87VMAX Cieng0000
87VMAX Cieng0000
Digital Multimeter
                                                                                                       Calibration Information
             Introduction
                                                                  XWWarning
                            To prevent electric shock or injury, do not perform the
                            performance tests or calibration adjustment procedures unless
                            qualified to do so.
                            The information provided in this document is for the use of
                            qualified personnel only.
                        This document provides adjustment and performance test procedures for the
                        Fluke 87V MAX Digital Multimeter (the Meter).
                        See the 87V MAX Users Manual for complete operating instructions.
             Contact Information
                        To contact Fluke, call one of the following telephone numbers:
                        • Technical Support USA: 1-800-44-FLUKE (1-800-443-5853)
                        • Calibration/Repair USA: 1-888-99-FLUKE (1-888-993-5853)
                        • Canada: 1-800-36-FLUKE (1-800-363-5853)
                        • Europe: +31 402-675-200
                        • Japan: +81-3-6714-3114
                        • Singapore: +65-6799-5566
                        • China: +86-400-921-0835
                        • Brazil: +55-11-3530-8901
                        • Anywhere in the world: +1-425-446-5500
                        Or, visit Fluke's website at www.fluke.com.
                        To register your product, visit http://register.fluke.com.
                        To view, print, or download the latest manual supplement, visit
                        http://us.fluke.com/usen/support/manuals.
             Safety Information
                        General Safety Information is in the printed Safety Information document that
                        ships with the Product and at www.fluke.com. More specific safety information is
                        listed where applicable.
January 2020
© 2020 Fluke Corporation. All rights reserved. Specifications are subject to change without notice.
All product names are trademarks of their respective companies.                                                             1
87V MAX
Calibration Information
             General Specifications
             Maximum voltage between any
             terminal and earth ground ........................................... 1000 V rms
             Fuse Protection for mA or μA inputs .......................... 0.44 A, 1000V, IR 10kA
             Fuse Protection for A inputs........................................ 11 A, 1000 V, IR 17 kA
             Display
                Digital .......................................................................... 6000 counts, updates 4/sec / 19,999 counts in high-resolution
                                                                                                   mode
                Bargraph ..................................................................... 33 segments; updates 40/sec
             Altitude
                Operating .................................................................... 2000 meters
                Storage ....................................................................... 10 000 meters
             Temperature
                Operating .................................................................... -15 °C to 55 °C, to -40 °C for 20 minutes when taken from 20 °C
                Storage ....................................................................... -55 °C to 85 °C (without battery)
                                                                                                -55 °C to 60 °C (with battery)
             Temperature coefficient ............................................... 0.05 X (specified accuracy) / °C (<18 °C or >28 °C)
             Safety ............................................................................. IEC 61010-1: Pollution Degree 2
                                                                                                  IEC 61010-2-033: CAT III 1000 V, CAT IV 600 V
             Electromagnetic Compatibility (EMC) ......................... In an RF field of 3 V/m, accuracy = specified accuracy +20
                                                                           counts, except 600 μA dc range total accuracy = specified
                                                                           accuracy +60 counts. Temperature not specified.
                International ................................................................ IEC 61326-1: Portable Electromagnetic Environment
                                                                                               CISPR 11: Group 1, Class A
                             Group 1: Equipment has intentionally generated and/or uses conductively coupled radio frequency energy
                             that is necessary for the internal function of the equipment itself.
                             Class A: Equipment is suitable for use in all establishments other than domestic and those directly connected
                             to a low-voltage power supply network that supplies buildings used for domestic purposes. There may be
                             potential difficulties in ensuring electromagnetic compatibility in other environments due to conducted and
                             radiated disturbances.
                             Caution: This equipment is not intended for use in residential environments and may not provide adequate
                             protection to radio reception in such environments.
                             Emissions that exceed the levels required by CISPR 11 can occur when the equipment is connected to a test
                         object.
