58XC Operation Manual
58XC Operation Manual
Pulsed/CW Microwave
Frequency Counters
Operation Manual
Certification
Phase Matrix, Inc. certifies this instrument to be in conformance with the specifications noted
herein at time of shipment from the factory. Phase Matrix, Inc. further certifies that its calibration
measurements are traceable to the United States National Institute of Standards and Technology.
To maintain the technical accuracy of the manual, it may be necessary to provide new or additional
information with the manual. In these cases, the manual is shipped with a Manual Update. Please
be sure to incorporate the information as instructed in the Manual Update.
iii
TM
SAFETY
The Phase Matrix 585C & 588C are designed and tested according to international safety
requirements, but as with all electronic equipment, certain precautions must be observed. This
manual contains information, cautions, and warnings that must be followed to prevent the
possibility of personal injury and/or damage to the instrument.
WARNING
A WARNING denotes a hazard to personnel. It calls attention to a
procedure or practice, which, if not correctly performed or adhered
to, could result in personal injury.
CAUTION
A CAUTION denotes a hazard to the equipment. It calls attention to
an operating procedure or practice, which, if not correctly performed
or adhered to, could result in damage to or destruction of part or all
of the product.
Dangerous Voltage
Toxic Substance
Static-Sensitive Component
Fire Hazard
iv
OVERALL SAFETY CONSIDERATIONS
WARNING
Before this instrument is switched on, its protective earth terminals
must be connected to the AC power cord’s protective conductor. The
main plug must only be inserted in a socket/outlet that has a
protective earth contact. The protective action must not be negated
by using an extension cord (power cable) or adapter that does not
have a protective earth (grounding) conductor.
WARNING
Use only fuses of the type specified with the required current and
voltage ratings. Never use repaired fuses or short-circuited
fuseholders, as doing so causes a shock and/or fire hazard.
WARNING
Whenever it is likely that electrical protection is impaired, the
instrument must be made inoperative and be secured against any
unintended operation.
WARNING
All protective earth terminals, extension cords, autotransformers,
and other devices connected to this instrument must be connected
to a socket/outlet that has a protective earth contact. Any
interruption of the protection causes a potential shock hazard that
can result in personal injury.
WARNING
The power supply is energized whenever AC power is connected to
this instrument. Disconnect the AC power cord before removing the
covers to prevent electrical shock. Internal adjustments or servicing
that must be done with the AC power cord connected must be
performed only by qualified personnel.
v
TM
WARNING
Since the power supply filter capacitors may remain charged after
the AC power cord is disconnected from the equipment,
disconnecting the power cord does not ensure that there is no
electrical shock hazard.
WARNING
Some of the components used in this instrument contain resins and
other chemicals that give off toxic fumes if burned. Be sure to
dispose of these items properly.
WARNING
Beryllia (beryllium oxide) is used in the construction of the YTF
assembly. This material, if handled incorrectly, can pose a health
hazard. NEVER disassemble the microwave converter assembly.
CAUTION
Static sensitive components are used in the YTF assembly. These
components can be damaged if handled incorrectly.
CAUTION
Before connecting power to the instrument, ensure that the correct
fuse is installed and the voltage-selection switch on the instrument’s
rear panel is set properly. Refer to Section 2, Installation.
CAUTION
Excessive signal levels can damage this instrument. To prevent
damage, do not exceed the specified damage level. Refer to the
instrument specifications in Section 1 of this manual.
vi
TABLE OF CONTENTS
WARRANTY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
CERTIFICATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
MANUAL CHANGE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iii
SAFETY. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
SAFETY AND HAZARD SYMBOLS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . iv
OVERALL SAFETY CONSIDERATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . .v
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-1
OPERATING CONDITIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
VENTILATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
STORAGE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1-2
SECTION 2: INSTALLATION
UNPACKING . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
INSTALLATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-1
FUSE REPLACEMENT . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-2
VOLTAGE SELECTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
INCOMING OPERATIONAL CHECKOUT . . . . . . . . . . . . . . . . . . . . . . . . . . 2-3
SERVICE INFORMATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
PERIODIC MAINTENANCE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
COUNTER IDENTIFICATION. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-4
FACTORY SERVICE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
SHIPPING INSTRUCTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 2-5
SECTION 3: OPERATION
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-1
FRONT PANEL CONTROLS, INDICATORS, AND CONNECTORS . . . . . . . . . . . 3-1
STATUS DISPLAY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-2
GPIB STATUS INDICATORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-3
SIGNAL INPUT CONNECTORS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-4
REAR PANEL CONTROLS AND CONNECTORS . . . . . . . . . . . . . . . . . . . . . . 3-4
INSTRUMENT DEFAULT SETTINGS. . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-6
KEYBOARD OPERATION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
NUMERIC ENTRY KEYS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
TERMINATOR KEYS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3-7
1
TM
TM
SECTION 4: ACCESSORIES
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-1
CARRYING CASE
M585C-ACC010 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-2
RACK-MOUNT KIT WITH HANDLES
M585C-ACC021 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-3
RACK-MOUNT KIT WITHOUT HANDLES
M585C-ACC022 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-4
SERVICE KIT (ADJUSTMENT AND TROUBLESHOOTING TOOL KIT)
M585C-ACC043 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-5
CHASSIS-SLIDE KIT (INCLUDES RACK-MOUNT KIT WITH HANDLES)
M585C-ACC101 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-7
CHASSIS-SLIDE KIT
(INCLUDES RACK-MOUNT KIT WITHOUT HANDLES)
M585C-ACC102 . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4-9
SECTION 5: PROGRAMMING
2
SECTION 6: OPERATIONAL VERIFICATION TESTS
INTRODUCTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
EQUIPMENT REQUIREMENTS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-1
OPERATIONAL VERIFICATION TEST PROCEDURES . . . . . . . . . . . . . . . . . . 6-2
BAND 0 RANGE AND SENSITIVITY TEST (CW ONLY) . . . . . . . . . . . . . . . . 6-2
BAND 1 RANGE AND SENSITIVITY TEST . . . . . . . . . . . . . . . . . . . . . . . 6-4
BAND 2 RANGE AND SENSITIVITY TEST . . . . . . . . . . . . . . . . . . . . . . . 6-5
BAND 2 AMPLITUDE DISCRIMINATION TEST . . . . . . . . . . . . . . . . . . . . 6-8
BAND 3 SUBBAND 1 RANGE AND SENSITIVITY TEST
(588C OPTION 5804 ONLY) . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-9
OPERATIONAL TEST RECORD . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6-11
3
TM
TM
4
1
GENERAL INFORMATION
INTRODUCTION
The Phase Matrix 585C and 588C Pulsed Microwave Frequency Counters are microprocessor-
based, multifunction instruments used for both CW and pulsed microwave measurements. They
can automatically measure the frequency of repetitive pulse signals as narrow as 50 ns. Both models
can also automatically measure pulse widths from 50 ns to 1 second and pulse periods from 250 ns
to 1 second, to a 10 ns resolution. Additionally, with an optional built-in delaying pulse generator,
the 585C and 588C can profile pulsed or chirped signals with measurement windows as narrow as
15 ns. No manual switching is required to measure CW or pulsed signals.
The frequency range of the 585C is 100 Hz to 20 GHz. The frequency range of the 588C is 100 Hz
to 26.5 GHz, and is extendible, by option, up to 170 GHz. Band 0, 100 Hz to 250 MHz, is for CW
measurements only.
The unique YIG preselector, available only on Phase Matrix counters, enables frequency selective
measurements and provides superior burnout protection.
All major functions are controlled through the 25-button, functionally grouped keyboard.
Information is presented for viewing on a twelve-digit sectionalized display and a 20-message
annunciator bank.
Microprocessor control and the unique architecture employed offer all the major counter functions,
such as frequency offsets, frequency range limits, and averaging capabilities, plus a variety of special
functions including internal diagnostics, calibration and test aids, and sophisticated operational
enhancements.
All front panel controls (except the POWER switch) and all background functions are externally
programmable via the IEEE 488 - 1978 standard GPIB (General Purpose Interface Bus) port. The
instrument output status and all displayed information are accessible via the GPIB.
1-1
TM
OPERATING CONDITIONS
This instrument is designed to be operated at temperatures not exceeding 0 to 50° C at relative
humidity not to exceed 95% (75% for temperatures greater than 25° C and 45% for temperatures
greater than 40° C) non-condensing. This instrument performs to specifications at altitudes not
exceeding 10,000 ft (3050 m) and tolerates vibration not exceeding 2 g. It is fungus resistant. The
chassis is not designed to provide protection from mechanical shock or falling water particles and
is intended for normal bench use in an environmentally uncontaminated area.
VENTILATION
Air circulates through the vents in the rear panel of the counter. These vents must not be
obstructed or the temperature inside the counter may increase enough to reduce counter stability
and shorten component life.
STORAGE
Store the instrument in an environment that is protected from moisture, dust, and other
contaminants. Do not expose the instrument to temperatures below –55° C or above 75° C, nor to
altitudes above 40,000 ft (12,000 m).
SPECIFICATIONS
GENERAL
1-2
1/GENERAL INFORMATION
SPECIFICATIONS (Continued)
BAND 1
( 4 ) ( PP )
Measurement Time (Pulse) ① MT = --------------------------------- + 0.05
( GW ) ( RES )
4
Measurement Time (CW) ① MT = ------------- + 0.05
RES
AQ = ⎛⎝ ------------------------⎞⎠ + 0.05
1
Acquisition Time ①
MINPRF
0.07
Gate Error in Hz ① GE = ± -----------
GW
0.03
Distortion Error in Hz ① DE = ± ------------------------------------------
–8
PW – ( 3 x 10 )
1-3
TM
SPECIFICATIONS (Continued)
BAND 2
( PP )
Measurement Time (Pulse) ① MT = --------------------------------- + 0.05
( GW ) ( RES )
1
Measurement Time (CW) ① MT = ------------- + 0.05
RES
0.01
Gate Error in Hz ① GE = ± -----------
GW
0.03
Distortion Error in Hz ① DE = ± ------------------------------------------
–8
PW – ( 3 x 10 )
1-4
1/GENERAL INFORMATION
SPECIFICATIONS (Continued)
0.03
Gate Error in Hz ① GE = ± -----------
GW
0.02
Distortion Error in Hz ① DE = ± ------------------------------------------
–8
PW – ( 3 x 10 )
70
Acquisition Time (CW) AQ = ------------------------ + 0.2
MINPRF
( 4 ) ( PP )
Measurement Time (Pulse) ① MT = --------------------------------- + 0.05
( GW ) ( RES )
4
Measurement Time (CW) MT = ------------- + 0.05
RES
Center Frequency Signal must be within ±2 GHz of specified center frequency.
PULSE PERIOD
1-5
TM
SPECIFICATIONS (Continued)
PULSE WIDTH
Frequency 10 MHz
Aging Rate < 1 x 10–7/month
Short Term Stability < 1 x 10–9 RMS for one second averaging time
Temperature Stability < 1 x 10–6 over the range 0° to 50° C
Line Variation < 1 x 10–7 (±10% line voltage change)
Warm-up Time < 5 minutes
Output Signal 10 MHz, 1 V p-p minimum into 50 Ω
External Time Base Input 10 MHz, 1 V p-p minimum into 1 kΩ
Frequency 10 MHz
Aging Rate < 5 x 10-10/day (after 72-hour warm-up), 1 x 10–7/year
Short Term Stability < 1 x 10-10 RMS for one second averaging time
Temperature Stability < 3 x 10-8 over the range 0° to 50° C
Line Variation < 2 x 10-10 (±10% line voltage change)
Warm-up Time Within 5 x 10–9 10 min after turn-on at 25° C
Within 1 x 10–9 30 min after turn-on at 25° C
1-6
1/GENERAL INFORMATION
SPECIFICATIONS (Continued)
Pulse Output
Level TTL ➂ into 50 Ω
Connector BNC
Trigger Input
Level TTL ➂
Polarity Positive or Negative (see special functions 12 and 13)
Impedance 1 kΩ
Connector BNC
Trigger Output
Level TTL ➂ into 50 Ω
Pulse Width 50 to 100 ns
Connector BNC
Pulse Width
Minimum 24 ns
Maximum 800 ms
Resolution 2 ns
Pulse Delay ➃
Minimum 74 ns
Maximum 800 ms
Resolution 2 ns
Pulse Period
Minimum 100 ns
Maximum 800 ms
Resolution 50 ns
1-7
TM
SPECIFICATIONS (Continued)
REMOTE SENSORS
OPTION DESCRIPTION
5803 Rear Panel Input Connectors
5804 Band 3 Frequency Extension Module. Available on Model 588C only.
Required for frequencies between 26.5 GHz and 170 GHz. Frequency
Extension Cable Kit 890 and appropriate remote sensors are also required.
5809 Ovenized High Stability Time Base (Aging Rate: < 5 x 10–10/day)
5810 Delaying Pulse Generator
ACCESSORY DESCRIPTION
010 Carrying Case
101 Chassis-Slide kit(includes Rack Mount kit with handles)
102 Chassis-Slide kit(includes Rack Mount kit without handles)
890 Frequency Extension Cable kit
031 Operation Manual (one supplied with each instrument)
032 Operation and Service Manual
022 Rack-Mount kit without handles
091 Remote Sensor 26.5 GHz - 40 GHz (WR-28)
092 Remote Sensor 40 - 60 GHz (WR-19)
093 Remote Sensor 60 GHz - 90 GHz (WR-12)
094 Remote Sensor 90 GHz - 110 GHz (WR-10)
1-8
1/GENERAL INFORMATION
ACCESSORY DESCRIPTION
095 Remote Sensor 50 GHz - 75 GHz (WR-15)
096 Remote Sensor 33 GHz - 50 GHz (WR-22)
097 Remote Sensor 26.5 GHz - 50 GHz (2.92 mm coax, SMA compatible)
098 Remote Sensor 110 GHz - 170 GHz (WR-6)
021 Rack-Mount kit with handles
043 Service Kit
NOTES
① All formulas are written in fundamental units (Hz, sec, etc.).
MV is the maximum video amplitude in dBm.
SL is the input signal level in dBm.
FV is the frequency component of the video in Hz.
GW is the logical “AND” of pulse width and inhibit signal minus 30 ns.
PW is pulse width of the incoming signal in seconds.
PP is the period of the input signal in seconds.
RES is the resolution in Hz up to 1 MHz. Above 1 MHz only the displayed resolution changes; the
internal resolution remains at 1 MHz.
AVG is the number of measurements averaged.
FH is the difference between Frequency Limit High and Frequency Limit Low in Hz.
