Fluke 170 Series True-Rms Digital Multimeter: Extended Specifications
Fluke 170 Series True-Rms Digital Multimeter: Extended Specifications
           Model Differences
           (all other specifications are the same for each model)
            Feature                                                    Model
                                                    175                177                 179
            Backlight                                                   •                   •
            Temperature Measurement                                                         •
            Basic dc voltage accuracy            0.15%               0.09%               0.09%
           Nominal Specifications
            Function                                             Absolute Range or Description
            AC Voltage, True-rms             0.1 mV to 1000V (1 kHz)
            DC Voltage                       0.1 mV to 1000V
            Continuity                       Beeper guaranteed on < 25 Ω, guaranteed off > 250 Ω;
                                             detects opens or shorts of 250 µs or longer.
            Resistance                       0.1Ω to 50.00 MΩ
            Diode Test                       2.400V
            Capacitance                      1 nF to 9999 uF
            AC Current, True-rms             0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
            DC Current                       0.01 mA to 10.00A (20.00 A over-range for 30 seconds)
            Frequency                        2 Hz to 50 kHz
            Temperature (179 Only)           -40 ºC to +400 ºC; -40 ºF to +752 ºF
            Basic dc voltage accuracy        0.15% (175)                              0.09% (177 & 179)
            Basic ac voltage accuracy        1.0%
           Features
            Feature                          Description
            Digital Display                  6000 counts, updates 4 x second
            Analog Bargraph Display          33 segments, updates 40 x second
            Backlight (177 & 179 only)       Automatically turns off after 2 minutes to save battery life
                                             The timeout feature can be disabled with a power-up option
            HOLD & Auto HOLD                 HOLD: freezes the display at the push of a button
                                             Auto HOLD: Display holds present reading until it detects new stable input,
                                             then the meter beeps and displays new reading
            MIN MAX AVG                      Minimum, maximum, and average reading memory
            Manual or auto ranging           In auto range, the meter selects the range with the best resolution for
                                             the present measurement value
            Fast continuity/open detection   The beeper sounds with a stretched pulse for opens or shorts as
                                             brief as 250 µs
            Test lead alert                  The message “LEAd” appears briefly on the display when the rotary switch
                                             is moved to or from any A (Amps) position
            Power-up options                 (1) Turn on all LCD segments, (2) Disable beeper, (3) Disable sleep mode,
                                             (4) Enable smoothing, (5) Disable backlight timeout (Models 177 & 179 only)
            Closed-case calibration          No internal adjustments needed
            Probe holders                    The instrument comes with built-in probe holders for probe storage and
                                             for convenience when making measurements
            Battery access door              Battery replacement without voiding calibration
            High-impact overmold case        Integrated overmolded protection provides superior impact protection for
                                             your meter
Page 2 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
©2000 Fluke Corporation
Rev. A-11/2000
           General Specifications
            Accuracy is specified for 1 year after calibration, at operating temperatures of 18 °C to 28 °C, with relative
            humidity of 0 % to 75 %. Accuracy specifications take the form of: ± ( [ % of Reading ] + [ Counts ] )
            Maximum voltage between any             1000V DC or AC RMS
            terminal and earth ground
            Surge Protection                        8 kV peak per IEC 61010
            Fuse for mA inputs                      440 mA, 1000 V FAST Fuse
            Fuse for A input                        11A, 1000V FAST Fuse
            Display                                 Digital: 6,000 counts, updates 4/sec
                                                    Bar Graph: 33 segments, updates 40/sec
                                                    Frequency: 9,999 counts
                                                    Capacitance: 9,999 counts
            Altitude                                Operating: 2000 m; Storage: 12000 m
            Temperature                             Operating: -10 °C to +50 °C
                                                    Storage: -30 °C to +60 °C
            Temperature coefficient                 0.1 X (specified accuracy / °C)
                                                    (< 18 °C or > 28 °C)
            Electromagnetic Compatibility           In an RF field of 3 V/M, accuracy = specified accuracy except in
            (EN 61326-1:1997)                       temperature: specified accuracy ± 5 °C, ± 9 °F
            Relative Humidity                       0 % to 90 % @ 0 °C to 35 °C;
                                                    0 % to 70 % @ 36 °C to 50 °C
            Relative Humidity in                    0 % to 80 % @ 0°C to 35°C;
            50 MΩ Range                             0 % to 70 % @ 36 °C to 50 °C
            Battery Life                            Alkaline: ~200 hrs typical
            Size, with Holster (H x W x L)          4.3 cm x 9 cm x 18.5 cm
            Weight                                  420g
            Safety Compliances                      ANSI/ISA S82.02.01, CSA C22.2-1010.1, IEC 61010 to 1000 V Overvoltage
                                                    Category III, 600 V Overvoltage Category IV
            Certifications                          CSA, TÜV (EN61010),
                                                    Australian (N10140)
Page 3 of 3, Document #2155
170 Series True-rms DMMs
Extended Specifications
©2000 Fluke Corporation
Rev. A-11/2000
           Detailed Specifications
                                                                       Accuracy ± ( [ % of Reading ] + [ Counts ] )
            Function              Range 1                Resolution    Model 175           Model 177             Model 179
            Continuity            600Ω                   1Ω            Meter beeps at < 25 Ω, beeper turns off at > 250 Ω;
                                                                       detects opens or shorts of 250 ms or longer.
