Mult Adj
Mult Adj
AX
Service
Adjustment Instructions
English
Print No.: RXB1-140.071.01.04.02 Doc. Gen. Date: 07.01
Replaces: RXB1-140.071.01.03.02
0-2 Revision
Chapter Page Revision
all all 01
all all 02
all all 03
all all 04
Disclaimer
The installation and service of equipment described herein is to be performed by qualified per-
sonnel who are employed by Siemens or one of its affiliates or who are otherwise authorized by
Siemens or one of its affiliates to provide such services.
Assemblers and other persons who are not employed by or otherwise directly affiliated with or
authorized by Siemens or one of its affiliates are directed to contact one of the local offices of
Siemens or one of its affiliates before attempting installation or service procedures.
Page
1 _______Requirements/Notes ____________________________________________ 1 - 1
Page
Adjusting the Horizontal SID to the Bucky Wall Stand (only for ACSS, ACSS P) . . .2 - 26
Adjusting the R2 Longitudinal Potentiometer (only in Tomo Versions) . . . . . . . . .2 - 27
Replacing the Support Cable in the Column Stand . . . . . . . . . . . . . . . . . .2 - 28
Patient Table . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 1
Longitudinal Roller Bearings . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 2
Bucky Longitudinal Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Adjusting the S111, S112 Safety Limit Switches . . . . . . . . . . . . . . . . . . 3 - 3
R11 Potentiometer . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 3
Adjusting the Toothed Belt Tension . . . . . . . . . . . . . . . . . . . . . . . . 3 - 4
Lift Drive (only for versions with table lift) . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
Manually Adjusting the Lift Drive . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
Adjusting the R10 Lift Potentiometer . . . . . . . . . . . . . . . . . . . . . . . . 3 - 5
S10 / S11 Safety Limit Switch for the Lift Drive (with toothed belt) . . . . . . . . . 3 - 5
Adjusting the Pinion Rack for Height Sensing (R10 Potentiometer) . . . . . . . . 3 - 6
Adjusting the R10 Lift Potentiometer (with pinion rack). . . . . . . . . . . . . . . 3 - 6
Adjusting the S10 / S11 Safety Limit Switches for the Lift Drive (with pinion rack) . 3 - 7
Adjusting the Table Lift Guide . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 8
Bucky . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 9
General Adjustments . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3 - 9
Grid Blur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 10
Grid Drive . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 11
Checking the Safety Circuit. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 12
Tabletop Brakes . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 13
Adjusting the Air Gap . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 13
Programming the Table Type . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 13
Programming the Buzzer Status with a Floating Tabletop . . . . . . . . . . . . .3 - 14
Setting a Customer-specific Table Lift Stop Position (not for FIX) . . . . . . . . .3 - 14
Tomo Device (only for the tomo version) . . . . . . . . . . . . . . . . . . . . . . . .3 - 15
Adjustment of the Tomo Device . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 15
Adjusting the Catapult Bucky - Center . . . . . . . . . . . . . . . . . . . . . . .3 - 15
Adjustment of the Stand - Table Middle . . . . . . . . . . . . . . . . . . . . . .3 - 15
Test Conditions . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 16
Brief Description of the Test . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 16
Description of the Tomo Test Phantom . . . . . . . . . . . . . . . . . . . . . . . . .3 - 17
Sequence of the Checks . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 19
Evaluation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Accuracy of the Tomo Height Display . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Adjustment of the Tomo Height. . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Optical Resolution Quality . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 20
Tomo Figure and Even Movement of the Tomo Device . . . . . . . . . . . . . .3 - 21
Overall Evaluation of the Tomo Device . . . . . . . . . . . . . . . . . . . . . . . . .3 - 22
Form and Movement of the Tomo Figure . . . . . . . . . . . . . . . . . . . . . .3 - 22
Evenness of the Blur . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 23
Tomo Symmetry . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .3 - 24
Adjusting the Tomo Height Light Localizer (Option with Tomo Versions) . . . . . . . .3 - 25
Page
4 _______VERTIX PRO/TOP ______________________________________________ 4 - 1
Safety Information 1
When performing the work steps and checks, the product-specific safety information con-
tained in this document as well as the general safety information contained in Register 2
must be observed.
Tolerances 1
These tolerances apply for all dimensions listed in these instructions providing no other
tolerance is expressly listed after the value.
A tolerance of ±10% is permitted for torque values.
When checking tightness, use the appropriate torque value for hardness rating 8.8.
A tolerance of ± 10% is permitted for torque values.
NOTICE All tools, test equipment and aids, with the exception of the "stan-
dard installation tool kit" are listed and specified in the Service
Tools Catalogue.
Text Emphasis 1
WARNING All texts labeled with "WARNING" call attention to potential risks
to health or life.
Abbreviations Used 1
Many adjustments are performed using the service software. A help text is available for
the individual input masks. Additional details for the individual adjustment steps are listed
in the help texts.
Select help text: "F1" key
Close help text: "Esc" key
The terms ”length” and ”width” used and insertion of cassettes should be understood
as illustrated in the sketch:
Length
Tabletop
24 cm x 30 cm cassette
*1
Width X
30 cm x 24 cm cassette
Patient
*1 The first digit
always indicates the
length!
Y
Special Symbols 1
NOTICE Prior to making any adjustments on the column, the table and the
rails must be bolted to the floor to the required torque value.
Fig. 1
• Move the column in to the normal position; to do this, position the support arm above the
Bucky.