MINPRF is the specified instrument MINPRF in Hz up to 1 kHz. Above 1 kHz, MINPRF is 1 kHz.
➁ In Band 3, if FM/Chirp is > 150 MHz and nonsymmetrical, the measured frequency is a function of
average frequency and geometric center frequency.
➂ TTL levels have a maximum low of 0.4 and a minimum high of 2.4.
➃ In Internal Trigger Mode, pulse delay is measured from the rising edge of the Internal Trigger Output
to the Pulse Output. In External Trigger Mode, pulse delay is measured from either the rising or the
falling edge (software selectable) of the External Trigger Input to the Pulse Output.
1-9
TM
1-10
2
INSTALLATION
UNPACKING
The Phase Matrix 585C and 588C series Pulsed Microwave Frequency Counters arrive ready for
operation. Carefully inspect the shipping carton for any sign of damage. If the carton is damaged,
immediately notify shipper’s agent.
Remove the packing carton and supports, being careful not to mar or damage the instrument. Make
a complete visual inspection of the counter, checking for any damage or missing components.
Check that all switches and controls operate mechanically. Report any damage to Phase Matrix
immediately.
INSTALLATION
There are no special installation instructions for these counters. They are self-contained bench or
rack-mount instruments, which only require connection to a standard, single-phase power line for
operation.
CAUTION
To prevent damage to the counter, verify that the voltage value
visible through the window in the fuse drawer (see Figure 2-1) is
correct for the available AC power-input voltage, and inspect the fuse
to ensure that it is the correct type and that it matches the rating
specified in Table 2-1.
WARNING
Disconnect the AC power cord before removing the fuse or changing
the voltage selector setting.
2-1
TM
FUSE REPLACEMENT
The fuse is located inside the rear-panel power-input-module assembly (see Figure 2-1). Use only
the fuse types listed in Table 2-1.
FUSE
DRAWER VOLTAGE VOLTAGE
SLOT SELECTOR SELECTOR
WINDOW INSERT
SLOT
100
VA MAX
.
100
VA M
A X.
120
120
LINE
SELE
CT 1
FUS 00 V
E: or 11
5/12
FRE 1.5 A 0V
Q: 220
SB V o r2
50-6 30/2
LINE 0.75 40 V
0 Hz
SEL A SB
ECT
FUS 100 50-6
E: V or1 0 Hz
15/1
FRE 20 V
Q: 1.5 A 22 0
SB V or
230/2
50-6 40 V
0 Hz 0.75
A SB
50-6
0 Hz
FUSE VOLTAGE
SELECTOR
INSERT
FUSE
DRAWER
To access the fuse or the voltage selector, insert a screwdriver into the fuse-drawer slot, gently push
to the right, and pull the fuse drawer out (see Figure 2-2). To reinsert the fuse drawer, push it slowly
back into the power-input module until it snaps into place.
2-2
2/INSTALLATION
100 100
VA M VA M
A X. A X.
GENTLY PUSH
TO THE RIGHT
120
120
LINE
SELE LINE
CT 1 SELE
FUS 00 V CT 1
E: or 11 FUS 00 V
5/22 E: or 11
FRE 1.5 A 0V 5/12
Q: SB 220 FRE 1.5 A 0V
V or Q: 220
230/2 SB V or
50-6 40 V 230/2
0 Hz 0.75 50-6 40 V
A SB 0 Hz 0.75
A SB
50-6
0 Hz
GENTLY PULL 50-6
0Hz
FORWARD
VOLTAGE SELECTION
The voltage-selector insert is located inside the rear-panel power-input module (see Figure 2-1).
The insert must be installed correctly for the available AC line voltage. To change the voltage,
remove the fuse drawer as described on page 2-2, insert a screwdriver in the slot on the side of the
voltage-selector insert, and gently pry the voltage-selector insert out while pulling the tab on the
voltage-selector insert towards you. Reinstall the voltage-selector insert so the correct voltage is
visible on the tab on the insert. Slowly push the fuse drawer back into the power-input module until
it snaps into place.
1. Before connecting power to the instrument, ensure the correct fuse is installed and that the
voltage-selector insert is set properly.
2. Connect the power cord to an appropriate single-phase power source. The ground terminal on
the power cord plug must be properly grounded.
3. Turn the POWER switch on. All LEDs and annunciators should light for about two seconds.
The counter should then display all zeros indicating that the automatic self-check has been
successfully completed.
SPECIAL
SPECIAL
5. Press: 9 0 2 The display should read all 8s. All annunciators and decimal points
FUNC should be lit..
2-3
TM
SPECIAL
6. Press: 9 0 3 Each display segment should light in turn (adjustable by the front-
FUNC panel SAMPLE RATE control).
SPECIAL
7. Press: 9 0 4 Each digit should light in turn (adjustable by the front-panel
FUNC SAMPLE RATE control).
SERVICE INFORMATION
PERIODIC MAINTENANCE
No periodic maintenance is required. However, to maintain accuracy, it is recommended that the
counter be recalibrated every 12 months. The specific calibration interval may be redefined
depending upon the measurement accuracy required. For sample measurement error calculations
for both 6- and 12-month calibration intervals, see “Timebase Error” in Section 3.
CAUTION
Do not attempt to repair or disassemble the microwave converter,
millimeter wave converter, or time-base-oscillator assemblies. Such
actions void the counter’s warranty. Contact Phase Matrix or your
sales representative if you think these assemblies require service.
WARNING
Do not attempt to repair or dissemble the microwave converter
assembly. This assembly contains beryllia (beryllium oxide), which,
if handled improperly, could pose a health hazard.
COUNTER IDENTIFICATION
This counter is identified by three sets of numbers: a model number (585C or 588C), serial number,
and a configuration control number (CCN). These numbers are located on a label affixed to the
frame at the rear of the counter (some units may have separate CCN and serial-number labels). The
model number, CCN, serial number, and the numbers of any options must be included in any
correspondence regarding your counter. A typical serial-number label is shown in Figure 2-3.
2-4
2/INSTALLATION
FACTORY SERVICE
If the counter is being returned to Phase Matrix for service or repair, be sure to include the following
information with the shipment:
SHIPPING INSTRUCTIONS
Wrap the counter in heavy plastic or kraft paper, and repack in original container if available. If the
original container cannot be used, use a heavy (275 pound test) double-walled carton with
approximately four inches of packing material between the counter and the inner carton. Seal
carton with strong filament tape or strapping. Mark the carton to indicate that it contains a fragile
electronic instrument. Clearly mark the carton to indicate that the contents were manufactured in
the U.S.A. Ship the carton to the Phase Matrix address on the cover of this manual.
2-5
TM
2-6
3
OPERATION
INTRODUCTION
This section lists the counter controls, connectors, and indicators, explains how each counter
function operates, and provides some general measurement considerations.
NOTE
A carrying case, a tool kit, and rack-mounting kits that are available from
Phase Matrix for the 585C and 588C counters are described in Section 4.
B2
POWER
• POWER switch — selects ON or STBY. In the standby position power is applied to the oven
oscillator, if installed, and to the primary side of the power transformer.
3-1
TM
WARNING
Whenever AC power is connected to this instrument, the power
supply is energized. Therefore, exercise caution whenever the
covers are removed.
• SAMPLE RATE/HOLD control — varies time between measurements from 0.1 to 10 seconds
(nominal). The last reading is retained indefinitely in HOLD.
• GATE indicator — lights when the signal gate is open and a measurement is being made.
• SEARCH indicator — lights when the counter is not locked to an input signal.
• Data display — 12-digit numeric display provides direct read-out of measurement data.
Frequency data is displayed in a fixed position format that is sectionalized in GHz, MHz, kHz
and Hz. When active, pulse parameter data (pulse width and pulse period) is displayed in the
three least significant digits of the frequency display in a floating point format.
• Status display — A series of annunciators provided to indicate current operating status of the
counter (see Figure 3-2).
• Keyboard — Both data entry and function selection are controlled through the keyboard (see
Keyboard Operation on page 3-7).
STATUS DISPLAY
ms B1 PP HI HLD TLK
3-2
3/OPERATION
• PP (pulse period) — lights when the pulse period display function is selected.
• DLY (delay) — lights when the pulse generator (optional) display delay function is selected.
• LO (frequency limit low) — lights when a low limit other than the factory default is being used.
• HI (frequency limit high) — lights when a high limit other than the factory default is being
used.
• CF (center frequency) — lights when the center frequency mode of operation is enabled.
• GEN (generator) — lights when optional delaying pulse generator is active.
• PRF (minimum pulse repetition frequency) — lights when a MINPRF other than the factory
default is being used.
• HLD (hold) — lights when measurement updating is disabled.
• SPC (special function) — lights when a special function is enabled.
• EXT (external reference) — lights when the counter is set to an external timebase reference.
NOTE
For proper counter operation: when the EXT function is selected, a 10 MHz
external reference MUST be applied to the rear panel input connector.
3-3
TM
BAND 0
100 Hz - 250 MHz INPUT POWER
BAND0,1 +7 dBm Maximum +27 dBm DAMAGE
! BAND2
BAND3
+7dBm Maximum +45 dBm DAMAGE
+5 dBm Maximum +10 dBm DAMAGE
BAND 2
50OHM 950 MHz - 26.5 GHz BAND 3
26.5 GHz - 170 GHz
BAND 1
250 MHz - 1 GHz
REMOTE SENSOR
50 OHM
50 OHM
• BAND 0 (BNC female) — has a nominal input impedance of 50 Ω and is used for CW
measurements in the range of 100 Hz to 250 MHz.
• BAND 1 (BNC female) — has a nominal input impedance of 50 Ω and is used for
measurements in the range of 250 MHz to 1 GHz.
• BAND 2 (precision N female for Model 585C, GPC 3.5 female for Model 588C) — has a
nominal input impedance of 50 Ω and is used for measurements in the range of 1 GHz to 20
GHz (26.5 GHz for 588C).
• BAND 3 (Optional – for Model 588C only, SMA female) — Used for measurements in the
range of 26.5 to 170 GHz. This input is used in conjunction with the Model 890 Frequency
Extension Cable Kit and a remote sensor.
100 VA MAX.
BAND 3 BAND 2
BAND 0
10 MHz
IN/OUT
THRESH
OUT
GATE
OUT GPIB IEEE488/1978
(CW ONLY) 120
WARNING:
TO AVOID ELECTRIC SHOCK:
DO NOT REMOVE COVERS.
NO USER SERVICABLE PARTS INSIDE.
REFER ALL SERVICING TO QUALIFIED PERSONNEL.
THIS UNIT MUST BE GROUNDED.
BAND 1
LINE SELECT 100 V or 115/120 V 200 V or 220/240 V
FREQ: 50 - 60 Hz 50 - 60 Hz
• BAND 0, BAND 1, BAND 2, and BAND 3 — are provided on instruments with optional rear
panel inputs (Option 5803).
3-4
3/OPERATION
• THRESH OUT — is the digitized pulse envelope. When the counter has a converter lock and
a signal is present, the output is a TTL high into 50 Ω. Without a converter lock, the output is
a TTL low. The Thresh Out and the Gate Out, together, show exactly where in the pulse the
measurement is being taken.
• GATE OUT — represents the gate to the Count Chain board. The gate output follows the
actual gate, not the gate enable. When the gate is active, the output is a TTL high into 50 Ω;
otherwise, the output is a TTL low. The Threshold Out and the Gate Out, together, show
exactly where in the pulse the measurement is being taken.
• 10 MHz IN/OUT — provides a 10 MHz square wave output at 1 V p-p, AC coupled into 50 Ω,
when the counter’s internal timebase is enabled. Accepts a 10 MHz 1 V p-p signal into 1000 Ω
for external timebase operation. Special Functions 08 and 09 are used to select either the
external or internal timebase.
• INHIBIT IN — used to delay frequency measurements. TTL level input is used to control
when the counter can perform a measurement. A TTL high on this input allows counter to
search, center, and perform measurements on incoming signals. A TTL low on this input
prevents the counter from detecting an incoming signal, inhibiting signal measurements.
• GPIB — connects the instrument to the IEEE 488 - 1978 bus.
• AC Power Connector — accepts the power cord supplied with the counter.
• DELAYING PULSE GENERATOR (OPTIONAL)
• TRIGGER IN — TTL input signal for triggering the internal delaying pulse generator. Input
impedance is approximately 1 kΩ. Trigger polarity is selectable using Special Functions 12 and
13.
• TRIGGER OUT — TTL output signal, 50 to 100 ns in duration, with a low-to-high transition
occurring at delay time equal to zero. This output is capable of driving 50 Ω.
• PULSE OUT — TTL output pulse from the internal delaying pulse generator. This output is
capable of driving 50 Ω.
3-5
TM
Models 585C and 588C both offer a feature that enables the user to customize the state of the
instrument at turn-on. For more information on this feature, see Special Functions 72 and 73.
3-6
3/OPERATION
KEYBOARD OPERATION
The keyboard consists of 25 push-button keys that control the major functions of the counter (see
Figure 3-5).
7 8 9
FUNC
PULSE CENTER DF MHz/ms
Hz 4 5 6
WIDTH FREQ
PULSE GENERATOR PULSE FREQLIMIT kHz/us
1 2 3
WIDTH DELAY PERIOD MODE PERIOD LOW HIGH
INIT DISPLAY CLEAR RESET TRIG MIN Hz/ns
± 0 "
TERMINATOR KEYS
Four keys (GHz/s, MHz/ms, kHz/us, and Hz/ns dB) act as terminators for the input of frequency
and time parameters. The CLEAR DISPLAY and CLEAR DATA keys are also considered
terminator keys.
3-7
TM
• Press CLEAR DISPLAY to abort the sequence and return to normal operation without
changing the value of the called parameter.
• Press CLEAR DATA to abort the sequence and assign the default value to the called
parameter.
BAND
This key controls the frequency measurement range. Select the appropriate band according to
Table 3-2.
KEYBOARD EXAMPLES:
B AND C LE AR
B AND
3-8
3/OPERATION
GPIB EXAMPLE:
This parameter controls the frequency measurement range of Band 3. It is set according to the
remote sensor being used. Select the appropriate subband per Table 3-3:
KEYBOARD EXAMPLES:
BAND CLEAR
BAND
BAND CLEAR
GPIB EXAMPLES:
Enter: Send 0, 18, “BAND 3, SUBBAND 4", NLend to select Band 3, subband 4.
Enter: Send 0, 18, “SUBBAND 4", NLend to select subband 4 (if counter is already in Band 3).
This command does not automatically set counter to Band 3.
