CREATININE
Kinetic-spectrophotometric
ALKALINE PICRATE
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument: BS-200 ( = Accent 200)
Reagent preparation
Working Reagent: Pour the contents of a Reagent B bottle into a Reagent A bottle. Mix gently.
Stable for 1 month at 2-8ºC.
Parameters
Test name Crea R1 200l
Test nº R2 0 l
Full name Creatinine Sample Volume 20 l
Standard nº R1 Blank ----
Analy.Type Fixed Time Mixed read Blank ----
Pri.Wave 510 Linearity range 0.03 - 20
Secon.Wave - Linearity limit
Trend Ascending Substrate Limit
Reac time 1 - 6 Factor
Incuba Time 3 □ prozone check q1 q2 q3 q4
Unit mg / dL
Precision 0.01 Pc Abs
Reference
N.o Gender Sample type Age Low.Lim Up.Lim
Men Serum 0.9 1.3
Women Serum 0.6 1.1
(*) values entered by user
Calibration
Rule
Rule One Point Linear Calibrators
Sensivity 1 Name Concentration
Replicates 3
Interval 99 days Std 1 (*)
Difference
Limit ----
SD 0
Blank response ----
Erros limit ----
(*)enter the
Coefficient 0 Standard value
Vers 0901