NE 1010 510 511 User Manual
NE 1010 510 511 User Manual
www.k echnology.it
         HIGH PRESSURE
   MODEL NE-1010/NE-510/NE-511
NE-1000 FAMILY OF PROGRAMMABLE SYRINGE PUMPS
                       Firmware Version V 3.923
                                                  WARNING
                                                    NOT FOU R
                                                        A L   SE
                                                  CLINIC ANS
                                                  ON HUM
  •   When the display shows               , the pump is set for continuous pumping. Pressing any up arrow key will change
      the display to 0.
  •   For continuous pumping: Set the volume to 0.
  •   For a Volume to be Dispensed: Set the volume. The default units are set according to the syringe diameter.
  Set the Pumping Direction
  •   When the ‘Withdraw’ LED is lit, the pump is set for withdrawing. When not lit, the pump is set for infusing. Use the
      ‘     ’ key to change the pumping direction.
  Load the Syringe
  •   Press in the white drive-nut button to move the pusher block.
  •   Insert the syringe plunger in the pusher block slot.
  •   Insert the syringe barrel flange in the flange brackets with the syringe barrel holder on the syringe. Tighten the flange
      brackets onto the syringe flange. Tighten the pusher block screw.
  Prime / Purge: Press and hold the ‘Start/Stop’ key for one second. Release to stop.
  Start the Pump: Press and release the ‘Start/Stop’ key to start or stop the pump.
  When Pumping
  •   The pumping rate can be changed.
  •   With continuous pumping, the pumping direction can be changed.
38035(6(7 Press and hold the right-most up-arrow key while turning on power to the pump.
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                                                             New Era Pump Systems™
        Table of Contents
        1. GENERAL INFORMATION .............................................................................................. 1
ii
                                                                      www.k echnology.it
NE-1010/510/511 High Pressure                                                                                               New Era Pump Systems™
          7.5     CHANGING THE PUMPING RATE AND DIRECTION WHILE PUMPING ........................ 13
          7.6     VOLUME DISPENSED..................................................................................................... 13
          7.7     RESUMING WHEN PAUSED ........................................................................................... 14
          7.8     PUMP STALLED ............................................................................................................. 14
        8. SETUP CONFIGURATION .............................................................................................. 14
          8.1     POWER FAILURE MODE ............................................................................................... 15
          8.2     LOW NOISE MODE ........................................................................................................ 15
          8.3     AUDIBLE ALARM ENABLE ............................................................................................ 15
          8.4     TTL I/O OPERATIONAL TRIGGER DEFAULT CONFIGURATION ................................ 16
          8.5     TTL I/O DIRECTIONAL CONTROL INPUT CONFIGURATION ..................................... 16
          8.6     PUMP MOTOR OPERATING TTL OUTPUT CONFIGURATION ..................................... 17
          8.7     KEYPAD LOCKOUT ....................................................................................................... 17
          8.8     KEYPAD AND NOTIFICATION BEEP ENABLE ............................................................... 17
          8.9     RS-232 PUMP NETWORK CONFIGURATION ................................................................ 17
        9. PUMPING PROGRAM...................................................................................................... 18
          9.1 HOW TO ENTER PUMPING PROGRAMS ....................................................................... 18
            9.1.1 PUMPING PROGRAM PHASE NUMBER ..................................................................... 19
          9.2 PUMPING PROGRAM EDIT FUNCTIONS ....................................................................... 19
          9.3 PROGRAM FUNCTION DESCRIPTIONS ......................................................................... 20
            9.3.1 ‘RATE’: RATE FUNCTION ....................................................................................... 20
            9.3.2 ‘FILL’: FILL FUNCTION .......................................................................................... 20
            9.3.3 ‘INCR’: INCREMENT RATE FUNCTION.................................................................... 20
            9.3.4 ‘DECR’: DECREMENT RATE FUNCTION ................................................................. 21
            9.3.5 ‘STOP’: STOP PUMPING OPERATION AND END THE PROGRAM.............................. 21
            9.3.6 ‘JP:NN’: JUMP TO PHASE ........................................................................................ 21
            9.3.7 'PR:IN': SUB-PROGRAM SELECTION INPUT............................................................. 21
            9.3.8 'PR:NN': SUB-PROGRAM START LABEL ................................................................... 22
            9.3.9 ‘LP:ST’: DEFINE STARTING PHASE OF LOOP......................................................... 22
            9.3.10 ‘LP:EN’: DEFINE CONTINUOUS LOOP END ........................................................... 22
            9.3.11 ‘LP:NN’: DEFINE LOOP END AND LOOP REPETITIONS ........................................... 22
            9.3.12 'PS:NN': PAUSE PUMPING ....................................................................................... 22
            9.3.13 'IF:NN': JUMP TO PHASE IF EXTERNAL TRIGGER ................................................... 23
            9.3.14 'ET:NN': SETUP EVENT TRIGGER JUMP PHASE ....................................................... 23
            9.3.15 'ES:NN': SETUP EVENT SQUARE WAVE TRIGGER JUMP PHASE .............................. 23
            9.3.16 'ET:RS': EVENT RESET ............................................................................................ 23
            9.3.17 ‘CLR.D’: CLEAR TOTAL VOLUME DISPENSED ........................................................ 23
            9.3.18 ‘TR:AA’: OVERRIDE OPERATIONAL TRIGGER CONFIGURATION ............................. 24
            9.3.19 'OUT.N': SET TTL OUTPUT PIN .............................................................................. 24
            9.3.20 'BEEP': BEEP .......................................................................................................... 24
          9.4 PUMPING PROGRAM EXAMPLES.................................................................................. 24
            9.4.1 EXAMPLE 1: 2 STEP RATE ...................................................................................... 24
            9.4.2 EXAMPLE 2: REPEATED DISPENSES WITH SUCK BACK .......................................... 25
            9.4.3 EXAMPLE 3: RAMPING THE FLOW RATE ................................................................ 26
            9.4.4 EXAMPLE 4: COMPLEX DISPENSES WITH EXTERNAL SYNCHRONIZATION ............ 27
            9.4.5 EXAMPLE 5: CONTROL FROM A HIGH-LOW PRESSURE SENSOR ............................ 28
            9.4.6 EXAMPLE 6: AUTOMATED DISPENSING WITH SYNCHRONIZATION........................ 29
            9.4.7 EXAMPLE 7: SUB-PROGRAMS ................................................................................ 30
            9.4.8 EXAMPLE 8: DISPENSING WITH COMPLEX SYNCHRONIZATION ............................. 31
            9.4.9 EXAMPLE 9: AUTOMATIC REFILL .......................................................................... 32
        10.     RS-232 COMMUNICATIONS ....................................................................................... 32
          10.1       CONNECTION AND NETWORKING ............................................................................ 32
iii
                                                              www.k echnology.it
NE-1010/510/511 High Pressure                                                                                                 New Era Pump Systems™
iv
                                                               www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
        1. General Information
        Thank you for purchasing the NE-1000 Multi-Phaser™ Programmable Syringe Pump. With the NE-1000
        syringe pump you will be able to perform simple infusions or implement a complex automated dispensing
        system.
        Please familiarize yourself with the NE-1000’s operation by reading this user's manual. For future
        reference, record the serial number, located on the rear of the pump, and the date of purchase.
       New Era Pump KF  Technology
                    Systems           Srl in
                            Inc., located  - Via Amedeo NY
                                             Farmingdale, Bocchi,
                                                             USA, 84/8
                                                                   can be- contacted
                                                                           00125 Romaat: - Italy
       Ph: +39Phone: (631) 249-1392
               06.454.34.179  - Mobile: FAX:+39(707) 248-2089 -Email:
                                                 339.533.03.22    Fax +39 INFO@SYRINGEPUMP.COM
                                                                             06.972.53.131 - Skype: kft2002
                                        WWW.SYRINGEPUMP.COM
                                  info@kftechnology.it    - www.kftechnology.it
        This Operating Manual, and the NE-1000’s hardware, electronics and firmware are copyrighted.
        Copyright 1999-2014, all rights reserved.
        1.2 Disclaimer
        New Era Pump Systems Inc. makes no representations or warranties, expressed, statutory or implied,
        regarding the fitness or merchantability of this product for any particular purpose. Further, New Era Pump
        Systems Inc. is not liable for any damages, including but not limited to, lost profits, lost savings, or other
        incidental or consequential damages arising from ownership or use of this product, or for any delay in the
        performance of its obligations under the warranty due to causes beyond its control. New Era Pump
        Systems Inc. also reserves the right to make any improvements or modifications to the product described
        in this manual at any time, without notice of these changes.
        New Era Pump Systems Inc. products are not designed, intended, or authorized for use in applications or
        as system components intended to support or sustain human life, as a clinical medical device for humans,
        or for any application in which the failure of the product could create a situation where personal injury or
        death may occur.
        All brand and product names used in this manual are the trademarks of their respective owners.
                                                      www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
        1.3 Warranty
        New Era Pump Systems Inc. warranties this product and accessories for a period of two years, parts and
        labor, from the date of purchase. The repaired unit will be covered for the period of the remainder of the
        original warranty or 90 days, whichever is greater. Return shipping charges are not included.
        A return authorization number must be obtained from New Era Pump Systems Inc. before returning a unit
        for repair. Warranty covered repairs will not be performed without a return authorization number. At the
        option of New Era Pump Systems Inc., a defective unit will be either repaired or replaced.
        This warranty does not cover damage by any cause including, but not limited to, any malfunction, defect
        or failure caused by or resulting from unauthorized service or parts, improper maintenance, operation
        contrary to furnished instructions, shipping or transit accidents, modifications or repair by the user, harsh
        environments, misuse, neglect, abuse, accident, incorrect line voltage, fire, flood, other natural disasters,
        or normal wear and tear. Changes or modifications not approved by New Era Pump Systems Inc. could
        void the warranty. Wearable parts, such as drive nuts, are not covered by the warranty.
        The foregoing is in lieu of all other expressed warranties and New Era Pump Systems Inc. does not
        assume or authorize any party to assume for it any other obligation or liability.
        2. Overview
        The model NE-1000 is a general purpose single syringe pump capable of infusion and withdrawal. It is
        controlled from a microcontroller based system which drives a step motor, allowing a large range of
        pumping rates configured to the inside diameter of the loaded syringe. The syringe is driven from a drive-
        screw and drive-nut mechanism.
        Features:
        ♦   Infusion and withdrawal pumping of syringes         ♦   RS-232 bi-directional control from a computer
            up to 60 mL. 140 mL partially filled.
                                                                ♦   Built-in pump network driver. Pump network
        ♦   Pumping rates from 0.73 μL/hr with a 1 mL               supports up to 100 pumps and other devices.
            syringe to 2120 mL/hr with a 60 mL syringe.
                                                                ♦   Two modes of RS-232 control, Basic and Safe.
        ♦   Stall detection: Automatically stops pump               Safe mode provides communication error
            when pumping is impeded.                                detection, loss of communication detection, and
                                                                    automatic transmitting of alarm conditions.
        ♦   Infusion and withdrawal volumes separately
            accumulated.                                        ♦   TTL logic I/O with firmware filtered control
                                                                    inputs to eliminate glitches and ringing on the
        ♦   Programmable dispense volumes.                          control inputs.
        ♦   Programmable Phases allowing complex                ♦   Configurable and programmable TTL
            pumping applications and interaction with               operational trigger for flexible logic control.
            external devices.
                                                                ♦   Power Failure Mode: Restarts the Pumping
        ♦   Program Sub-Programs, selectable by the user.           Program after a power interruption.
        ♦   Non-volatile memory of all operating                ♦   Audible Alarm.
            parameters and Pumping Program.
                                                                ♦   Many more features!