• Loosen the mounting screws for the 4 eccentric bearings (1/Fig.1)
• Adjust each eccentric bearing so that the bearings are flat against the contact surfaces
and so that when the carriage is moved over its entire range, the bearings can just still be
turned by hand.
Fig. 2
• Move the column into the normal position; to do this, position the support arm above the
Bucky.
• Adjust the 3 rear eccentric bearings (1/Fig.2) by loosening and turning them so that the
bearings are flat against the contact surfaces and can just still be turned.
• Secure the eccentric bearings in position.
• Rotate the column 90° both to the left and to the right and in these positions check, and if
needed, correct the adjustment of the bearings.
Fig. 3
Fig. 4
Fig. 5
NOTICE The position of the guide rail is currently secured in place by roll
pins and should be changed only if absolutely necessary.
• Check the vertical position of the column over the entire movement range:
To do this, place on a spirit level as shown in Fig. 5 and move the column across the rail
and observe any changes in the position of the bubble.
• Shift the position of the guide rail to compensate for any deviations; to do this, loosen the
mounting screws (1/Fig. 5).
• Temporarily remove the switch strike plates to gain access to the external mounting
screws. Be sure to mark their positions!!
• With rails that are mounted using roll pins, remove the safety roll pins:
To do this, remove the mounting screws and carefully pry up the guide rail until the roll
pins can be pulled or tapped out. Turn the mounting screws back in but do not tighten
them yet.
• Move the guide rail accordingly and check the adjustment with the spirit level.
• Check this adjustment at all mounting screws on the guide rail and, if needed, repeat the
above procedure.
• Tighten the screws. Torque 20 Nm.
• Check the correctness of this adjustment and, if necessary, readjust.
• With tomo versions, confirm the correctness of the adjustment by making a tomo
exposure.
• Drill new 4 mm holes next to the old holes for the roll pins.
• Tap in the roll pins until they are flush.
Fig. 6
Adjusting the Column Stop Position (required only for versions without
tomo) 2
Fig. 7
• If not already done, remove the front cover panel on the column. To do this, remove the
top cover panel (1/Fig. 7) on the column and remove the front cover panel (2/Fig.7)
towards the top.
• Adjust the bearings (4/Fig.7) on the column longitudinal carriage (3/Fig.7) so that the
carriage moves centered and parallel in the column.
• To do this, loosen the mounting screws for the bearings and move the bearings (4/Fig.7)
laterally as needed.
• Adjust the contact pressure so that the bearings (4/Fig.7) can just be turned over the
entire movement path.
Fig. 8
Support Arm 2
• Use a feeler gauge to adjust the air gap of 0.3, ± 0.1mm on the rotation brake (1/Fig 8):
- Remove the bottom cover panel on the support arm.
- Pivot the tube assembly so that the Allen-head set screws (3) (2/Fig.8) are accessible.
- Loosen each of the Allen-head set screws approx. 1 turn and adjust the air gap.
- Check this adjustment at the two other openings of the air gap and, if needed, correct
them (parallelism).
- Retighten the Allen-head set screws.
- Check the air gap of 0.3, ± 0.1 mm at all 3 accessible openings.
Fig. 9
DHA Stop 2
Fig. 10
• If not already done, remove the front cover panel on the column. To do this, remove the
top cover panel (1/Fig.10) on the column and remove the front cover panel (2/Fig.10)
towards the top.
• Move the support arm into the top position (reference position).
• Select the following routine in the service software:
- Adjustment – Calibration
- Function groups: Column
- Adjustments: SID Potentiometer
• The display of the reference value must be 0, +/-2.
• To make a correction, use a size 12 open-end wrench to loosen the R21 potentiometer
mount and pull the potentiometer with the pinion gear towards the front so that the pinion
gear can be turned. Then turn the pinion gear until a zero adjustment is achieved with a
tolerance of +/- 2.
• When the value is within tolerance, a signal sounds on the Service PC.
• Press the potentiometer back against the toothed belt and secure it in place.
Fig. 11
• Pivot the tube assembly into the 90° left position and the 90° right position and check the
angle display in the control panel display field.
• If the displays are not symmetrical (e.g. 89° & 91°), the display can be readjusted by
adjusting the R51 potentiometer in the particular position.
• Check the display in the opposite direction and, if needed, repeat the adjustment.
Fig. 12
• The rotation brake must be in the reference position. To do this, turn the tube assembly
in the uncontrolled mode so that the cable outlet is positioned at the middle of the rotation
brake (brake assembly) on the top.
• Check the R50 potentiometer adjustment (1/ Fig.12) and, if needed, correct it.
• To do this, select the following routine in the service software:
- Adjustment – Calibration
- Function groups: Column
- Adjustments: Initial rotation position potentiometer
• The service software display value must be 0, ± 2.
• If needed, loosen the set screws on the pinion gear (drive side) and turn the pinion gear
(at the potentiometer shaft) until the zero adjustment is reached with a tolerance of ± 2.
• When the value is within tolerance, a signal sounds at the Service PC.
• Resecure the pinion gear at the drive with the screw.
3
2
Fig. 13 Fig. 14
• Position the tube assembly in the vertical beam direction approx. 1 m above the floor.
• Switch on the light localizer and line up the marks on the floor with the light field.
• Use the knobs (1/Fig. 13) to collimate a narrow light field in the longitudinal direction.
