RIGOL
Data Sheet
                                              DS1000E, DS1000D Series
                                                   Digital Oscilloscopes
                                               DS1102E, DS1052E, DS1102D, DS1052D
Product Overview
The DS1000E, DS1000D series instruments are
economical, high-performance digital oscilloscopes.
The DS1000E series scopes feature dual channels
and 1 external trigger channel.
The DS1000D series scopes feature dual channels
and 1 external trigger channel as well as a 16
channel logic analyzer.
Applications                                          Easy to Use Design
   Electronic Circuit Test                              Built-in help menu makes the features
   Circuit Functional Test                               more convenient
   Logical Signal Comparison                            Multiple Language User Interface, support
   Component R & D or Failure Analysis                   Chinese & English input
   Mixed Signal Circuit Test                            Supports USB disk and local files storage
   Education & Training                                 Adjustable waveform intensity
                                                         One-Button signal display with AUTO key
                                                         Easy to read and access menus
                                                         Powerful UltraScope PC application software
Main Features                                            Built-in FFT function
   Dual analog channels and 16 channels logic           Adjustable digital filter settings
    analyzer, 100MHz maximum bandwidth,                  High resolution delayed scan function
    1GSa/s maximum real-time Sampling rate and           Pass/Fail detection function enables
    25GSa/s maximum equivalent Sample rate                isolated output for production applications
   5.6 inch, 64 k TFT color LCD makes the               Math operations available to multiple waves
    waveform display clear and vivid                     Standard configuration interface: USB
   Variety of trigger types: Edge, Pulse Width,          Device, USB Host , RS-232 and support USB
    Video, Slope, Alternate, Pattern and                  disk storage and PictBridge printer standards
    Duration                                             “Key Lock” function meets the needs of
   Unique, adjustable trigger sensitivity enables        industrial production
     triggering in high noise signal environments        Support for remote command control using
   Able to measure 20 types of wave                      VISA drivers via USB or RS-232 and standard
    parameters and track measurements with                SCPI instruction set
    cursors automatically
   Unique waveform record and replay
                                         2010 Update
                                  RIGOL Technologies USA, Inc.
    Automatically Measure 20 Wave Parameters                                    Cursor Measure
                              DS1000E & DS1000D series oscilloscopes
                              provide 20 types of wave parameters for
                              automatic measurements, which contain
                              10 voltage and 10 time parameters.
                              In cursor mode, users can easily
                              measure and track signals. Additionally,
Automatic measure             3 types of cursor measurements are               FFT cursor measure
                              available: Manual, Track and Auto.
   Multiple Trigger
                            Both DS1000E and DS1000D series scopes feature multiple trigger modes:
                             Edge trigger, Pulse Width trigger, Video trigger, Slope trigger,
                             Alternate trigger, Pattern trigger (DS1000D), Duration trigger
                             (DS1000D)
                            The Duration trigger is a new trigger type combining pattern and pulse width
                            triggering. The adjustable trigger sensitivity is ideal for ignoring
                            noise in signals that would otherwise cause false triggering.
   Pattern trigger
   16 Channel Logic Analyzer
Equipped with a 16 channel logic analyzer, the DS1000D series mixed
signal oscilloscopes achieve mixed signal measurements coordinated
with the 2 analog channels.
Each channel can be turned on or off independently, or in groups of
                                                                          Digital channels setup
8 (D7-D0 and D15-D8); also, you can adjust waveform size and threshold
types or change the display positions on screen for the digital channels.
    Waveform Recording
  With the waveform recording function on the DS1000E and DS1000D
series oscilloscopes, not only the outputs from the two channels can be
recorded, but also the output of the Pass/Fail test. In total, up to 1000
frames of waves can be recorded. Additionally, users can view waves            Waveform recording
and independently recall or save specific waveforms for further analysis.
     Pass/Fail Testing
The Pass/Fail function monitors changes in a signal by comparing
whether the input signal is within the pre-defined mask. The test results
can be displayed on screen, output by isolated pass/fail port, and/or set to
generate an alarm via the system speaker.                                        Pass/Fail testing
                              UltraScope Software
                            RIGOL provides UltraScope, a powerful PC software application, which
                            enables users to: Capture and measure waves; Perform local or remote
                            operations; Save waves in “.bmp” format; Save files as “.txt” or “.xls”
                            format; Print waveforms.
Measurement window
                                                                                                      1
Specifications
All specifications apply to DS1000E, DS1000D Series Oscilloscopes unless where noted. To come up to
these specifications, two conditions must be met:
   The instrument must have been operated continuously for 30 minutes under the specified
    operating temperature.
   Perform Self-Calibration operation through the Utility menu if the range of operating
    temperature variations up to or more than 5°C.
NOTE: All specifications are guaranteed except where marked “typical”.