            Ohms                  600.0Ω                 0.1Ω          0.9 % + 2           0.9 % + 2             0.9 % + 2
                                  6.000 kΩ               0.001 kΩ      0.9 % + 1           0.9 % + 1             0.9 % + 1
                                  60.00 kΩ               0.01 kΩ       0.9 % + 1           0.9 % + 1             0.9 % + 1
                                  600.0 kΩ               0.1 kΩ        0.9 % + 1           0.9 % + 1             0.9 % + 1
                                  6.000 MΩ               0.001 MΩ      0.9 % + 1           0.9 % + 1             0.9 % + 1
                                  50.00 MΩ               0.01 MΩ       1.5 % + 3           1.5 % + 3             1.5 % + 3
            Diode test            2.400V                 0.001V        1%+2
            Capacitance           1000 nF                1 nF          1.2 % + 2           1.2 % + 2             1.2 % + 2
                                  10.00 µF               0.01 µF       1.2 % + 2           1.2 % + 2             1.2 % + 2
                                  100.0 µF               0.1 µF        1.2 % + 2           1.2 % + 2             1.2 % + 2
                                  9999 µF 3              1 µF          10 % typical        10 % typical          10 % typical
            AC Amps               60.00 mA               0.01 mA       1.5 % + 3           1.5 % + 3             1.5 % + 3
            (True-rms)            400.0 mA               0.1 mA
                                  (600 mA for 18 hrs)    0.001A
                                  6.000A
            (45 Hz to 1 kHz)      10.00A                 0.01A
                                  (20A for 30s)
            DC Amps               60.00 mA               0.01 mA       1.0 % + 3           1.0 % + 3             1.0 % + 3
                                  400.0 mA               0.1 mA
                                  (600 mA for 18 hrs)
                                  6.000A                 0.001A
                                  10.00A                 0.01A
                                  (20A for 30s)
            Hz                    99.99 Hz               0.01 Hz       0.1 % + 1           0.1 % + 1             0.1 % + 1
            (AC- or DC-           999.9 Hz               0.1 Hz
            coupled,              9.999 kHz              0.001 kHz
            V or A 4, 5 input )   99.99 kHz              0.01 kHz
            Temperature           -40 °C to +400 °C      0.1 °C        NA                  NA                    1 % + 1.0 °C
                                  -40 °F to +752 °F      0.1 °F                                                  1 % + 1.8 °F
            MIN MAX AVG           For DC functions, accuracy is the specified of the measurement function ± 12 counts for changes
                                  longer than 275 ms in duration.
                                  For AC functions, accuracy is the specified of the measurement function ± 40 counts for changes
                                  longer than 1.2 s in duration.
            1. All AC voltage and AC current ranges are specified from 5 % of range to 100 % of range.
            2. Crest factor of ≤ 3 at full scale up to 500 V, decreasing linearly to crest factor ≤ 1.5 at 1000 V.                   Fluke. Keeping your world
            3. In the 9999 µF range for measurements to 1000 µF, the measurement accuracy is 1.2 % for all models.                          up and running.
