Manual Supplement
2009-2013 Fluke Corporation. All rights reserved.   
   Manual Title:  114,115, 116, and 117 Calibration   Supplement Issue: 5   
Print Date:  September 2006  Issue Date:       4/13 
Revision/Date: 1, 10/08  Page Count:       3               
    This supplement contains information necessary to ensure the 
accuracy of the above manual. 
114, 115, 116, and 117 Calibration Information  Manual Supplement 
11/09  1 
Change #1 
On page 10, delete step 30. 
Change #2 
On page 4, replace Table 2 with the following: 
Table 2. Accuracy Specifications 
Function  Range  Resolution 
Accuracy 
 ([% of Reading] + [Counts]) 
Model
[1] 
DC millivolts  600.0 mV  0.1 mV  0.5 % + 2  114, 115, 116, 117
DC Volts  
6.000 V 
60.00 V 
600.0 V 
0.001 V 
0.01 V 
0.1 V 
0.5 % + 2  114, 115, 116, 117
  DC, 45 to 500 Hz 500 Hz to 1 kHz   
Auto-V LoZ
[2]
 True-rms  600.0 V  0.1 V  2.0 % + 3  4.0 % + 3  114, 116, 117 
  45 to 500 Hz  500 Hz to 1 kHz   
AC millivolts
[2]
 True-rms  600.0 mV  0.1 mV  1.0 % + 3  2.0 % + 3  114, 115, 116, 117
AC Volts
[2]
 True-rms 
6.000 V 
60.00 V 
600.0 V 
0.001 V 
0.01 V 
0.1 V 
1.0 % + 3  2.0 % + 3  114, 115, 116, 117
Continuity  600   1  
Beeper on <20 , off >250 ; detects opens 
or shorts of 500 s or longer. 
114, 115, 116, 117
Ohms 
600.0  
6.000 k 
60.00 k 
600.0 k 
6.000 M 
40.00 M 
0.1  
0.001 k 
0.01 k 
0.1 k 
0.001 M 
0.01 M 
0.9 % + 2 
0.9 % + 1 
0.9 % + 1 
0.9 % + 1 
0.9 % + 1 
5.0 % + 2 
114, 115, 116, 117
Diode test  2.000 V  0.001 V  0.9 % + 2  115, 116, 117 
Capacitance  1000 nF 
10.00 F 
100.0 F 
9999 F 
1 nF 
0.01 F 
0.1 F 
1 F 
1.9 % + 2 
1.9 % + 2 
1.9 % + 2 
100 F - 1000 F: 1.9 % + 2 
>1000 F: 5 % + 20 
115, 116, 117 
Lo-Z Capacitance 
(Power-up option) 
1 nF to 500 F  10 % +2  typical 
115, 116, 117 
AC Amps True-rms
[2] 
(45 Hz to 500 Hz) 
6.000 A  
10.00 A
[4] 
20 A overload for 30 
seconds maximum, 10 
minutes rest minimum. 
0.001 A 
0.01 A 
1.5 % + 3  115, 117 
DC Amps 
6.000 A 
10.00 A
[4]  
20 A overload for 30 
seconds maximum, 10 
minutes rest minimum. 
0.001 A 
0.01 A 
1.0 % + 3  115, 117 
Manual Supplement    114, 115, 116, and 117 Calibration Information 
  2  4/13  
Temperature (Type K 
thermocouple) 
-40 C to 400 C 
-40 F to 752 F 
0.1 C 
0.2 F 
1.0 % + 10
[5] 
1.0 % + 18
[5] 
116 
AC Amps True-rms
[2] 
(45 Hz to 1 kHz) 
600.0 A  0.1 A 
1.5 % + 3 (2.5 % + 3 > 500 Hz)  116 
DC Amps  600.0 A  0.1 A  1.0 % + 2  116 
Hz (V or A input)
[3] 
99.99 Hz 
999.9 Hz 
9.999 kHz 
50.00 kHz 
0.01 Hz 
0.1 Hz 
0.001 kHz 
0.01 kHz 
0.1 % + 2   115, 116, 117 
Notes: 
[1]  Models listed in this column also refer to the C version of the model. For example, those rows containing model 115 are applicable to the 115C 
as well. 
[2]  All ac ranges except Auto-V LoZ are specified from 1 % to 100% of range. Auto-V LoZ is specified from 0.0 V. Because inputs below 1 % of range 
are not specified, it is normal for this and other true-rms meters to display non-zero readings when the test leads are disconnected from a circuit 
or are shorted together. For volts, crest factor of 3 at 4000 counts, decreasing linearly to 1.5 at full scale. For amps, crest factor of 3. AC volts is 
ac-coupled. Auto-V LoZ, AC mV, AC amps, and AC amps are dc-coupled. 
[3]  AC Volts Hz is ac-coupled and specified from 5 Hz to 50 kHz. AC amps Hz is dc-coupled and specified from 45 Hz to 5 kHz. 
[4]  >10 A unspecified. 
[5]  Temperature uncertainty (accuracy) does not include the error of the thermocouple probe. 
Change #3 
On page 2, under Safety Information: 
Change:  Do not use the Meter around explosive gas or vapor. 
To:  Do not use the Product around explosive gas, vapor, or in damp or wet environments 
Change #4, 63151, 63405 
On page 3, replace Table 1 with: 
Table 1. Electrical Symbols 
Symbol  Description  Symbol  Description 
  AC (Alternating Current)  =  Fuse 
  DC (Direct Current)  _  Double Insulated 
@  Hazardous voltage    Important Information; Refer to manual  
Battery (Low battery when shown on the 
display.) 
}  Earth ground 
_  Do not dispose of this product as unsorted municipal waste. Go to Flukes website for recycling information. 
@  Conforms to relevant South Korean EMC Standards 
CAT II 
Measurement Category II is applicable to test and measuring circuits connected directly to utilization points 
(socket outlets and similar points) of the low-voltage MAINS installation. 
CAT III 
Measurement Category III is applicable to test and measuring circuits connected to the distribution part of 
the buildings low-voltage MAINS installation. 
CAT IV 
Measurement Category IV is applicable to test and measuring circuits connected at the source of the 
buildings low-voltage MAINS installation.  
114, 115, 116, and 117 Calibration Information    Manual Supplement  
4/13  3 
On page 3, under General Specifications, delete Surge Protection, Safety Compliance and 
Certifications and add: 
Safety ............................................................. IEC 61010-1: 600 V CAT III, Pollution Degree 2 
Electromagnetic Environment ..................... IEC 61326-1: Portable   
Electromagnetic Compatibility .................... Applies to use in Korea only: 
Class A Equipment (Industrial Broadcasting & Communication 
Equipment) 
[1] 
[1] This product meets requirements for industrial (Class A) 
electromagnetic wave equipment and the seller or user should 
take notice of it. This equipment is intended for use in business 
environments and is not to be used in homes.