3-9
TM
This key is used to turn on the optional internal delaying pulse generator and to select the trigger
source.
KEYBOARD EXAMPLES:
CLEAR
GPIB EXAMPLES:
Enter: Send 0, 18, “GMODE 1", NLend to trigger from external source.
Enter: Send 0, 18, “GMODE 2", NLend for continuous internal trigger.
Enter: Send 0, 18, “GMODE 3", NLend to trigger from input RF signal threshold.
3-10
3/OPERATION
RES
This key controls the frequency measurement resolution. Select the desired resolution according
to Table 3-4.
KEYBOARD EXAMPLES:
RES CLEAR
RES
RES
Press: 9 to select resolution 9 (1 GHz).
GPIB EXAMPLE:
Enter: Send 0, 18, “RESOLUTION 2", NLend to select resolution 2 (100 Hz).
TEST
This key calls the 200 MHz self-test. It is functionally the same as activating Special Function 01.
For a full description of Special Function 01, see “ACTIVATION OF SPECIAL FUNCTIONS,”
on page 3-43.
3-11
TM
SPECIAL FUNC
This key is used to call any of the various special functions listed in the “SPECIAL FUNCTION
DIRECTORY” on page 3-41.
KEYBOARD EXAMPLES:
SPECIAL CLEAR
SPECIAL
SPECIAL
Press: 0 4 to activate Special Function 04, scan digits test.
FUNC
GPIB EXAMPLE:
Enter: Send 0, 18, “SPECIAL 01", NLend to activate Special Function 01, 200 MHz self-test.
CENTER FREQ
This key controls the center of the frequency range in which the counter searches for a signal. The
ΔF key controls the search range. This function is used to reduce the acquisition time or to search
for a particular signal in a multiple signal environment. This function is only available in bands 2
and 3.
Select Band 2 CENTER FREQ in the range of 1 GHz to 20 GHz for Model 585C, and in the range
of 1 GHz to 26.5 GHz for Model 588C. The counter searches and locks onto signals within ±ΔF
from the entered value. Depending on input-signal power and frequency, the counter can lock on
signals outside the ΔF frequency range. The actual locking range is determined by the entered ΔF
frequency, input signal level, the maximum frequency range of Band 2, and the bandpass of the YIG
filter located at the input to Band 2.
Select Band 3 CENTER FREQ in the range of the subband currently selected. The counter locks
on signals ±2 GHz from the entered center frequency. The ΔF key is not active in Band 3. The
counter does not reject signals outside this range. If a signal more than ±2 GHz from the entered
center frequency is applied, an erroneous reading may result.
The entered center frequency value is truncated to 10 MHz resolution. The number can be entered
in any fixed-point format; the units terminator determines the scale of the input number.
3-12
3/OPERATION
KEYBOARD EXAMPLES:
CENTER CLEAR
Press: to disable center-frequency operation.
FREQ DATA
CENTER GHz/s
CENTER MHz/ms
Enter: Send 0, 18, "CENTERFREQ 14.8 GHZ", NLend to select a center frequency of 14.8 GHz.
This key controls the low end of the frequency window that is searched for a signal in Band 2. Select
frequency limit low in the range of 900 MHz to 20.5 GHz for Model 585C, and in the range of 900
MHz to 26.7 GHz for Model 588C. The value entered is truncated to 10 MHz resolution. This
function is only available in Band 2. The frequency limit low must always be less than the frequency
limit high. The number can be entered in any fixed-point format; the units terminator determines
the scale of the input number.
KEYBOARD EXAMPLES:
FREQ LIMIT CLEAR
GPIB EXAMPLE:
Enter: Send 0, 18, “LOWLIMIT 2.35 GHZ”, NLend to select a 2.35 GHz value.
This key controls the high end of the frequency window that is searched for a signal in Band 2.
Select the high frequency limit in the range of 900 MHz to 20.5 GHz for Model 585C, and in the
range of 900 MHz to 26.5 GHz for Model 588C.
The value entered is truncated to 10 MHz resolution. This function is only available in Band 2.
Frequency limit high must always be greater than frequency limit low.
3-13
TM
The number can be entered in any fixed-point format; the units terminator determines the scale of
the input number.
KEYBOARD EXAMPLES:
Enter: Send 0, 18, “HIGHLIMIT 3.2 GHZ”, NLend to select a 3.2 GHz value.
MIN PRF
This key controls the minimum pulse repetition frequency of the pulsed signals that can be acquired
and measured by the counter. For example, if a MIN PRF of 500 Hz is selected, the counter only
measures signals with a minimum pulse repetition frequency of 500 Hz or greater. This parameter
affects the acquisition speed indirectly by affecting two internal processes: the time of waiting for a
pulse at each frequency step in the frequency range search, and the time of waiting for a pulse when
taking measurements before declaring a “signal lost” condition. CW signals can always be acquired
and counted regardless of the specified MIN PRF.
Select MIN PRF in the range of 1 Hz to 100 kHz, depending on the minimum pulse repetition
frequency of the signal being measured. The number can be entered in any fixed-point format; the
units terminator determines the scale of the input number.
NOTE
If the PRF of the input signal is less than the MIN PRF chosen, erroneous
measurements may be displayed.
KEYBOARD EXAMPLES:
MIN CLEAR
MIN Hz/ns
3-14
3/OPERATION
GPIB EXAMPLE:
Enter: Send 0, 18, “MINPRF 500 HZ”, NLend to select a minimum pulse repetition frequency
value of 500 Hz.
This key is used to control the delay from trigger of the output pulse from the optional internal
delaying pulse generator. The delay can be set in increments of 2 ns, from a minimum of 74 ns
(values smaller than 74 ns may be entered, but the minimum guaranteed is 74 ns) to a maximum
of 800 ms using either the keyboard or GPIB. Entries made to a resolution finer than 2 ns are
rounded down to the closest 2 ns value.
KEYBOARD EXAMPLES:
PULSE GEN ns
GPIB EXAMPLES:
Enter: Send 0, 18, "GDELAY 100 NS", NLend to set delay to 100 ns
This key is used to control the period of the output signal from the internal delaying pulse
generator. In MODES 1 and 3, the period of the signal is determined by the trigger source, so this
parameter only applies when the pulse generator is in the continuous trigger mode (MODE 2). The
pulse period can be set in increments of 50 ns, from a minimum of 100 ns to a maximum of 800 ms
using either the keyboard or GPIB. Entries made to a resolution finer than 50 ns are rounded down
to the closest 50 ns value.
KEYBOARD EXAMPLES:
PULSE G EN CLEAR
PULSE GEN ns
PULSE GEN ms
3-15
TM
GPIB EXAMPLES:
Enter: Send 0, 18, “GPERIOD 100 MS”, NLend to set period to 100 ms.
Enter: Send 0, 18, “GPERIOD 100 US”, NLend to set period to 100 μs.
This key is used to control the output pulse width for the optional internal delaying pulse generator.
The pulse width can be set in 2 ns increments, from a minimum of 24 ns to a maximum of 800 ms
using either the keyboard or GPIB. Entries made to a resolution finer than 2 ns are rounded down
to the closest 2 ns value.
KEYBOARD EXAMPLES:
PULSE GEN ns
Press: 5 0 to set width to 50 ns.
WIDTH
GPIB EXAMPLES:
Δ FREQ
This key controls the search range for the center frequency function. It defaults to a range of ±50
MHz, but can be set from 0 to the full range of the selected band in increments of 10 MHz. ΔF is
only available in Band 2.
The counter searches ±ΔF around the center frequency. Setting ΔF to zero is a special case. If ΔF
is zero the counter does not search for the signal. The YIG filter is moved to the entered center
frequency, and if a signal is present, it is measured. Since the counter does not search for the signal,
when ΔF equals zero, the signal acquisition time is effectively eliminated.
The remaining keys are called one-shot action keys. These are DISPLAY DELAY, INIT/LOCAL,
PULSE PERIOD, PULSE WIDTH, RESET, and TRIG.
• DISPLAY DELAY — causes the counter to display the current delay time of the internal
delaying pulse generator in the parameter display.
• INIT/LOCAL — when the counter is in local mode, this key causes the counter to be
initialized to the power-on state. When the counter is in remote mode, the INIT/LOCAL key
causes a return to local mode (unless a GPIB local lockout is active).
3-16
3/OPERATION
• PULSE WIDTH — turns the pulse width measurement on or off. The result is displayed in the
parameter display.
• PULSE PERIOD — turns the pulse period measurement on or off. The result is displayed in
the parameter display.
NOTE
When DISPLAY DELAY, PULSE WIDTH, or PULSE PERIOD are enabled,
the frequency display is limited to 100 kHz resolution and the parameter is
displayed in the three rightmost digits (the parameter display). If Special
Function 69 is active when the parameter is enabled, frequency is not
displayed and the parameter occupies the entire display.
• RESET — resets the converter and restarts the signal acquisition process. If a signal is found, a
measurement is taken, even if the counter is in HOLD.
• TRIG — begins a new measurement cycle. If a measurement cycle is in progress, it is aborted
and a new cycle is begun.
THEORY OF OPERATION
The Phase Matrix 585C/588C Pulsed Microwave Frequency Counters are able to automatically
measure the frequency of repetitive pulsed signals as well as CW signals. Pulse widths from 50
nanoseconds to 1 second and pulse periods from 250 nanoseconds to 1 second are measured to a
resolution of 10 ns. Using an inhibit input, the instruments can profile pulsed or chirped CW signals
using gates as narrow as 15 ns. No manual switching is required to select CW or pulse signal
measurement. The frequency range of the 585C is 100 Hz to 20 GHz. The frequency range of the
588C is 100 Hz to 26.5 GHz and can be extended by option up to 170 GHz.
Microprocessor control and the unique architecture permit not only the major counter functions
such as frequency offsets, frequency range limits, and averaging capabilities, but also a variety of
special functions such as internal diagnostics, calibration and test aids, and sophisticated
operational enhancements. All primary and background functions are fully programmable.
FUNCTIONAL DESCRIPTION
In the following description, the Phase Matrix 585C/588C is divided into two major portions: the
Basic Counter and the RF Converter. The RF Converter is further divided into the RF Band 0
Converter, the RF Band 1 Converter, and the RF Band 2 Microwave Converter. Theory of
operation for the Band 3 mm wave converter is provided in the OPTIONS section.
3-17
TM
BASIC COUNTER
The Basic Counter, shown in Figure 3-6, receives input signals from all four bands and performs
both frequency measurements and pulse parameter measurements (pulse width and pulse period).
The basic counter can directly measure the frequency of signals from 100 Hz to 250 MHz.
VCO
FROM
400-505 MHz DIVIDE BY 4
BAND 2
CONVERTER PRESCALER
WAVESHAPING COUNT
NETWORK CHAIN
FROM RF/IF
SIGNAL DELAY
CONDITIONER LINE
FROM IF
THRESHOLD GATE GATE
DETECTOR GENERATOR ACCUMULATOR
Overall operation of the counter is controlled by the CPU assembly. This assembly contains a
Motorola 68B09 microprocessor, its control logic, the system memory, and the circuitry for the
GPIB interface. It communicates with all other assemblies in the counter via a triple bus system:
the data bus, address bus, and control bus. Each assembly (except for the Signal Conditioner and
Gate Control) contains a peripheral interface adapter (PIA) that provides the interface between
the bus system and the counter hardware. The basic counter performs frequency measurements by
comparing the unknown signal to a reference frequency, namely the timebase. The standard
timebase is a 10 MHz temperature compensated crystal oscillator (TCXO). An optional high-
stability ovenized oscillator is also available for improved frequency accuracy. For coherence with
system clocks, the counters have the capability of accepting an external 10 MHz reference via the
rear panel 10 MHz IN/OUT connector.
A frequency measurement is made by generating a time interval (gate time) consisting of a number
of cycles of the reference. This gate time is used as an interval during which the input signal is
counted by the Count Chain assembly. This process is considerably more difficult for pulsed signals
than it is for CW signals and must be accomplished as a two-step operation. The first step is to
supply a gate to the Count Chain that is present only when an input signal is also present. The
second step is to accumulate the total time during which the gate is applied, until the desired gate
time is reached.
The first step requires that the gate begin after the signal is present at the Count Chain and end
prior to the end of the signal. This is accomplished by generating a gate approximately 30 ns shorter
than the RF signal. The arrival time at the Count Chain of the IF from the converter is then
controlled by a delay line so the gate signal falls entirely within the IF pulse width.
3-18
3/OPERATION
The second step is accomplished by counting reference clock pulses whenever the gate is open until
1
a total time equal to ---------------------------
Resolution is obtained. This requires that each gate opening be an exact integral
-
number of clock pulses. Since an 80 MHz clock is being used, the gate is always an exact multiple
of 12.5 ns.
Pulse widths are measured by detecting the signal and counting the number of zero crossings of the
phase locked VCO signal (prescaled by 4) that occur while the signal is present. The
microprocessor then calculates the pulse width by multiplying the number of zero crossings by the
period of the VCO signal. Pulse period measurements are made using a similar technique, except
that the counter counts zero crossings during the time from the rising edge of one pulse to the rising
edge of the next pulse.
RF CONVERTER
BAND 0 INPUT
AMPLIFIER/ BAND BASIC
(CW ONLY) AMPLIFIER
LIMITER SELECT COUNTER
100 Hz to 250 MHz
Signals between 100 Hz and 250 MHz are counted directly. The gate width is set according to a
specified resolution (from resolution 0 to resolution 6).
Band 1 RF Converter
ATTEN
MICROPROCESSOR
Signals between 250 MHz and 1 GHz are prescaled by four before reaching the basic counter. The
gate is therefore made four times longer to properly count the prescaled signal. During signal
acquisition, the counter monitors the RF detector, and when a signal is detected, the RF gain is
3-19
TM
adjusted to set the level of the IF 6 dB above threshold. The signal is then counted. After every
measurement, the RF gain is readjusted to ensure rapid tracking of a moving signal (see Figure 3-9).