            4. In mA and A ranges, frequency measurement is specified to 30 kHz.
            5. Frequency < 10 kHz are not specified in 600 mV AC, 60 mA AC, and 6 A AC ranges.
                                                                                                                                     Fluke Corporation
            Function Overload Protection 1 Input Impedance            Common Mode Rejection Ratio Normal Mode Rejection
                                           (Nominal)                  (1 kΩ Unbalanced)                                              PO Box 9090, Everett, WA USA 98206
            Volts AC     1000 V RMS or DC       > 10 MΩ < 100 pF      > 60 dB @ DC, 50 or 60 Hz                                      Fluke Europe B.V.
                                                                                                                                     PO Box 1186, 5602 BD
            Volts DC     1000 V RMS or DC       > 10 MΩ <100 pF       >120 dB @ DC, 50 or 60 Hz           > 60 dB @ 50 Hz or 60 Hz   Eindhoven, The Netherlands
                                                Open Circuit Test     Full Scale Voltage To:              Short Circuit Current      For more information call:
                                                Voltage               6.0 MΩ          50 MΩ                                          U.S.A. (800) 443-5853 or
            Ohms         1000V RMS or DC        < 1.5 V DC            < 600 mV DC     < 1.5 V DC          < 500 µA                   Fax (425) 446-5116
                                                                                                                                     Europe (31 40) 2 675 200 or
            Diode test 1000V RMS or DC          2.4 to 3.0 V DC       2.4 V DC                            < 1.2 mA typical           Fax (31 40) 2 675 222
            1. 10 7 V-Hz maximum.                                                                                                    Canada (905) 890-7600 or
                                                                                                                                     Fax (905) 890-6866
                                                                                                                                     Other countries (425) 446-5500 or
                                                                                                                                     Fax (425) 446-5116
                                                                                                                                     Web access: http://www.fluke.com
                                                                                                                                     ©2001 Fluke Corporation. All rights reserved.
                                                                                                                                     Specifications subject to change without notice.
                                                                                                                                     Printed in U.S.A. 3/2001
                                                                                                                                     Printed on recycled paper.
Page 1 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
                      Fluke 187/189
         True-rms Digital Multimeter
                                                     Extended Specifications
Features +34 to +60 0.01 dB 0.1 dB 0.2 dB 0.1 dB Not specified Not specified
    Feature                     Description
                                                                                                                                                               Accuracy
    Dual Displays               50,000 count primary display
                                5,000 count secondary display                                                                     45 Hz-            20 Hz-             1 kHz-           20 kHz-
                                                                                            Function     Range         Resolution 1 kHz             45 Hz              20 kHz           100 kHz
    Backlight with 2             Bright white backlight for clear readings
    brightness selections        in poorly lighted areas                                    AC µA        500.00 µA     0.01 µA      0.75% + 20       1% + 20           0.75% + 20       6% + 40
    Fast Autorange              Meter automatically selects best range -                                 5000.0 µA     0.1 µA       0.75% + 5       1% + 5             0.75% + 10       2% + 40
                                instantly                                                   AC mA        50.000 mA     0.001 mA     0.75% + 20       1% + 20           0.75% + 20       9% + 40
    AC+DC True RMS, ac           Choices for AC only, AC and DC dual                                     400.00 mA     0.01 mA      0.75% + 5        1% + 5            1.5% + 10        4% + 40
    rms specified to 100 kHz     display, or AC+DC readings
                                                                                            AC A         5.0000A       0.0001A      1.5% + 20        1.5% + 20         6% + 40          Not specified
    dBm, dBV                    User selectable impedance references
                                for dBm                                                                  10.000A4      0.001A       1.5% + 5        1.5% + 5           5% + 10          Not specified
    AutoHOLD                    Holds readings on display
    Continuity/Open Test        Beeper sounds for Ohms readings below                                                               Accuracy                  Accuracy Dual Display AC or AC+DC5
                                threshold or for momentary
                                open circuit indication                                                                                                       20 Hz-        45 Hz-        1kHz-
                                                                                            Function     Range         Resolution DC                          45 Hz         1 kHz         20 kHz
    Fast Bar Graph              51 Segments for peaking and nulling
                                                                                            DC mV        50.000 mV     0.001mV      0.1% +20                  2% + 80       0.5% + 40
    Duty Cycle/Pulse Width      Measure time a signal is on or off in %                                                                                                                   6% + 40
                                or milliseconds                                                          500.00 mV     0.01 mV      0.03% + 2                 2% + 80       0.5% + 40
    MIN MAX/Fast MIN MAX        Record Maximum, Minimum, and Average                                     3000.0 mV     0.1 mV       0.025% + 5                2% + 80       0.5% + 40
    with elapsed and            values. Real Time for MAX or MIN, elapsed                   DC V         5.0000V       0.0001V      0.025% + 103              2% + 80       0.5% + 40     2% + 40
    Real Time Stamp             time for AVG.