                Korea (KCC) ............................................................... Class A Equipment (Industrial Broadcasting & Communication
                                                                                            Equipment)
                             Class A: Equipment meets requirements for industrial electromagnetic wave equipment and the seller or user
                             should take notice of it. This equipment is intended for use in business environments and not to be used in
                        homes.
               USA (FCC).................................................................. 47 CFR 15 subpart B. This product is considered an exempt
                                                                                           device per clause 15.103. In an RF field of 3 V/M, accuracy =
                                                                                           specified accuracy +20 counts, except 600 μA dc range total
                                                                                           accuracy = specified accuracy +60 counts. Temperature not
                                                                                           specified
             Relative Humidity .......................................................... 0 % to 95 % (0 °C to 35 °C)
                                                                                           0 % to 70 % (35 °C to 55 °C)
       2
                                                                                                                                          Digital Multimeter
                                                                                                                                        Detailed Specifications
Detailed Specifications
For all detailed specifications:
Accuracy is specified for 2 years after calibration, at operating temperatures of 18 °C to 28 °C, with relative humidity at
0 % to 95 %. Accuracy specifications take the form of ±([% of Reading] + [Number of least-significant digits]).
For the 4 ½-digit mode, multiply the number of least-significant digits (counts) by 10.
AC Voltage
AC conversions are ac-coupled and valid from 3 % to 100 % of range.
                                                                                            Accuracy
 Range         Resolution
                                45 – 65 Hz          15 – 200 Hz          200 – 440 Hz        440 Hz – 1 kHz               1 – 5 kHz             5 – 20 kHz
600.0 mV        0.1 mV
                               ±(0.7 % + 4)                                                                          ±(2 % + 4)              ±(2 % + 20) [2]
6.000 V         0.001 V
60.00 V         0.01 V                                                   ±(1.0 % + 4) [1]                                                    Unspecified
                                                                                                                     ±(2 % + 4) [3]
600.0 V         0.1 V                                                                                                                        Unspecified
1000 V          1V             ±(0.7 % + 2)                                                                         Unspecified              Unspecified
                                                                          +1.0 % + 4
      Low-Pass Filter                             ±(1.0 % + 4) [1]                              Unspecified         Unspecified              Unspecified
                                                                         -6.0 % - 4 [4]
[1]   Below 30 Hz, use smoothing function. Below 20 Hz add 0.6 %.
[2]   Below 10 % of range, add 12 counts.
[3]   Frequency range: 1 to 2.5 kHz
[4]   Specification increases from -1 % to -6 % at 440 Hz when filter is used.
                                                                                                                                                               3
87V MAX
Calibration Information
Temperature
                     Range                                          Resolution                                     Accuracy [1,2]
 -200 °C to +1090 °C                                 0.1 °C                                                          ±(1.0 % + 10)
 -328 °F to +1994 °F                                 0.1 °F                                                          ±(1.0 % + 18)
 [1] Does not include error of the thermocouple probe.
 [2] Accuracy specification assumes ambient temperature stable to ± 1 °C. For ambient temperature changes of ± 5 °C, rated accuracy applies after 2
     hours.
AC Current
                                                                                                                 Accuracy[1]
  Function                 Range                  Resolution           Burden Voltage                          (45 Hz – 2 kHz)
DC Current
       Function                  Range                       Resolution             Burden Voltage                          Accuracy
Capacitance
                     Range                                          Resolution                                        Accuracy
 10.00 nF                                            0.01 nF
                                                                                                                   ±(1.0 % + 2) [1]
 100.0 nF                                            0.1 nF
 1.000 μF                                            0.001 μF
 10.00 μF                                            0.01 μF
                                                                                                                    ±(1.0 % + 2)
 100.0 μF                                            0.1 μF
 9999 μF                                             1 μF
 [1]    With a film capacitor or better, using the rel mode to zero residual.