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
       To facilitate and enhance your understanding of the NE-1000’s operation, please take the time to familiarize
       yourself with the basic concepts below:
4 5 6 7 8 9
3 10
                                                                                                      11
                                                                                                      12
           2
                                                                                                      13
                                    16           15                         14
                                               3DUWVRIWKH3XPS
           1) Pusher Block                                                 10) Syringe Retainer Bracket
           2) End Plate                                                    11) Syringe Holder Block
           3) Anti-Siphon Plate Adjustment Knob                            12) ‘V’ Slot (on Syringe Holder Block)
           4) Power On/Off Switch                                          13) Syringe Clamp
           5) Drive-Nut Button                                             14) Keypad / User Interface
           6) Anti-Siphon Plate                                            15) Guide Rod Collar Clamp
           7) Drive-Screw                                                  16) Guide Rod (2 guide rods)
           8) Syringe Retainer Thumbscrew (2, one on each side)
           9) Hex Wrench (in tool holder)
3DUWVRID6\ULQJH
            Plunger
       
            Flange
       
       
                            Plunger                         Barrel Flange               Barrel
                                                   www.k echnology.it
NE-1010/510/511 High Pressure                                                                      New Era Pump Systems™
       7HUPLQRORJ\
        0RPHQWDU\3UHVV        A quick press, less then 1 second, then release of a key on the keypad.
        'LVSOD\%OLQN          A momentary blanking of the LCD display. This indicates that the new data
                                entered by the user is valid and has been stored.
        3URJUDP(QWU\          The mode where the Program Phase and Program function are selected and
        0RGH                   modified. In this mode the ‘Program Phase #’ and the ‘Program Function’
                                modes of the ‘Rate’ and ‘Volume’ keys are relevant.
        3XPSLQJ3URJUDP       The sequence of automated operations entered into the pump. This could be as
                                simple as a single function to pump at a single infusion rate continuously.
        3XPSLQJ3URJUDP        When the pump is started with the ‘Start’/’Stop’ key, or any other source, the
        2SHUDWLQJ              pump begins performing the operations in the Pumping Program until the
                                Pumping Program either stops automatically or the pump is stopped when the
                                ‘Start/Stop’ key is pressed, or from any other source. While performing the
                                operations defined in the Pumping Program, the Pumping Program is referred to
                                as operating.
                                While Operating, the motor can be pumping or stopped, according to the
                                Pumping Program.
        3XPSLQJ3URJUDP        The motor is stopped and the pump is not operating the Pumping Program.
        6WRSSHG
        3XPSLQJ3URJUDP        The Pumping Program has been stopped, but can be resumed at the point where
        3DXVHG                it was stopped.
        3XPSLQJ3URJUDP        Continuing a Pumping Program that was Paused before the completion of the
        5HVXPHG                Pumping Program. The Pumping Program continues at the point where the
                                Pumping Program was stopped.
        ([HFXWHG              The pump has performed a single operational Phase as defined in the Pumping
                                Program.
        3URJUDP3KDVH          A single defined operation in the Pumping Program.
        3KDVH1XPEHU           A Program Phase’s numerical sequence location in the Pumping Program.
        &XUUHQWO\6HOHFWHG     Each Pumping Program Phase instructs the pump to perform a particular
        )XQFWLRQ               operation. Only one Program Phase is selected at any one time. This is the
                                current Phase. Each Phase is set to one function. The set function of the current
                                Phase is the currently selected function.
        3XPSLQJ5DWH           Each Pumping Program function instructs the pump to perform a particular
        )XQFWLRQ               operation. If the Phase’s operation instructs the NE-1000 to pump, then
                                associated with that Phase is the Phases’ pumping information. When ‘Program
                                Entry Mode’ is exited, the ‘Rate’, ‘Volume’, and pumping direction keys refer
                                to the currently selected Program Phase’s function. The Program functions that
                                are associated with pumping information are referred to as Pumping Rate
                                functions.
        )XQFWLRQ               Certain functions, which do not instruct the NE-1000 to pump, require
        3DUDPHWHU              additional data. This additional data, displayed with the function, is the
                                function’s parameter.
        6WDUW7ULJJHU          The Pumping Program may be started, or stopped, from multiple sources.
                                These are the keypad’s ‘Start’/’Stop’ key, the TTL I/O ‘Operational Trigger’
                                input, or from a command received through the RS-232 connection.
                                                 www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
       3. Setup
       ♦    Place the pump on a stable surface.
       ♦    Plug the round connector end of the supplied power supply adapter into the power plug located on the
            lower right of the pump's rear. See section 11, Logic Interface: TTL Input and Output, for a diagram of
            the rear of the pump. Plug the other end of the power supply adapter into an appropriate electrical outlet.
            The pump will be powered when the bottom of the power switch, located on the upper right of the rear of
            the pump, labeled ‘1’, is pressed. The red indicator on the switch is visible when the power switch is in
            the ‘on’ position. After power is applied to the pump, the pump’s display will flash.
       ♦    Next the Pumping Program can be entered. Before the Pumping Program can be operated, the pump
            needs the measurement of the inside diameter, in millimeters, of the syringe that will be loaded. The
            syringe diameter can be entered using the keypad on the front panel of the pump.
       ♦    Finally, the syringe can be loaded and the pump started.
       4. Loading Syringes
       The syringe is loaded by securing the barrel and the pusher flange as follows:
       1:   Loosen the 2 thumbscrews on the syringe retainer bracket.
       2:   Press in fully the white drive-nut button on the pusher block, releasing the block. Taking care not to
            drag the drive-nut on the drive-screw, slide the block away from the syringe holder, providing sufficient
            space for the loaded syringe. Then release the white button.
       3:   Lift the syringe clamp above the syringe holder block. Turn it 1/4 turn and then lower it onto the
            syringe holder block. The syringe clamp should be out of the ‘V’ slot.
       4:   Load the syringe with the barrel over the syringe holder and the syringe plunger towards the middle of
            the pump. Place the barrel on the syringe holder block, in the ‘V’ slot, with the barrel flange inserted
            between the syringe holder block and syringe retainer bracket.
       5:   On the pusher block, turn the thumbscrew to make the slot large enough for the plunger flange. Press in
            fully the white drive-nut button on the pusher block, releasing the pusher block. Then slide the block
            towards the syringe plunger. Place the syringe plunger flange into the slot and against the anti-siphon
            plate. When the flange is positioned in the slot, release the white drive-nut button.
       6:   Lift the syringe clamp to slightly above the height of the syringe barrel and turn the syringe clamp 1/4 turn
            back to its original position and then lower it onto the syringe barrel.
       7:   Firmly push in the syringe retainer bracket against the syringe barrel flange and tighten the 2 thumbscrews
            on the syringe retainer bracket. On the pusher block, turn the thumbscrew to move the plate against the
            plunger flange.
        To unload the syringe, reverse the instructions for syringe loading.
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       6. User Interface
                     RS-232            Units      Volume Dispensed
                                                                   Pumping Direction Motor
                     Indicator         Indicators Indicator                          Operating
                                                                   Indicator
                                                    Syringe ID Volume
                                                      mm       Dispensed
                                                      mL       min
                                                      μL       hr       Withdraw           Pumping
                    Decimal Point: Press 2 keys Program Entry Keys                   Pumping Direction Key
       Figure 1: Front Panel
       6.3 LEDs
       To the right of the LCD are 8 red, round, LED (Light Emitting Diode) indicators. The first 2 columns display
       the units of the displayed values. Units are expressed using 1 or 2 LEDs. For instance, ‘mL / hr’ is expressed
       by lighting the ‘mL’ and the ‘hr’ LEDs.
       ‘Dispensed’ indicates that the displayed volume is the total ‘Volume Dispensed’ or pumped.
       When ‘Pumping’ is lit (not blinking), the motor is operating, either infusing or withdrawing. If blinking, the
       motor is not operating, and the Pumping Program is paused. When the pump is restarted, the Pumping
       Program will resume at the point where the Pumping Program was interrupted. When not lit (not blinking) the
       pump is stopped, but the Pumping Program may be operating a pause Phase. Starting the pump, when the
       Pumping Program is stopped, will start the Pumping Program from the beginning (Phase 1).
       ‘Withdraw’ indicates that the pumping direction is set for withdrawing. If not lit, then the pumping direction
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       “Volume Dispensed” refers to the volume withdrawn. If not lit, the “Volume Dispensed” refers to the volume
       infused.
            LED                         Description
                                                                         Pumping rate units are expressed using 2 LEDs:
       mm            Millimeters
       mL            Milliliters
       min           Minutes                                                 ‘mL/min’ =
       μL            Microliters
       hr            Hours
                                                                             ‘mL/hr’ =
       Dispensed     Volume dispensed displayed
       Withdraw      Pumping Direction:
                        Lit:      Withdraw
                                                                             ‘μL/min’ =
                        Not lit:  Infuse
                        Blinking: “Sticky Direction”
       Pumping       Lit:          Motor is operating
                                                                             ‘μL/hr’   =
                     Not lit:      The Pumping Program is stopped
                                   or executing a Pause Function
                     Blinking: The Pumping Program is paused
                                                        www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
       Arrow and Decimal Point Key). The ‘mm’ LED will blink while the diameter is being changed.
       If the ‘Diameter’ key is pressed and held, ‘Setup’ mode will be entered. (See sec. 6.12, ‘Setup’ ).
       rate limits, the display will show      nn   , where ‘nn’ indicates the currently selected Phase Number.
       Pressing any key clears the message and returns to the previous pumping rate.
       6.6.1 Pumping Rate Units
       The pumping rate units can only be changed when the Pumping Program is not operating. If the currently
       selected function allows selection of rate units (‘RATE’ function), a momentary press of the ‘Rate’ key will
       enter Rate Units Change mode. The 2 LEDs representing the units will blink and the display will show:
                   .
       Each press of any up arrow key selects the next rate units, as indicated by the blinking units LEDs. When the
       required rate units are blinking, press any non-arrow key or wait 2 seconds. The display will blink, indicating
       the rate units are stored. The rate units are stored in the currently selected Program Phase. The rate units can
       be independently set for each Phase with a ‘RATE’ function.
       6.6.2 Program Entry Mode
       While the Pumping Program is stopped, “Program Entry Mode” can be entered by pressing and holding the
       ‘Rate’ key. Release the key when the display shows the current Program Phase number:              nn , where
       ‘nn’ indicates the current Program Phase number.
       With the current Program Phase number displayed, if the currently selected Program Phase is set to a pumping
       rate function, a momentary press of this key will exit “Program Entry Mode” and return to the rate display.
       Arrow and Decimal Point Key). The units of the volume are set according to the syringe diameter, but can be
       changed. The new “Volume to be Dispensed” is stored in the current Program Phase. If the “Volume to be
       Dispensed” is disabled (‘off’), pressing any up arrow key will change the display to 0.0. The “Volume to be
       Dispensed” can now be set using the up arrow keys.
       While pumping, pressing this key will switch between displaying the current “Volume Dispensed” and
       “Volume to be Dispensed”.
       6.7.1 Disabling “Volume to be Dispensed”
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
To disable the “Volume to be Dispensed”, i.e. continuous pumping, set the “Volume to be Dispensed” to 0.0.
       After being stored, the display will show            , indicating the “Volume to be Dispensed” is off. In this
       mode, the pump will not stop at a set volume and will pump continuously until the pump is stopped, or an
       “event trigger”, programmed into the Pumping Program, occurs.
       6.7.2 Clearing “Volume Dispensed”
       With the Pumping Program stopped, display the “Volume Dispensed”. Pressing and holding any up arrow key
       for one second will reset the infusion and withdrawal dispensed volumes to 0.
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       number will be displayed as         nn , where ‘nn’ is the current Program Phase number.
       When the Program Phase number is displayed and the current Phase’s function is a rate function, a momentary
       press of the ‘Program Phase #’ key exits ‘Program Entry Mode, and displays the pumping rate.
       To change the current Program Phase number, press the arrow keys below the Phase number’s digits. The
       maximum Phase number is 41. To reset to Phase number 1, press and hold the ‘Program Phase #’ key until the
       Phase number is 1.
       When a new Program Phase number is selected, the current value of all settings will be that of the currently
       selected Program Phase.
       the pump. The display will show:      n.nn      , where ‘n.nn’ is the firmware version number. Pressing any key
       will clear the display.
10
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       power to the pump. The display will show                  . Pressing any key will clear the display.
       With a pump with as many complex features as the NE-1000, it is easy for a novice user experimenting with
       the pump's setup to get the pump into a 'weird' state. Performing this reset function will bring the pump out of
       a 'weird' state.