                                Fast MIN MAX captures peaks to 250 µsec.                                 50.000V       0.001V       0.03% + 33                2% + 80       0.5% + 40
    Closed Case Calibration     No internal adjustments needed                                           500.00V       0.01V        0.1% + 23                 2% + 80       0.5% + 40     Not specified
    Battery/Fuse                Battery or fuse replacement without                                      1000.0V       0.1V         0.1% + 23                 2% + 80       0.5% + 40     Not specified
    Access Door                 voiding calibration
                                                                                            DC µA        500.00 µA     0.01 µA      0.25% + 20                1% + 20       1% + 20       2% + 40
    Hi-Impact Overmold Case Integrated Protective Holster provides
                            superior impact protection for your meter                                    5.000 µA      0.1 µA       0.25% + 2                 1% + 10       0.75% + 10 2% + 40
                                                                                            DC mA        50.000 mA     0.001 mA     0.15% + 10                1% + 20       0.75% + 20 2% + 40
1
  For the 5,000 count mode, divide the number of least significant digits (counts) by 10.                400.00 mA     0.01 mA      0.15% + 2                 1% + 10       1% + 10       3% + 40
2
  A residual reading of 8 to 180 digits with leads shorted, will not affect stated          DC A         5.0000A       0.0001A      0.5% + 10                 2% + 20       2% + 20       6% + 40
  accuracy above 5% of range.
3
  20 counts in dual display DC or AC + DC                                                                10.000A4      0.001A       0.5% + 2                  1.5% + 10     1.5% + 10     5% + 40
4
  10A continuous up to 35°C, less than 10 minutes 35° to 55°. 20A overload for
    30 seconds maximum
5
    See AC conversion notes for AC mV and V.
Above specifications are subject to change without notice.
Page 2 of 3, Document #2153
187/189 Extended Specifications
©2000 Fluke Corporation
Rev. A-8/00
   Input Impedance
    Function                                    Input Impedance (Nominal)
    Volts, mV                                         10 MΩ, < 100 pF
                         Common Mode Rejection Ratio                     Normal Mode Rejection
    DC Volts, mV        >100 dB at dc, 50 Hz or 60 Hz ± 0.1% >90 dB at 50 Hz or 60 Hz ± 0.1%
    AC Volts, mV        > 90 dB dc to 60 Hz
                                                                              Full-Scale Voltage
                            Open Circuit Test Voltage
                                                               To 5 MΩ               30 to 500 MΩ + nS
    Ohms                < 5V                                   500 mV                3.1V
    Diode Test          < 5V                                                   3.1000V
                                                Typical Short-Circuit Current
                          500Ω         5 kΩ          50 kΩ      500 kΩ           5 MΩ            30 MΩ
    Ohms                   1 µA        100 µA        10 µA        1 µA           0.1 µA          0.1 µA
    Diode Test                                         1.0 mA typical
Accuracy Specifications
                                                                                       Accuracy                                   [1]
           Function                   Range               Resolution                                                      Model
                                                                             ± ([% of Reading] + [Counts])
DC millivolts                 600.0 mV                  0.1 mV                           0.5 % + 2                  114, 115, 116, 117
                              6.000 V                   0.001 V
DC Volts                      60.00 V                   0.01 V                           0.5 % + 2                  114, 115, 116, 117
                              600.0 V                   0.1 V
                                                                       DC, 45 to 500 Hz        500 Hz to 1 kHz
               [2]
Auto-V LoZ True-rms           600.0 V                   0.1 V              2.0 % + 3                 4.0 % + 3      114, 116, 117
                                                                         45 to 500 Hz          500 Hz to 1 kHz
                [2]
AC millivolts True-rms        600.0 mV                  0.1 mV             1.0 % + 3                 2.0 % + 3      114, 115, 116, 117
                              6.000 V                   0.001 V
         [2]
AC Volts True-rms             60.00 V                   0.01 V             1.0 % + 3                 2.0 % + 3      114, 115, 116, 117
                              600.0 V                   0.1 V
                                                                       Beeper on <20 Ω, off >250 Ω; detects opens
Continuity                    600 Ω                     1Ω                                                          114, 115, 116, 117
                                                                       or shorts of 500 μs or longer.