Specifications
Bandwidth
DS1102E                     DS1052E                 DS1102D                   DS1052D
100MHz                      50MHz                   100MHz                    50MHz
Acquisition
Sample Modes                Real-Time Sample        Equivalent Sample
                                                    DS1102X                    DS1052X
Sample Rate                 1GSa/s, 200MSa/s[1]
                                                    25GSa/s                    10GSa/s
                            The waveform will be displayed one time while all the channels finish N
Averages
                            times Sample, N could be selectable from 2, 4, 8, 16, 32, 64, 128 and 256
Inputs
Input Coupling                             DC, AC, GND
Input Impedance                            1MΩ± 2%, the input capacity is 18pF± 3pF
Probe Attenuation Factors                  1X, 5X, 10X, 50X, 100X, 500X,1000X
                                           400V (DC+AC Peak, 1MΩ input impedance)
Maximum Input Voltage
                                           40V (DC+AC Peak) [1]
Time Delay between Channel (typical)       500ps
Horizontal
                               Real-Time: 13.65Sa/s-1GSa/s
Sample Rate Range
                               Equivalent: 13.65Sa/s-25GSa/s
Waveform Interpolation         Sin(x)/x
                               Channel                          Memory Depth    Memory Depth
                                              Sample rate
                               Mode                             (normal)          (long memory)
                               Single
                                              1GSa/s            16kpts          N.A.
                               channel
                               Single         500MSa/s
Record Length                                                   16kpts          1Mpts
                               channel        or lower
                               Dual           500MSa/s
                                                                8kpts           N.A.
                               channel        or lower
                               Dual           250MSa/s
                                                                8kpts           512kpts
                               channel        or lower
                               2ns/div~50s/div, DS1102X
Scanning Speed Range
                               5ns/div~50s/div, DS1052X
(Sec/div)
                               1-2-5 Sequence
Sample Rate and Delay
                               ± 50ppm (any interval ≥1ms)
Time Accuracy
Delta Time
                               Single: ± (1 Sample interval + 50ppm × reading + 0.6 ns)
Measurement Accuracy
                               >16 averages: ± (1Sample interval + 50ppm × reading + 0.4 ns)
(Full Bandwidth)
Vertical
                                                                                                   2
A/D Converter                      8-bit resolution, all channels sample simultaneously[2]
Volts/div Range                    2mV/div~10V/div (at the input terminal connecting to BNC)
                                   Maximum input voltage on analog channel
                                   CAT I 300Vrms, 1000Vpk; instantaneous overvoltage 1000Vpk
                                   CAT II 100Vrms, 1000Vpk
Maximum Input                      RP2200 10:1:CAT II 300Vrms
                                   RP3200 10:1:CAT II 300Vrms
                                   RP3300 10:1:CAT II 300Vrms
                                   ± 40V (250mV/div~10V/div)
Offset Range
                                   ± 2V (2mV/div~245mV/div)
                                   100MHz (DS1102D,DS1102E)
Analog Bandwidth
                                   50MHz (DS1052D, DS1052E)
                                   100MHz (DS1102D, DS1102E)
Single-shot Bandwidth
                                   50MHz (DS1052D, DS1052E)
Selectable Analog
                                   20MHz
Bandwidth Limit (typical)
Lower Frequency
                                   ≤5Hz (at input BNC)
Response (AC –3dB)
Rise Time at BNC (typical)         <3.5ns, <7ns, respectively at 100MHz, 50MHz
                                   2mV/div-5mV/div:
                                   ± 4% (In Normal or Average acquisition mode)
DC Gain Accuracy
                                   10mV/div-10V/div:
                                   ± 3% (In Normal or Average acquisition mode)
                                   When vertical displacement is zero, and N ≥16:
                                   ± (DC Gain Accuracy× reading+0.1div+1mV)
                                   When vertical displacement is not zero, and N ≥16:
DC Measurement                     ± [DC Gain Accuracy× (reading+vertical displacement)+(1% of
Accuracy, Average                  vertical displacement) + 0.2div]
Acquisition Mode                   When vertical scale is between 2mV/div and 245mV/div, add 2mV
                                   more for setting value.
                                   When vertical scale is between 250mV/div and 10V/div, add 50mV
                                   more for setting value.
