C-REACTIVE PROTEIN (CRP)
Turbidimetry
                                                                                                                                                      LATEX
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by
the instrument. It is advisable to validate the use when there is any change in software or reagent versions.
                                                      Instruments:           AU 400 / 640
Reagent preparation
Working Reagent: Pour the contents of a Latex vial into a Diluent bottle.
Shake the Latex before pipetting. Mix thoroughly. Stable for 20 days at 2-8 ºC.
                                                                   Instrument settings
              Specific Test Parameters                                            Min OD:                              Max OD:
              Sample:         Volume 2 µL             Dilution 0 µL                   L: -0.1                            H: 2.5
              Reagents: R1 Volume 300 µL              Dilution 0 µL               Reagent OD limit:
                         R2 Volume µL                                                                   First L -0.1 First H 1.5
              Wavelength:       Pri. 540              Sec.                                              Last L -0.1 Last H 1.5
              Method:                END
              Reaction slope:         +                                           Dynamic Range:
              Measuring Point 1: First 1              Last 7                                                   L     0 H 150
              Measuring Point 2: First 0              Last 0                      Correlation Factor:
              Linearity:                 %                                                                     A     1 B     0
              No Lag Time:                                                        On-board stability period:
                                                                                  L                            H
              7. None selected
              8. Out of range                                                     0                            5.0
                                                      L                     H
              Panic Value                                                         Unit: mg/L                           Decimal places: 1
              Calibration Specific
              Calibration Type:       AB              Formula: Y=AX+B             Counts:      2                       Process: CONC.
                                            Cal. Nº              OD               CONC                Factor/OD-L          Factor/OD-H
              Point 1:                         #                                    *                    -0.1                      2.5
              * Enter the assigned standard value
              # User defined
                                                                                                                                  Version 0409