                              600.0 Ω                   0.1 Ω                            0.9 % + 2
                              6.000 kΩ                  0.001 kΩ                         0.9 % + 1
                              60.00 kΩ                  0.01 kΩ                          0.9 % + 1
Ohms                                                                                                                114, 115, 116, 117
                              600.0 kΩ                  0.1 kΩ                           0.9 % + 1
                              6.000 MΩ                  0.001 MΩ                         0.9 % + 1
                              40.00 MΩ                  0.01 MΩ                          1.5 % + 2
Diode test                    2.000 V                   0.001 V                          0.9 % + 2                  115, 116, 117
Capacitance                   1000 nF                   1 nF                             1.9 % + 2                  115, 116, 117
                              10.00 μF                  0.01 μF                          1.9 % + 2
                              100.0 μF                  0.1 μF                           1.9 % + 2
                              9999 μF                   1 μF                   100 μF - 1000 μF: 1.9 % + 2
                                                                                   >1000 μF: 5 % + 20
Lo-Z Capacitance (Power-                                                                  20% + 2                   115, 116, 117
                                              1 nF to 500 μF
up option)                                                                           (10% +2 typical)
                              6.000 A                    0.001 A
                                        [4]
                      [2]     10.00 A                    0.01 A
AC Amps True-rms
                              20 A overload for 30                                       1.5 % + 3                  115, 117
(45 Hz to 500 Hz)
                              seconds maximum, 10
                              minutes rest minimum.
                              6.000 A                    0.001 A
                                        [4]
                              10.00 A                    0.01 A
DC Amps                       20 A overload for 30                                       1.0 % + 3                  115, 117
                              seconds maximum, 10
                              minutes rest minimum.
                                                                                                     [5]
Temperature (Type K           -40 °C to 400 °C           0.1 °C                         1.0 % + 10
                                                                                                     [5]            116
thermocouple)                 -40 °F to 752 °F           0.2°F                          1.0 % + 18
                        [2]
AC μAmps True-rms             600.0 μA                   0.1 μA
                                                                             1.5 % + 3 (2.5 % + 3 > 500 Hz)         116
(45 Hz to 1 kHz)
DC μAmps                      600.0 μA                   0.1 μA                          1.0 % + 2                  116
4
                                                                                                                                                  Digital Multimeters
                                                                                                                                                  Basic Maintenance
Input Characteristics
                                                                                   Common Mode Rejection Ratio
        Function               Input Impedance (Nominal)                                                                                 Normal Mode Rejection
                                                                                       (1 kΩ Unbalanced)
Volts AC                              >5 MΩ <100 pF                                     >60 dB at dc, 50 or 60 Hz
Volts DC                             >10 MΩ <100 pF                                    >100 dB at dc, 50 or 60 Hz                          >60 dB at 50 or 60 Hz
Auto-V LoZ                             ~3 kΩ <500 pF                                    >60 dB at dc, 50 or 60 Hz
                               Open Circuit Test Voltage                                    Full Scale Voltage                             Short Circuit Current
Ohms                                                                              To 6.0 MΩ                         40 MΩ
                                          <2.7 V dc                                                                                                <350 μA
                                                                                  <0.7 V dc                       <0.9 V dc
Diode Test                                <2.7 V dc                                              2.000 V dc                                        <1.2 mA
Basic Maintenance
Testing the Fuse (115 & 117 only)
To test the fuse:
            1. Set the rotary switch to Ω.
            2. Plug a test lead into the v jack and touch the probe to the 10A jack, as shown in Figure 1.
If the display shows a resistance value in the range of that shown in Figure 1, the fuse is good.
If the display reads 0L, replace the fuse and test again.
If the display shows any other value, have the Meter serviced. See “Service Information” earlier in this document.
                                                                                                                                                                       5
                                                                                                RIGOL
DC Characteristics
[1]    Specifications are for 60 minute warm–up, selecting 5 3/4 reading resolution and calibration temperature
       18 ℃ – 28 ℃.