                                   Under same setting and condition, the voltage difference (△ V)
Delta Volts
                                   between any two points in the waves coming from the average of
Measurement Accuracy
                                   more than 16 waves have been acquired: ± (DC Gain
(Average Acquisition Mode)
                                   Accuracy× reading + 0.05 div)
Trigger
Trigger Sensitivity                0.1div~1.0div (adjustable)
                                   Internal     ± 6 divisions from center of screen
Trigger Level Range
                                   EXT          ± 1.2V
Trigger Level Accuracy (typical)   Internal     ± (0.3div × V/div)(± 4 divisions from center of screen)
applicable for the signal of
                                   EXT           ± (6% of setting + 200 mV)
rising and falling time ≥20ns
                                   In Normal mode: pre-trigger (memory depth/ 2*Sample rate),
Trigger Offset                     delayed trigger 1s
                                   In Slow Scan mode: pre-trigger 6div, delayed trigger 6div
Trigger Holdoff Range              500ns~1.5s
Set Level to 50% (typical)         When input signal frequency ≥50Hz
Edge Trigger
Edge trigger slope                 Rising, Falling, Rising + Falling
Pulse WidthTrigger
Trigger Condition                  (>, <, =) Positive pulse width, (>, <, =) Negative pulse width
Pulse Width Range                  20ns ~10s
                                                                                                      3
Video Trigger
Video Standard                       Support for standard NTSC, PAL and SECAM broadcast systems. Line
Line Frequency                       number range: 1~525 (NTSC) and 1~625 (PAL/SECAM)
Slope Trigger
Trigger Condition                    (>, <, =) Positive slope, (>, <, =) Negative slope
Time Setting                         20ns~10s
Alternate Trigger
Trigger on CH1                       Edge, Pulse Width, Video, Slope
Trigger on CH2                       Edge, Pulse Width, Video, Slope
Pattern Trigger[1]
Pattern Type                         D0~D15 select H, L, X,         ,
Duration Trigger[1]
Pattern Type                         D0~D15 select H, L, X
Qualifier                            >, <, =
Time Setting                         20ns~10s
Measurements
                                   Voltage difference between cursors (∆V)
                       Manual      Time difference between cursors (∆T)
                                   Reciprocal of ∆T in Hertz (1/∆T)
Cursor                             Voltage value for Y-axis waveform
                       Track
                                   Time value for X-axis waveform
                       Auto        Cursors are visible when measure automatically
                       Vpp, Vamp, Vmax, Vmin, Vtop, Vbase, Vavg, Vrms, Overshoot, Preshoot, Freq,
Auto Measure           Period, Rise Time, Fall Time, +Width, -Width, +Duty, -Duty, Delay1→2 ,
                       Delay1→2
Remarks:
[1] For DS1000D Series;
[2] Only one channel is available when the Sample rate is 1GSa/s.
                                                                                                    4
General Specifications
Display
Display Type                     5.7inch. (145mm) diagonal TFT Liquid Crystal Display
Display Resolution               320 horizontal × RGB× 234 vertical pixels
Display Color                    64k color
Display Contrast (typical)       150:1
Backlight Brightness (typical)   300 nit
Probe Compensator Output
Output Voltage (typical)         Approximately 3Vpp (peak to peak value)
Frequency (typical)              1kHz
Power Supply
Supply Voltage                   100 ~ 240VACRMS, 45~440Hz, CAT II
Power Consumption                Less than 50W
Fuse                             2A, T level, 250 V
Environmental
                                 Operating 10℃~ 40℃
Ambient Temperature
                                 Non-operating -20℃~ +60℃
Cooling Method                   forced cooling by fan
                                 below +35℃: ≤90% relative humidity
Humidity
                                 +35℃~ +40℃: ≤60% relative humidity
                                 Operating at 3,000 m or below
Altitude
                                 Non-operating at 15,000 m or below
Mechanical
                                 Width                              303mm
Dimensions                       Height                             154mm
                                 Depth                              133mm
                                 Without package                    2.3kg
Weight
                                 Packaged                           3.5kg
IP Protection
IP2X
Calibration Interval
The recommended calibration interval is one year
                                                                                        5
Ordering Information                                Warranty
Name of Product                                     Thank you for choosing RIGOL products!
RIGOL DS1000E, DS1000D series oscilloscopes         RIGOL Technologies, Inc. warrants that this
Standard Accessories                                product will be free from defects in materials and
                                                    workmanship from the date of shipment. If a
   Probe× 2 (1.5m), (1:1 or 10:1 adjustable)       product proved defective within the respective
    Passive Probes                                  period, RIGOL will provide repair or replace
   Power Cord                                      as described in the complete warranty statement.
   USB Cable
   A Data Cable (DS1000D series)                   For the copy of complete warranty statement or other
   An Active Logic Head (DS1000D series)           maintenance information, please contact us directly
   20 Logic Testing Clips (DS1000D series)         at our local RIGOL sales and service office.
   20 Logic Testing leads (DS1000D series)
   A CD-ROM (including User’s Guide and
                                                    RIGOL does not provide any other warranties
    Application Software)
                                                    except the one being provided by this summary
   A Quick Guide
                                                    and the warranty statement.
Optional Accessories
   BNC Cable
   RS232 Cable
   USB-GPIB Adapter
   DS1000E, DS1000D soft carrying case
Contact Us
If you have any questions, concerns, or thoughts about our products, documentation, or services,
please contact RIGOL Technologies USA, Inc. or your local distributor or representative.
To contact our US office:
Tel: (440) 232-4488
Toll Free: 877-4-RIGOL-1
Service & Support
8:30 am –5: 00 pm Eastern Time, Monday to Friday
Or by e-mail at:
info@rigol.com
Or mail to:
RIGOL Technologies USA, Inc.
7401 First Place Suite N
Oakwood Village, OH 44146
Outside of North America: Contact the local RIGOL distributors or sales office.
For the latest product information and service, visit our website: http://www.rigolna.com