• Switch on the light localizer and move the tube assembly to the max. SID; when doing
this, check the drift of the light field. The drift may not exceed 5 mm.
Correction:
• Remove the tube assembly cover panel (1/Fig. 14) on the collimator (2 paneling screws).
• Loosen the 2 Allen screws (2/Fig. 14) approx. 2 turns.
• Loosen the 2 set screws (3/Fig.14).
• Adjust the tube assembly around the longitudinal axis.
• Retighten the 2 Allen screws and the set screws.
• Recheck the light field drift.
• Reinstall the tube assembly cover panel.
3 2
Fig. 15 Fig. 16
• Position the tube assembly in the vertical beam direction approx. 1 m above the floor.
• Switch on the light localizer and line up the marks on the floor with the light field.
• Use the knobs (1/Fig. 15) to collimate a narrow light field in the transverse direction.
• Switch on the light localizer and move the tube assembly into the max. SID; when doing
so, check the drift of the light field. The drift may not exceed 5 mm.
Correction:
- Remove the cover panel on the tube support arm (1/Fig. 16).
- Loosen the M6 nut (2/Fig. 16).
- Adjust the eccentric bearing (3/Fig. 16); to do this, release the rotation brake at the
keyboard.
- Retighten the nut (2/Fig. 16).
- Reinstall the cover panel.
• Mark the center of the tabletop in the longitudinal direction (e.g. using a stretched out
piece of string).
• Switch on the light localizer.
• Adjust the light localizer centering cross with the mark.
• Move the column over the entire movement range and determine the amount of drift in
the longitudinal direction (tolerance must be less than / equal to 5 mm).
1 3
2 3
Fig. 17
To adjust the focus-near plates, use the template included in the shipment.
(The point (see 2/Fig. 17) with the template is only 2 mm wide.)
• Check the plate adjustment using the adjustment gauge that is provided for this (see 1/
Fig. 17).
• To do this, place on the adjustment gauge and let it be up against the side flange.
• Completely close the plates.
• The focus-near plates (3/ Fig. 17) must be touching the point (see 2/ Fig. 17).
• If necessary, adjust the plates by bending them. When doing this, it is absolutely
necessary to secure the guide for the plate to prevent damaging the drive.
• When the plates are completely open, the focus-near plates may not be symmetrically
flat against the sides of the adjustment gauge.
Fig. 18 Fig. 19
• If needed, the automatic format blanking in the ACSS mode can be reduced.
• To do this, select the following routine per the service software:
- Menu: Adjustment - Calibration
- Function groups: Aperture
- Adjustments: Adjustment of the collimator correction values
Program the type of cassettes used on side and the SID displays to be uniform in "cm" or
"inch".
• To do this, select the following routine per the service software:
- Menu: Adjustment
- Function group: Aperture
- Adjustment: Adjustment of display.
For additional options, see the screens and Help texts for the collimator SSW
beginning with version VB01A
Length
Light field
Marker
Side marker
Fig. 20
Checking and Adjusting the Coincidence of the Radiation Field to the Light
Field 2
Y1
Film
Light field
X2 X1
Radiation field
Y2
Fig. 21
1 2
2
Fig. 22
1 2 3
Fig. 23
• The collimators used in the MULTIX PRO / TOP system are shown in the illustration, Fig.
23.
- 1. Collimator 04 68 215 G052G (ACSS)
- 2. Collimator 04 68 207 G052G (manual)
- 3. Collimator RM 70 06 625 GH001 (manual)
Checking the Lamp Voltage for Collimator Model G052G
(1 and 2/Fig. 23)
• The light localizer lamp in collimator G052G (1 and 2 Fig. 23), beginning with Serial
Number 3001, can be controlled in the controlled or uncontrolled mode beginning with
introduction of a new power supply in the MULTIX table.
Programming is possible using the S1.4 switch in collimator G052G!
- Controlled mode = voltage stabilization (S1.4 = ON)
- Uncontrolled mode = direct power supply from the transformer (S1.4 = OFF)
NOTICE To increase the operating life of the light localizer lamp, if possi-
ble, it should be operated in the uncontrolled mode.
Prerequisite for the uncontrolled mode of collimator G052G:
1. In the M8 T1 transformer, four additional taps for the uncon-
trolled mode of the collimator must be present:
35, 36, 37, 38, see (1/Fig. 25)
The 16 = 34 Volt tap may used only for the controlled mode!!!
2. Set switch S1.4 in the collimator to OFF
(voltage regulator off).
Measure a lamp voltage of 23.5 V,+/- 0,5 directly at the lamp base.
Fig. 24 Fig. 25
• Check for correct connection of the collimator at M8.T1.X3 in the transformer for MULTIX
(1/Fig. 25):
NOTICE The lamp voltage with a line voltage of 400/440/480 Volt should be
as close as possible to 23.5, +/- 0.5 Volt. Select the tap at M8.X3 so
that the required value is almost reached.
The upper limit of 24 Volt may not be exceeded under any circum-
stances.
Fig. 26 Fig. 27
• Check for correct connection of the collimator at M8.T1.X3 in the transformer for MULTIX
(1/Fig. 25):
- MULTIX PRO M8X3.35 = 26 Volt (AC) factory setting
- Reserve M8X3.36 = 27 Volt (AC)
- MULTIX TOP M8X3.37 = 29 Volt (AC) factory setting
- Reserve M8X3.38 = 30 Volt (AC)
• Switch the collimator light localizer and measure the voltage (23.5 Volt) directly at the
lamp base (1/Fig. 27).