NO NO
YES
IF OUTPUT BASIC
ATTEN SELECT COUNTER
YIG
FILTER
DETECTOR
BAND 2
INPUT REF DET IF
1 GHz to 20 HARMONIC RF AMPLITUDE THRESHOLD
(26.5) GHz GENERATOR DISCRIMINATION
CIRCUIT
POWER VCO
AMP (LO)
YTF
DRIVER AFC
CPU
3-20
3/OPERATION
The actual process of signal measurement in the microwave band is accomplished using a series of
steps:
START
ADJUST VCO
SEARCH FOR SO THAT
SIGNAL FIF = 120 MHz
DETERMINE COUNT IF
LARGEST AND OUTPUT
SIGNAL F = NFVCO + FIF
CENTER YIG
FILTER ON
SIGNAL SIGNAL
PRESENT? NO
CALCULATE YES
N AND VCO
FREQUENCY
During the search routine the counter selects the largest signal present within the selected, or
default, range of the band 2 input. During this routine, the electronically tunable microwave
bandpass YIG filter is being continuously stepped from its low to high limits. The output from the
YIG filter is applied to the mixer, which is used as an RF detector. The output from the mixer is
applied to the video amplifier which feeds a flash A/D converter, with a resolution of approximately
5 dB. When the counter detects a signal, the YIG DAC setting and relative amplitude of the signal
(output from the flash A/D) is stored in memory, and the search routine continues. If other signals
are detected, their relative amplitudes are compared with the stored information. If the new signal
3-21
TM
is higher in amplitude, the memory is updated with information on the new signal. After searching
the entire band, the YIG DAC setting and relative amplitude of the highest signal present are
stored in memory.
The next step is to precisely center the YIG on the selected signal. This process begins by moving
the YIG to the signal selected during the search routine. The YIG is then stepped in 2 MHz steps
around the signal until four points are found: the points on either side of the peak 1.25 dB down
from the peak and the points on either side of the peak 5 dB down from the peak. From these points
the approximate “center of mass frequency” of the signal is found, and the YIG filter is set to that
frequency. The “center of mass” algorithm compensates for pulsed signals that deviate from perfect
sinX/X shape and for nonsymmetries in the YIG filter.
After the YIG is centered on the signal, the harmonic number N is calculated based on the setting
of the YIG filter using the following formula:
The resulting N is rounded up to the next higher integer. At this point low side mixing
(NFVCO < FIN) is assumed, and the proper VCO frequency is calculated using the formula:
If the results yield a VCO frequency which is less than 400 MHz (the minimum VCO frequency),
high side mixing is assumed, and FVCO is recalculated using the formula:
At this point the IF gain is adjusted to set the signal level approximately 6 dB above signal
threshold.
Since FYIG is only approximately equal to FIN, the IF frequency (FIF) is not exactly 120 MHz. ,
therefore, the next step is to adjust the FVCO to shift FIN to the center of the IF passband at
120 MHz. This is done by counting FIF and recalculating FVCO based on the following formula:
FYIG ± FIF
FVCO =
N
Where +FIF is used if high side mixing and – FIF is used if low side mixing.
3-22
3/OPERATION
Once the VCO corrections have been made, the counter counts the IF and calculates the input
frequency using the following formula and then displays the results:
Where:
N = Harmonic number
FVCO = VCO frequency
± = + for low side mixing and – for high side mixing
After each measurement, new frequencies for the YIG and VCO are calculated to maintain the IF
at 120 MHz, and the IF gain is readjusted to keep the signal 6 dB above threshold. This method
provides rapid tracking of a signal being tuned.
SIGNAL MEASUREMENTS
AUTOMATIC FREQUENCY MEASUREMENTS
The Phase Matrix 585C and 588C Pulsed Microwave Frequency Counters can automatically
measure the frequency of CW and repetitive pulse signals having pulse widths as narrow as 50 ns.
To measure the frequency of a CW signal, apply the signal to the input connector that corresponds
to the frequency being measured and select the appropriate band. The counter then automatically
finds the signal, measures it, and displays the measured frequency.
The average frequency of repetitive pulse signals is measured in much the same way as CW signals.
The only difference is that for pulse signals with pulse repetition frequencies of less than 2 kHz, the
minimum pulse repetition frequency must be entered into the counter using the MINPRF key on
the front panel. If the MINPRF is not set at or below the minimum pulse repetition frequency of
the signal to be measured, the counter is unable to lock on the signal.
As an example, consider the signal shown in Figure 3-12 (page 3-24). The signal is a 2 GHz signal
with a pulse width of 1 ms and a pulse period of 2 ms. Since the pulse repetition frequency is the
reciprocal of pulse period, the minimum pulse repetition frequency of the signal shown is 500 Hz.
Since this is less than 2 kHz, it must be entered into the counter. To enter a minimum pulse
repetition frequency of 500 Hz into the counter, press the MIN PRF key followed by the 5 key, the
0 key, and the 0 key; then terminate the sequence with the Hz terminator key. If the signal at this
point is applied to the Band 2 input connector and Band 2 is selected, the counter would
automatically find the signal and display the frequency on the front panel.
3-23
TM
Pul se Period 2 ms
Pul s e 1 ms
W i dth
These counters can also automatically measure both the pulse width and the pulse period of the
incoming signal to a resolution of 10 ns. This is accomplished by pressing either the PULSE
WIDTH key to measure the pulse width or the PULSE PERIOD key to measure the pulse period.
In actual microwave environments, there are often multiple signals present. In a multi-signal
environment, the counter automatically finds and measures the largest signal (as specified by
amplitude discrimination).
In Band 2, the counter can also measure signals other than the largest signal present. This is
accomplished by setting frequency limits around the desired signal. Figure 3-13 (page 3-25) shows
an example of the frequency limits feature.
3-24
3/OPERATION
INPUT POWER
(dBm)
+20 FREQUENCY FREQUENCY
LIMIT LOW LIMIT HIGH
+10
+10
0
-5
- 10
- 20
- 20
- 30 FREQUENCY
(GHz)
5 6 6.3 6.6 7
DISPLAYED
FREQUENCY
If the signals shown in Figure 3-13 are applied to Band 2, the counter automatically finds the signal
at 6 GHz since it is the largest signal. If it is desired to measure the signal at 6.3 GHz set the
frequency limit low to 6.2 GHz and the frequency limit high to 6.4 GHz. This prevents the counter
from seeing either the signal at 6 GHz or the signal at 6.6 GHz.
The counter also provides a center frequency mode which is used in combination with ΔF. In this
mode, the counter searches around the entered center frequency ± the entered ΔF. For example,
if the center frequency in Figure 3-14 (page 3-26) is set to 8 GHz, and ΔF is set to 1 GHz, the
counter searches ±1 GHz around the center frequency and measures the signal at 8 GHz.
3-25
TM
INPUT POWER
(dBm)
+20 SEARCH RANGE
+10
-ΔF +ΔF
- 10
- 20
- 30 FREQUENCY
(GHz)
7 8 8.4 9
CENTER
FREQUENCY
PULSE PROFILING
Automatic pulsed frequency measurements determine the average frequency across a pulse. In
some cases, however, additional information may be required. For example, a pulsed magnetron
may exhibit substantial frequency shift near the leading and trailing edges of the pulse. Other
specialized pulsed signals are intentionally ramped in frequency. Repetitive frequency variations
across a pulsed signal can easily be measured using the counter and either an external delaying pulse
generator or the optional internal delaying pulse generator. Figure 3-15 (page 3-27) shows a typical
setup for profiling frequency changes across a pulsed signal.
3-26
3/OPERATION
DELAYING
PULSE GENERATOR*
TRIGGER TRIGGER
FROM SOURCE INPUT
OUTPUT
PULSE
INHIBIT
INPUT
INPUT
SIGNAL
* External pulse generator is not required if Option 5810, Internal Delaying Pulse
Generator, is installed.
The output pulse of the signal generator is used as an enable input to the counter. As the pulse
delay is varied, the measurement window can be “walked” through the pulse. A plot of frequency-
versus-delay gives the frequency-versus-time profile of the pulse directly, as shown in Figure 3-16.
The width of the measurement window is determined by the width of the pulse generator output.
Measurement windows as narrow as 15 ns can be used, although wider windows yield higher
accuracy.
INPUT SIGNAL
. . .
PROFILE
MEASUREMENT
WINDOW
FREQUENCY PROFILE
FREQUENCY f AVERAGE
TIME
3-27
TM
With the optional internal delaying pulse generator, frequency profiling is further simplified. For
example, consider the repetitive pulsed signal shown below, which is ramping from 5 GHz to
5.2 GHz.
Using the delaying pulse generator in the 585C/588C counter, this signal can be profiled either
manually or automatically. When profiling a pulsed signal, the change in frequency across the
measurement window must be less than 20 MHz to ensure that the signal passes through the input
YIG filter. The signal shown in Figure 3-17 is changing frequency at a rate of 2 MHz per μs, so the
maximum width of the measurement window should be less than 10 μs.
Pulse 100 μ s
Width
To profile the signal shown in Figure 3-17 manually, perform the following steps:
The counter automatically finds the signal and displays the average frequency over the 5 μs
measurement window beginning 75 ns after the rising edge of the pulse. The frequency can be
measured anywhere along the pulse by changing the delay time of the pulse generator.
To profile the signal shown in Figure 3-17 automatically, perform the following steps:
3-28
3/OPERATION
5. Set up the pulse generator to trigger on the external signal by pressing the Pulse Generator
Mode key and entering 1.
6. To minimize signal acquisition time, set frequency limit low at 4.5 GHz and frequency limit
high at 5.5 GHz.
7. Put the counter in hold by turning the sample rate control fully clockwise.
8. Enable the auto profile function by pressing Special Function 77. The counter displays “Inc”
(increment) in the three least significant digits of the display.
9. Enter a measurement window increment of 5 μs.
10. The counter displays “End” in the three least significant digits of the display. Enter an end
delay of 99 μs.
The counter automatically finds the signal and displays the average frequency over the 5 μs
measurement window beginning 75 ns after the rising edge of the pulse. Press the “Trig” key to
increment the measurement window by 5 μs. Each time the “Trig” key is pressed, the measurement
window is incremented by 5 μs until the end delay is reached. Press the “Display Delay” key to
display the delay.
Frequency profiling can be easily automated using Special Function 77 and a computer.
Many complex signals are not pulsed at all, but are continuous signals with frequencies that vary
repetitively over time. One example is a settling time measurement of a voltage controlled oscillator
(VCO). When a voltage step is applied to the tuning voltage input on a VCO its output frequency
changes to reflect the voltage change on the tuning input. However, as shown in Figure 3-18, it
takes the VCO a finite amount of time to settle in at the new frequency. The amount of time it
takes for the VCO to settle in at the new frequency within some predetermined limits is specified
as its settling time. A typical VCO settling time specification would require that the frequency
output be within ±10 MHz of the settled frequency within 1 ms after the voltage step is applied to
the tuning input on the VCO.
INPUT SIGNAL
PROFILE
MEASUREMENT • • •
WINDOW
FREQUENCY PROFILE
AVERAGE
3-29
TM
VCO settling time measurements can easily be made using the counter and a delaying pulse
generator(either external or the optional internal delaying pulse generator), as shown in
Figure 3-19.
VCO
TO BE 585C/588C COUNTER INHIBIT
TESTED INPUT
INPUT
FREQUENCY MODULATED SIGNAL
MODULATION MICROWAVE
INPUT OUTPUT
SIGNAL GENERATOR
DELAYING
PULSE GENERATOR*
MODULATION
TRIGGER OUTPUT
OUTPUT
TRIGGER TTL OUTPUT
INPUT PULSE
* External pulse generator is not required if Option 5810, Internal Delaying Pulse
Generator, is installed.
With the equipment set up as shown in Figure 3-19, the signal generator provides a repetitive
frequency step, square wave modulation, to the VCO under test. The trigger output from the signal
generator, triggers the delaying pulse generator which in turn provides a measurement window to
the counter. The counter measures the VCO during this measurement window. Changing the
delay time of the delaying pulse generator, moves the measurement window and allows the
frequency of the VCO to be measured at any point in time relative to the trigger from the signal
generator.
Using the optional internal delaying pulse generator further simplifies the setup. For this setup, the
trigger from the signal generator is connected directly to the trigger input on the counter and the
internal delaying pulse generator provides the measurement window.
Another type of measurement is that of a repetitive sequence of pulses that differ in frequency. In
this case, it is desirable to measure the frequency of each pulse in the sequence separately. The same
test setup as shown in Figure 3-19 is required, with the trigger pulse synchronous with the sequence.
In this measurement, the input inhibit is used to discriminate between pulses. The enabling pulse
can be slightly wider than the pulse to be measured. By shifting the delay time of the enabling pulse,
the user can measure each input pulse of the sequence separately.
3-30
3/OPERATION
TIMING CONSIDERATIONS
The internal timing usually should be of no concern to the user. However, in applications where a
few nanoseconds are significant, two factors of internal operation must be considered. These
involve two areas. One factor is the measurement window width, and the other is synchronization
with external signals.
The measurement window width is the period during which the gate is actually open to enable the
counting of a signal. This gate width is typically 30 ns narrower than the pulse applied to the
INHIBIT IN connector. The width of the gate is always an integral number of clock periods (12.5
ns). For applications where the measurement window must be known to an accuracy better than
20 ns, it is recommended that the gate output on the rear panel be observed on a high speed
oscilloscope. The desired gate width may be set by varying the input inhibit pulse width. For
accurate pulse representation, the oscilloscope input should be terminated in a 50 Ω load.
Synchronization
When it is necessary to measure the signal frequency at a precise point in time, the internal delays
of the measuring instrument can be significant. In the Phase Matrix 585C and 588C counters, the
total delay between the time a signal is applied to an input connector and the time it is available to
be counted is nominally 60 ns. The signal threshold output on the rear panel typically occurs 20 ns
after the signal is applied. The gate signal at the rear panel occurs at the measurement time with
virtually no delay. In other words, when absolute time positioning of a signal is required, it is
necessary to consider that the gate signal (representing the measurement period) is actually making
a measurement of the signal which appeared at the input connector 60 ns earlier. If the signal
threshold output is used as an indication of input signal, then it occurs 40 ns prior to measurement.
Figure 3-20 (page 3-32) shows the relative timing of these signals for a pulsed input signal. Timing,
however, is not a function of input signal characteristics.
3-31
TM
INPUT
MICROWAVE
PULSE
t pw
SIGNAL
THRESHOLD
t pw
20 ns
60 ns t pw
DELAYED
IF PULSE
5 to 10 ns
GATE
t pw -30 ns
ACCURACY
When making any type of measurement, some degree of measurement error exists. In Phase
Matrix’s CW type frequency counters, as with most other CW counters, these errors are limited to
a combination of timebase error, gate phasing error (±1 count), and gate width error. In making
frequency measurements on pulsed RF signals, the preceding errors, along with one additional error
due to distortion of the pulsed RF signal, affect measurement accuracy. To minimize these errors
and to properly interpret the results of the measurements, the magnitude of these errors must be
known.