       Program Menu and the display will show the first default program:             , Reciprocating Pumping
       program. Us the up arrow keys to select the next program. When the Reciprocating Pumping program is
       selected, the following program will be loaded and stored in the Pumping Program memory, plus the RS-232
       communications mode will be set to Reciprocating Mode. The default program will over write any other
       program stored in memory and it can then be modified as needed.
       The Reciprocating Pumping Program, plus cable Part# CBL-DUAL-3, sets the pump for use with a second
       pump to create a continuous infusion system.
       following will be displayed:           n, currently, the only parameter. The “Program Entry Mode Lockout”,
       when enabled, prevents inexperienced users from entering “Program Entry Mode” from the keypad. Mode
       Disabled: ‘n’ = 0 (default). Mode Enabled: ‘n’ = 1. When enabled, only Phase 1 ‘Rate’, ‘Volume’ and
       Pumping Direction can be changed. Cannot be enabled when the Pumping Program is currently programmed
       with a multiple Phase Program.
                        An out of range error occurred at Pumping Program Phase number ‘nn’, or the value just
               nn       entered is out of range. Check the pumping parameters and syringe diameter.
                        A Pumping Program error was encountered at Pumping Program Phase number ‘nn’.
               nn       The indicated Phase is invalid in the context of the entire Pumping Program.
                        Key pressed is not currently applicable.
11
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
                        An error was detected during power up, where ‘n’ indicates the error. If n=1, then the
                n       values stored in the pump’s non-volatile memory were invalid and were reset. If n=2,
                        then the non-volatile memory may need to be replaced.
                        Pump settings are locked out from the keypad. The lockout key is needed to change
                        settings. Lockout can also be reset with the reset function.
       6.15 6WDWXV0HVVDJHV
                        Indicates pumping rate units change mode. The units LED's will also be blinking.
                        Indicates that the Pumping Program has paused and is waiting for the user to press
                        ‘Start’, or for an external operational trigger, to continue.
                        Indicates that the pump is busy completing a long operation.
                        Indicates that the “Volume to be Dispensed” is 0.00, and is turned off. This is the
                        continuous pumping mode.
                        Indicates that the pump is purging. Displayed while holding down the 'Start/Stop' key.
                        Indicates that the Pumping Program paused and is waiting for the user to select a sub-
             nn         program.
                        Indicates that the pump’s RS-232 communications is set for either Reciprocating or Dual
                        pumping modes. One of these messages will be briefly displayed while the pump is
                        searching for the secondary pump. Normally, seeing one of these messages would
                        indicate that the secondary pump is not attached or communication cannot be established.
                        Indicates that the Pumping Program entry mode has been entered, possibly after the
                        ‘Rate’ key was pressed and held. A momentary press of the ‘Rate’ key will return the
                        display to the pumping rate. This display may also indicate that the first Pumping
                        Program Phase is not a ‘Rate’ function.
       7. 2SHUDWLRQ
                         Before the pump can be operated, the pumping data must be setup. At minimum, the
                         syringe inside diameter and a non-zero pumping rate needs to be set. The operation of the
                         pump can then be started from the keypad, TTL I/O connector, or from RS-232 control.
                         From the keypad, pressing the ‘Start / Stop’ key will start the pump operation.
       7.1 6\ULQJH,QVLGH'LDPHWHU
       The syringe inside diameter can only be set while the Pumping Program is stopped. Use the up arrow keys to
       set the diameter value. While the diameter value is being set, the ‘mm’ LED will blink. The new diameter
       value is stored after pressing any non-arrow key, or after a 2 second delay.
       Valid syringe diameters are from 0.1 mm to 50.0 mm. If the diameter is out of this range, the display will
       show ‘oor’. Pressing any key restores the diameter display to its previous value. Changing the syringe
       diameter will not zero any current settings. Section 12.7, “Syringe Diameters and Rate Limits”, is a
       representative list, for reference, of syringe diameters for various syringe manufacturers and syringe sizes.
       7.1.1 Default Volume Units
       The units of the accumulated infusion and withdrawal volumes and the “Volume to be Dispensed” are set
       according to the diameter setting. NOTE: A change in the volume units will affect all “Volume to be
       Dispensed” settings in the Pumping Program. If the default volume units are changed (see next section), the
       selected volume units will remain in effect until a reset function is performed.
                From 0.1 to 14.0 mm          Syringes smaller than 10 mL:               Volume units are ‘μL’
                From 14.01 to 50.0 mm        Syringes greater than or equal to 10 mL: Volume units are ‘mL’
12
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       Now, pressing any up arrow key will change the display to                 and the current Volume Units will
       blink.
       Then, press any up arrow key to switch the Volume Units between ‘mL’ and ‘μL’. Press any non-arrow key or
       wait 2 seconds to enter the new Volume Units. The display will blink when entered. The selected Volume
       Units will remain in affect and override the default Volume Units. Changing the diameter will no longer
       change the Volume Units. Performming a system reset will cancel the override and allow the Volume Units to
       change to the default Volume Units when setting the syringe diameter.
       7.4 Purging
       To purge the syringe, with the Pumping Program stopped, press and hold the 'Start/Stop' key. The Pumping
       Program will start then, after one second, purge will begin. The pump will pump at its top speed in the
       currently set direction. Purging will continue until the 'Start/Stop' key is released, and then the pump will stop.
       pump, the display will show         nn .    Pressing any key clears the error message.
       The pumping direction can be changed while pumping if the “Volume to be Dispensed” is 0.0 (off). Pressing
       the direction key will immediately change the pumping direction and store the pumping direction in the current
       Program Phase. Also changing the pumping direction changes the accumulated “Volume Dispensed”
       according to the new pumping direction.
13
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
       Dispensed can be displayed by pressing the ‘Volume’ key one, two, or three times, depending on the current
       display.
       The volume is accumulated separately for infusion and withdrawal. When the pump changes direction, the
       “Volume Dispensed” changes to the accumulated volume for the current pumping direction.
       The “Volume Dispensed” accumulations for infusion and withdrawal are reset to 0 when:
             A) With the pump stopped, pressing and holding any up arrow key while displaying the “Volume
                Dispensed”.
             B) A sub-program is selected when the Pumping Program executes a Program Selection function.
             C) The syringe diameter is changed.
             D) From the RS-232 clear “Volume Dispensed” command (CLD) or Clear Dispense function.
             E) The accumulated Volume Dispensed rolls over from 9999 to 0.
             F) The pump is powered on.
       occur. The display will show              , the ‘Pumping’ LED will blink, and the buzzer will sound
       continuously if alarms are enabled. Also, if the RS-232 Safe Mode is enabled, an auto-alarm message will be
       sent to an attached computer.
       Pressing any key will stop the buzzer and clear the alarm. When the problem causing the pump motor to stall
       has been corrected, the Pumping Program can be resumed from any start trigger source: ‘Start’/’Stop’ key,
       TTL input, or RS-232 command.
       8. Setup Configuration
       To change or view the setup configuration, the Pumping Program must be stopped. Press the
       ‘Diameter’/‘Setup’ key until the first parameter, ‘PF’ is displayed. After 2 seconds, or when any non-arrow
       key is pressed, the next parameter will be displayed (see sec. 6.12, ‘Setup’ Key). Pressing an arrow key under
       a value will increment, select, or scroll through the valid values for the parameter. The Setup Configurations
       will be displayed in the following order:
                                 Display TTL I/O external logic connector settings. Press any arrow key to select.
                                 If TTL is selected, the following TTL logic settings will be displayed:
14
                                                   www.k echnology.it
NE-1010/510/511 High Pressure                                                                    New Era Pump Systems™
                         n       *** The "Lockout Disable Key" needs to be inserted to display this setting. ***
                                 Locks out changing settings from the keypad. 'n' is the current setting.
                        n      Keypad and notifications beep enable, where ‘n’ is the current setting.
               If standard communications mode with a computer is set, then the following are displayed:
       )      CAUTION: The Pumping Program will start operating from the beginning of the Pumping Program
       (Phase 1), regardless of what part of the Pumping Program was operating when the power was disrupted.
       When the NE-1000 syringe pump is used as a component in an automated infusion/withdrawal dispensing
       system, a Pumping Program can be designed to automatically synchronize the pusher block at the start of the
       Pumping Program. This would be accomplished using attached sensors that send signals to the Pumping
       Program.
15
                                                  www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
      The setting names are relevant to a 2 pump system, whereby the ‘Directional Control Input’ TTL pin is
      attached to the second pump’s ‘Pumping Direction Output’ TTL pin.
      In addition, “Directional Control Input” will affect how the “Sticky Direction” function operates. See “ Sticky
      Direction”, sec: 6.8.1
16
                                                       www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
                   will be displayed.
       Auto-Run Mode: When used in conjuction with the Sub-Program Select programming function as Phase 1, the
       pump will enter Auto-Run Mode. In this mode, on power up, the Pumping Program will immediately begin to
       execute and the user would be prompted to enter a Sub-Program number.
       The "Lockout Disable Key" connects the "Program Input", TTL connector pin 6, to Ground, pin 9. The
       "Lockout Disable Key" is available as an accessory item.
       Lockout can also be disabled by performing a system reset, see sec. 6.13.2, "Reset Pumping Program."
17
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                      New Era Pump Systems™
       In Reciprocating or Dual Pumps Modes, the secondary pump must be left in its default communications setting
       of Address Mode, Address 0, and 19,200 baud rate. See the documentation with CBL-DUAL-3 cable for more
       detailed information.
       9. Pumping Program
       A Pumping Program is simply a pre-defined sequence of actions, or functions, which guarantees consistent and
       precise operation of the pump, automatically, and with or without any user intervention. A Pumping Program
       can be as simple as continuous pumping at a fixed infusion rate. Or a Pumping Programs could consist of a
       pumping rate and direction of pumping for a specified volume, then switch to another pumping rate. Also a
       Program can interact with external devices through the TTL I/O connector, make decisions, or stop pumping
       for a period of time.
       Programs are broken into individual operations called Phases. Each Phase consists of a function that can be a
       control function or pumping function. A pumping function, such as ‘RATE’, consists of a pumping rate,
       optional “Volume to be Dispensed”, and the pumping direction.
       Complex dispensing systems can be designed, involving multiple liquids, each dispensed from a different
       pump, plus other equipment and sensors. Pumping Programs can be designed for each pump which enables
       multiple pumps to synchronize with each other, and the other equipment and sensors, using a cable connected
       to the TTL I/O connectors of each pump.
       When the Pumping Program is started, either from the keypad, TTL I/O connector, or from RS-232, the
       Pumping Program will begin with Phase 1 of the Program. After the completion of each Phase, the pump will
       immediately start the next consecutive Phase. This linear sequence of Phases can be altered by certain
       functions that direct the Pumping Program to continue operation with a different Phase number. Some
       functions can change the order of operation conditionally based on external events.
display will show , where ‘01’ refers to Phase 1. The pump will now be in “Program Entry
18
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
       Mode”. If the current Phase is not 1, press and hold the ‘Rate’/’Program Phase #’ key until the display is as
       shown. The pump will now be in Phase 1.
       When in “Program Entry Mode”, with the display showing the Program Phase number, pressing the
       ‘Volume’/’Program Function’ key will display the current “Program Function” for this Phase. If the current
       function is ‘RATE’, then a pumping infusion or withdrawal can be setup for this Phase.
       To change the “Program Function” selected, use the arrow keys to scroll through the functions until the
       required function is displayed. If the function has an associated parameter, enter the parameter after the
       function has been stored.
       Momentarily pressing the ‘Volume’/’Program Function’ key again will exit “Program Entry Mode” and
       display the “Volume to be Dispensed.” The pumping rate data, which includes the pumping rate, “Volume to
       be Dispensed” and pumping direction, can now be setup as previously described.
       When finished setting up the pumping rate data for the current Phase, enter “Program Entry Mode” again to
       select the next Program Phase. Press and hold the ‘Rate’/’Program Phase #’ key until the Phase number is
       displayed. Then use the arrow keys to set the Phase number to the next Phase to be setup. Pressing the right-
       most arrow once will set the Phase to Phase 2. Now all pumping data will refer to Phase 2. The second Phase
       can now be setup as described above for Phase 1.