NOTICE With line voltages of 400/440/480 Volt, the lamp voltage should be
as close as possible to 23.5, +/- 0.5 Volt. Select the tap at M8.X3 so
that the required value is almost reached.
The upper limit of 24 Volt may not be exceeded under any circum-
stances.
2
1
4 3
Fig. 28
• To adjust the laser light localizer, first only slightly loosen the Allen screw (1/Fig. 28)
(pivot point) (2/Fig. 28); then the laser light localizer must be aligned laterally to the table
or vertically to the VERTIX.
• If the laser light localizer is not straight, slightly loosen the Allen screw (3/Fig. 28) and
turn the laser (4/Fig. 28) accordingly.
3
1
Fig. 29
• Move the stand carriage to approx. 40 mm in front of the mechanical stop at both the left
and right ends.
• Adjust the position of the switch strike plates for the safety limit switches (1/Fig. 29) in the
long holes so that the S1 and S2 switches are positively actuated in these positions.
• Move the column stand up against the left or right mechanical stop.
• Determine the middle of the toothed belt between the column carriage and the
tensioning piece and hook on a spring balance at this location.
• Deflect the toothed belt with a force of 20 N vertically to the toothed belt.
• When this force is applied, the toothed belt should be deflected 20 mm.
• To make a correction, adjust the position of the tensioning pieces (2/Fig. 29) in the
mounting holes.
• If this means of correction is not sufficient, the toothed belt can be moved at the clamping
location (3/Fig. 29).
15
1 m
Fig. 30
Adjusting the Vertical Reference SID of 115 cm (only for ACSS,ACSS P)2
• Use the tape measure in the collimator to set an SID of 109.5 cm to the top surface of the
tabletop (the distance of the film to the top surface of the tabletop is 5.4 cm).
• Select the following routine per the service software:
- Function group: Column
- Adjustment: Adjustment of the Reference SID
- Carry out the adjustment per the service software.
• Completely insert the cassette tray
➪ the set SID will be displayed.
2m
2 00
m
Fig. 31 Fig. 32
Adjusting the Horizontal SID to the Bucky Wall Stand (only for ACSS,
ACSS P) 2
• Position the tube assembly to the Bucky wall stand as shown in Fig. 31.
• Use the tape measure in the collimator to set an SID of 195.9 cm to the front cover on the
Bucky wall stand (the distance of the film to the top surface of the front cover is 4.1 cm).
• Select the following routine per the service software:
- Function group: Column
- Adjustment: Adjustment of the SID to the Bucky wall stand
- Enter 200 at the Service PC for the horizontal SId and confirm the adjustment with OK.
• If necessary, correct the 2.0 m SID adhesive label and the marker line (from the
accessory pack) as shown in Fig. 32.
R2 poten.
Fig. 33
Fig. 34 Fig. 35
4
1
2 3
2
Fig. 36 Fig. 37
• After carefully removing the safety plate and the bolt (2/ Fig. 36), remove the support
cable on the counterweight (1/ Fig. 36).
• Remove the screw (1/ Fig. 37) and the safety plate.
• If the support arm is secured with a wooden block or with another object, pull the anti-
drop brake lever (2/ Fig. 37) in the direction of the compression spring and bolt (3/ Fig.
37) forward.
• Disconnect the cable (4/ Fig. 37) from the anti-drop brake.
Fig. 38
• Remove the snap ring with a pair of snap ring pliers at the head piece (1/ Fig. 38) and
remove the cable after removing the two safety screws (2/ Fig. 38).
• Insert a new cable into the head piece and secure it to the anti-drop brake and
counterweight. Reinstall the head piece; when doing this, make sure it is properly
seated.
• Reinstall the anti-drop brake lever on the support arm. Only slightly lift the support arm
again; when doing this, release the anti-drop brake by pulling on the support arm in the
head-end direction or by pressing the anti-drop brake against the compression spring
and slowly releasing the support arm.
• Remove the safety screws (1/ Fig. 34) on the counterweight.
1
2
Fig. 1
Patient Table 3
2
1
2
2
1 1
Fig. 2 Fig. 3
• Adjusting the 4 eccentric bearings (1/Fig. 3) for longitudinal movement of the tabletop.
- Remove the cover panel (1/Fig. 2) with the expansion spring (2/Fig. 2).
- Remove the rubber bumper with set screws and nuts (3,4,5/Fig. 3).
- Move the tabletop back so that the eccentric bearings (2/Fig. 3) can be adjusted.
- Loosen the Allen-head set screw (1/Fig. 3) (2mm Allen wrench).
- Adjust the eccentric bearing (2/Fig. 3) so that can just still be turned by hand.
- Retighten the Allen-head set screw (1/Fig. 3).
• Carry out this adjustment for all 4 eccentric bearings.
• Check for easy movement of the tabletop over the entire movement path.
NOTICE The eccentric bearings have been optimally adjusted if the eccen-
tric bearings can just still be stopped when the tabletop is moved.
Fig. 4
• Adjusting the switch strike plate (1/Fig. 4) for the S111 safety limit switch:
- Move the Bucky up against the end stop (2/Fig. 4).
- Adjust the strike plate (1/Fig. 4) so that the safety limit switch (3/Fig. 4)
is positively actuated approx. 1 cm in front of the Bucky end position.
