Test Equipment Depot - 800.517.8431 - 99 Washington Street Melrose, MA 02176 - TestEquipmentDepot.
com
                         Fluke 170 Series
               True-rms Digital Multimeter
                                    Extended Specifications
Model Differences
(all other specifications are the same for each model)
 Feature                                                    Model
                                         175                177                 179
 Backlight                                                                      
 Temperature Measurement                                                         
 Basic dc voltage accuracy            0.15%               0.09%               0.09%
Nominal Specifications
 Function                                             Absolute Range or Description
 AC Voltage, True-rms             0.1 mV to 1000V (1 kHz)
 DC Voltage                       0.1 mV to 1000V
 Continuity                       Beeper guaranteed on < 25 , guaranteed off > 250 ;
                                  detects opens or shorts of 250 s or longer.
 Resistance                       0.1 to 50.00 M
 Diode Test                       2.400V
 Capacitance                      1 nF to 9999 uF
 AC Current, True-rms             0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
 DC Current                       0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
 Frequency                        2 Hz to 50 kHz
 Temperature (179 Only)           -40 C to +400 C; -40 F to +752 F
 Basic dc voltage accuracy        0.15% (175)                              0.09% (177 & 179)
 Basic ac voltage accuracy        1.0%
Features
 Feature                          Description
 Digital Display                  6000 counts, updates 4 x second
 Analog Bargraph Display          33 segments, updates 40 x second
 Backlight (177 & 179 only)       Automatically turns off after 2 minutes to save battery life
                                  The timeout feature can be disabled with a power-up option
 HOLD & Auto HOLD                 HOLD: freezes the display at the push of a button
                                  Auto HOLD: Display holds present reading until it detects new stable input,
                                  then the meter beeps and displays new reading
 MIN MAX AVG                      Minimum, maximum, and average reading memory
 Manual or auto ranging           In auto range, the meter selects the range with the best resolution for
                                  the present measurement value
 Fast continuity/open detection   The beeper sounds with a stretched pulse for opens or shorts as
                                  brief as 250 s
 Test lead alert                  The message LEAd appears briefly on the display when the rotary switch
                                  is moved to or from any A (Amps) position
 Power-up options                 (1) Turn on all LCD segments, (2) Disable beeper, (3) Disable sleep mode,
                                  (4) Enable smoothing, (5) Disable backlight timeout (Models 177 & 179 only)
 Closed-case calibration          No internal adjustments needed
 Probe holders                    The instrument comes with built-in probe holders for probe storage and
                                  for convenience when making measurements
 Battery access door              Battery replacement without voiding calibration
 High-impact overmold case        Integrated overmolded protection provides superior impact protection for
                                  your meter
General Specifications
Accuracy is specified for 1 year after calibration, at operating temperatures of 18 C to 28 C, with relative
humidity of 0 % to 75 %. Accuracy specifications take the form of:  ( [ % of Reading ] + [ Counts ] )
Maximum voltage between any             1000V DC or AC RMS
terminal and earth ground
Surge Protection                        8 kV peak per IEC 61010
Fuse for mA inputs                      440 mA, 1000 V FAST Fuse
Fuse for A input                        11A, 1000V FAST Fuse
Display                                 Digital: 6,000 counts, updates 4/sec
                                        Bar Graph: 33 segments, updates 40/sec
                                        Frequency: 9,999 counts
                                        Capacitance: 9,999 counts
Altitude                                Operating: 2000 m; Storage: 12000 m
Temperature                             Operating: -10 C to +50 C
                                        Storage: -30 C to +60 C
Temperature coefficient                 0.1 X (specified accuracy / C)
                                        (< 18 C or > 28 C)
Electromagnetic Compatibility           In an RF field of 3 V/M, accuracy = specified accuracy except in
(EN 61326-1:1997)                       temperature: specified accuracy  5 C,  9 F
Relative Humidity                       0 % to 90 % @ 0 C to 35 C;
                                        0 % to 70 % @ 36 C to 50 C
Relative Humidity in                    0 % to 80 % @ 0C to 35C;
50 M Range                             0 % to 70 % @ 36 C to 50 C
Battery Life                            Alkaline: ~200 hrs typical
Size, with Holster (H x W x L)          4.3 cm x 9 cm x 18.5 cm
Weight                                  420g
Safety Compliances                      ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage
                                        Category III, 600 V Overvoltage Category IV
Certifications                          CSA, TV (EN61010),
                                        Australian (N10140)
Detailed Specifications
                                                           Accuracy  ( [ % of Reading ] + [ Counts ] )
Function              Range 1                Resolution    Model 175           Model 177             Model 179
AC Volts 2            600.0 mV               0.1 mV        1.0 % + 3           1.0 % + 3             1.0 % + 3
                      6.000V                 0.001V        (45 Hz to 500 Hz)   (45 Hz to 500 Hz)     (45 Hz to 500 Hz)
                      60.00V                 0.01V
                      600.0V                 0.1V
                      1000V                  1V            2.0 % + 3           2.0 % + 3             2.0 % + 3
                                                           (500 Hz to 1 kHz)   (500 Hz to 1 kHz)     (500 Hz to 1 kHz)
DC mV                 600.0 mV               0.1 mV        0.15 % + 2          0.09 % + 2            0.09 % + 2
DC Volts              6.000V                 0.001V        0.15 % + 2          0.09 % + 2            0.09 % + 2
                      60.00V                 0.01V
                      600.0V                 0.01V
                      1000V                  1V            0.15 % + 2          0.1 % + 2             0.1 % + 2
Continuity            600                   1            Meter beeps at < 25 , beeper turns off at > 250 ;
                                                           detects opens or shorts of 250 ms or longer.
