PDMM01
PYLE®
Digital Manometer
WWW.PYLEAUDIO.COM
Introduction
The device measures Gauge/Differential Pressure from -55 H 2 0 to +55 H 2 0
(+2 psi) . It features 11 selectableunitsofmeasure:in H 2 0, psi, bar, mbar, Kpa,
InHg, mmHg, ozin 2 , FtH 2 0, cmH 2 0, kgcm,Additional features include Data
Hold and Auto Power Off disabled. Careful use of this meter will provide years
of reliable service.
5
Specifications
Function Range Resolution Function Range Resolution
inH 2 0 ±55.40 0.01 ozin 2 32.00 0.01
psi ±2 .000 0 .001 FtH 2 0 4 .612 0 .001
Mbar ±137.8 cmH 2 0 140.6 0.1
kPa ±13.79 Kgcm 2 0.140 0 .001
inHg ±0.407 0.001 Bar 0.137 0.001
mmHg ±103.4
Display Dual LCD
Accuracy ±0.3%FSO(25°C)
Repeatability ±0.2%(Max+/-0.5%FSO)
Linearity/Hysteresis ±0.29% FSO
Maximum Pressure 10psi
Response Time 0 .5Seconds typical
Low Battery Indicator Yes
Over Range Indicator Err.1
Under Range Indicator Err.2
Operating Conditions 32°F to 122°F(0 to 50°C)
Storage Conditions 14°F to 140FO(-1 to 60°C)
Power Supply 1X 9V Battery (included) or External 9VDC
Standard Accessories Pouch &.Tubing
Button
The meter has six buttons.
1. UNITS
2. DIF
3. HOLD
3 4. ON/OFF
A. REC
B. BACKLIGHT
1_ --t=1lIOOID
Display Description
1. Hold Function 3
2. Differential Mode D
3. Primary Data Screen
4. Record Mode
5.
6.
7.
8.
MAX/MI N/AVG
Low Battery Indicator
Pressure Unit Indicator
Relative Time Clock
4
5
6 -
w _ •~
~
I
rnlIA\'ii'
IIOOO~
1:1:
~
~ IIIIIlIm' IIIIIlIIIIiMI(g
• !1J!i]{J. CID WHI.CID mn1){J. CID
8
2
Operation
1. Power
Press the ON/OFF button to turn power on or off. The meter will perform
a short self-test when turned on. The meter defaults to the setting used during
the last operation.
2. Zero Adiust and Offset
Before use and without fittings attached to the meter, press the HOLD button
for more than 2 seconds to zero the meter. The meter will reset and display
"000" from right to left. 0 .000 will be display.
If it is desired to offset the meter and display, press the DIF button while
taking a measurement subsequent reading will be the difference between the
current measurement and the measurement that was on the display when the
DIF button was pressed.
3. Data Hold
Press the HOLD button momentarily to freeze the displayed reading the
HOLD icon will appear on the upper left-hand side of the display. press the
HOLD button again to return to normal operation.
4 Record
The MAX/MIN mode allows the user to view only the highest (MAX), lowest
(MIN) or anerage (AVG) readings with relative time stamp.
1 . Press the REe button once and REe appears on the display (All other
Functions are locked out except for Power & Backlight) . The relative time
clock also appears in the secondary display and starts counting .
2 . Press the REe button again and MAX appears on the display. The
Display is now showing the maximum pressure and relative time it was
recorded .
3
3 . Press the REC button to display the minimum MIN reading with the relative
time.
4. Press the REC button to display the average value AVG reading with the
relative time.
5 . Press the REC button again to continue recording MINIMAXI AVG
measu rements.
6. To exit the MINIMAXI AVG mode, press the REC button for 3 seconds to
return to normal operation.
7. Backlight Display
Press the Backlight button to turn on the backlight. It will remain on for
40 seconds.
8. Automatic Power OFF
The meter will automatically shut off after 20 minutes to conserve battery life.
To disable Auto Power OFF, hold down the hold button and turn on the meter,
an "n" will appear in the display indicating that Auto Power OFF mode is
disabled . The meter will return to normal operation when it is turned off.
9. Units
Press the Unit button momentarily and the meter will cycle through the 11
available units of measure as indicated by the cursor on the button of
the display.
Error Codes
An error message will appear on the display if the meter fails an internal
diagnostic test.
1. Err. 1: Pressure value is over the range.
2. Err. 2: Pressure value is below the range.
3. Err. 3: DIF function, the differential pressure value is over range .
4. Err. 4: DIF function, the differential pressure value is below range.
BaHery Replacement
When the battery power falls low, "BAT" will appear on the LCD . Replace
the 9V battery.