NOTE
A Microsoft® Excel® spreadsheet that automatically calculates pulse
accuracy and acquisition/measurement times is available for download
from the Phase Matrix Web site. To download the spreadsheet, point your
Web browser to < http://www.phasematrix.com/prodpages/58XC.html >,
and click the link for the Pulse Calculator (note where you select to have
the file saved on your hard disk). When the download is complete, run
eippulse.exe. Select the location on your hard drive where you want to
save the Pulse Calculator spreadsheet, open the spreadsheet in Excel,
and follow the directions.
3-32
3/OPERATION
CW MEASUREMENT ACCURACY
Timebase error causes an error in the measured frequency proportional to the error in the timebase
oscillator. For example, if the 10 MHz oscillator is off frequency by 3 Hz, the corresponding
measurement error on a 1 GHz signal would be 300 Hz. For an 18 GHz signal, the same 3 Hz error
in the timebase would cause a measurement error of 5.4 kHz. The maximum error in the timebase
is the sum of the various possible errors, such as aging rate and temperature stability.
The second type of error, ±1 count, is due to the lack of phase coherence between the gate and
the signal. Simply stated, if an event occurs every 400 ms (F = 2.5 Hz), a counter could measure
either 2 or 3 events in a one second interval.
The above note “based on measurement averaging” is included due to a random instrumentation
error in the counter. This error can be virtually eliminated by averaging measurements.
Each of the sources of CW measurement error contributes to the overall error in pulsed frequency
measurements, along with gate error and distortion error. For narrow pulses, the gate phasing error,
referred to as averaging error, and gate error can become the dominant sources of error for pulse
measurements. The following list describes the source of potential measurement errors when using
the Phase Matrix 585C and 588C counters.
TIMEBASE ERROR
Given:
3-33
TM
Calculation:
= ± 12 kHz
Counter measurement, after a six-month calibration interval, could have an error of ±12 kHz in
measuring a 20 GHz signal.
Given:
Calculation:
= ± 24 kHz
Counter measurement after the recommended 12-month calibration interval could have an error
of±24 kHz in measuring a 20 GHz signal due to timebase aging.
These examples are to illustrate error due to the timebase aging rate only. Actual calculations of
measurement error must include the other sources of error discussed in the following text.
AVERAGING ERROR
This error is caused by the relative timing between the gate and the incoming signal and results in
an uncertainty of ±1 count in the least significant digit of each measurement. If the counter
resolution is set to 10 kHz, then the potential error is ±10 kHz. On signals having pulse widths less
than the required gate time (determined by the resolution), the counter generates more than one
gate per measurement cycle. If the counter generates N number of gates, then an uncertainty of
±N counts is possible, though very unlikely. The resultant averaged measurement follows the rules
of statistics in that, on successive gates, the ±1 count error varies randomly. In fact, most of the
3-34
3/OPERATION
readings (63%) fall between ± the square root of N, where N is the number of gates required to
accumulate the required gate time. This is called the RMS averaging error. In the following
RES-
formulas, N = ------------
GW
. It should be noted that the total gate time is typically 30 ns narrower than the
input pulse. The RMS averaging error, in Hz, can be calculated using the following formulas:
RES
Bands 1 and 3: averaging error (RMS) = ± 2 -----------------
GW
RES
Band 2 : averaging error (RMS) = ± -----------------
GW
Where RES is the specified instrument resolution in Hz, up to 1 MHz. Above 1 MHz, RES is always
1 MHz. GW is the logical AND of the pulse width and the inhibit signal minus 30 ns. See
Figure 3-21 for a graphic description of the logical AND function.
A RF PULSE
B INHIBIT SIGNAL
(ENABLE ON TTL HIGH)
GATE ERROR
When narrow pulses are counted, the gate is opened and closed many times in order to accumulate
enough gate time to provide the required resolution. Each time the gate opens and closes, there is
a small but finite error. The total error is proportional to the number of times the gate is cycled
during a measurement, and is inversely proportional to the gate width. This error is also related to
both temperature and input frequency.
In the 585C and 588C counters, the worst-case gate error, including all variables, is specified as:
0.07
Band 1: gate error = ± ---------------
GW
0.01
Band 2: gate error = ± ---------------
GW
0.03
Band 3: gate error = ± ---------------
GW
3-35
TM
Where GW, in seconds, is the logical AND of the pulse width and the inhibit signal minus 30 ns
(see Figure 3-21). Unlike averaging error, which is random, gate error is systematic, and is not
reduced by averaging.
DISTORTION ERROR
During the first and last few nanoseconds of a pulse, phase distortion caused by impedance
mismatches or video effects can occur, resulting in shifts in time of the zero crossing. On wide
pulses, distortion error is insignificant; however, on narrow pulses it may become a dominant source
of error. To reduce the effect of distortion error on count accuracy, the 585C and 588C counters
automatically adjust the gate to start 15 ns after the pulse begins, and to end 15 ns before the end
of the pulse. The specified maximum distortion error for all three bands can be calculated from the
following formula:
0.03
Maximum Distortion Error = ± --------------------------------
(PW–30 ns)
Where PW = pulse width (minimum pulse width is 50 ns).
TIMEBASE CALIBRATION
A frequency error in the internal timebase oscillator results in a proportional error in the frequency
reading for either CW or pulsed signals. The aging rate of the standard internal timebase is specified
to be less than 1 x 10–7 parts per month. This means that if the oscillator were set precisely on
frequency at the beginning of the month, it could be 1 Hz off frequency at the end of the month.
On a frequency measurement of 18 GHz, a 1 Hz error in the 10 MHz timebase would cause a
measurement error of 1.8 kHz. Other errors can result from changes in ambient temperature.
Measurement errors caused by the timebase can be reduced by adjusting the timebase (at the
temperature it is used) using a standard of known accuracy. Another error reduction method is to
use an external 10 MHz timebase with a known degree of accuracy, such as a 10 MHz frequency
standard.
Averaging error is reduced to ±1 count whenever the gate width (GW) is greater than RES, where
RES is the counter resolution in Hz. Since the averaging error is random in nature, it can also be
reduced by increasing the number of individual gates. This can be accomplished by increasing the
resolution of the counter and/or averaging a number of individual measurements. The counter
allows 1 Hz maximum resolution and can automatically average up to 99 individual measurements
internally. With the GPIB and a controller, the user can average a larger number of individual
measurements, which virtually eliminate averaging error. The following formulas can be used to
3-36
3/OPERATION
determine the averaging error (RMS) when averaging a number of individual measurements.
RES
Bands 1 and 3: averaging error (RMS) = ± 2 ----------------------------------
(GW) (AVG)
RES -
Band 2: averaging error (RMS) = ± ---------------------------------
(GW) (AVG)
Where RES is the specified instrument resolution in Hz, up to 1 MHz. Above 1 MHz, RES is always
1 MHz. GW is the logical AND of the pulse width and the inhibit signal minus 30 ns. AVG is the
number of individual measurements to be averaged.
Gate error at any given frequency and pulse width can also be virtually eliminated by comparing a
CW frequency measurement to a simulated pulsed frequency measurement and computing a
correction factor due to gate error. This correction factor can then be added to, or subtracted from,
the indicated pulsed measurement to obtain the corrected frequency. The CW signal should be the
same frequency (within 25 MHz) as that of the actual pulsed signal to be measured. To simulate a
pulsed signal, apply an enable signal (of the same width as the pulse to be measured) to the
INHIBIT IN connector on the rear panel. A single measurement contains both averaging error and
gate error. Averaging measurements reduces averaging error by the square root of the number of
measurements averaged. If 100 measurements are averaged, the averaging error is reduced by a
factor of 10. Gate error, and any residual averaging error, is the difference in reading between the
pulsed and non-pulsed measurement of the same CW signal.
EXAMPLE:
Pulse width = 2 ms
1. Apply a CW signal to the counter at 2 GHz ±25 MHz and record the displayed frequency.
This frequency is called F1.
2. Apply a TTL signal with a pulse width of 2 ms at the INHIBIT IN connector on the rear panel.
Set the 585C/588C counter to average 99 readings. The frequency displayed on the counter is
called F2.
3. Gate Error = F2 – F1
NOTE
This procedure avoids errors associated with pulsed signal distortion and
any possible pulling of the signal source. It should be noted that by using
Special Function 92, gate error can also be automatically calibrated out of
the system for a given pulse width and frequency. However, the calibration
procedure may result in additional errors for other pulse widths or
frequencies. For additional information on Special Function 92, see the
Service Manual.
3-37
TM
Since distortion error is most significant on the edges of the pulse, it may be reduced by using the
counter inhibit feature to measure only in the middle of the pulse; however, measuring only the
middle of the pulse narrows the gate, and gate error increases. For pulses less than 70 to 80 ns, this
may add more error than it removes. The performance test section of the service manual describes
a method of determining the magnitude of this error that can be used to determine the
improvement in accuracy achieved by using the inhibit function.
The following is a sample calculation for determining the maximum specified measurement error
for a typical pulse frequency measurement.
Given:
Frequency: 18 GHz
Pulse width: 530 ns
Resolution: 100 kHz
• TIMEBASE ERROR (TBE) (Based on 6 Hz error from 10 MHz timebase) - - - - - - - ±10.8 kHz
NOTE
The direction of the timebase error is not specified, so it is not known
whether the timebase error caused the indicated reading to be higher or
lower. If the actual frequency of the timebase was 6 Hz high, then its period
would be reduced and the counter would indicate a lower frequency.
100 E3
AE (RMS) = ± ----------------------------------- = ±45 kHz
(500E-9) (99)
NOTE
To reduce the averaging error for this example, the measurement
averaging feature of the counter was used. If it had not been used, the
averaging error would have been ±450 kHz.
3-38
3/OPERATION
NOTE
The total measurement error, as calculated above, is the worst-case error.
The errors that make up the total error would not under normal
circumstances be additive. Errors in opposite directions would offset one
another, with the effect of reducing the total error.
The following worksheet can be used to determine the maximum specified measurement error for
a particular application. To determine the specified maximum error, select the desired operating
parameters and use the formulas given to determine the magnitude of each type of error.
Variables:
Frequency (F):
Pulse width (PW):
Counter resolution (RES):
This error can be determined by accurately measuring the frequency at the rear panel 10 MHz
IN/OUT connector. The frequency measured (F mea) is then used in the following formula to
determine measurement error.
10 MHz – F mea
TBE = -------------------------------------------- X F
10 MHz
where F mea = measured timebase frequency and F = input frequency
3-39
TM
RES -
Band 2 : AE (RMS) = ± ---------------------------------
(GW) (AVG)
Where RES = specified counter resolution in Hz up to 1 MHz. Above 1 MHz resolution, the
counter’s internal resolution remains at 1 MHz. GW = (pulse width AND inhibit signal) – 30 ns
and AVG = number of measurements averaged.
NOTE
1
If GW is > ------------ then AE = ±1 count.
RES
0.01
Band 2: GE = ± ---------------
GW
0.03
Band 3: GE = ± ---------------
GW
Where GW = (pulse width AND inhibit signal) –30 ns
3-40
3/OPERATION
The following special functions are used to verify proper operation. Special Function 01 is the most
comprehensive because it tests most of the major subassemblies contained in the counter.
There are a wide range of special functions for counter setup and performance enhancement. The
most commonly used setup special functions include 08, 72, 74, 81, and 90. The most commonly
used performance enhancement special functions include 61, 63 and 77.
3-41
TM
The following special functions set up the counter in special modes to aid in troubleshooting or to
initiate calibration routines. These functions should only be used by qualified technicians for
calibration and repair.
3-42
3/OPERATION
CAUTION
Executing Special Function 46, 91, 92 or 93 can cause a loss of
calibration data. To prevent this from occurring, access to these
functions is blocked by an internal memory protect feature.
Attempting to access these functions with the memory protected
causes the counter to display “ERROR 53”.
Special functions can be activated through both the front panel keyboard and the GPIB interface.
To activate a special function through the keyboard, press the SPECIAL key followed by two digit
keys. To activate a special function through the GPIB interface, enter the word SPECIAL followed
by a two-digit number. Activating special functions does not alter any previously entered
parameters unless specifically stated. To terminate all previously activated special functions: press
the SPECIAL key and then the 0 key followed by the 0 key again; or, press the SPECIAL key
followed by the CLEAR DATA key. To terminate all special functions using GPIB, issue the
command SPECIAL 00.
This function verifies that the count chain, gate generator, and VCO are operational.
NOTE
The front-panel TEST button provides a convenient shortcut for accessing
Special Function 01.
The display shows the frequency measurement results. These results are output to the GPIB
interface when frequency readings are requested. The measurement result should be 200 MHz ±1
count.
3-43
TM
This function verifies that all the digit segments and annunciator LEDs are operational. When this
function is activated, all digit segments and all annunciators are turned on. The GATE and the
SEARCH annunciators are both on for the duration of the special function.
Each segment in all the digits and banks of annunciators is turned on sequentially by this function
to test the display segment drivers. The scan rate is determined by the setting of the SAMPLE
RATE control.
Each digit and each bank of annunciators is turned on sequentially by this function to check the
display digit driver. The scan rate is determined by the setting of the SAMPLE RATE control.
After this function is activated the counter stops normal operations and the display shows the key
code of the last key pressed. When a new key is pressed, the display is updated to show the code of
the new key. When the GPIB controller requests a key code, the code of the last key pressed is
output. (If the controller requests a key code, the counter outputs to the GPIB interface the code
of the last key pressed even if Special Function 05 is not activated). If the counter is in LOCAL,
this function must be terminated by the CLEAR DISPLAY key. If it is in remote, this function can
be terminated by any device-dependent command.
3-44
3/OPERATION
This function generates the checksum for the PROM in the counter and compares it with the
checksum table stored in the firmware. If the checksum generated is correct, the counter displays
the word “PASSEd” on the front panel. If the checksum is incorrect, an error message is output to
the display. At the same time, the error-condition-status bit in the GPIB serial poll status byte is
set. During checksum generation, “SPECIAL 06” is displayed.
This function enables the user to find out whether the counter is configured as a 585C or 588C
counter. It displays the appropriate model number on the front panel.
Selecting this function configures the counter to external timebase input mode. The EXT
annunciator is lit when the counter is in external timebase mode.
3-45
TM
Selecting this function configures the counter to internal timebase mode. The EXT annunciator is
turned off when the counter is in internal timebase mode.
This function disconnects the pulse generator from the counters inhibit input, allowing the pulse
generator to be used as a completely separate instrument. When this function is activated, the
internal pulse generator has no affect on the operation of the counter, but the pulse generator’s rear
panel connections are still active.
This function differs from the Pulse Generator ON/OFF key in that the ON/OFF key turns off the
pulse generator and the pulse generator’s rear panel connections.