• Adjust the strike plate for the S112 safety limit switch to correspond to the S111 safety
limit switch.
R11 Potentiometer 3
320 mm
1
3 22 mm
2
F = 2N
Fig. 5
NOTICE Remove the cover panel (3/Fig. 5) and check whether there is a
red adjustment screw for adjusting the toothed belt tension.
• Adjusting the toothed belt tension (1/Fig. 5) for the Bucky drive:
- Move the Bucky (1/Fig. 5) all the way to the right (arrow/Fig. 5).
- Loosen the 2 countersunk Allen screws (2/Fig. 5) approx. 1 turn.
- Press the holder with the toothed belt gear in the direction of the arrow, Fig.5
or
if there is an adjustment screw, use it to adjust the tension.
- Adjust the toothed belt tension as indicated in Fig.5: F = 2N
3
1
2
+1
m
m
Fig. 6 Fig. 7
• If there is a malfunction, the lift drive can be moved manually (see Fig. 6).
The wrench attachment needed for this is located underneath the lift motor.
S10 / S11 Safety Limit Switch for the Lift Drive (with toothed belt) 3
2 6 7
4
1
3
Fig. 8
• Use the lift drive to carefully position the table so that the switch ring (1/Fig.8) dips into
the hole.
• Adjust the switch ring (1/Fig.8) so that it is centered in the hole:
- To do this, shift the position of the bracket (2/Fig.8) and the assembly (3/Fig.8).
• Adjust a play of 0.1, +0.1 mm between the pinion rack (4/Fig.8) and the pinion (5/Fig.8):
- Loosen the 2 screws (6/Fig.8) approx. 1 turn.
- Insert a piece of paper that is 0.1, +0.1 mm thick (7/Fig.8) between the plastic guide
and the pinion rack (see detailed enlargement, Fig.8).
- Press the plastic guide against the pinion rack and when doing so, retighten the 2
screws (6/Fig.8).
- Remove the piece of paper.
• Use the lift drive to raise the table the complete distance and when doing so, check the
function of the pinion rack.
2, 4
+1
m m
2
+1
m
2 m
Fig. 9
Adjusting the S10 / S11 Safety Limit Switches for the Lift Drive (with pinion
rack) 3
NOTICE Beginning with service software VB01A, table lift (software stop)
at the top and bottom can each be limited by 20 mm in the SSW.
NOTICE If the safety limit switches can be actuated when table lift is initi-
ated, the cause could be the factory settings or errors in measure-
ment. For exact adjustment, proceed according to the two
Service Instructions, AX067/99/I and AX090/00/I.
Fig. 10
Bucky 3
General Adjustments 3
Wiring Diagram G051G is the reference document for configuration of the Dip switches
and jumpers:
S3 DIP switch:
S3.1 = OFF i.e. Automatic mode: specification of the advance time for the grid is
made in the service software. Also select this position for the manual
MULTIX versions because here, too, the advance time is set in the
service software.
NOTICE Allocation of the grid speed to the type of grid used is critical for
quality exposures.
If the grid type is changed (even later), a change of the grid speed
is absolutely necessary.
The MANUAL MODE programming is not relevant in this system. In this system, program-
ming is always made in the software.
Carry out a check for every application unit with a moveable grid to determine at which
exposure time, tg, grid stripes are no longer visible on the film.
HINWEIS Werkseinstellung:
Vorhaltezeit (Default) ist 34 ms für POLYDOROS SX/LX/Lite
Wird die Anlage mit einem POLYDOROS IT geliefert ist dieser
Wert (als Ausgangsbasis) auf 0 ms zu ändern.
Procedure:
• At the generator, select small focus, 80%, 40 kV and middle IONTOMAT chamber.
• Set an SID of 115 cm.
• Insert a 18 x 24 cm cassette without film.
• Trigger exposure beginning with 40 kV until the exposure times of approx. 3 ms result
(increasing the kV value -> exposure time becomes briefer).
• Insert a 18 x 24 cm cassette with film.
• Trigger the exposure with the previously determined kV value.
• Using a magnifier at the light box, determined at which switch time grid stripes are no
longer visible on the film.
• If the grid advance time must be changed, carry out the test setup and the measurement
as shown in Fig. 37. The generator delay time can be changed using the Service PC as
described in the chapter “Programming the System”.
- Grid stripes at exposure times of ≥ 3 ms are inadmissible.
The relationship between advance time and the grid movement and thus the problem of
grid stripes is illustrated in the following graphic.
Exposure!
Programmed generator delay time
corresponds to the actual measured
Grid path delay time of the generator being used
Turning point
34 ms
Middle position
Grid start
Park position
t
Delay time in the SSW 180 ms 90 ms 0 ms
Channel 1 Channel 2
Radiation enable
Generator delay time
U1 measured
HW signal
radiation enable Physical occurrence of
radiation
Channel 1: Test points on backplane D54 X109 (U1) and X110 (ground)
Channel 2: Radiation measured with measuring chamber, Part No.: 96 60 754 RE999
Fig. 11
Top diagram (Fig. 11): temporal progression of grid movement with the time at which radi-
ation is triggered.
Bottom diagram (Fig. 11): Enable signal relative to grid movement.
Example:
Very brief exposure times and slow grid speed (grid type 17/70) require reducing the
advance time to almost 0 ms to place the grid focused to the film.
Prior to checking the electrical function of the safety circuit, it is recommended that the
mechanical adjustment of all limit and safety limit switches be completed.