       Continue selecting Phase numbers and entering the infusion or control setup for each Phase of the Pumping
       Program. The entire Pumping Program will be stored in non-volatile memory.
       Use the ‘STOP’ function to stop the pump and end the Pumping Program. If the Pumping Program does not
       operate the pump continuously, the last Phase of the Pumping Program must be a ‘STOP’ function (unless the
       last Phase number is the maximum Phase number).
       When the Pumping Program is started, with the ‘Start / Stop’ key, TTL I/O input, or RS-232 command, the
       Pumping Program will begin operating from Phase 1.
       Very complex dispensing Programs can be created with the Program functions available. Section 9.3 contains
       a detailed description of all the functions.
       9.1.1 Pumping Program Phase Number
       To set the current Program Phase number, enter “Program Entry Mode” and display the current Program Phase
       number.
       Using the right-most 2 arrow keys, change the selected Program Phase number. The displayed Program Phase
       number now becomes the currently selected Program Phase number. All function and pumping rate data will
       now refer to the currently selected Program Phase number.
       If the maximum Program Phase number, 41, is exceeded while changing the Phase number, the displayed
       Phase number will automatically be set to the maximum Program Phase number.
19
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
                    Defines a pumping function with a fixed pumping rate. This function defines a pumping setup
       consisting of the pumping rate, optional “Volume to be Dispensed”, and pumping direction. Use the ‘Rate’,
       ‘Volume’, and ‘Pumping Direction’ keys to set or review the pumping setup. For continuous pumping, set the
       “Volume to be Dispensed” to 0.0 (off). The “Volume to be Delivered” is disabled when the display reads
       [ off].
       9.3.2 ‘FILL’: Fill Function
                   The Fill function reverses the pumping direction and withdraws or dispenses the volume
       dispensed or withdrawn. After a series of arbitrary dispenses, the Fill Function uses the Volume Dispensed to
       automatcally refill the syringe. The Volume Dispensed is cleared when the Fill Function begins.
       The pumping rate can be set, as with the Rate Function.
       Fill at the current dispense/withdraw rate: If the Fill function’s pumpng rate is set to 0.0, the previous
       function’s pumping rate will be used for the Fill rate.
                     The increment and decrement functions operate the same as the ‘RATE’ function, except that the
       specified rate is added (‘INCR’) or subtracted (‘DECR’) from the current pumping rate. The current pumping
       rate when the function is executed is the base pumping rate for the function. If no base pumping rate exists,
       such as when executing a pause function or when the Pumping Program is first started, a Program error will
       occur and the Program will stop.
       The pumping rate units will be the same as the base pumping rate, and therefore cannot be set, nor are they
       displayed, with the pumping rate increment or decrement value. As with the ‘RATE’ function, a “Volume to
       be Dispensed” and pumping direction can be specified for the increment and decrement functions.
20
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       When used within a Program loop, the pumping rate can be incremented or decremented in small step
       intervals.
       9.3.4 ‘DECr’: Decrement Rate Function
                    The decrement function subtracts the specified rate from the current pumping rate. For a full
       description, see section 9.3.3, ‘InCr’: Increment Rate Function..
       9.3.5 ‘StoP’: Stop Pumping Operation and End the Program
                   Stops the pumping operation and stops the Pumping Program. The Pumping Program will begin
       at Phase 1 when started again. An implicit ‘Stop’ function is executed when the Program exceeds the
       maximum Phase number during operation.
       If alarms are enabled, the buzzer will beep continuously when the Pumping Program stops.
       9.3.6 ‘JP:nn’: Jump to Phase
             nn   The ‘Jump’ function alters the consecutive operation of Program Phases. When executed, the
       Pumping Program will continue operation with Phase ‘nn’.
       9.3.7 'Pr:In': Sub-Program Selection Input
       The Pumping Program can be broken into sub-programs which can be selected by the user.
       Auto-Run Production Mode: When Phase 1 is set to Program Selection Input, and Lockout Mode is selected,
       the pump will enter Auto-Run Mode. When the pump is powered on, it will immediately begin executing the
       Pumping Program. The user will immediately be prompted to enter a sub-program number. This permits a set
       of production dispenses to be programmed into the pump. The user then would only be able to select from one
       of these production dispenses and not make any changes.
       Foot Switch Trigger Feature: If the Pumping Program was Paused, and the Paused Phase was a continuous
       pumping Rate function, then starting the pump with a foot switch will cancel the Pause and restart the
       Pumping Program Phase 1 causing the current program selection to be re-executed.
When the Program Selection Input function is executed, the Pumping Program pauses and
       again. If the selected label is not found, the 'out of range' error message            is displayed. Pressing
       any key returns the display to the Program Selection Input display.
       If more than one Phase is defined with the same label, then execution continues with the first matching label
       encountered. The last selected program label is stored in non-volatile memory and becomes the default label
       the next time the current Program Phase is executed. More than one Program Selection Input function can be
       defined and placed at any Program Phase needed. To cancel the Program Select Input and stop the Pumping
       Program, turn the power to the pump off and on.
       If alarms are enabled, the buzzer will beep continuously while waiting for the start trigger.
21
                                                      www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       The Sub-Program Start Label function             defines the start of a Pumping Program sub-program that
       can be selected by the user during Pump Program execution.
       After selecting the function, change '00', if needed, to a unique Program Label, from 0 to 99. See sec: 9.3.7,
       'Pr:In': Sub-Program Selection Input for a full description. Place a Sub-Program Start Label, with a unique
       number for each Sub-Program, from 0 to 99, at the starting Phase of each sub-program section.
       When the Pumping Program encounters a Sub-Program Start Label in normal execution, it will be interpreted
       as “Jump to Phase 1” [JP:01], ending the sub-program, and, assuming Phase 1 is a Sub-Program Select
       function, [Pr:In], the user will be immediately prompted to select a sub-program.
                    Defines the start of a Program loop. For a full description of Program looping, see sec. 9.3.11,
       ‘LP:nn’: .
       9.3.10 ‘LP:EN’: Define Continuous Loop End
                    Loops to the most recently executed, unpaired, ‘loop start’ Phase, or Phase 1 if none. This
       function allows a section of the Program to be repeated continuously. For a full description of Program
       looping, see sec. 9.3.11, ‘LP:nn’: .
       9.3.11 ‘LP:nn’: Define Loop End and Loop Repetitions
       executed, the display will show         nn , with ‘nn’ decrementing to indicate the number of seconds until
       the next Program Phase is executed. After the pause interval, the next Program Phase will be executed.
       To set a pause time in tenths of seconds, select the decimal point between the digits. To select the decimal
       point, press and hold the right-most arrow key until the right-most digit scrolls to 9. After 9, the decimal point
       between the 2 digits will toggle on and off              . Release the key when the decimal point is displayed,
       or cleared, as required. Now enter the required pause time from 0.1 to 9.9 seconds. While executing a pause
       time set in tenths of seconds, the display will only show '01' seconds during the pause.
       For pauses longer than the ‘99’ second maximum pause for this function, put the pause function within a
       Program loop. A Program section with the following functions in consecutive Phases:
                [LP:ST] [LP:ST] [PS:60] [LP:60] [LP:24],
22
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                            New Era Pump Systems™
                 nn   The ‘IF’ function conditionally alters the Pumping Program’s execution based on an external
       signal.
       When executed, if the TTL I/O Program Input pin (pin 6) is low level, then the Pumping Program continues
       operation with Phase number ‘nn’. Otherwise, the Pumping Program continues operation with the next Phase.
       9.3.14 'Et:nn': Setup Event Trigger Jump Phase
                 nn   The ‘Event’ function sets a background event trap that is triggered by an external signal.
       This one time background trap, or interrupt, stays set during the Pumping Program’s entire execution until it is
       triggered, redefined, or reset. This function has no other effect on the operation of the pump until it is
       triggered.
       The event is triggered with either:
            1) A falling edge (high to low TTL transition) on the TTL ‘Event Trigger’ input (pin 4).
            2) A low level on the 'Event Trigger' input pin of at least 200 ms at the time the function is executed.
            3) The RS-232 ‘RUN E’ command.
       When triggered, the current operation of the pump and the Pumping Program is interrupted, and the Pumping
       Program immediately continues operation (jumps to) with Phase number ‘nn’.
       After being triggered, the event trigger is reset. If an event trigger function is executed (either 'Et' or ES'
       function) while another event trap is still set, the new event trigger will replace the previous event trap. Only
       one event can be defined at any time.
       9.3.15 'ES:nn': Setup Event Square Wave Trigger Jump Phase
                 nn  The 'Event Square Wave' function operates the same as the 'ET' 'Event' function, with the
       exception of the triggering conditions.
       The event is triggered with either:
               1) The rising or falling edge of the TTL ‘Event Trigger’ input (pin 4).
               2) The RS-232 ‘RUN E’ command.
       Therefore, a square wave function on the inupt pin can be used to toggle the pump between 2 sections of a
       Pumping Program. An example of this would be a Pumping Program that switched between a slow and fast
       pumping rate, controlled by a square wave input.
                      ’Event Reset’ cancels a previously set event trap by either the 'ES' or 'ET' function.
       9.3.17 ‘CLr.d’: Clear Total Volume Dispensed
Sets the total volume dispensed to 0. Both infusion and withdraw volumes are reset.
23
                                                       www.k echnology.it
NE-1010/510/511 High Pressure                                                                                     New Era Pump Systems™
‘Button Trigger Event Trap’ redirects a press of the ‘Stop’ key to the background event trap.
‘Trigger Event Trap’ redirects a stop pump Operational Trigger to the background event trap.
       Instead of stopping the pump, an event defined by the ‘Event’ or ‘Event Square Wave’ function will be
       triggered, causing the Pumping Program to continue execution at the Phase number defined in the Event
       function.
       If an Event Trap has not been defined, then the Pumping Program continues execution at the next program
       phase. This is useful if the pump is executing a continuous infusion Phase. The pump can continue to infuse
       until a foot switch is pressed. The pumping program can then jump to a different Phase, such as a withdraw
       Phase, then stop the pump.
       The Trigger Event Trap is a one-time function. After triggering an Event Trap, the Trigger Event Trap is
       cleared. The Trigger Event function will not affect the current trigger configuration.
       9.3.19 'OUt.n': Set TTL Output Pin
                             n
                    Set ‘Program Output’ TTL I/O output (pin 5) to level ‘n’. If ‘n’ = 0, the output pin will be set
       low. If ‘n’ = 1, the output pin will be set high.
       9.3.20 'bEEP': Beep
                       500
         ml/h r Rate
                       2.5
                        0
                             05         30 ml Dispensed
       Infuse 5.0 mL at 500 mL/hr, and then infuse 25.0 mL at 2.5 mL/hr. Then stop the pump.
       Phase                  Function       Rate          Volume       Direction
       1                      RATE           500 mL/hr     5.0 mL       Infuse
       Phase                  Function       Rate          Volume       Direction
       2                      RATE           2.5 mL/hr     25.0 mL      Infuse
       Phase                  Function
       3                      STOP
24
                                                                 www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
                        750
           ml/hr Rate
                         0                     
                              0   :30
                                  5:00   Time
        Dispense 2.0 mL with a 5 minute pause between dispenses. In addition, after each dispense, a volume of 0.25
        mL is sucked back to prevent dripping. Also, 30 seconds before the end of the pause interval, a beep is
        sounded to alert the operator to prepare for the next dispense.
        Starting with the second dispense, 0.25 is added to the volume dispensed to compensate for the sucked back
        volume of the previous dispense. By changing the last Phase to a [LP:nn] function, the total number of
        dispenses can be set.
        When entering a function with associated data, such as with the ‘Pause’ in Phase 5, or the ‘Loop’ in Phase 6,
        the function is entered in 2 steps. First select the function and store it. Then enter the associated data.