This function reverses the action of Special Function 10, internally reconnecting the output of the
pulse generator to the inhibit input of the counter. By default, the pulse generator output is
connected to the inhibit input.
This function causes the Internal Delaying Pulse Generator to trigger on the falling edge of the
pulse generator trigger input.
This function causes the Internal Delaying Pulse Generator to trigger on the rising edge of the pulse
generator trigger input.
3-46
3/OPERATION
This function verifies coarse calibration of the Band 2 YIG DAC offset and YIG DAC slope
adjustments.
When this function is activated, the counter waits for the user to enter the new YIG calibration
frequency. The previously entered frequency number and “Fr” are displayed in the frequency
section and the pulse parameter section of the display, respectively. The special function stops in
this state until the user enters a new frequency value, or, if the previously entered frequency value
is the required frequency, pushing the kHz key allows the special function to continue.
After the number has been entered, the YIG DAC is set to the entered frequency number. The
display shows “CAL DAC” plus a number from 0 to 29 that corresponds to the information on the
Band 2 power discrimination circuitry. The counter outputs the power discrimination circuitry
information when requested by the GPIB to OUtput LEvel.
When this function is activated, the counter waits for the user to enter the start frequency of the
YIG sweep. The previously entered start frequency and “F1” are displayed in the frequency section
and the pulse parameter section of the display, respectively. The special function stops in this state
until the user enters a new start frequency or, if the previously entered frequency is the required
start frequency, the user presses the kHz key to tell the special function to continue.
After the start frequency is entered, the counter waits for the user to enter the stop frequency of
the YIG sweep. The previously entered stop frequency and “F2” are displayed in the frequency
section and the pulse parameter section of the display, respectively. The special function stops in
this state until the user enters a new stop frequency or, if the previously entered frequency is the
required stop frequency, the user presses the kHz key to tell the special function to continue.
When both the start and stop frequencies have been entered, the display reverts to displaying
SPECIAL “40”. The YIG DAC sweeps continuously from F1 to F2 until the function is terminated.
If F1 and F2 are equal, the YIG DAC is set to the frequency corresponding to F1 and F2.
To activate this function in remote, the user programs the controller to output SPECIAL 40. The
start and stop frequencies used are the frequencies specified in the GPIB commands Y1FREQ and
Y2FREQ (where Y1FREQ and Y2FREQ correspond to F1 AND F2 respectively). If the start or stop
frequency required is different from that specified in Y1FREQ or Y2FREQ respectively, the number
in that frequency register must be updated before Special Function 40 is activated.
3-47
TM
After this function is activated, the counter waits for the user to enter the start frequency of the
VCO sweep. The previously entered start frequency and “F1” are displayed in the frequency section
and the pulse parameter section of the display, respectively. The special function stops in this state
until the user enters a new start frequency or, if the previously entered frequency is the required
start frequency, the user presses the kHz key to tell the special function to continue.
After the start frequency is entered, the counter waits for the user to enter the stop frequency of
the VCO sweep. The previously entered stop frequency and “F2” is displayed in the frequency
section and the pulse parameter section of the display, respectively. The special function stops in
this state until the user enters a new stop frequency or, if the previously entered frequency is the
required stop frequency, the user presses the kHz key to tell the special function to continue.
When both the start and stop frequencies have been entered, the display reverts to “SPECIAL 41”.
The VCO sweeps continuously from F1 to F2 in 100 kHz steps until the function is terminated.
The sweep rate is controlled by the sample rate. Maximum sweep rate may be obtained by disabling
the sample rate (Special Function 63) before calling this function. If F1 and F2 are equal, the VCO
is set to that particular frequency. The VCO power amplifier is turned on during this function.
To activate this function in remote, the user instructs the controller to output SPECIAL 41. The
start and stop frequencies used are the frequencies specified in the GPIB commands V1FREQ and
V2FREQ (where V1FREQ and V2FREQ correspond to F1 and F2 respectively). If the start or stop
frequency required is different from that specified in V1FREQ or V2FREQ, the number in that
frequency register must be updated before SPECIAL 41 is activated.
After this function is activated, the counter waits for the user to enter the start frequency of the
VCO sweep. The previously entered start frequency and “F1” is displayed in the frequency section
and the pulse parameter section of the display, respectively. The special function stops in this state
until the user enters a new start frequency or, if the previously entered frequency is the required
start frequency, the user presses the kHz key to tell the special function to continue.
After the start frequency is entered, the counter waits for the user to enter the stop frequency of
the VCO sweep. The previously entered stop frequency and “F2” is displayed in the frequency
section and the pulse parameter section of the display respectively. The special function stops in
this state until the user enters a new stop frequency or, if the previously entered frequency is the
required stop frequency, the user presses the kHz key to tell the special function to continue.
When both the start and stop frequencies have been entered, the display reverts to “SPECIAL 42”.
The VCO sweeps continuously from F1 to F2 in 100 kHz steps until the function is terminated.
The sweep rate is controlled by the sample rate. Maximum sweep rate may be obtained by disabling
3-48
3/OPERATION
the sample rate (Special Function 63) before calling this function. If F1 and F2 are equal, the VCO
is set to that particular frequency. The VCO power amplifier is turned off during this function.
To activate this function in remote, the user instructs the controller to output SPECIAL 42. The
start and stop frequencies used are the frequencies specified in the GPIB commands V1FREQ and
V2FREQ (where V1FREQ and V2FREQ correspond to F1 and F2, respectively). If the start or stop
frequency required is different from that specified in V1FREQ or V2FREQ, the number in that
frequency register must be updated before SPECIAL 42 is activated.
This function prevents the counter from performing the normal converter lock and measurement
cycles. It freezes the counter in the state it was in at the moment the function was activated. The
display shows “PAUSE” and the STOP ON/OFF status bit is set when this function is active.
Special Function 44 remains activated until terminated through Special Function 45 or by pressing
the SPECIAL key followed by the CLEAR DATA key, or by pressing the SPECIAL key followed
by the 0 key and then the 0 key again.
This function reverses the action taken when Special Function 44 is activated. The function
returns the counter to normal operation. A reset is generated and the STOP ON/OFF status bit is
cleared when this function is activated.
CAUTION
Care must be used when operating Special Function 46. Although the
counter cannot be damaged by this function, improper operation of
it can affect the counter calibration. For this reason, access to this
function is blocked by an internal memory protect feature.
Attempting to access this function with the memory protected
causes the counter to display “ERROR 53”.
This function allows the user to display and/or alter any memory location. The counter continues
its normal operations when performing this function unless Special Function 44 has previously been
activated.
3-49
TM
In the local mode, the keys on the keyboard take on different meanings after Special Function 46
is activated. The following are the definitions of the keys when this function is enabled.
After activating Special Function 46, the user can do one of the following:
• Exit the function by issuing a CLEAR DISPLAY command (pressing the CLEAR
DISPLAY key).
• Alter the content of the memory location by entering a two-digit hexadecimal number.
• Display the next memory location by issuing an INCREMENT command (pressing the
DISPLAY DELAY key).
• Display the previous memory location by issuing a DECREMENT command (pressing the
DISPLAY TEST key).
• Enter another memory address by first issuing an ADDRESS command (pressing the
CLEAR DATA key).
If the content of a memory location is altered, the new content of that memory location is displayed
in the pulse parameter section of the display. If the ADDRESS command is issued, the display
changes to show “Addr_ _ _ _”. While the address is being entered, the hexadecimal digits keyed
in replace each blank sequentially. After the memory address is entered, the content of that
memory location is displayed in the pulse parameter section of the display. This function must be
terminated by the CLEAR DISPLAY command.
In the remote mode, a memory content can be interrogated by using the OUTPUT MEMORY
command. When the counter is addressed to talk, the last memory address accessed is output. A
memory location can be accessed using the MEMORY 0HHHH command (where H is a
hexadecimal digit). The content of a memory location can be altered using the MEMORY 0HHHH
3-50
3/OPERATION
0HH command. In the remote mode, Special Function 46 need not be activated when accessing
and altering memory locations. Those operations can be done by the controller in the background.
This function provides the user with the means to measure the frequency of the IF signal, present
at the input of the count chain assembly, without having the counter converter locked on the
signal. The counter does not measure pulse parameters when this function is activated.
When Special Function 47 is activated, the counter stops the normal converter lock and
measurement cycles. The VCO, YIG, and all microprocessor-controlled hardware switches are left
at the state they were in when the function was activated. The counter then starts measuring the
frequency of the IF signal present at the input to the count chain assembly. The measurement
results are displayed on the front panel and are also output via the GPIB interface if frequency
readings are requested.
NOTE
This function does not check periodically for the presence of a signal as in
the normal operation of the counter.
This function configures the counter to skip the execution of the input signal tracking function that
normally occurs after every measurement cycle. This function shortens the measurement cycle
time, but prohibits the counter from tracking a moving signal.
The action taken with this function can be reversed by activating Special Function 62.
This function reverses the action taken with Special Function 61.
3-51
TM
This function configures the counter to ignore the local and the remote sample rate controls. The
counter measurement cycle rate is maximized, which shortens the measurement cycle time.
The action taken with this function can be reversed by activating Special Function 64.
This function reverses the action taken with Special Function 63.
This function prevents the display of measurement results on the front panel. When Special
Function 65 is activated, the front panel displays only a row of dots. When the user enters
parameters through the keyboard, the display responds normally. This function shortens the
measurement cycle time and provides security in systems used with classified frequencies.
The action taken with this function can be reversed by activating Special Function 66.
This function reverses the action taken by Special Function 65. When this function is activated,
the display is immediately updated with the last measurement results.
This function configures the counter to display the pulse period measurements as a frequency. It
has no effect on pulse width measurements.
After this function is turned on, frequency measurements are not displayed on the front panel. The
pulse period is displayed to the maximum available resolution, using the pulse parameter display as
3-52
3/OPERATION
the 100 Hz, 10 Hz and 1 Hz digits. Since the PRF is derived mathematically from the period, the
resolution is a function of the period measurement resolution per the formula:
1 1
Resolution ( Hz ) = --------------------------------------- – -----------------
Period ± 10 ns Period
When requested by the GPIB bus controller to output a period measurement, the counter outputs
the period measurement instead of the PRF of the input signal.
If pulse period measurements are enabled, Special Function 67 has a higher priority than Special
Function 69. That is, the front panel is configured according to Special Function 67 if both Special
Function 67 and Special Function 69 are activated.
The action taken with this function can be reversed by activating Special Function 68.
This function reverses the action taken with Special Function 67.
When this function is activated, frequency measurements are not displayed on the front panel.
Instead, the 12 digits on the front panel are devoted exclusively to displaying pulse parameter
measurements to 10 ns resolution.
If the pulse period function is on, this special function has a lower priority than Special Function
67. That is, the front panel is configured according to Special Function 67 if both Special Function
67 and Special Function 69 are activated.
The action taken with this function can be reversed by activating Special Function 70.
This function returns the counter to the normal mode of displaying measurement results, reversing
the action taken by Special Function 69.
3-53
TM
This function serves two purposes. Its primary use is to store the present counter setup in the
storage register specified. When this function is activated, the counter requests the user to enter
the register number by displaying “REG _” on the front panel. The counter remains in this state
until the user enters a number between 0 and 9. After the register number is entered, the function
stores the current counter setup in the register specified. During this time, “REG N” is displayed on
the front panel (where N is the register number entered).
This function also can be used to customize the default values used by the counter. The default
values determine the state of the instrument at turn-on. This is accomplished by setting the
instrument up in the desired turn-on condition and storing it in register 0. The information stored
in register 0 is used to determine the power-on state of the counter. To clear the instrument back
to the factory-set default values, select Special Function 72 and press the CLEAR DATA key.
This function recalls the counter setup stored in the storage register specified.
When this function is activated, the counter requests the user to enter the register number by
displaying “REG _” on the front panel. The counter remains in this state until the user enters a
number between 0 and 9. After the register number is entered, the function proceeds to set up the
counter according to the information stored in the register specified. During this time, “REG N” is
displayed on the front panel (where N is the register number entered).
When the counter finishes setting the counter up, a reset is generated.
When this function is activated, the counter assigns a negative value to the last input frequency
reading and enters it into the frequency offset register (overwriting any previously entered
frequency offset). The last input frequency in this case means the actual frequency of the input
signal, not the frequency displayed on the front panel, which may be affected by a frequency
multiplier or another special function. The counter displays the difference between the last input
frequency and the current one, subject to any other functions activated. It continues to do so until
the SPECIAL FUNC 83 and CLEAR DATA keys are pressed. The OFS annunciator is turned on
when this function is activated.
3-54
3/OPERATION
When this function is activated, the counter assigns a negative value to the local oscillator (LO)
frequency and enters it into the frequency offset register (overwriting any previously entered
frequency offset). The counter then subtracts the LO frequency from the input frequency and
displays the resulting IF frequency. It continues to do so until the SPECIAL FUNC 83 and CLEAR
DATA keys are pressed. The OFS annunciator is turned on.
This special function calculates a checksum based on the data contained in the EEPROM and
compares the calculated checksum to the checksum stored in the EEPROM. If the calculated
checksum is the same as the stored checksum, the counter displays “PASSEd”.
If the counter displays “ERROR 94", the calculated checksum is different from the stored
checksum. Although pressing clear display does clear the error, the counter should not be used; it
must be recalibrated to ensure proper operation.
This function is used to automatically profile the frequency of a repetitive pulsed microwave signal.
When this function is activated, the counter displays “000000.000 Inc”. At this point the desired
measurement increment is entered. After a valid measurement increment has been entered, the
counter displays “000000.000 End”. At this point, the maximum delay time, relative to the trigger,
is entered.
This function only sets up the measurement increment and last measurement time. The front panel
Pulse Generator control keys are used to select the trigger source (Mode) and measurement
window (Width).
This function turns off automatic pulse profiling, reversing the action of Special Function 77.
3-55
TM
This function connects a low pass filter to the Band 0 input. The 3 dB point on the filter is at
approximately 30 kHz which means the attenuation of signals above 30 kHz increases as a function
of frequency.
When measuring signals below 50 kHz, if the count seems unstable, enabling the low pass filter may
stabilize the count by attenuating the higher frequency components of the input signal.
This function disconnects the Band 0 Low Pass Filter, reversing the action of Special Function 79.
The Band 0 Low Pass Filter is, by default, disconnected.
This function allows multiple measurements to be mathematically averaged. Use this function to
select the number of averages in the range of 01 to 99. Entering 01 or CLEAR DATA clears the
averaging function back to the default value of 1.