Diode
                     Range                                          Resolution                                        Accuracy
 2.000 V                                             0.001 V                                                        ±(1.0 % + 1)
          4
                                                                                                                                    Digital Multimeter
                                                                                                                                   Detailed Specifications
Frequency
                  Range                                         Resolution                                     Accuracy
199.99 Hz                                          0.01 Hz
1999.9 Hz                                          0.1 Hz
                                                                                                           ±(0.005 % + 1) [1]
19.999 kHz                                         0.001 kHz
199.99 kHz                                         0.01 kHz
>200 kHz                                           0.1 kHz                                                      Unspecified
[1]   From 0.5 Hz to 200 kHz and for pulse widths >2 μs.
Input Characteristics
                                     Input                  Common Mode
                Overload
Function                          Impedance               Rejection Ratio                        Normal Mode Rejection
              Protection[1]
                                   (nominal)             (1 kΩ unbalance)
                                                        > 120 dB at dc, 50 Hz
              1000 V rms                                                                        > 60 dB at 50 Hz or 60 Hz
                                                               or 60 Hz
                                10 MΩ <100 pF
                                                        > 120 dB at dc, 50 Hz
              1000 V rms                                                                        > 60 dB at 50 Hz or 60 Hz
                                                               or 60 Hz
                               10 MΩ < 100 pF
              1000 V rms                               > 60 dB, dc to 60 Hz
                                 (ac-coupled)
                                                                                                                                                        5
87V MAX
Calibration Information
             Basic Maintenance
                                                                     XWWarning
                             To prevent possible electrical shock, fire, or personal injury:
                             •     Remove the input signals before you clean the Product.
                             •     Do not operate the Product with covers removed or the case
                                   open. Hazardous voltage exposure is possible.
                             •     Use only specified replacement parts.
                             •     Have an approved technician repair the Product.
             General Maintenance
                   Periodically wipe the case with a damp cloth and mild detergent. Do not use
                   abrasives or solvents.
                         Dirt or moisture in the terminals can affect readings and can falsely activate the
                         Input Alert feature. Clean the terminals as follows:
                         1. Turn the Meter off and remove all test leads.
                         2. Shake out any dirt that may be in the terminals.
                         3. Soak a clean swab with mild detergent and water. Work the swab around in
                            each terminal. Dry each terminal using canned air to force the water and
                            detergent out of the terminals.
       6
                                                                                    Digital Multimeter
                                                                                     Basic Maintenance
 h Static Awareness h
             Semiconductors and integrated circuits can be damaged by electrostatic
 X           discharge during handling. This notice explains how to minimize damage to
             these components.
            1. Understand the problem.
            2. Learn the guidelines for proper handling.
            3. Use the proper procedures, packaging, and bench techniques.
             Follow these practices to minimize damage to static sensitive parts.
                                           XW Warning
                       To prevent electric shock or personal injury. De-
                       energize the product and all active circuits before
                       opening a product enclosure, touching or handling
                       any PCBs or components.
                                                                                                    7
87V MAX
Calibration Information
             Fuse Test
                   As shown in Figure 1, with the Meter in the  function, insert a test lead into
                   the  jack and place the probe tip on the other end of the test lead against the
                   metal of the current input jack. If “” appears in the display, the probe tip has
                   been inserted too far into the amps input jack. Back the lead out a bit until the
                   message disappears and either OL or a resistance reading appears in the
                   display. The resistance value should be as shown in Figure 1. If the tests give
                   readings other than those shown, have the Meter serviced.
                                                          XWWarning
                          To prevent possible electrical shock, fire, or personal injury:
                          •   Replace a blown fuse with exact replacement only for continued
                              protection against arc flash.
                          •   Use only specified replacement fuses.
                                                                                                         gaq105.emf
                                                      Figure 1. Current Fuse Test
       8
                                                                                  Digital Multimeter
                                                                                      Basic Maintenance
                                                                                                     9
87V MAX
Calibration Information
                      7. Replace the fuse compartment door by aligning the arrow on the fuse door
                         with the arrow on the case bottom and lowering the door into the fuse
                         compartment.