        Phase                 Function       Rate        Volume    Direction
        1                     RATE           750 mL/hr   2.0 mL    Infuse
        Phase                 Function       Rate        Volume    Direction
        2                     RATE           750 mL/hr   0.25 mL   Withdraw
        Phase                 Function
        3                     LP:ST
        Phase                 Function
        4                     LP:ST
        Phase                 Function
        5                     PS:90
        Phase                 Function
        6                     LP:03
        Phase                 Function
        7                     BEEP
        Phase                 Function
        8                     PS:30
        Phase                 Function       Rate        Volume    Direction
        9                     RATE           750 mL/hr   2.25 mL   Infuse
        Phase                 Function       Rate        Volume    Direction
        10                    RATE           750 mL/hr   0.25 mL   Withdraw
        Phase                 Function
        11                    LP:EN
25
                                                             www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
                                        Phases              Phases
                       250              5-8                 9-11
          ml/hr Rate
200
150
                              0
                                  Time           Note: Graph is representative only.
                                                 Pumping rate increments and decrements in steps.
       Continuously ramp up and down the pumping rate. Starting at 200 mL/hr, the pumping rate will increment to
       250 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Then the pumping rate will decrement to
       150 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed. Finally, the pumping rate is incremented
       back to 200 mL/hr in 1.0 mL/hr steps after every 0.1 mL has been dispensed, then the process is repeated.
       Phase                 Function      Rate              Volume       Direction
       1                     RATE          200 mL/hr         0.1 mL       Infuse
       Phase                 Function
       2                     LP:ST
       Phase                 Function      Rate              Volume       Direction
       3                     INCR          1.0               0.1 mL       Infuse
       Phase                 Function
       4                     LP:50
       Phase                 Function
       5                     LP:ST
       Phase                 Function      Rate              Volume       Direction
       6                     DECR          1.0               0.1 mL       Infuse
       Phase                 Function
       7                     LP:99
       Phase                 Function      Rate              Volume       Direction
       8                     DECR          1.0               0.1 mL       Infuse
       Phase                 Function
       9                     LP:ST
       Phase                 Function      Rate              Volume       Direction
       10                    INCR          1.0               0.1 mL       Infuse
       Phase                 Function
       11                    LP:50
       Phase                 Function
       12                    JP:02
26
                                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                                      New Era Pump Systems™
                                                                                         Start over
          Withdraw (ml/hr) Infuse
750
                                                                               
                                     0
                                          0                        :60   :60     :60
       A more complex dispensing example, this Program contains different pumping requirements, including
       dispenses with multiple pumping rates. The first set of 3 dispenses drops down to a lower pumping rate during
       the dispense. When each dispense is completed, the buzzer beeps to alert the operator, then the pump waits for
       a start trigger before starting the next dispense.
       The next set of 3 dispenses have a fixed time interval of 60 seconds between dispenses. After the last set of
       dispenses, the syringe is refilled by the amount infused, 17.25 mL. Then the buzzer beeps, to alert the operator
       to the start of the first set of dispenses. The process is then repeated.
       Phase                              Function       Rate                  Volume   Direction
       1                                  RATE           750.0 mL/hr           0.5 mL   Infuse
       Phase                              Function       Rate                  Volume   Direction
       2                                  RATE           300.0 mL/hr           1.5 mL   Infuse
       Phase                              Function
       3                                  BEEP
       Phase                              Function
       4                                  PS:00
       Phase                              Function
       5                                  LP:02
       Phase                              Function       Rate             Volume        Direction
       6                                  RATE           750.0 mL/hr      0.5 mL        Infuse
       Phase                              Function       Rate             Volume        Direction
       7                                  RATE           300.0 mL/hr      1.5 mL        Infuse
       Phase                              Function
       8                                  BEEP
       Phase                              Function
       9                                  LP:ST
       Phase                              Function
       10                                 PS:60
       Phase                              Function       Rate             Volume        Direction
       11                                 RATE           500.0 mL/hr      3.75 mL       Infuse
       Phase                              Function
       12                                 LP:03
27
                                                                                 www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
                        25
                       ml/hr
          Pumping
          Rate
10
       This example demonstrates a Pumping Program whose control depends on an external sensor. Assuming a
       pressure sensor that is configured to detect a high pressure point and a low pressure point, the Pumping
       Program individually selects whether it will react to the high or low pressure point.
       The “Program Output” pin on the TTL I/O connector (pin 5) is used to select the high or low pressure point.
       When low, the low pressure point is selected (PH:01), and when high, the high pressure point is selected
       (PH:05). The Program begins by infusing continuously at 10.0 mL/hr (PH:02), while a background trap is set
       for the low pressure point (PH:03). To create a delay when the pressure sensor is switched from high pressure
       to low pressure when the “Program Output” pin is set, a small volume is pumped (PH:02, 06) before the
       background traps are set.
       When the low pressure trap is triggered, the pump sets the high pressure trap (PH:07) and begins to increment
       the flow rate. The flow rate is incremented in 1.0 mL/hr steps with every 0.25 mL dispensed (PH:08-10). If
       the high pressure trap hasn’t as yet been triggered, the flow rate will max out at 25.0 mL/hr while waiting for
       the high pressure trap (PH:11). When the high pressure point is reached, the pump immediately will drop
       down to 10.0 mL/hr (PH:02), and once again wait for the low pressure point.
       Phase        Function
       1            OUT.0
       Phase        Function   Rate           Volume           Direction
       2            RATE       10.0 mL/hr     0.005            Infuse
       Phase        Function
       3            EV:05
       Phase        Function   Rate           Volume           Direction
       4            RATE       10.0 mL/hr     0.0 mL (off)     Infuse
       Phase        Function
       5            OUT.1
       Phase        Function   Rate           Volume           Direction
       6            RATE       10.0 mL/hr     0.005            Infuse
28
                                                   www.k echnology.it
NE-1010/510/511 High Pressure                                                                                    New Era Pump Systems™
       Phase     Function
       7         EV:01
       Phase     Function
       8         LP:ST
       Phase     Function        Rate               Volume               Direction
       9         INCR            1.0                0.25 mL              Infuse
       Phase     Function
       10        LP:14
       Phase     Function        Rate               Volume               Direction
       11        RATE            25.0 mL/hr         0.0 mL (off)         Infuse
                   'Power Failure
                   Mode' Pumping
                   Program Restart
                   at Power Up
                                                                                                      
       The following is an automated dispensing Program for a NE-1000 syringe pump equipped with a ‘Syringe
       Filled Sensor’ attached to the TTL I/O connector, and a valve system to refill the syringe from a reservoir. It is
       also assumed that the ‘Power Failure’ mode is enabled.
       After a power fail restart, the pusher block is in an unknown position, making it impossible for an automated
       dispensing system to regain synchronization. With the Syringe Filled Sensor, the following Pumping Program
       will automatically synchronize the dispensing system, and then continue with the normal dispense.
       The first 2 Phases set an event trap for the Syringe Filled Sensor and refills the syringe until the sensor is
       triggered. When the sensor triggers the event, the pump’s pusher block will be synchronized with the Pumping
       Program. It is assumed that the sensor is positioned to refill the syringe with 60 mL. A withdraw volume of
       61 mL is set as a safety feature.
       After the syringe is refilled, one 5 mL dispense is made every 5 hours. After 12 dispenses, the syringe is
       refilled using the sensor again.
       Phase     Function
       1         EV:03
       Phase     Function        Rate                 Volume               Direction
       2         RATE            1000.0 mL/hr         61 mL                Withdraw
       Phase     Function
       3         LP:ST
       Phase     Function        Rate                 Volume              Direction
       4         RATE            200.0 mL/hr          5.0 mL              Infuse
       Phase     Function
       5         LP:ST
       Phase     Function
       6         LP:ST
29
                                                              www.k echnology.it
NE-1010/510/511 High Pressure                                                                     New Era Pump Systems™
       Phase     Function
       7         PS:60
       Phase     Function
       8         LP:60
       Phase     Function
       9         LP:05
       Phase     Function
       10        LP:12
       Phase     Function
       11        JP:01
30
                                                   www.k echnology.it
NE-1010/510/511 High Pressure                                                                    New Era Pump Systems™
                                                      Repeat if
                                Infuse until          Program Input
                                Event Trigger         low
          Infuse
                                                           
                                                     :01
                                                                     
          Withdraw
                                                           :10       :10
                     TTL High TTL Low
                     TTL Program Ouput
                                                               Event Trigger
Event Trigger
       This example demonstrates a complex interaction with external equipment, such as synchronizing with another
       syringe pump. The Program includes a variety of interactions with external equipment, which demonstrates
       the various control possibilities of the NE-1000.
       The Pumping Program begins by canceling any previous event traps (PH:01) and raising the ‘Program Output’
       TTL line (PH:02). After 5.0 mL has been dispensed at 800 mL/hr (PH:03), the ‘Program Output’ TTL line is
       lowered (PH:04), sending a synchronization signal to another device.
       The pump then continues to pump at 800 mL/hr (PH:06) until a synchronization signal is received at the
       ‘Event Trigger’ TTL input, causing the Program to jump to Phase 7 (PH:05).
       The pump then withdraws 0.25 mL (PH:07), pauses for 1 second (PH:08), then repeats this process if the
       Program Input TTL line is low (PH:09), otherwise it continues with the next Phase.
       Next, the pump pauses for 10 seconds (PH:10). Then it pauses again for the lesser of another 10 seconds
       (PH:12) or until an Event Trigger occurs (PH:11). The Program then restarts (PH:13).
       Phase         Function
       1             ET:RS
       Phase         Function
       2             OUT.1
       Phase         Function     Rate               Volume                 Direction
       3             RATE         800.0 mL/hr        5.0 mL                 Infuse
       Phase         Function
       4             OUT.0
       Phase         Function
       5             ET:07
31
                                                            www.k echnology.it
NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
       Phase     Function
       1         TR:ET         Redirect trigger input to the Event Trap
       Phase     Function
       2         ET:05         Set Trigger Event Trap to start syringe refill
       Phase     Function      Rate             Volume             Direction
       3         RATE          1000.0 mL/hr     1.5 mL             Infuse       Begin bolus dispense
       Phase     Function      Rate             Volume             Direction
       4         RATE          500.0 mL/hr      0.0 mL (OFF)       Infuse       Then infuse continuously
                  On the rear of the pump are 2 square RJ-11 (“phone jack” style) sockets.
       Connect the RS-232 cable into the socket labeled “To Computer”. Connect the other end to the serial port on
       the computer, or other control device. Turn power off to the pump and the computer before connecting cables.
       If the pump is part of a pump network, connect a pump network cable between the socket labeled “To
       Network”, on the first pump, and the socket labeled “To Computer” on the next pump in the network. Repeat
       for each pump in the network, connecting the “To Network” socket of one pump to the “To Computer” socket
                                                              32
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
       on the next pump in the network. Up to 100 pumps can be networked together with a computer. See section
       11, Logic Interface: TTL Input and Output, for a diagram of the rear of the pump. When communicating with
       a pump in a multi-pump network, each preceding pump in the network must be powered on.
       Each pump in the network needs a unique network address to identify the pump to the computer. Network
       addresses are from 00 to 99. If the network consists of only 1 pump, set the pump’s address to 0. Also, each
       pump needs to be set to the same baud rate as the computer. Use the ‘Setup’ function on the keypad to set the
       network address and the baud rate. See section 6.12, ‘Setup’ . The '*ADR' command can also be used to set
       the network address.
       The supported baud rates are 300, 1200, 2400, 9600, and 19200. The trade-off on baud rates is
       communications speed versus noise immunity. For most environments, 19200 would be acceptable. But in
       environments that are electrically noisy and/or over long cables, the communications signal may degrade or be
       disrupted, causing communications errors. In these situations, a lower baud rate may improve the reliability of
       the communications.
33
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                           New Era Pump Systems™
       <float> => <f> [ <float> ]                              Floating point number. Maximum of 4 digits plus 1
                                                               decimal point. Maximum of 3 digits to the right of the
                                                               decimal point.