When measuring pulsed signals, one of the sources of error is averaging error. Special Function 81
is used to average measurements which reduces averaging error. Refer to the measurement
accuracy section of the manual for more information on averaging error.
This function allows the measured frequency to be multiplied by a fixed value prior to being
displayed or returned over GPIB. The multiplier value defaults to a value of 1, but any integer value
between 01 and 99 can be used.
When this function is activated, the counter displays the current multiply factor. A new factor can
then be entered. When a factor is entered, the SPC (special) annunciator is lit. Entering 01 or
CLEAR DATA clears the multiply function back to the default value.
3-56
3/OPERATION
This function allows the measured frequency to be offset by a fixed value prior to being displayed
or returned over GPIB. Frequency offset defaults to zero, but may be any positive or negative value
in the range of –99.999999 GHz to +99.999999 GHz. The maximum resolution on frequency offset
is 1 kHz.
When this function is activated, the counter displays the current frequency offset. When an offset
is entered, the SPC (special) annunciator is lit. Pressing CLEAR DATA or entering “0" followed
by any frequency terminator resets the Frequency Offset value back to the default value of zero.
When this function is activated, the counter displays the current address of the GPIB interface. If
the address need not be changed, the function may be terminated by pressing the CLEAR
DISPLAY or CLEAR DATA keys.
After this function has been activated, the GPIB address can be changed by entering a two-digit
decimal number between 01 and 99. The function is terminated and the display returned to
displaying measurement results after the second digit key is released.
(Refer to the GPIB interface section on page 5-15 for the meanings of GPIB addresses above 31.)
CAUTION
Care must be used when operating Special Function 91. Although the
counter cannot be damaged by this function, improper operation of
it can affect the counter calibration. For this reason, access to this
function is blocked by an internal memory protect feature.
Attempting to access this function with the memory protected
causes the counter to display “ERROR 53”.
This function is used to calibrate the Band 2 input filter. Refer to the service manual for more
information.
3-57
TM
CAUTION
Care must be used when operating Special Function 92. Although the
counter cannot be damaged by this function, improper operation of
it can affect the counter calibration. For this reason, access to this
function is blocked by an internal memory protect feature.
Attempting to access this function with the memory protected
causes the counter to display “ERROR 53”.
This function is used for calibration of the counter’s measurement gate. Refer to the service manual
for more information.
CAUTION
Care must be used when operating Special Function 93. Although the
counter cannot be damaged by this function, improper operation of
it can affect the counter calibration. For this reason, access to this
function is blocked by an internal memory protect feature.
Attempting to access this function with the memory protected
causes the counter to display “ERROR 53”.
The EEPROM is non-volatile memory which stores calibration data along with other instrument
parameters. After performing YIG DAC and/or Gate Error calibrations, this function commands
the counter to compute a new checksum based on the new data and store the new checksum in the
EEPROM. At turn on, the counter automatically calculates the checksum based on the stored data
and compares the calculated checksum to the stored value. If the two checksums are not the same,
the counter displays Error 94.
Although pressing clear display does clear the error, the counter should not be used; it must be
recalibrated to ensure proper operation.
3-58
3/OPERATION
ERROR MESSAGES
When an error occurs, an error number is displayed. The probable cause of each error is listed
below.
ERROR NUMBERDEFINITION
ERROR NUMBERDEFINITION
3-59
TM
35 Word misspelled
36 Missing space
3-60
3/OPERATION
ERROR NUMBERDEFINITION
51 Special Functions 72 and 73 not supported via GPIB (use store and fetch
commands)
52 Illegal entry
60 RAM fault
3-61
TM
ERROR NUMBERDEFINITION
99 No IF detected
3-62
EXT FREQ ASSY - OPTION 5804 A107 - PULSE GENERATOR - OPTION 5810 REAR PANEL
J1 J3
BAND 3 INPUT REMOTE CABLE LO X12 TRIGGER IN
DIPLEXER DIVIDE 100 MHz
26.5 - 170 GHz SENSOR PWR AMP MULTIPLIER
BY N OSC TRIGGER
LO = 5 TO 6 GHz@ +13dBm J2
IF = 600 MHz @ -50 dBm IF IF IF IF SELECT
AMP AMP TRIGGER OUT
ATTEN FILTER
22 dB 40 dB
THRESH OUT
FINE
F/F
DELAY
BAND 3 IF
INHIBIT IN
GATE OUT
A108 - SIGNAL CONDITIONER
(CL)
CONTROL J5
AFC
LOGIC ATT AMP
BAND 0 INPUT
I/O
100Hz - 250MHz CONTROL VIDEO A108 - GATE GENERATOR
LINES (CL) CONTROL
TEMP MUX
J1
GATE
BUFF ACCUM I/O
CONTROL
REAR PANEL
COLUMN
DRIVERS
Figure 3-22. Functional Block Diagram for Models 585C and 588C
3
3-63
This Page Intentionally Left Blank
3-64
4
ACCESSORIES
INTRODUCTION
The accessories available for use with the Phase Matrix 585C and 588C counters are listed in Table
4-1. Part numbers and brief descriptions of the accessories are included.
Table 4-1. Accessories
Accessory Phase Matrix
Number Description Part Number
ACC010 Carrying Case — ——
ACC021 Rack-Mount kit with handles (includes hardware) 2011121-01
ACC022 Rack-Mount kit without handles (includes hardware) 2011122-01
ACC031 Operation Manual (one supplied with each instrument) 5585065-04
ACC032 Operation and Service Manual 5585107-01
ACC043 Molded case and tools for adjustments and troubleshooting 2011125-02
ACC091 Remote Sensor 26.5 GHz - 40 GHz (WR-28) * 2010404-02
ACC092 Remote Sensor 40 - 60 GHz (WR-19) * 2030029-00
ACC093 Remote Sensor 60 GHz - 90 GHz (WR-12) * 2030030-00
ACC094 Remote Sensor 90 GHz - 110 GHz (WR-10) * 2030031-00
ACC095 Remote Sensor 50 GHz - 75 GHz (WR-15) * 2030038-00
ACC096 Remote Sensor 33 GHz - 50 GHz (WR-22) * 2030059-00
ACC097 Remote Sensor 26.5 GHz - 50 GHz (2.92 mm coax, SMA compatible) * 2030122-00
ACC098 Remote Sensor 110 GHz - 170 GHz (WR-6) * 2030121-00
ACC101 Chassis-Slide kit (includes Rack-Mount kit with handles) 2011123-01
ACC102 Chassis-Slide kit (includes Rack-Mount kit without handles) 2011124-01
ACC890 Frequency Extension Cable Kit 2000043-02
* Remote Sensors ACC091 through ACC098 are used in conjunction with model 588C counters that have Option
5804 (Extended Frequency Capability, Band 3) installed.
4-1
TM
CARRYING CASE
M585C-ACC010
The carrying case is shown in Figure 4-1. It is designed to protect the instrument when transporting
it or using it in the field. It is a molded case with die-cut foam inserts that hold and protect the
instrument from shock and vibration.
4-2
4/ACCESSORIES
4-3
TM
4-4
4/ACCESSORIES
4-5
TM
4-6
4/ACCESSORIES
4-7
TM
4-8
4/ACCESSORIES
CHASSIS-SLIDE KIT
(INCLUDES RACK-MOUNT KIT WITHOUT HANDLES)
M585C-ACC102
This chassis-slide kit provides all the parts and hardware required to rack-mount the instrument on
slides for easier access. The rack-mount brackets do not have handles. Figure 4-6 shows one rack
mount, one slide, and the mounting hardware provided in the kit. Two rack mounts, two slides, and
mounting hardware for both sides of the instrument are included.
4-9
TM
4-10
5
PROGRAMMING
REMOTE PROGRAMMING
GENERAL PURPOSE INTERFACE BUS
The GPIB interface of the 585C/588C counters conforms to the IEEE Code and Format
conventions and the IEEE 488-1978 Standards. With the GPIB interface, the counter can respond
to remote control instructions and can output measurement results via the IEEE 488-1978 bus
interface. At the simplest level, the counter can output data to other devices, such as a thermal
printer. In more sophisticated systems, an instrument controller or computer can program the
counter remotely, trigger measurements, and read results. A quick reference list of GPIB commands
is located in Appendix A at the end of this manual.
5-1
TM
The 585C/588C counters thus have the capacity to provide the following capabilities in remote
operation:
5-2
5/PROGRAMMING
• Header only
• Header and argument
• Header and argument and terminator
Where the header is a reserved word, the argument is a number or a reserved word, and the
terminator is a reserved word.
Messages can be concatenated with a comma (,) or semicolon (;) as separators. A message chain
can be terminated with CR LF or NL or EOI.
NOTE
Any device-reserved word is recognized by at least the two first characters,
with the exception of RESEt, which requires the first four letters to be
entered, and DELta, which requires the first three letters to be entered.
These first two characters are printed in large boldface type in the following
command lists and in program examples to promote user familiarity with
the shortened form of the command. Spelling of more characters (up to the
full word) is optional for user program readability.
A <number> can be sent in any of the defined IEEE formats (NR1, NR2, NR3).
Example: 12000, 12000.00, 001.2e4, and .12000E+5 are all recognized equivalently.
The reserved word DEFAULT can replace a numeric argument for default value assignment.
The terminator in the parameter messages group is optional, and defaults to Hz or seconds.
A command message having more than one word (e.g., PERIOD ON) should have a space between
words. However, this is optional if the second word is a number (OFFSET4.3e9 and OFFSET 4.3E9
are recognized equivalently). Additional spaces in front of words, between words, or after a message
are optional, and are ignored. Nulls and CRs are ignored anywhere. Both upper case and lower case
characters are equally acceptable.
Following are the possible GPIB command messages for the 585C/588C series of counters.
5-3
TM
CONTROL MESSAGES
Control, mode, and parameter messages are all used with the controller in the talker mode to send
instructions and data, as shown in Table 5-2, Table 5-3, and Table 5-4.
NOTE
In the local mode, SAMPLE RATE and HOLD are controlled via the front
panel control, but in remote the front panel control has no effect. In the
remote mode, both SAMPLE RATE and HOLD are under software control.
Refer to GPIB SAMPLE RATE and HOLD commands.
5-4
5/PROGRAMMING
5-5
TM
Header Description
OUTPUT BAND Outputs the number of the last specified band.
OUTPUT CENTERFREQ Outputs the center frequency last specified.
Outputs a 42-character string that shows the revision level and date code of the
OUTPUT DATE software. The format is: #######-## REV.X MM-DD-YY.
OUTPUT DEFAULT Outputs displayed data.
OUTPUT DELTAF Outputs the last specified ΔF value.
OUTPUT ERRORNUMBER Outputs the number of the last error. See listing of error numbers in Section 3.
OUTPUT FREQUENCY Controls which measurement results to output. (Note: More than one measurement
(AND WIDTH) result is optional. The order of the results is preserved in the output. These three,
(AND PERIOD) frequency, width, and period, can be used in any combination.)
5-6
5/PROGRAMMING
none
OUTPUT GPERIOD none Outputs the last pulse period
SYNTAX DEFINITION
In the instructions that follow, | means “or” and N|S means “null or space”. The format used for
the examples is generic for HP controllers. Sample formats can easily be adapted for other
controllers.
5-7
TM
Example: To instruct the instrument to begin a new signal acquisition process, the following
statement could be used: Send 0, 18, “RESET”, NLend
Example: To instruct the instrument to an external reference, the following statement would
be used: Send 0, 18, “EXTERNAL ON”, NLend
Parameter 1 BAND|SUBBAND|RESOLUTION|SPECIAL|
AVERAGE|MULTIPLIER|SQRMASK
Argument DEFAULT|<number>
Number <NULL|+|><mantissa><exponent>
Example: To instruct the instrument to accept an averaging value, the following statement
could be used: Send 0, 18, “AVERAGE 70", NLend
Parameter 2 OFFSETFREQ|HIGHLIMIT|LOWLIMIT|
MINPRF|CENTERFREQ|Y1FREQ|Y2FREQ|
Y3FREQ|V1FREQ|V2FREQ|DELTAF
5-8
5/PROGRAMMING
Example: To instruct the instrument to set a frequency high limit value of 12.3 GHz, the
following statement could be used: Send 0, 18, “HIGHLIMIT 12.3 GHz”
Example: To instruct the instrument to set the sample rate to 0.1 Hz, the following
statement could be used: Send 0, 18, “SAMPLERATE 100 MSEC”, NLend
Example: To instruct the instrument to change memory location 99AF to 3B, the following
statement could be used: Send 0, 18, “MEMORY O99AF 03B”, NLend
Example: To instruct the instrument to store or recall a counter setup in a specified storage
register, these statements could be used:
Send 0, 18, “STORE 03", NLend or Send 0, 18, ”FETCH 03", NLend
5-9
TM
Example: To request the controller to display the width and frequency, in that order, the
operator enters:
The following program segment illustrates how controllers function with the counter. This program
sets the counter up in a sample configuration and programs it to make a measurement of a 12.5 GHz
pulsed signal with a 13.258 ms period. The talk and listen address of the counter is assumed to be
18.
Dimension A$(36)
Send 0, 18, “IN”, NLend
Wait 4000
Send 0, 18, “WI ON, RE 4", NLend
Send 0, 18, ”HI 17.5 GHZ, LO 1.1 GHZ", NLend
Wait 1000
Send 0, 18, “OUTPUT WI AND FR”, NLend
Wait 1000
Receive 0, 18, A$, StopEnd
Print A$
5-10
5/PROGRAMMING
This program initializes the counter, provides a resolution value and a high and low frequency limit,
and instructs it to output pulse width to the counter display and pulse width and frequency to the
controller display. The controller display would appear something like this:
0.0000132580 12500000000
The controller can be instructed to output any ordered combination of the three measurements, no
matter what is displayed on the front panel.
OUTPUT PERIOD
The format of each output message can be controlled using the following:
NOTE
Terminator takes over the exponential role if both SCIENTIFIC and
HEADER are ON.
5-11
TM
Example: The counter is measuring a 12.34 GHz pulsed signal with 98 ns width and 14.567 ms
period. The operator enters the following messages through the controller:
RESOLUTION 6
OUTPUT FREQ AND WIDTH AND PERIOD
The output for each message format is listed in Table 5-7(b = blank).
Parameter Output
Default bbbbb12340000000,bbbbb0.000000100,bbbbb0.000014570 CR LF
SCIENTIFIC on bbbbbbbb12.340E+9,bbbbbbbbbb100E-9, bbbbbbbb14.57E-6 CR LF
DYNAMIC on 34E9,100E-9,14.57E-6 CR LF
SEPARATE on 2.34E9 CR LF
100E-9 CR LF
14.57E-6 CR LF
If the counter is searching, zero data is output to the controller on all results once every search loop.
