CHOLESTEROL
Enzymatic-spectrophotometric
CHOLESTEROL OXIDASE/PEROXIDASE
The application parameters comprised here constitute a guide to facilitate the validation of our reagents by the
instrument. It is advisable to validate the use when there is any change in software or reagent versions.
Instrument: BS-200 ( = Accent 200)
Reagent preparation
Reagent is ready to be used.
Parameters
Test name Chol R1 300µl
Test nº R2
Full name Cholesterol Sample Volume 3µl
Standard nº R1 Blank ----
Analy.Type End-point Mixed read Blank ----
Pri.Wave 510 Linearity range 0.3 - 1000
Secon.Wave - Linearity limit
Trend Ascending Substrate Limit
Reac time 0 - 18 Factor
Incuba Time □ prozone check q1 q2 q3 q4
Unit mg / dL
Precision 1 Pc Abs
Reference
N.o Gender Sample type Age Low.Lim Up.Lim
Serum 0 200
(*) values entered by user
Calibration
Rule
Rule One point Linear Calibrators
Sensivity Name Concentration
Replicates 3
Interval 99 days Std 1 (*)
Difference
Limit ----
SD 0
Blank response ----
Erros limit ----
(*)enter the
Coefficient 0 Standard value
Vers 0902