       <volume units> => UL                                    μL (microliters)
                         ML                                    mL (milliliters)
       <TTL level> =>          1                               TTL high level
                               0                               TTL low level
       <on-off> =>             1                               On, enabled
                               0                               Off, disabled
       <phase data> => <n> [<n>]                               Program Phase number. Valid values: 1 to 41
       <count data> => <n> [<n>]                               Valid values: 1 to 99
       <number data> => <n> [<n>]                              Valid values: 0 to 99
       <text> => "any printable character" [<text>]
       <f> => { <n> | . }                                      Floating point digits
       <n> => { 0 | 1 | 2 | 3 | 4 | 5 | 6 | 7 | 8 | 9 }        Digits
       <byte> => “one byte of any data”
       ()                                                      One byte of data expressed as (0xhh), where ‘hh’ is the
                                                               data in hexadecimal.
       =>                                                      Is defined by. Syntax expands to next level of expansion.
       <>                                                      Non-terminal syntax expansion
           []                                                  Optional syntax
       {}                                                      Required syntax
       |                                                       Or. Choose one of the syntax options.
       λ                                                       None. Syntax expands to nothing (lambda production).
       ""                                                      Description of syntax expansion
       10.2.2 RS-232 Protocol: Basic Mode
       Command syntax (to pump):
            <basic command protocol> => <command data> <CR>
       Response syntax (from pump):
              <basic response protocol> => <STX> <response data> <ETX>
       In the “Basic” communications mode, a master-slave protocol is used, whereby the pump will only transmit in
       response to a received command.
       When the pump receives the <basic command protocol>, <command data> will automatically be stripped of
       all space and control characters, and all text will be converted to upper case. This simplifies communications
       with the pump when commands are being manually typed in from a generic terminal emulator.
       To return the pump to Basic mode when in the Safe mode, send the following packet to the pump:
                (0x2) ( 0x8) SAF0 (0x55) (0x43) (0x3)
       10.2.3 RS-232 Protocol: Safe Mode
       Command syntax (to pump):
            <safe command protocol> => <STX> <length> <command data> <CRC 16> <ETX>
       Response syntax (from pump):
              <safe response protocol> => <STX> <length> <response data> <CRC 16> <ETX>
       Safe mode uses a more structured protocol including detection of corrupted communications, communications
       time outs, and auto-alarm responses. Safe mode is enabled using the ‘SAF’ command whose parameter setting
       is stored in the non-volatile memory.
34
                                                          www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
       Safe mode uses a modified master-slave protocol, whereby the pump transmits in response to a received
       command. But, the pump also automatically transmits a status packet when an alarm condition occurs.
       Corrupted communications is detected using the 16 bit CCITT CRC algorithm computed over <transmitted
       data>. Packets transmitted and received include the CRC within the packets.
       The parameter sent with the “SAF” command is the communications time out. This time out, in seconds, is the
       time between the reception by the pump of consecutive valid communications packets. Each time a valid
       communications packet is received, the time out is reset. If the time out elapses, a pump alarm will occur,
       stopping the pump and the Pumping Program. The pump will display                  , and the buzzer will sound,
       if alarms are enabled, alerting the user. The communications time out timer will not restart until the next
       reception of a valid packet.
       In addition, there is a 0.5 second packet inter-byte time out. While receiving a communications packet, and
       before its complete reception, if a delay of 0.5 seconds occurs between bytes, the incomplete packet will be
       discarded.
       With the Auto-Alarm feature, whenever a pump alarm occurs, such as a pump stall, a response packet with the
       alarm status information will automatically be transmitted.
       Until the Safe Mode is disabled, each time power is applied to the pump, the pump defaults to the Safe mode
       of communications, but the communications time out timer will not be enabled until the first reception of a
       valid packet.
       Although the communications time out timer is not enabled, the Auto-Alarm feature will be enabled.
       Therefore, the pump will be in an Auto-Alarm only communications mode.
       When power is applied to the pump, or if the system should reset, a system reset alarm occurs. The Auto-
       Alarm feature, therefore, alerts the host computer that a pump reset has occurred.
       Also, when the user changes the baud rate, the communications time out timer is disabled until the next valid
       communications packet.
       10.2.4 RS-232 Protocol: Basic and Safe Mode Common Syntax
       <transmitted data> => { <command data> | <response data> }
       <command data> => [<address> | * ] [<command>]                        To pump
       <response data> => <address> <status> [ <data> ]                      From pump
       <status> => { <prompt> | <alarm> }                                    Operational state of pump
       <prompt> =>
                I                                   Infusing
                W                                   Withdrawing
                S                                   Pumping Program Stopped
                P                                   Pumping Program Paused
                T                                   Timed Pause Phase
                U                                   Operational trigger wait (user wait)
                X                                   Purging
       <alarm> => A ? <alarm type>                  Alarm
       <alarm type> =>
               R                                      Pump was reset (power was interrupted)
               S                                      Pump motor stalled
               T                                      Safe mode communications time out
               E                                      Pumping Program error
               O                                      Pumping Program Phase is out of range
       <address> => <n> [ <n> ]                       Pump network address, 0 to 99
       *                                              System command (overrides network address)
       <data> => <text>                               Response to command
       <CR> => (0x0D)                                 Carriage return
       <STX> => (0x02)                                Start of packet transmission indicator
       <ETX> => (0x03)                                End of packet transmission indicator
       <CRC 16> => <byte> <byte>                      16 bit CCITT CRC of <transmitted data> (high byte, low byte)
       <length> => <byte>                             Number of bytes remaining in packet, including this byte
35
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
       Example: Simultaneously change the pumping rates of 3 pumps on a pump network as follows:
              Pump 0: 100 mL/hr
              Pump 1: 250 mL/hr
              Pump 2: 375 mL/hr
       Assuming that the current pumping rate units of all 3 pumps are currently mL/hr, send the following
       command, followed by a carriage return <CR>, (spaces are optional):
                 0 rat 100 * 1 rat 250 * 2 rat 375 *
36
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                      New Era Pump Systems™
37
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                       New Era Pump Systems™
       PUMPING RATE
       RAT [ C | I ] [ <float> [ <rate units> ] ]
                Set/query pumping rate.
                <rate units> =>    UM               = μL/min
                                   MM               = mL/min
                                   UH               = μL/hr
                                   MH               = mL/hr
                Applicable only with “Rate Data Functions”.
                When setting the pumping rate, if the current Phase’s function is not ‘RATE’, then <rate units> is not
                applicable.
                While pumping, the pumping rate can only be set if the current Phase function is ‘RATE’ and the next
                Program Phase’s function to be executed is not ‘INCR’ or ‘DECR’. Also, while pumping, <rate
                units> cannot be set.
                The new pumping rate will only be stored in non-volatile memory if the Pumping Program is not
                operating.
                When the pumping rate is queried while pumping, the response will be the current pumping rate and
                units. Otherwise, the response will be the rate setting and units, if applicable. With the ‘INCR’ and
                ‘DECR’ functions, these two responses are not the same.
                RAT C <float> [ <rate units> ]
                Allows the Pumping Program to continue after the pumping rate is changed. The RAT C command
                will not clear a Paused Pumping state after the rate is changed when the Pumping Program is paused.
                Normally, the Pumping Program would re-start from Phase 1 after the rate is changed while paused.
                RAT I <float> [ <rate units> ]
                Changes the pumping rate only if the pump is infusing. If the pumping direction is set to withdraw,
                the rate change will be ignored.
       VOLUME TO BE DISPENSED AND SET VOLUME UNITS
       VOL [ <float> | <volume units>]
                Set/query volume to be dispensed: <float>
                Applicable only with “Rate Data Functions”. Can only be set when the Pumping Program is not
                operating. The volume units are set according to the current syringe diameter setting. Do not send
                the volume units when setting the volume.
                Example: VOL 12.45        Sets the current phase “Volume to be Dispensed” to 12.45.
                Set volume units: <volume units>
                Overrides the default volume units set when the syringe diameter is set.
                Example: VOL UL           Sets all volume units to μL.
                Query response:
                          <float> <volume units>
       PUMPING DIRECTION
       DIR [ INF | WDR | REV | STK ]
                Set/query pumping direction
                INF       = Infuse
                WDR       = Withdraw
                REV       = Reverse pumping direction
                STK       = “Sticky Direction” (See “Sticky Direction”, sec: .6.8.1)
                Applicable with all Program Phase functions. Cannot be set when the Pumping Program is operating
                and the “Volume to be Dispensed” is non-zero.
                The pumping direction cannot be changed if an alarm condition exists.
38
                                                       www.k echnology.it
NE-1010/510/511 High Pressure                                                                      New Era Pump Systems™
               If the Pumping Program is paused, the stop command will cancel the pause and reset the Pumping
               Program to Phase 1.
       VOLUME DISPENSED
       DIS  Queries volume dispensed only. Set not applicable.
            Response:
                        I <float> W <float> <volume units>
               Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
               volume.
       CLEAR VOLUME DISPENSED
       CLD { INF | WDR }
               Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
               Program is not operating. Query is not applicable.
               INF      = Infusion volume
               WDR      = Withdrawn volume
               Query is not applicable.
       10.4.3 Configuration and Setup Commands
       New settings for any of the following commands will be stored in the non-volatile memory.
       LOW MOTOR NOISE
       LN [ <on-off> ]
               Set/query low motor noise mode. Increases motor stepping resolution.
39
                                                  www.k echnology.it
NE-1010/510/511 High Pressure                                                                      New Era Pump Systems™
               If the Pumping Program is paused, the stop command will cancel the pause and reset the Pumping
               Program to Phase 1.
       VOLUME DISPENSED
       DIS  Queries volume dispensed only. Set not applicable.
            Response:
                        I <float> W <float> <volume units>
               Where: “I <float>“ refers to the infusion volume dispensed, and “W <float>” refers to the withdrawn
               volume.
       CLEAR VOLUME DISPENSED
       CLD { INF | WDR }
               Sets the Infused or withdrawn volume dispensed to 0. Command is only valid while the Pumping
               Program is not operating. Query is not applicable.
               INF      = Infusion volume
               WDR      = Withdrawn volume
               Query is not applicable.
       10.4.3 Configuration and Setup Commands
       New settings for any of the following commands will be stored in the non-volatile memory.
       LOW MOTOR NOISE
       LN [ <on-off> ]
               Set/query low motor noise mode. Increases motor stepping resolution.
40
                                                  www.k echnology.it
NE-1010/510/511 High Pressure                                                                     New Era Pump Systems™
               ‘Volume’ and Pumping Direction can be changed. Cannot be enabled when the Pumping Program is
               currently programmed with a multiple Phase Program.
       SET KEYPAD AND NOTIFICATION BEEP
       BP [ <on-off> ]
               Set/query keypad and notification beep mode. Set beep enables or disables beep mode.
       10.4.4 General Control and Status Commands
       TTL I/O OUTPUT SETTING
       OUT <n> <TTL level>
               Sets TTL level on user definable output pin on the ‘TTL I/O’ connector.
               <n> Indicates pin number on ‘TTL I/O’ connector
                        Valid value: 5 (Program Output pin)
              Query is not applicable.
       TTL INPUT QUERY
       IN <n>
              Queries TTL level of pin on ‘TTL I/O’ connector. Set is not applicable.
               <n> Indicates pin number on ‘TTL I/O’ connector
                        Valid values: 2, 3, 4, and 6.
               Query response: <TTL level>
       BUZZER
       BUZ [ 0 | { 1 [ < n > ] } ]
               Sets / queries buzzer
               Set:     0 = Turn buzzer off;
                        1 = Turn buzzer on
                                if <n> specified
                                         If <n> = 0, buzzer beeps continuously,
                                         otherwise buzzer beeps <n> times
                                if <n> not specified, buzzer sounds continuously
               Query response: { 0 | 1 }
                       0 = Buzzer off
                       1 = Buzzer is on continuously or beeping.
       10.4.5 System Commands
       SET PUMP NETWORK ADDRESS AND BAUD RATE
       * ADR [ <address> [ B { 19200 | 9600 | 2400 | 1200 | 300 } ] ]
             [DUAL | RECP | ALTR]
               Set/query pump network address
               <address> => <n> [<n>]
               <address> Valid range: 0 to 99
               B { 19200 | 9600 | 2400 | 1200 | 300 } will change the baud rate as indicated. NOTE: The command
                 response and all further communications will be at the specified baud rate.
               Special communications modes
               DUAL Sets the pump to Dual Pumping mode with a secondary pump.