If the counter has found a signal, a measurement result is output only once. When the instrument
is in HOLD, therefore, the user must trigger the counter before sending another talk address.
Otherwise, since it has no data to output, the counter holds indefinitely.
STATUS BYTE
Both the 585C and 588C counters maintain a one-byte register that contains current information
on the status of the instrument. This register, called the status byte, can be accessed through the
GPIB using the serial poll command. When serial polled, the counter responds by returning a
numeric value between 0 and 256. This value is the weighted sum of the status bits which are set.
The status byte structure is shown in Figure 5-1. For example, a value of 132 indicates that the
GPIB input buffer is empty and the counter is in the search mode.
5-12
5/PROGRAMMING
bit7 6 5 4 3 2 1 bit0
(128) (64) (32) (16) (8) (4) (2) (1)
For example, to instruct the counter to generate an SRQ whenever it has valid data available or an
error condition exists, send the following service request mask:
This would tell the counter to generate an SRQ whenever bit 0 and bit 5 of the status byte are set.
Since bit 0 corresponds to valid measurement result ready and bit 5 corresponds to an error
condition, the counter would generate an SRQ whenever either an error condition exists or a valid
measurement is available.
The following items should be included in any program using the SRQ feature:
1. Tell the counter when to generate an SRQ. That is, tell the counter which events should
generate an SRQ. This is done using the SRQMASK command.
2. Tell the controller to monitor the SRQ line on the GPIB. The SRQ is a maskable interrupt and
the controller needs to know if it should respond to the interrupt.
3. Tell the controller what to do when it receives an SRQ interrupt.
4. Serial poll the counter after an SRQ is generated to clear the interrupt. When the counter
generates an SRQ it sets bit 6 in the status byte. Serial polling the instrument clears the SRQ
bit and allows the instrument to generate a new SRQ upon the next occurrence of the
conditions specified in the SRQ Mask.
5. It may also be necessary to clear the SRQ register in the controller. Consult your manual on
the controller for more information on clearing the SRQ register in the controller.
5-13
TM
5-14
5/PROGRAMMING
NOTE
Address is composed of the binary value of the choices +32.
5-15
TM
It takes a specific amount of time for the counter to process input data (error checking, formatting,
changing the mode of operation, etc.). To prevent the data rate of the bus from slowing down while
the counter is processing input data, the data is accepted as soon as it is available on the bus and is
temporarily stored in a 256-character storage memory.
It is necessary to be aware of the difference between accepting data and complying with it. If the
counter is asked to output a reading before it is finished processing the input data, the output does
not reflect the newly entered data. To prevent this, sufficient programmed delays must be provided
(see the sample program formats on page 5-7). Bit 7 in the status byte can be used to determine if
the counter has completed the processing of the GPIB command messages. Refer to the section on
the status byte.
OUTPUT SPEED
Several options have been provided in the GPIB interface for the user who wants to increase the
output speed of the counter. Each of the following conditions increases the measurement cycle rate.
The fastest measurement cycle time is achieved with all of the following conditions set:
• HEADER OFF: Outputs the numeric results without header or terminator (default).
• SCIENTIFIC OFF: Outputs fixed point results which are shorter than exponential notations
(default).
• DYNAMIC ON: Suppresses leading blanks. Note: The controller has to have free field
capability.
• SPECIAL 61: Disables the tracking feature, thus saving the time required for YIG and VCO
corrections.
• SPECIAL 63: Disables sample rate control, thus deleting any delay between gates. (For counter
in local mode.)
• SPECIAL 65: Disables the LED results display thus saving the time required for display,
formatting and output.
• SAMPLERATE 0: Same as SPECIAL 63 for counter in remote mode.
READING MEASUREMENTS
The Phase Matrix 585C/588C counters provide a choice of methods for taking readings. When the
command HOLD is ON, the counter takes one reading then waits for a RESET command or a
device trigger GPIB command. In this condition, the counter is sent a RESET command or a device
trigger and (when addressed to talk) outputs a new reading to the bus. The counter holds that
particular reading on the display until another RESET command or device trigger is received.
When the HOLD command is off, data is read out to the bus in the normal way. The display is
automatically updated at the specified sample rate, and the counter outputs successive
measurement readings without requiring a RESET command or device trigger each time.
5-16
6
OPERATIONAL VERIFICATION TESTS
INTRODUCTION
This section contains information for verifying proper operation of the counter. Although these
tests are not comprehensive, they do ensure, to a high degree of confidence, that the instrument is
operating properly. The tests can be useful for incoming inspection and should be performed after
any servicing to ensure proper operation of the counter. All tests can be performed without
removing the instrument covers. A test report form is included at the end of this section that can
be used to provide a test record. If the application is especially critical in nature, more extensive
testing may be required and is covered in the performance verification test section of the service
manual. Because of the high cost and specialized nature of frequency sources above 40 GHz, testing
above this frequency is not covered. Also, for the purpose of operational verification tests,
simulated pulsed signals are used in Bands 1 and 3.
EQUIPMENT REQUIREMENTS
Equipment required for the operational verification tests on the Phase Matrix 585C or 588C
counter is listed in Table 6-1. The critical parameters are the minimum use specifications required
for the performance of the procedures, and are included to assist in the selection of alternative
equipment. Satisfactory performance of alternative items should be verified prior to use. All
applicable equipment must bear evidence of current calibration. For many of the following tests, an
Phase Matrix 578B counter is used to source-lock the microwave sweeper, thus providing a stable
source for testing. This combination may be replaced by a frequency synthesizer.
6-1
TM
This test verifies counter operation from 100 Hz to 250 MHz at –20 dBm (0. 063 V p-p into 50 Ω).
The oscilloscope is used to set signal levels below 10 MHz, and the power meter is used to set signal
levels at 10 MHz and above. Test setup 1 covers the frequency range from 100 Hz to 10 MHz and
test setup 2 covers the range from 10 MHz to 250 MHz.
EQUIPMENT
• Frequency synthesizer (Agilent 3325A)
• Synthesized Sweeper (Agilent 83640B)
• Power meter (Agilent 437B)
• Power sensor (Agilent 8481A)
• Oscilloscope (Tektronix 2445)
6-2
6/OPERATIONAL VERIFICATION TESTS
TEST SETUP 1
PROCEDURE
1. Connect equipment as shown in Figure 6-1.
BAND 0 OSCILLOSCOPE
INPUT
50 Ω
TERMINATION
Figure 6-1. Band 0 Range and Sensitivity Test Setup (100 Hz to 10 MHz)
585C/588C COUNTER
POWER METER
BAND 0
SYNTHESIZED SWEEPER INPUT
POWER
SENSOR
Figure 6-2. Band 0 Range and Sensitivity Test Setup (100 MHz to 250 MHz)
6-3
TM
This test verifies counter operation from 250 MHz to 1 GHz at –20 dBm for both CW and simulated
pulsed signals. The pulse generator is used to simulate a pulsed signal by applying a 50 ns TTL high
with a 1 MHz repetition rate to the INHIBIT IN connector on the rear panel of the 585C/588C
counter. The power meter is used to set signal levels.
EQUIPMENT
• Synthesized Sweeper (Agilent 83640B)
• Power meter (Agilent 437B)
• Power sensor (Agilent 8481A)
• Pulse generator (Wavetek 801)
• Oscilloscope (Tektronix 2445)
INHIBIT
IN 585C/588CICOUNTER
SYNTHESIZED SWEEPER
50IOHM
TERMINATION
PROCEDURE
1. Connect equipment as shown in Figure 6-3.
2. Set the 585C/588C counter to Band 1, and select resolution 3.
3. Set the synthesized sweeper to 250 MHz.
4. Using the power meter, set the output signal level from the synthesized sweeper to –20 dBm.
5. Apply the 250 MHz signal to the counter, verify proper reading, and record the frequency
reading.
6-4
6/OPERATIONAL VERIFICATION TESTS
6. Repeat steps 3, 4, and 5 at 300 MHz, 400 MHz, 500 MHz, 600 MHz, 700 MHz, 800 MHz,
900 MHz and 1 GHz.
7. Using the oscilloscope, set up the pulse generator to output a 50 ns wide TTL high signal with
a 1 MHz pulse repetition rate. Apply the signal to the INHIBIT IN connector to simulate a
pulsed signal.
8. Repeat steps 3, 4, 5, and 6 for the simulated pulsed signal and record the results.
This test verifies counter operation from 1 to 20 GHz (26. 5 GHz for the 588C counter). The first
part of the test verifies operation in the CW mode. Next, the counter is tested in the pulse mode
using the pulse modulators to modulate the microwave source. Attenuators are necessary on the
input and output of the pulse modulators to reduce frequency pulling of the microwave source. To
be able to accurately set the power level of the pulsed signal, it is necessary to compensate for the
insertion loss of the pulse modulators by applying a constant enable signal to the pulse modulator
and adjusting the sweeper at each test frequency until the output power from the modulator is at
the required level.
EQUIPMENT
• Synthesized Sweeper (Agilent 83640B)
• Power meter (Agilent 437B)
• Power sensor (Agilent 8485A)
• Pulse generator (Wavetek 801)
• Pulse modulator (Agilent 8731B)
• Pulse modulator (Agilent 11720A)
• Pulse modulator (Narda S214DS)
• Directional coupler (Narda 4226-10)
• Directional coupler (Narda 4017C-10)
• 3 dB attenuator (2) (Weinschel 9-3)
6-5
TM
TEST SETUP 1
585C/588C COUNTER
POWER METER
BAND 2
SYNTHESIZED SWEEPER INPUT
POWER
SENSOR
PROCEDURE
1. Connect equipment as shown in Figure 6-4.
2. Set the 585C/588C counter to Band 2, and select resolution 3.
3. Set the synthesized sweeper to 1 GHz.
4. Using the power meter, set the output signal level from the synthesized sweeper to –20 dBm.
5. Apply the 1 GHz signal to the 585C/588C counter, verify proper reading, and record the
frequency reading.
6. Repeat steps 3, 4, and 5 at 3 GHz, 6 GHz, 10 GHz, 12. 4 GHz (at –25 dBm), 15 GHz, 18 GHz,
and 20 GHz (at –20 dBm). For Model 588C counters, also test at 22 GHz, 24 GHz, and 26. 5
GHz (at –15 dBm).
6-6
6/OPERATIONAL VERIFICATION TESTS
TEST SETUP 2
585C/588CICOUNTER
BAND 2
INPUT
POWERIMETER
POWER
SENSOR
PULSE GENERATOR
OSCILLOSCOPE
OUTPUT
50IOHM
TERMINATION
Figure 6-5. Band 2 Pulsed Signal Range and Sensitivity Test Setup
PROCEDURE
For this test, three pulse modulators and two directional couplers are used. Use the devices that
correspond to the frequency under test. See Table 5-1 for a listing of frequency ranges of the pulse
modulators and directional couplers.
6-7
TM
8. Repeat steps 4, 5, 6, and 7 at 3 GHz, 6 GHz, 10 GHz, 12. 4 GHz, 15 GHz, 18 GHz, and
20 GHz. For Model 588C counters, change the pulse width from the modulator to 500 ns and
test at 22 GHz, 24 GHz, and 26. 5 GHz.
This test verifies that the counter accurately measures the larger of two signals differing in
amplitude by 15 dB or more.
EQUIPMENT
• Synthesized Sweeper (Agilent 83640B)
• Synthesized Sweeper (Agilent 83624B)
• Spectrum analyzer (Agilent 8566B)
• Power splitter (Agilent 11667B) (used as signal combiner)
RF OUT
BAND 2
INPUT
HP 11667B
SPECTRUM ANALYZER
SIGNAL
SYNTHESIZED SWEEPER COMBINER
RF OUT
RF IN
PROCEDURE
1. Connect equipment as shown in Figure 6-6.
2. Set signal generator 1 to 3. 0 GHz at 0 dBm and set signal generator 2 to 3. 1 GHz at +6 dBm.
3. Using the spectrum analyzer, adjust the generator power levels so that the signal amplitude
difference is 15 dB.
4. Verify that the counter correctly measures the frequency of the higher power signal source.
5. Repeat steps 2, 3, and 4 at the following signal generator 1 and signal generator 2 frequencies:
6 and 6. 1 GHz, 12 and 12. 1 GHz, 17. 9 and 18 GHz.
6-8
6/OPERATIONAL VERIFICATION TESTS
This test verifies counter operation from 26. 5 GHz to 40 GHz at –20 dBm for both CW and
simulated pulsed signals. The pulse generator is used to simulate a pulsed signal by applying a 50 ns
TTL high with a 1 MHz repetition rate to the INHIBIT IN connector on the rear panel of the
counter. The power meter is used to set signal levels.
EQUIPMENT
• Synthesized Sweeper (Agilent 83640B)
• Power meter (Agilent 437B)
• Power sensor (Agilent R8486A)
• Pulse generator (Wavetek 801)
• Oscilloscope (Tektronix 2445)
• Remote sensor (Phase Matrix 091)
• Cable kit (Phase Matrix 890)
INHIBIT
SWEEP GENERATOR IN 585C/588CICOUNTER
BAND 3
INPUT
POWER METER
REMOTE
SENSOR
POWER
SENSOR
PULSE GENERATOR
OSCILLOSCOPE
OUTPUT
Figure 6-7. Band 3 Range and Sensitivity Test Setup (Model 588C Option 5804 Only)
PROCEDURE
1. Connect equipment as shown in Figure 6-7.
2. Set the 585C/588C counter to Band 3, sub-band 1, and resolution 3.
3. Set the output frequency from the synthesized sweeper to 26. 5 GHz.
6-9
TM
4. Using the power meter, set the output signal level from the sweeper to –20 dBm.
5. Apply the 26. 5 GHz signal to the remote sensor, verify proper reading, and record the results.
6. Repeat steps 3, 4, and 5 at 30 GHz, 35 GHz, and 40 GHz.
7. Set up the pulse generator to output a 50 ns wide TTL high signal with a 1 MHz repetition
rate and apply the signal to the INHIBIT IN connector on the rear panel of the counter. This
signal gates signal threshold inside the counter and is used to simulate a pulsed signal.
8. Repeat steps 3, 4, 5, and 6 for the simulated pulsed signal and record the results.
6-10
6/OPERATIONAL VERIFICATION TESTS
1 GHz 15 GHz
3 GHz 18 GHz
6 GHz 20 GHz
10 GHz (588C)
12.4 GHz 22 GHz
24 GHz
26.5 GHz
F1 F2 PASS FAIL
6-11
TM
6-12