[2]    20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range.
[3]    Specifications are for 4–wire resistance function, or 2–wire resistance using Math Null. Without Math Null,
       add 0.2 Ω additional errors in 2–wire resistance function.
[4]    For each additional volt over ± 500 VDC add 0.02 mV of error.
[5]    For current terminal, > 7A DC or ACRMS for 30 seconds ON and 30 seconds OFF.
[6]    Accuracy specifications are for the voltage measured at the input terminals only. 1 mA test current is
       typical.
       Variation in the current source will create some variation in the voltage drop across a diode junction.
Settling Considerations
Reading settling times are affected by source impedance, cable dielectric characteristics, and input signal
changes. Typically, settling time <1.5s when source impedance less than 1kΩ.
AC Characteristics
28 ℃ to 55 ℃
[1]    Specifications are for 60 minute warm–up and select 5 3/4 reading resolution.
[2]    20% over range on all ranges, except DCV 1000V, ACV 750V, DCI and ACI 10A range.
[3]    Specifications are for sine wave input >5% of range. For inputs from 1% to 5% of range and <50 kHz,
       add 0.1% of range additional error. For 50 kHz to 100 kHz, add 0.13% of range.
       750 Vac range limited to 100 kHz or 8x107 Volt-Hz.
[4]    Specifications are for sine wave input >5% of range. Add 0.1% of the range for the sine wave input is
       1%~5% of the range.
[5]    For current terminal, > 7A DC or ACRMS for 30 seconds ON and 30 seconds OFF.
[6]    Typically 30% of reading error at 100kHz.
Frequencies greater than these filter settings are specified with no additional errors.
Settling Considerations
Applying >300VRMS (or >1ARMS) will cause self-heating in signal-conditioning components. These errors are
included in the instrument specifications. Internal temperature changes due to self-heating may cause additional
error on lower ac voltage ranges. The additional error will be less than 0.02% of reading and will generally
dissipate within a few minutes.
28 ℃ to 55 ℃
Measurement Considerations
All frequency counters are susceptible to error when measuring low–voltage, low–frequency signals. Shielding
inputs from external noise pickup is critical for minimizing measurement errors.
Settling Considerations
Errors will occur when attempting to measure the frequency or period of an input following a DC offset voltage
change. The input blocking RC time constant must be allowed to fully settle (up to 1 sec) before the most
accurate measurements are possible.
Capacitance Characteristics
                                        Accuracy Specifications (% of reading + % of range)[1]
                                                                                    Temperature
                                                               1 Year               Coefficient
 Function            Range[2]          Test Current
                                                               23℃± 5℃               0 ℃ to 18 ℃
28 ℃ to 55 ℃
[1]    Specifications are for 60 minute warm–up using Math Null. Additional errors may occur for non–film
       capacitors.
[2]    Specifications are for 1% to 120% of range on the 4nF range and 10% to 120% of range on all other
       ranges.
Measuring Characteristics
DC Voltage
 Input Resistance:
   400mV, 4V ranges          Selectable 10MΩ ± 2% or >10GΩ
   40V, 400V, 1000V ranges   10MΩ ± 2%
Resistance
 Measurement Method:         Selectable 4-wire or 2-wire.
                             Current source referenced to LO input.
 Open-circuit Voltage:       Limit in <7V.
 Max. Lead Resistance:       10% of range per lead for 400Ω, 1kΩ per lead
 (4-wire ohms)               on all other ranges.
 Input Protection:           1000V on all ranges.
DC Current
 Shunt Resistor:             0.025Ω for 1A, 10A
                             1.025Ω for 200mA
                             11.025Ω for 2mA, 20mA
 Input Protection:           Externally accessible 10A, 250V fuse
                             Internal 12A, 250V fuse
Capacitance
 Measurement Method:                Current input with measurement of resulting ramp.
 Connection Type:                   2-wire
CAUTION:
Real-time Clock
 Precision:                         1min/month (Environment Temperature >0℃)
 Clock battery Life:                2 years
Math Functions
Null, Min/Max/Average, dBm, dB, Limit Test (with TTL output)
Other Functions
Reading Hold, Ratio Measurement
High-speed Sampling
50kSa/s (In Datalog function)
Reading Resolution
480,000 Count, >5 3/4 digits