The function of the safety circuit is checked by pressing all switches in the circuit. When
the safety circuit is interrupted, the k2 safety breaker in the table base is deenergized and
the red control lamp on the table goes on.
• Press the following switches, one after the other, and check the function of the safety
circuit.
- Lift sensing S10 und S12
- Emergency stop switch
• With the tomo version, also check the following switches. In addition, select any tomo
program at the control console.
- Bucky longitudinal drive S111 and S112 (for this remove the Bucky cover
panel)
- Column carriage. long. drive S1 and S2 (switches are located under the column
stand)
NOTICE When the S1 key is pressed on the D54, the safety circuit is
jumpered.
Tabletop Brakes 3
3
2
1
2
1 1
Fig. 12 Fig. 13
NOTICE Prerequisite for correct adjustment of the air gap is that the bear-
ings for longitudinal tabletop movement (see patient table) be
exactly adjusted.
• Remove the stops on the tabletop (1/2/3 Fig.13) and move the tabletop to the side.
• Use a spacer gauge to check the air gap between the piston brake and the rail.
The air gap should be 0.3, + 0.2 mm.
• To make any correction, remove the side cover panel.
• Loosen the mounting screws (1/Fig.12) on the mounting bracket (2/Fig.12) for the piston
brakes (3/Fig.12) and correct the position of the piston brakes inside the holes.
The kick switch for the table longitudinal-transverse brake can be configured as a "dead-
man’s switch" or as a sensor.
• To do this, select the following routine per the service software:
- Adjustment - Calibration
- Functiongroup: HOST Table
- Adjustment: Adjustment of the table sensor method
In discussion with the customer, it is possible to adjust the table lift individually (adaptation
to the existing bed height, etc.). This position (working position) can be moved into auto-
matically from down or from up.
To do this, select the following routine per the service software:
- Adjustment-Calibration
- Function group: Host table
- Adjustment: Adjustment of customer-specific lift stop position.
Fig. 14
• Move the ceiling stand into the longitudinal and transverse stops.
• The LED´s in the operating keys are lit, i.e. the ceiling stand is located in the centered
position to the middle of the cassette (check using the line light localizer!)
• To do this, select the following routine per the service software
- Function groups: Stand
- Adjustment: Adjustment of stand - table middle (longitudinal/transverse)
- Carry out the adjustment of the stand - table middle per the service software.
1. Confirm the mask with "OK" if the longitudinal/transverse potentiometers are
adjusted
2. Confirm the mask "Enter" after carrying out the steps.
Test Conditions 3
• Free adjustment data so that an optical density that can be evaluated well is achieved.
• Small focus and fine-resolution screen.
• Tomo height approx. 115 mm (depending on the tomo test phantom).
• Make an exposure for each: 40°; 2 sec and 20°; 0.6 sec.
• Compare the exposure quality by comparing the tomo exposures with the "Bucky
exposure" made under the same conditions.
• Move the unit into the tomo position, center the tomo block in the beam path and
collimate to the tomo block.
• Measure the notch height of the lead strip tests and record the results, set the tomo
height to the measured value.
• Trigger the tomo exposures and develop the film (18cm x 24cm format).
• Make a Bucky exposure under the same conditions.
• Evaluate the exposures:
- Accuracy of the tomo height display
- Visual resolution (comparison with the Bucky exposure)
- Form and movement of the tomo figure
- Evenness of the blur
Anode side
20°
2mm
115, ± 1 mm
SHa
Fig. 15
The test phantom is comprised of a hard foam block with two slots and two lead strip
tests.
The lead strip tests are inserted into the slots of the hard foam test phantom so that:
- the notches are pointing towards the outside edge
- the lead strip tests do not overlap
- the two notches are positioned exactly at the same height over the tabletop.
The test phantom is positioned on the tabletop so that:
- one line test is on the anode side
- the notch of the other line test is turned towards the tomo height line light localizer.
- one axis of the test phantom is parallel to the longitudinal table axis.
The holes of the line test that are positioned at a 20º oblique angle all have a height differ-
ence of 2 mm to each other. This makes it possible to exactly determine the tom height
that appears on the exposure.
Holes, Notch
- Tomo height (tomo plane)
- Tomo figure
- Evenness of the tomo movement
(synchronous movement, smoothness of move-
ment)
- Dose distribution over the exposure
- Parallelism of focus and film path
Edge Appearance
- Continuity of blur
(even or uneven tomo movement)
Evaluation 3
The sharply focused hole of the focus test is in the tomo plane.
If the hole by the notch appears sharply focused, the numerically selected tomo height
SHa matches the actual tomo height SH; i.e.,. ∆ SHa = 0.
• If there are differences, determine the height difference ∆ SHa in mm between the
sharply focused hole and notch; there is a height difference of 2 mm from hole to hole.
If the actual tomo height is above the notch, ∆ SHa is positive.
If the actual tomo height is below the notch, ∆ SHa is negative.
The tomo height can be corrected using the S1 Dip switch on the D50 host board accord-
ing to the following table.
S1
8 Sign in front for tomo height correction: ON=positive / OFF=negative
7 23 on on on on on on on on off off off off off off off off
3 n.a. 0 1 2 3 4 5 6 7 8 9 10 11 12 13 14 15 mm
The evaluation of the resolution of tomo exposures is made by comparing them with a
Bucky exposure made under comparable conditions. This way, all influences which deter-
mine the resolution of a radiographic exposure (focal point, screen type, film, geometric
conditions, exposure data, etc.) can be ruled out and it is possible to make a clear state-
ment about the quality of the tomo device.