               RECP Sets the pump to Reciprocating Pumping mode with a secondary pump.
               ALTR Sets Alternating pumping mode.
                    All special communications modes use 19,200 baud rate.
       This is a special system command that will be accepted by the pump regardless of its current address or mode.
       Once set, the pump will only respond to commands with the set address and at the specified baud rate.
41
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
       Note: Once a special communications mode is set, the pump will only respond to commands that are preceded
       by the ‘*’ character. To exit special communication mode, reset the address: * ADR 0
       Example: *ADR                       Query current address setting
                  *ADR 3                   Set pump network address to 3. The pump will now only respond to
                                           commands with address 3
                  *ADR 5 B 1200            Set the pump network address to 5 and the baud rate to 1200. The
                                           command response will be at 1200 baud
       ENABLE SAFE COMMUNICATIONS MODE
       SAF [ <time out> ]
                Set/query Safe communications mode setting.
                <time out> => <n> [ <n> [ <n> ] ]
                <time out> Valid range: 0 to 255.
                If <time out> = 0 then Basic communication mode is set, disabling Safe mode,
                If <time out> > 0 then Safe communications mode is enabled. After the reception of this command,
                valid communications must be received every <time out> seconds.
       FIRMWARE VERSION QUERY
       VER  Response: NE<model>V <n> . <nnn>
                    where ‘<n>.<nnn>‘ is the current firmware version number.
                Set is not applicable.
       MASTER PROGRAM RESET
       * RESET
              Clears program memory and resets communication parameters to Basic mode and address 0.
       This is a special system command that will be accepted by the pump regardless of its current address.
42
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                              New Era Pump Systems™
                    12                                                                                         Pin
                                                                                                                1
                                                                     Pin
                                                                      9
43
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                             New Era Pump Systems™
44
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
         Stop Only:        Falling edge Stops the Pumping Program. This configuration only allows the
                           stopping/pausing of the Pumping Program. This would be useful, for example, with an end
                           of travel limit switch. Also, this switch can be used as a power on homing switch.
         Stop Only Reversed: Same as ‘Stop Only’, but operates on the Rising edge to stop the Pumping Program.
         Run on Low Level:      Starts the pump whenever the level is low
         Run on High Level:     Starts the pump whenever the level is high
         Stop on Low Level:     Stop the pump whenever the level is low
         Stop on High Level:    Stop the pump whenever the level is high
         Trigger off:           Trigger control is disabled
         Event Trap Trigger:    Redirects a Stop Operational Trigger to the Event Trap.
         Button Event Trigger: Redirects ‘Stop’ key to the Event Trap.
         Pump Motor Operating (Pin 7): This output provides an external signal indicating when the pump motor is
         operating. This pin is configured with the           n  setup command, or the "ROM" remote command.
         When set to 0, the output is only at logic high when the motor is operating (pumping). When set to 1, the
         output is logic high when the motor is operating or when the Pumping program is executing a pause timer.
         Otherwise, the output is a logic low.
         Pumping Direction Controls (Input: Pin 3; Output: Pin 8): Allows bi-directional control of the pumping
         direction. The input pin, when activated, emulates the pumping direction key, changing the pumping direction.
         This function, therefore, is only applicable where the pumping direction key would be applicable. The
         function of the input pin is configured with the       aa     setup command, or the "DIN" remote command.
         When the mode is set to reciprocating pumps ("rE" setup command), then if the current pumping direction is
         withdraw, a falling edge sets the direction to infuse. If the current pumping direction is infuse, a rising edge
         sets the direction to withdraw. Otherwise, this input pin has no effect.
         When the mode is set to dual pumps ("dU" setup command), then if the current pumping direction is withdraw,
         a rising edge sets the direction to infuse. If the current pumping direction is infuse, a falling edge sets the
         direction to withdraw. Otherwise, this input pin has no effect.
         Dual and reciprocating pumping systems are created using 2 pumps attached with the accessory cable CBL-
         TTL-1.
         The output pin provides an output signal to external devices indicating the direction of pumping. A logic low
         indicates withdraw, and a logic high indicates infuse. For example, this pin can be used to control an external
         valve, allowing the syringe to refill from a reservoir.
45
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                                         New Era Pump Systems™
       12. Appendix
       12.1 RS-232 Command Summary
       <command> =>
       DIA [ < float > ]                                  Syringe inside diameter
       PHN [ < phase data > ]                             Program Phase number
       FUN [ < phase function > ]                         Program Phase function
            < phase function > =>
                        RAT                               Pumping rate. ‘RATE’
                        FIL                               Fill syringe to volume dispensed. ‘FILL’
                        INC                               Increment rate. ‘INCR’
                        DEC                               Decrement rate. ‘DECR’
                        STP                               Stop pump. ‘STOP’
                        JMP <phase data>                  Jump to Program Phase. ‘JP:nn’
                        PRI                               Sub-Program Selection Input. 'Pr:In'
                        PRL <number data>                 Sub-Program Start Label definition. 'Pr:nn'
                        LPS                               Loop starting Phase. ‘LP:ST’
                        LOP <count data>                  Loop to previous loop start ‘nn’ times. ‘LP:nn’
                        LPE                               Loop end Phase. ‘LP:EN’
                        PAS <number data>                 Pauses pumping for ‘nn’ seconds. ‘PS:nn’
                        PAS [n.n]                         Pauses pumping for 'n.n' seconds. 'PS:n.n'
                        IF   <phase data>                 If Program input low, jump to Program Phase. ‘IF:nn’
                        EVN <phase data>                  Set event trigger. ‘Et:nn’
                        EVS <phase data>                  Set event square wave trigger. 'ES:nn'
                        EVR                               Event trigger reset. ‘Et:RS’
                        CLD                               Clear total dispense volume. ‘CLR.D’
                        TRG <nn>                          Override default operational trigger configuration ‘tr:aa’
                        BEP                               Sound short beep. ‘BEEP’
                        OUT { 0 | 1 }                     Set programmable output pin. ‘OUT.n’
       RAT [C | I ] [ <float> [ UM | MM | UH | MH ] ] Pumping rate
       VOL [ <float> | <volume units>]                    Volume to be Dispensed, or set Volume units
       DIR [ INF | WDR | REV | STK ]                      Pumping direction
       RUN [ <phase data> ]                               Starts the Pumping Program
            [E [<phase data>] ]                           Pumping Program event trigger
       PUR                                                Start purge
       STP                                                Stop/pauses the Pumping Program or purge
       DIS                                                                        Query volume dispensed
       CLD { INF | WDR }                                                          Clear volume dispensed
       SAF [ <n> [ <n> [ <n> ] ] ]                                                Safe communications mode
       LN [ 0 | 1]                                                                Low motor noise mode
       AL [ 0 | 1]                                                                Alarm buzzer mode
       PF [ 0 | 1]                                                                Power failure auto-restart mode
       TRG [ FT | FH | F2 | LE | ST | T2 | SP | P2 | RL | RH | SL | SH | OF ] Operational trigger default mode
       DIN [ 0 | 1]                                                               Directional input control mode
       ROM [0 | 1]                                 Pump Motor Operating TTL output mode
       LOC [ P ] [ 0 | 1]]                         Keypad lockout mode or Program Entry Mode lockout
       BP [ 0 | 1]                                 Key and Notification beep mode
       OUT 5 { 0 | 1 }                             Set TTL output level
       IN { 2 | 3 | 4 | 6 }                        Query TTL input level
       BUZ [ 0 | { 1 [ < n > ] } ]                 Buzzer control
       VER                                         Query firmware version
                             System Commands: Valid regardless of current network address
       *ADR [ <n> [<n>] [ B <baud-rate>] ]         Network address and baud rate
       *ADR [ DUAL | RECP | ALTR ]                        Set Reciprocating, Dual, or Alternating pumping mode
       *RESET                                             Resets pump. Clears program memory and resets setup.
       ------------------------------------------------------------------------------------------------------------------------------
       Network Command Burst => <n> <command> * [Network Command Burst]
       - Communicate simultaneously with multiple pumps on a pump network.
       Any command preceded by the asterisk symbol ‘*’ bypasses the pump network address and Safe mode.
46
                                                            www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
       12.3 Accessories
       12.3.1 ANA-BOX™
       Part #: ADPT-ANABOX, Analog voltage control interface.
       Allows the pump to be controlled by a variable voltage source. Creates a closed loop system allowing the
       pump to be controlled by an external sensor, such as a pressure sensor, or other variable voltage source. Start
       or stop the pump at a specific voltage level. Set the pumping rate to be proportional to the voltage input.
47
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                         New Era Pump Systems™
48
                                                     www.k echnology.it
NE-1010/510/511 High Pressure                                                                     New Era Pump Systems™
        Pump doesn’t stop after dispensing a set volume: The pump was previously setup with a multiple Phase
        Pumping Program. To simply dispense a fixed volume at a fixed pumping rate, the second Program Phase
        must be the ‘Stop’ function. See section 9.1, How to Enter Pumping Programs, for instructions on changing
        the Pumping Program.
        Pump stops pumping after a period of time: A dispense target volume has been set. Verify that the
        dispense target volume is set to 0.
        12.5 Specifications
                  Mechanical & Electrical                                       Operational
        Syringe sizes:          Up to 60 mL (140 mL           Maximum speed:             5.1005 cm/min
                                partially filled)
        Number of syringes:     1                             Minimum speed:             0.004205 cm/hr
        Motor type:             Step motor                    Maximum pumping            1699 mL/hr with a B-D
                                                              rate:                      60 mL syringe
        Motor steps per         400                           Minimum pumping            0.73 μL/hr with a B-D 1
        revolution:                                           rate:                      mL syringe
        Microstepping:          1/8 to 1/2 depending on       Maximum force:             35 lbs. at minimum
                                motor speed                                              speed, 18 lbs. at
                                                                                         maximum speed
        Advance per step:       0.2126 um to 0.8504 um        Number of Program          41
                                depending on motor speed      Phases:
        Motor to drive screw    15/28                         RS-232 pump network:       100 pumps maximum
        ratio:
        Drive screw pitch:      20 revolutions/”              RS-232 selectable baud     300, 1200, 2400, 9600,
                                                              rates:                     19200
        DC connector:           2.1 mm, center positive       Syringe inside diameter    0.100 to 50.00 mm
                                                              range:
        Voltage at DC           12V DC at full load
        connector:
        Amperage:               800 mA at full load
        Power supply type:      Unregulated linear external wall adapter, country and power source specific (or
                                compatible regulated power supply)
        Power supply output     12V DC @ 800 mA
        rating:
        Dimensions:             8 3/4” x 5 3/4” x 4 1/2” (LxWxH) (22.86 cm x 14.605 cm x 11.43 cm)
        Weight:                 3.6 lbs. (1.63 kg)
        Allen Wrench            3/32 Hex
49
                                                   www.k echnology.it
NE-1010/510/511 High Pressure                                               New Era Pump Systems™
           Publication #1200-01                            50
                                                            50                     04/06/16
                                                  www.k echnology.it
 NE-1010/510/511 High Pressure                                                                          New Era Pump Systems™
                          Phase ‘n+1’
                          Linear Function End Rate
                R
                a                                                        Linear Function Time
                t                                              Hour Minutes Seconds Tenths Seconds
                e                                                Phase ‘n’             Phase ‘n+1’
                          Phase ‘n’
                          Linear Function Start Rate
                                  Time
The Linear Flow Rate function is specified by using two Linear Functions          , in two consecutive Program Phases, defined by
four parameters:
                 1) Starting flow rate
                 2) Ending flow rate
                 3) Total time, expressed as "Hours : Minutes : Seconds : Tenth Seconds"
                 4) Pumping direction
The Linear Flow rate function is entered as follows, in two consecutive Pumping Program Phases:
        Time is entered using the "Volume" key. The "min" minutes LED will be lit when time is displayed.
        The units and pumping directions of the starting and ending flow rates must be the same.
After the set Target Volume is dispensed (or withdrawn), the pumping direction is reversed and the pump will withdraw (or infuse) the
set Target Volume. If the pumping rate is changed, the pump will continue pumping at the new rate when the pumping direction is
reversed.
With the addition of a dual check valve on the syringe, this function can be used to create an auto-refill system.