                      8. Replace the fuse compartment seal by aligning the tab on the seal with the
                         outline on the case bottom. Ensure the seal () is properly seated.
                      9. Refer to steps four through six under the Replacing the Batteries section
                         above to reinstall the battery door.
                              1
                                                   3                              4
                                                                                                     gaq10.emf
                                         Figure 2. Battery and Fuse Replacement
       10
                                                                                           Digital Multimeter
                                                                                            Performance Tests
Performance Tests
                                       XWWarning
           To prevent electric shock, do not perform the performance test
           procedures unless the Meter is fully assembled.
        The following performance tests verify the complete operation of the Meter and
        check the accuracy of each Meter function against its specifications.
        Performance tests should be performed bi-annually to ensure that the Meter is
        within accuracy specifications. If the Meter fails any part of the test, calibration
        adjustment and/or repair is indicated.
        In the performance tests, the Meter is referred to as the device under test (DUT).
Required Equipment
      Table 1 lists the equipment required to conduct a performance test on the Meter.
                                  DC Current                    350 μA to 2 A
                                                                ±0.05 %
                                  AC Volts                      0 to 1000 V
                                                                ±0.15 % @ 60 Hz to 20 kHz ±3 %
                                  AC Current                    350 μA to 2 A
                                                                ±0.39 % @ 60 Hz to 1 kHz
                                  Resistance                    1 Ω to 100 MΩ
                                                                ±0.06 %
                                  Capacitance                   9 to 900 μF
                                                                ±0.475 %
                                                                                                          11
87V MAX
Calibration Information
                                 Test
               Step                            Range         5520A Output              Display Reading
                              Function
8 60 V 33 V, 60 Hz 32.75 to 33.25
9 60 V 33 V, 20 kHz N/A
10 60 V 33 V, 5 kHz N/A
       12
                                                                                     Digital Multimeter
                                                                                       Performance Tests
               Test
 Step                       Range               5520A Output             Display Reading
             Function
30 60 kΩ 33 kΩ 32.92 to 33.08
33 50 MΩ 30 MΩ 29.69 to 30.31
40  60 mA 33 mA 32.89 to 33.11
51      VDC Peak Min/Max   6V          8 Vpp, 2 kHz Sq. Wave,     Min = -1.898 to -2.102
                                       DC offset 2 V              Max = 5.896 to 6.104
                                                                                                     13
87V MAX
Calibration Information
                                Test
             Step                                 Range                 5520A Output                                Display Reading
                              Function
             Calibration Adjustments
                          Perform the Calibration Adjustment Procedure if the Meter fails any performance
                          tests. If the adjustment routine is discontinued prior to completion, no changes
                          are made to the calibration constants that are stored in memory. The following is
                          an explanation of the pushbutton features and requirements to enter the CAL
                          mode.
             Cal Mode Pushbutton Functions
                   • The CAL mode will be initiated by holding down the MINMAX pushbutton at
                      power up and entering a four-digit password.
                          •     The AutoHOLD pushbutton will act as an "ENTER" key and will advance
                                through the CAL initiation and adjustment procedure steps.
                          •     The pushbuttons are used to select a four-digit password.
                          During initiation of the CAL mode, a display count will show how many times
                          calibration constants have been written to memory.
             Entering and Displaying the Four-Digit Password
                    When the Meter was manufactured it was given a default password of 1234. The
                    following pushbuttons are used to select the password. Each pushbutton
                    represents the indicated digit.
                             Yellow = 1       MINMAX = 2       RANGE = 3      AutoHOLD = 4
                             Backlight = 5    Continuity = 6 REL = 7          Hz = 8
                          After selecting the password, the user has two choices:
                          Pressing AutoHOLD will display "" which indicates correct password and
                          successful entry. You may now proceed with the first calibration step.
                          Pressing RANGE will display "", which indicates correct password, proceed
                          to select a new password.
                          If the password is incorrect, the concluding AutoHOLD or RANGE pushbutton
                          press will instead cause the Meter to double beep and the display will show
                          "". The password entry process was unsuccessful and can be tried again or
                          you may exit this mode by turning the Meter off.