• Determine the maximum optical resolution of the tomo exposures (magnifier min. 6 X).
• Record the Lp/mm which are visible on both line tests in the tomo plane.
• Determine the maximum optical resolution of the Bucky exposure.
The difference between the maximum optical resolution of the Bucky exposure and the
resolution of the tomo exposures may not be more than one line group. The resolution for
With a linear layer, the blur lines of the holes to the left and right of the notch must be par-
allel; if they are not, the focus and film paths are not parallel. This also results in reduced
resolution in the line test longitudinal to the tomo device.
If the blur patterns of the holes are wavy (jumpy, snake-like), there are unsynchronous
oscillations in the tomo system (loose or sticking locations, etc.). This also results in the
reduced resolution.
The blur patterns of the holes should have about the same density along the entire length.
If there are dotted blur patterns there is jumpy tomo movement (usually the catapult
Bucky). If there is a continuous density change of the blur patterns, there is changing
movement speed or changing dose rate.
Jumpy tomo movement can also be recognized because edges of the line test plates do
not have a continuous change in density but are in "steps".
The blur patters of the holes of the line tests show the form and movement of the tomo fig-
ure. If this is not sufficient in a single case for evaluation, a pin-hole exposure should also
be made for the particular tomo figure..
good Linear layer
satisfactory
Linear layer
NOTICE If the linear layer is poor, check the accuracy of the column move-
ment and, if necessary, carry out an adjustment of the guide rail
as described in Chapter 2.
satisfactory
unsatisfactory
The tomo exposures should be as symmetrical as possible to the point of the Bucky expo-
sure.
• For corrections, select the following routine per the SSW:
Function groups: Host table
Adjustments: Adjustment of the tomo mode lead time ( tomo time 0.4 sec)
Adjustment of the lead time, tomo mode 1 ( tomo time 0.6 sec)
Adjustment of the lead time, tomo mode 2 ( tomo time 0.8 sec)
Adjustment of the lead time, tomo mode 3 ( tomo time 1.2 sec)
Adjustment of the lead time, tomo mode 4 ( tomo time 2.0 sec)
HINWEIS Werkseinstellung:
Vorhaltezeit (Default) ist 90 ms für POLYDOROS SX/LX/Lite.
Wird die Anlage mit einem POLYDOROS IT geliefert ist dieser
Wert (als Ausgangsbasis ) auf 0 ms zu ändern.
Sind Schichtaufnahmen unsymmetrisch können diese durch Kor-
rekturen verbessert werden.
Fig. 16
• Determine the difference between the laser mark versus the display on the support arm;
to do this, use for example the tomo block.
• To make the correction, remove the front mirror; to do this, remove the 4 mounting
screws on the back of the tomo height light localizer.
• Move the S1 switch (1/Fig. 16) in the direction of the difference determined; to do this,
loosen the 2 mounting screws approx. one half turn.
• Carefully retighten the mounting screws (M2 thread!!)
• Check the adjustment; to do this, briefly switch off the tomo height light localizer
(disconnect the connector).
• Reinstall the front mirror.
Fig. 1
• Remove the cover panel (1/Fig. 1) by removing the two screws (2/Fig. 1).
• The R1 potentiometer is located at the back of the vertical carriage (7/Fig.2).
• The vertical carriage on the Bucky wall stand must be in the reference position. To do
this, move the vertical carriage to the top stop (if the R1 potentiometer is defective, it
must be replaced in this position).
• To do this, select the following routine per the service software:
- Adjustment – Calibration
- Function group: Bucky wall stand
- Adjustment: Adjustment of the Bucky W.S. height potentiometer
• Disengage the potentiometer shaft from the toothed belt pinion by loosening the set
screw on the pinion.
• Set a zero adjustment by turning the potentiometer shaft.
• When the value is within tolerance (+/- 2), a signal sounds on the Service PC.
2
3
4 1
7
6
Fig. 2
Adjusting the Stop for the 0° Bucky Position (only VERTIX TOP) 4
• Check the play in the S2 0o position and of the S3 90o position switches in the stop
positions.
• The particular switch must positively actuate in the particular position.
• An adjustment of the switch-point is possible by bending the switch strike levers
(3/Fig.2).
• The response threshold of the tilt brake is adjusted by adjusting the tension of the
Bowden cable using the adjustment screw (5/Fig.2).
2
1
Fig. 3
• To replace the complete vertical brake (1/Fig. 3), the vertical carriage must be moved up
and out of the column stand.
• Release the brake and while in this condition, secure it with cable ties (2/Fig. 3).
• Caution: the brake no longer works; secure the vertical carriage!
• Open the cable clamp (3/Fig. 3) on the brake lever and replace the cable.
1 2
Fig. 4 Fig. 5
• Turn the set screw (1/Fig. 4) so that the brake just begins to brake, then turn the set
screw approx. 3/4 to one turn to the left (counterclockwise).
• The braking force should correspond to approx. 250-300 N.
• Counter the position of the set screw using a hex nut (1/Fig. 5); for this, use a 10’ open-
end wrench as shown in (2/Fig. 4).
Fig. 6 Fig. 7
1 2
Fig. 8 Fig. 9
1 2
Fig. 10 Fig. 11
• Use a pencil to mark the position of the front panel (to make it easier to center it when it is
installed later).
