not to exceed
    the permissible peak ratings of 1000 mV peak or 2000 mV peak to peak maximum. The
    peak ratings by range are: 10 A peak on the 100 mV/A range, 100 A peak on the 10 mV/A
    range.
                                                                                        7
Agilent 1146A AC/DC Oscilloscope Current Probe
Maintenance
Maintenance
Be sure that mating surfaces of the jaw are free of dirt or foreign matter. If they
are rusted, gently clean with a soft, lightly oiled cloth. Do not leave excessive oil
residue.
Battery Replacement
When the probe is turned on, the green battery indication LED should light up.
If not, replace the 9 V battery (see Figure 1). To replace the battery, disconnect
the probe from the circuit and the oscilloscope. Turn the probe "OFF". Unscrew
the battery compartment screw and pull out the battery compartment cover.
Replace the battery and put the cover back on. Do not replace the battery while
probe is in use.
8
                                                 Agilent 1146A AC/DC Oscilloscope Current Probe
                                                                                    Calibration
           Calibration
           Repair and Calibration
           To guarantee that your instrument complies with the factory specifications, we
           recommend that the 1146A AC/DC Oscilloscope Current Probe be submitted to
           our factory service center at one year intervals for recalibration, or as required
           by other standards.
Equipment Required
           To calibrate the 1146A, carefully pop out the plastic dot located on the left side
           of the probe, to access the potentiometer.
           Turn the calibrator on and allow 15-30 minutes for it to warm up.
Figure 2
                                                                                              9
          Agilent 1146A AC/DC Oscilloscope Current Probe
          Calibration
               Positive                                    Negative
Measure        Minimum        Maximum         Measure      Minimum        Maximum
 10 A            98 mV         102 mV           -10 A       -98 mV        -102 mV
 30 A           294 mV         306 mV           -30 A      -294 mV        -306 mV
 60 A           576 mV         624 mV           -60 A      -576 mV        -624 mV
 90 A           837 mV         963 mV           -90 A      -837 mV        -963 mV
 100 A          910 mV          1.09 V         -100 A      -910 mV         -1.09 V
               Positive                                    Negative
Measure        Minimum        Maximum         Measure      Minimum        Maximum
  1A             98 mV         102 mV          -1 A         -98 mV        -102 mV
  3A            294 mV         306 mV          -3 A        -294 mV        -306 mV
          10
                                                  Agilent 1146A AC/DC Oscilloscope Current Probe
                                                                                     Calibration
                     Positive                                       Negative
       6A            588 mV             614 mV         -6 A         -588 mV         -612 mV
      10 A           980 mV              1.02 V       -10 A         -980 mV          -1.02 V
                   Measure             Range
                  +10 A ±2%           100 mV/A
                    +120 A            10 mV/A
                    -120 A            10 mV/A
                                                                                               11
Agilent 1146A AC/DC Oscilloscope Current Probe
Calibration
12
                         DECLARATION OF CONFORMITY
                        According to ISO/IEC Guide 22 and CEN/CENELEC EN 45014
Date: 06/28/2000
Name
For further information, please contact your local Agilent Technologies sales office, agent, or distributor.
Product Regulations
EMC              IEC 61326-1:1997+A1:1998 / EN 61326-1:1997+A1:1998
                 CISPR 11:1990 / EN 55011:1991
                 IEC 61000-4-2:1995+A1:1998 / EN61000-4-2:1995                            A
                 IEC 61000-4-3:1995 / EN 61000-4-3:1995                                   B
                 IEC 61000-4-4:1995 / EN 61000-4-4:1995                                   A
                 IEC 61000-4-5:1995 / EN 61000-4-5:1995                                   A
                 IEC 61000-4-6:1996 / EN 61000-4-6:1996                                   A
                 IEC 61000-4-11:1994 / EN 61000-4-11:1994                                 A
                 Canada: ICES-001:1998
                 Australia/New Zealand: AS/NZS 2064.1
                                    Performance Codes:
                                    A PASS - Normal operation, no effect.
                                    B PASS - Temporary degradation, self recoverable.
                                    C PASS - Temporary degradation, operator intervention required.
                                    D FAIL - Not recoverable, component damage.
Notes: (none)
ICES/NMB-001
This ISM device complies with Canadian ICES-001.
Cet appareil ISM est confomre à la norme NMB-001 du Canada.
Agilent Technologies
P.O. Box 2197
1900 Garden of the Gods Road
Colorado Springs, CO 80901
Notices                         with regard to this man-
                                ual and any information
                                                                  WARNING
                                contained herein, includ-         A WARNING notice
© Agilent Technologies, Inc.    ing but not limited to the        denotes a hazard. It
1996-2002                       implied warranties of             calls attention to an
No part of this manual may be merchantability and fit-
reproduced in any form or by    ness for a particular pur-        operating procedure,
any means (including elec-      pose. Agilent shall not be        practice, or the like
tronic storage and retrieval or liable for errors or for          that, if not correctly
translation into a foreign lan-                                   performed or adhered
guage) without prior agree-     incidental or consequen-
ment and written consent        tial damages in connec-           to, could result in
from Agilent Technologies,      tion with the furnishing,         personal injury or
Inc. as governed by United      use, or performance of            death. Do not proceed
States and international copy- this document or of any
right laws.                     information contained
                                                                  beyond a WARNING
                                herein. Should Agilent            notice until the
Manual Part Number
                                and the user have a sepa-         indicated conditions
01146-92002, September 2002                                       are fully understood
                                rate written agreement
Print History                   with warranty terms cov-          and met.
01146-92000, November 1996      ering the material in this
                                document that conflict
01146-92001, August 2000
                                with these terms, the war-
01146-92002, July 2002          ranty terms in the sepa-          CAUTION
Agilent Technologies, Inc.      rate agreement shall
1900 Garden of the Gods Road control.                             A CAUTION notice
Colorado Springs, CO 80907                                        denotes a hazard. It
USA                             Technology Licenses               calls attention to an
                                The hardware and/or soft-
Restricted Rights Legend        ware described in this docu-      operating procedure,
If software is for use in the   ment are furnished under a        practice, or the like
performance of a U.S. Govern- license and may be used or          that, if not correctly
ment prime contract or sub-     copied only in accordance
contract, Software is           with the terms of such license.
                                                                  performed or adhered
delivered and licensed as                                         to, could result in
“Commercial computer soft-                                        damage to the product
ware” as defined in DFAR                                          or loss of important
252.227-7014 (June 1995), or as
a “commercial item” as                                            data. Do not proceed
defined in FAR 2.101(a) or as                                     beyond a CAUTION
“Restricted computer soft-                                        notice until the
ware” as defined in FAR
52.227-19 (June 1987) or any                                      indicated conditions
equivalent agency regulation                                      are fully understood
or contract clause. Use, dupli-                                   and met.
cation or disclosure of Soft-
ware is subject to Agilent
Technologies’ standard com-
mercial license terms, and
non-DOD Departments and
Agencies of the U.S. Govern-
ment will receive no greater
than Restricted Rights as
defined in FAR 52.227-19(c)(1-
2) (June 1987). U.S. Govern-
ment users will receive no
greater than Limited Rights as
defined in FAR 52.227-14 (June
1987) or DFAR 252.227-7015
(b)(2) (November 1995), as
applicable in any technical
data.
Document Warranty
The material contained in
this document is provided
“as is,” and is subject to
being changed, without
notice, in future editions.
Further, to the maximum
extent permitted by appli-
cable law, Agilent dis-
claims all warranties,
either express or implied,
sA
Agilent Technologies
Printed in the Malaysia
*01146-92002*