       14
                                                                                  Digital Multimeter
                                                                               Calibration Adjustments
                                                                                                   15
87V MAX
Calibration Information
                                                                                                    S11        6
                                                          2X-2- 3001 REV 010
                                                                                                          S7
                                                                                       7
                                                                    5
TP1 TP2
3.5 - 5 V
                                                                                                                   gaq101.emf
                                           Figure 3. Revision 010 Board Layout
                      6. While shorting across keypad button S7, turn the rotary switch one position
                         clockwise. See Figure 3.
                      7. Short across keypad button S11. See Figure 3.
                      8. Turn the rotary switch one position counterclockwise, back to its original
                         position.
                      9. The default password is now restored.
                      10. Reassemble the Meter before performing adjustments or testing.
       16
                                                                             Digital Multimeter
                                                                           Calibration Adjustments
3.5 - 5 V
                                                                              gaq102.emf
                     Figure 4. Revision 011 Board Layout
5. Turn the rotary switch from OFF to  while holding down MIN MAX at the
   same time. The Meter should display CAL.
6. Short across keypad button S11 on the back of the PCB. See Figure 4. The
   Meter should beep.
7. Turn the rotary switch one position counterclockwise, back to the OFF
   position.
8. The default password is now restored.
9. Reassemble the Meter before performing adjustments or testing.
                                                                                               17
87V MAX
Calibration Information
                             Press and hold to test the present function. The measurement is not calibrated so it may be
                 (yellow)     inaccurate. This is normal.
                             Press to store the new calibration adjustment value and advance to the next step. This
                              pushbutton is also used to exit calibration adjustment mode after the calibration adjustment
                              sequence is complete.
       18
                                                                                 Digital Multimeter
                                                                               Calibration Adjustments
     Function
                       Adjustment Step                           Input Value
 (Switch Position)
                    C-09                      6.000 V
(DC Volts)           C-10                      60.00 V
                     C-11                      600.0 V
                    C-12                      600.0 mV
(DC Millivolts)      C-13                      60.00 mV
                    C-14                      600.0 Ω
(Ohms)               C-15                      6.000 kΩ
                     C-16                      60.00 kΩ
                     C-17                      600.0 kΩ
                     C-18                      6.000 MΩ
                     C-19                      0.000 Ω
                     C-20                      50.0 MΩ
                     C-21                      3.000 V
(Diode Test)
                    C-22                      6.000 A, 60 Hz
(Amps)               C-23                      6.000 A dc
                                                                                                   19
87V MAX
Calibration Information
       20
                                                                                 Digital Multimeter
                                                                              Reassembling the Meter
                                                                                                 21
87V MAX
Calibration Information
             Replacement Parts
                      Table 5 lists the Meter’s replaceable parts identified in Figure 5.
MP1
H1 (6)
MP2
MP3 (2)
BT1 (3)
                                       MP4                                                  H2 (6)
                                                                                             H3 (6)
                                   MP5
                                                                                               MP6
MP7
MP8 (2)
MP9
H4 (7) H4 (7)
H5 (4)
MP10
                                    F1
                                         F2
                                                                                          MP14
                                MP11
                               MP12                                                          DS1
                                       MP13                                                    MP15
MP16
MP17 MP18
MP19
MP20
                                                                                                          gaq106.emf
                                                 Figure 5. Exploded View of Meter
       22
                                                                                 Digital Multimeter
                                                                                  Replacement Parts
                                                                        Part
Item                            Description                                      Qty.
                                                                       Number
                                                                                                23
87V MAX
Calibration Information
                                                                                                             Part
                 Item                                      Description                                                          Qty.
                                                                                                            Number
[1] See www.fluke.com for more information about test leads and alligator clips available for your region.
       24
                                                                                                            Digital Multimeter
                                                                                                             Replacement Parts
5/07
                                                                                                                           25
87V MAX
Calibration Information
26