• Remove the bottom mounting screws on the front panel (1/Fig. 10) and remove the
toothed washers.
• Remove the top mounting screws on the front panel (1/Fig. 11) and take off the front
panel.
1 2
Fig. 12
• Remove the hex screw (2/Fig. 12) a remove the side cover panel (1/Fig. 12).
• Remove the cover panel (4/Fig. 12). To do this, remove the 4 screws (3/Fig. 12).
• For the following work steps, a 2nd person will be needed for a brief while or a suitable
object (chair, step stool) must be positioned in front of the VERTIX.
• Secure the Bucky against falling and remove the mounting screws (2/Fig.12).
• Remove the Bucky and hold it or place it down on the object placed in front of it.
• Add or remove additional weights (1/Fig. 8) in accordance with the previously
determined forces. To do this, loosen the bracket (2/Fig. 8) and add or remove weight(s).
• Reinstall the Bucky in the reverse order of the above instructions.
• When installing the front panel, align it with the marks.
• Check the function of the stop movements and pay special attention to sufficient spacing
between the grid and the front panel.
Fig. 13
• The support cable must be replaced after 3 years operating time or when:
12 or more broken single strands of the cable over 100 mm can be found, or 3 or more
broken single strands of the cable can be found over 10 mm.
• Even if only one single strand is broken, the operator must be informed that continued
use of the VERTIX poses a safety risk to patient and operator. The VERTIX may not be
used until the defective cable has been replaced.
• In both versions, the Vertix may not be removed to replace the cable.
- With the VERTIX PRO, the catapult Bucky and the support plate must be remove to
replace the cable.
- With the VERTIX TOP, the cable clamp is accessible when the catapult Bucky is tilted
into the horizontal position and the cover panel on the support plate is removed.
• Remove the cover panels on the VERTIX. Move the counterweight until the threaded
hole in the counterweight is lined up with the hole in the VERTIX (see also the Installation
Instructions). Secure the counterweight with an M12 screw and appropriate U-washer.
• Raise the catapult Bucky so that the load is released on the cable, and with the catapult
Bucky in this position, place a wooden block or a similar object under it to support it.
• Loosen the cable at the Bucky support. Loosen the components (Fig. 13) in the
sequence listed and pull up the mount with the cable pulley. The cable mount on the
counterweight will now be accessible.
• Remove the cable from the counterweight and replace it.
• Reinstall the mount with the cable pulley. Secure the cable on the Bucky support (make
sure that the U-plates, the snap ring on the two mounting points and the cable securing
device are reinstalled correctly).
• Remove the wooden block and remove the safety screw from the counterweight.
• Check for easy function and reinstall the cover panels.
Bucky 4
General Adjustments 4
Wiring Diagram G051G is the reference document for the configuration of Dip switches
and jumpers:
Jumper X4.1-2 "END NOTE" must be set if the Bucky is installed in the VERTIX.
Jumper X4.2-3 "NO END NOTE" must be set if the Bucky is installed in the MULTIX.
S3 DIP switch:
S3.1 = OFF i.e. Automatic mode: the setting the for the grid advance time is made
in the service software. This setting must also be set for manual
MULTIX versions because here, too, the advance time is set in the
service software.
NOTICE Allocation of the grid speed to the type of grid being used is basic
for quality exposures.
If the type of grid is changed (even later), a change of the grid
speed is absolutely necessary.
MANUAL MODE programming is not relevant in this system. Programming in this system
is always made in the service software.
A check must be carried out for every application unit with a moveable grid to determine at
which exposure time tg grid stripes can no longer be seen on the film.
Procedure:
• At the generator, select small focus, 80%, 40 kV and middle IONTOMAT chamber.
• Set a SID of 115 cm.
• Insert a 18 x 24 cm cassette without film.
• Trigger exposure beginning with 40 kV until exposure times of approx. 3 ms are obtained
(increasing the kV value -> exposure time becomes briefer).
• Insert a 18 x 24 cm cassette with film.
• Trigger the exposure with the previously determined kV value.
• Using a magnifier at the light box, determine at which switch time the grid stripes can no
longer be seen on the film.
• If the grid advance time needs to be changed, carry out the test setup and the
measurement as shown in Fig. 14. The generator delay time can be changed using the
Service PC as described in the chapter “Programming the System”.
- Grid stripes with exposure times of ≥ 3 ms are inadmissible.
The relationship between advance time and grid movement, and thus the problem of grid
stripes is illustrated in the following graphic.
Exposure!
The programmed generator delay
time corresponds to the actual measured
Grid path delay time of the generator being used
Turning point
34 ms
MIddle position
Grid start
Park position
t
Delay time in the SSW 180 ms 90 ms 0 ms
Channel 1 Channel 2
Radiation enable
Generator delay time
U1 measured
HW signal
radiation enable Physical occurrence of
radiation
Channel 1: Test points on backplane D54 X109 (U1) and X110 (ground)
Channel 2: Measure radiation using measuring chamber, Part No.: 96 60 754 RE999.
Fig. 14
Top diagram (Fig. 14): temporal progression of grid movement with the time point at which
radiation is triggered.
Bottom diagram (Fig. 14): enable signal relative to grid movement.
Example:
Very brief exposure times and slow grid speed (grid type 17/70) require reducing the
advance time to almost 0 ms to place this grid focused to the film.
TD PS 21 / Conrad/Tropia