Direct Bilirubin: Instrument

This document provides instrument settings and parameters for measuring direct bilirubin levels spectrophotometrically using a Prestige 24i analyzer. It specifies reagent preparation involving combining one vial of Reagent B-D into a Reagent A-D bottle. The calibration involves a linear 6-point calibration. The normal ranges listed are 0-0.30 mg/dL for males and females for both serum and urine samples.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
98 views1 page

Direct Bilirubin: Instrument

This document provides instrument settings and parameters for measuring direct bilirubin levels spectrophotometrically using a Prestige 24i analyzer. It specifies reagent preparation involving combining one vial of Reagent B-D into a Reagent A-D bottle. The calibration involves a linear 6-point calibration. The normal ranges listed are 0-0.30 mg/dL for males and females for both serum and urine samples.

Uploaded by

Enrique Duarte
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 1

DIRECT BILIRUBIN

Spectrophotometric
DIAZOTIZED SULFANILIC

The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.

Instrument: PRESTIGE 24i(=LABMAX 240, ZAPHYR 400)

Reagent preparation
Working Reagent: Transfer the contents of one Reagent B-D vial into a Reagent A-D bottle. Mix thoroughly.
Stable for 20 days at 2-8 ºC.

Instrument settings
Item Name BIL-D
CALIBRATION
DATA INFORMATION Type LINEAR
Units mg/dL STANDARD
Decimal 2 #1 * #4
#2 #5
ANALYSIS #3 #6
Type END
Main W. Length 1 546 NORMAL RANGE
Main W. Length 2 660 MALE FEMALE
Method LOW HIGH LOW HIGH
SERUM 0.00 0.30 0.00 0.30
CORR. URINE
Slope Inter
Y= 1 X + 0

ASPIRATION DATA PROCESS ABSORB LIMIT


KIND Double READ START END
VOLUME MAIN 41 43 Low -0.10
SAMPLE 30 µL SUB 28 30 High 0.35
REAGENT 1 VOL 200 µL
REAGENT 2 VOL 50 µL FACTOR ENDPOINT LIMIT 2.5
Third Mix. OFF Blank correction ---
R1 Blank Water Blank LINEAR CHECK (%) 0

Auto Rerun SW On Auto Rerun Condition (Absorbance)


Absorbance Range Lower On
Auto Rerun Range(Result) On Higher On
Lower Higher Prozone Range Off
Serum 0 4.00
Urine --- --- * assigned standard value
Version 0410