Ohms                  600.0                 0.1          0.9 % + 2           0.9 % + 2             0.9 % + 2
                      6.000 k               0.001 k      0.9 % + 1           0.9 % + 1             0.9 % + 1
                      60.00 k               0.01 k       0.9 % + 1           0.9 % + 1             0.9 % + 1
                      600.0 k               0.1 k        0.9 % + 1           0.9 % + 1             0.9 % + 1
                      6.000 M               0.001 M      0.9 % + 1           0.9 % + 1             0.9 % + 1
                      50.00 M               0.01 M       1.5 % + 3           1.5 % + 3             1.5 % + 3
Diode test            2.400V                 0.001V        1%+2
Capacitance           1000 nF                1 nF          1.2 % + 2           1.2 % + 2             1.2 % + 2
                      10.00 F               0.01 F       1.2 % + 2           1.2 % + 2             1.2 % + 2
                      100.0 F               0.1 F        1.2 % + 2           1.2 % + 2             1.2 % + 2
                      9999 F 3              1 F          10 % typical        10 % typical          10 % typical
AC Amps               60.00 mA               0.01 mA       1.5 % + 3           1.5 % + 3             1.5 % + 3
(True-rms)            400.0 mA               0.1 mA
                      (600 mA for 18 hrs)    0.001A
                      6.000A
(45 Hz to 1 kHz)      10.00A                 0.01A
                      (20A for 30s)
DC Amps               60.00 mA               0.01 mA       1.0 % + 3           1.0 % + 3             1.0 % + 3
                      400.0 mA               0.1 mA
                      (600 mA for 18 hrs)
                      6.000A                 0.001A
                      10.00A                 0.01A
                      (20A for 30s)
Hz                    99.99 Hz               0.01 Hz       0.1 % + 1           0.1 % + 1             0.1 % + 1
(AC- or DC-           999.9 Hz               0.1 Hz
coupled,              9.999 kHz              0.001 kHz
V or A 4, 5 input )   99.99 kHz              0.01 kHz
Temperature           -40 C to +400 C      0.1 C        NA                  NA                    1 % + 1.0 C
                      -40 F to +752 F      0.1 F                                                  1 % + 1.8 F
MIN MAX AVG           For DC functions, accuracy is the specified of the measurement function  12 counts for changes
                      longer than 275 ms in duration.
                      For AC functions, accuracy is the specified of the measurement function  40 counts for changes
                      longer than 1.2 s in duration.
1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
2. Crest factor of  3 at full scale up to 500 V, decreasing linearly to crest factor  1.5 at 1000 V.                   Fluke. Keeping your world
3. In the 9999 F range for measurements to 1000 F, the measurement accuracy is 1.2 % for all models.                          up and running.
4. In mA and A ranges, frequency measurement is specified to 30 kHz.
5. Frequency < 10 kHz are not specified in 600 mV AC, 60 mA AC, and 6 A AC ranges.
Function Overload Protection 1 Input Impedance            Common Mode Rejection Ratio Normal Mode Rejection
                               (Nominal)                  (1 k Unbalanced)
Volts AC     1000 V RMS or DC       > 10 M < 100 pF      > 60 dB @ DC, 50 or 60 Hz
Volts DC     1000 V RMS or DC       > 10 M <100 pF       >120 dB @ DC, 50 or 60 Hz           > 60 dB @ 50 Hz or 60 Hz
                                    Open Circuit Test     Full Scale Voltage To:              Short Circuit Current
                                    Voltage               6.0 M          50 M
Ohms         1000V RMS or DC        < 1.5 V DC            < 600 mV DC     < 1.5 V DC          < 500 A
Diode test 1000V RMS or DC          2.4 to 3.0 V DC       2.4 V DC                            < 1.2 mA typical
1. 10 7 V-Hz maximum.