With the addition of a second pump, attached to the first pump with the control cable, CBL-DUAL-3, the dual pump plumbing kit,
P-DKIT, and the setting of the Reciprocating communications mode, a continuous infusion system can be setup. One pump will
always be infusing while the other pump is refilling.
The dispense time is set using the "Volume" key and is expressed as "Minutes : Seconds". The "min" minutes LED will be lit when
time is displayed.
      nn                                         Stall Event
“Stall Event” allows a Pumping Program to relieve the pressure on a syringe after a pump stall by jumping to a Phase that reverses the
pumping direction.
“Stall Event” creates a background event interrupt trap for a pump stall. If the pump stalls, the Pumping Program will continue
execution with the Program Phase ‘nn’ defined by the “Stall Event” function. After the pump stops, the “Stall” message is cleared and
the Pumping Program continues execution at the Stall Event Phase defined by parameter ‘nn’.
After a “Stall Event” trap is executed, the event trap is cleared. The “Stall Event” function needs to be set again to trap another stall.
“Stall Event” is independent from the standard “Event” and “Event Square” functions.
The “Event Reset” function will clear the “Stall Event” trap along with any other Event trap.
These functions will operate normally, but when the next RET function is executed after the Program Branch function, program
operation will return to the Program Phase immediately following the Phase where the Pumping Program most recently branched
from, or the Program Phase that was interrupted for events:
IF:nn, JP:nn
After one of these functions cause a Program Phase Branch to Phase 'nn', a RET function will return the Program to the Phase
following the function that caused the most recent branch, thus converting these Pumping Program branch functions into program sub-
routine call functions.
ET:nn, ES:nn
These functions create a background Event trap. When the Event trap is triggered, the Pumping Program immediately jumps to
Program Phase 'nn'. A RET function will return to the Program Phase that was interrupted by the Event trap. NOTE: When the
Pumping Program returns to the interrupted Phase, the Program Phase starts over and not at the point that was interrupted. For
example, if the interrupted Phase specified a pumping volume, on return, the entire pumping volume will be dispensed, not just the
remainder of volume from the point of interruption.
RS-232Remote Support:
Return function command:           FUN RET
In addition, this will set the pump’s RS-232 communication to Continuous pump control with a second NE-1000X syringe pump.
Cable CBL-DUAL-3 is required to control the second pump.
4.      Example
Linear Ramp, Sub-Routines, Increment, Decrement
                                               Sub-Routine Calls
                                               to Linear Ramp
                                                                                    Repeat Program
                                                                                    Rate 100 ml/hr
Phase     Function   Rate         Volume   Direction      Start Infusing at 100 ml/hr for 1 ml
1         RATE       100 ml/hr    1.0 ml   Infuse
Phase     Function   Rate         Time     Direction      Increase rate 200 ml/hr and
2         INCR       200          00:10    Infuse         continue infusing for 10 seconds
Phase     Function                                        Call Linear Ramp Sub-Routine
3         JP:08
Phase     Function   Rate         Volume   Direction      Infuse at 500 ml/hr for 5 ml
4         RATE       500 ml/hr    5.0 ml   Infuse
Phase     Function   Rate         Time     Direction      Decrease rate 100 ml/hr and
5         DECR       100          00:10    Infuse         continue infusing for 10 seconds
Phase     Function                                        Call Linear Ramp Sub-Routine
6         JP:08
Phase     Function                                        Repeat Program
7         JP:01
Phase     Function   Rate         Time     Direction          Linearly ramp the infusion rate
8         LINR       0 ml/hr      00:10    Infuse             from 0 ml/hr to 1000 ml/hr over 10
Phase     Function   Rate         Time     Direction          minutes
9         LINR       1000 ml/hr   00:00    Infuse
Phase     Function   Rate         Time     Direction
                                                              Linearly ramp the infusion rate
10        LINR       1000 ml/hr   00:10    Infuse
                                                              from 1000 ml/hr to 0 ml/hr over 10
Phase     Function   Rate         Time     Direction          minutes
11        LINR       0 ml/hr      00:00    Infuse
Phase     Function
                                                              Return to Phase following calling
12        RET                                                 Program function
The above program would be entered from RS-232 with the following commands:
                                                       ‘X Firmware
                                         Display in
                   Mode                                  Version                             Description
                                           Setup
                                                           Only
          Dual Pumps                        dUAL                        Secondary pump duplicates the master pump
          Reciprocating Pumps /                                         Continuous infusion / auto-refill
                                            rECP
          Continuous Infusion
          Alternating Pump Control                           Extra      Second pump starts when first pump stops.
                                            ALtr
                                                            Features
          Next Generation                                               Continuous infusion while minimizing flow rate
                                            COnt               9
          Continuous Infusion                                           pauses and drop outs when changing directions
          Constant Flow Rate                                            Dual pumps inverse linear functions combine for a
                                            GrAd               9
          Gradient                                                      constant flow rate
                                Optional Infusion
     Syringe                    Overlap Volume
     Volume
                                                                            Infusing pump
                                       Infusion Pause and                   changing direction,
                                       Wait Position                        or reaching overlap
                                 Optional                                   volume, triggers
                                  Prime                                     paused pump to
                                 Volume                                     start infusing
The continuous pumping mode eliminates the problem of flow rate drop-off that is                              typical with
continuous syringe pump systems where one pump is refilling the syringe while the                              other infuses, and
then they switch directions.
In this system, the refilling pump refills at a faster rate than the infusing pump giving it time to prime the
syringe, then it pauses and waits for the infusing pump to empty.
Then, when the pumps switch directions, the refilled pump is primed and starts infusing at the set rate
immediately. Additionally, an overlap can be set, whereby the refilled syringe begins infusing before the
infusing pump is completely empty.
                                   RUN.O
                                           Pump 1
                                           Pump 2
                                                                           RUN.1     …
                                           Time
Maintains a constant total flow rate between 2 syringe pumps. A gradient (Linear function) dispensing
function is programmed into the Master pump. The secondary pump will be automatically programmed with
the inverse of the Master pump’s gradient function. The sum of the two pump’s flow rates will be a constant.
Both pumps will increase and decrease pumping rates in tandem.
2 Modes of Operation
- Single cycle: Single ramp up or ramp down, then stop.
          Set TTL setup setting RUN.0
- Continuous cycle: Ramp up, then ramp down, and repeat continuously.
         Set TTL setup setting RUN.1
Pump 2
Pump 1
   •   Starts alternate pump after either a Pause function or when the program stops.
Set the TTL setup setting: RUN.0, (Default setting)
Alternate pump will start on execution of Pause function.
Set the TTL setup setting: RUN.1
Alternate pump will start when the Pumping Program has stopped.
   •   Overlap start of alternate pump.
Pumping Program Function: OUT.0
Immediately sends start to alternate pump when OUT.0 function is executed.
Allows the creation of an overlap between the infusing and the alternate pump to allow the alternate pump to
prime the syringe before the infusing pump stops.
Simultaneously start both pumps: If Phase 1 is set to function OUT.0, the alternate pump will immediately
start when the Master pump starts.
Overrides sending a start command to the alternate pump when the pump stops.
   •   Stop Alternating Mode
Set TTL Trigger Mode to “Off”. When trigger is set to “Off”, start command will not be sent to alternate
pump.
Within a Pumping Program, the trigger mode can be changed with the Trigger Off function: (“tr:of”). This
allows a Pumping Program to limit the number of times that the alternate pump will execute its Pumping
Program.
, are displayed as 2 alternating messages. The indicated message will alternate with a message
        indicating the relevant Pumping Program Phase number.     nnnn      . Pressing any key will clear the
        messages.
        Any RS-232 command that takes a Pumping Program Phase as a parameter will take up to 4 digits to
        indicate the parameter. Responses will also indicate the Phase number as 4 digits:
        Examples
        Set the Pumping Program Phase to Phase 300, then set a jump to phase 100:
               PHN 300
               FUN JMP 100
A query of the Phase and Function will provide the following data responses:
               0300
               JMP0100
                   Synchronizes the operation of two pumps in one            Figure 1: Dual Pumps Cable CBL-DUAL-3
                   of the special communications modes.
Master Pump (connect to "To Computer") Secondary Pump (connect to "To Computer")
                              Display in                             Description
           Mode
                                Setup
  Address mode                  Addr       Communications with PC using cable CBL-PC-PUMP-7.
All modes (except Address) are set to 19,200 baud rate. In Address mode, the next setting is the
baud rate setting.
Except for Alternating Pumps mode, the secondary pump should remain in Address mode
[Ad:00] and 19,200 baud rate [1920] setting, which are the factory defaults. With Alternating
Pumps mode, the secondary pump can also be set to Alternating Pumps mode.
    Syringes: Each pump can have a different size syringe, but make sure that the pumping rates on
    the Master pump are within range of the syringe used on the Secondary pump. If an out-of-range
    pumping rate is sent to the Secondary pump, the rate will just be ignored by the Secondary pump
    without affecting the operation of the Master pump.
    Master Pump: Pumping rate and direction are only transmitted to the Secondary pump from the
    Master pump while the Master pump is pumping. If the Master pump starts or stops, the
    Secondary pump will start or stop. Linear Function pumping rates will not be transmitted.
    Secondary Pump: Changes to the pumping rate and direction on the secondary pump will not be
    transmitted to the Master pump. If the Secondary pump stops, the Master pump will also stop.
Pump Stall: If either pump stalls, then the other pump will also stop.
    Synchronization between pumps: The secondary pump will lag approximately 30 milliseconds
    behind the Master pump due to communications delay.
Pump 2
Pump 1
                                Optional Infusion
         Syringe                Overlap Volume
         Volume
                                                                         Infusing pump
                                       Infusion Pause and                changing direction,
                                       Wait Position                     or reaching overlap
                                 Optional                                volume, triggers
                                  Prime                                  paused pump to
                                 Volume                                  start infusing
      The continuous pumping mode eliminates the problem of flow rate drop-off that
      is typical with continuous syringe pump systems, where one pump is refilling the
      syringe, while the other infuses, and then they switch directions.
      In this system, the refilling pump refills at a faster rate than the infusing pump, giving it time to
      prime the syringe, then it pauses and waits for the infusing pump to empty.
      Then, when the pumps switch directions, the refilled pump is primed and starts infusing at the set
      rate immediately. Additionally, an overlap can be set, whereby the refilled syringe begins
      infusing before the infusing pump is completely empty.
      Requirements
         • 2 Pumps from the NE-1000X series of syringe pumps. Pumps can be different models.
         • 1 Dual pump cable, part #CBL-DUAL-3.
      Set communications mode to Continuous Mode [Cont]
      On the Master Pump only: In setup, when the network address is displayed [Ad:00], or other
      address mode, press the left-most up arrow key to select [COnt].
      Note: Do not change the default settings on the second pump: Address 0 [Ad:00] and 19,200
      baud rate [1920]. If unsure, reset the secondary pump: While turning on power, hold the left-
      most up arrow button. The display will show [rESt].
      Set Pumping Parameters on the Master Pump
      Set syringe diameter and pumping parameters on the Master pump. The syringe diameter and
      pumping parameters are transmitted to the second pump when the Master pump starts. Both
      syringes start empty, unless pre-filled syringes are specified.
                             Pumping Parameters
                   Phase # Function         Rate        Volume
                     1     RATE         Infusion       Syringe Volume
                     2     RATE         Refill         Overlap     ÍOptional parameters
                     3     RATE         Prime          Prime       ÍOptional parameters
                     4     RATE         Pre-filled Syringe Mode (Optional Parameter)
                                        00 = Both empty (default)        11 = Both pre-filled
                                        10 = Master only                 01 = Secondary only
      Set any optional parameter to 0.000 (off) if not used, or set the function to STOP. Pumping rates
      not set will default to the Infusion Pumping Rate.
      Hardware Setup
      Attach the cable, CBL-DUAL-3, to the “To Computer” connector on the back of both pumps.
      Operational Sequence
      - Both syringes start empty. Set Phase 4, pre-filled mode, to use pre-filled syringes.
      - Start the Master pump: Both pumps start filling, unless pre-filled syringes are specified.
      - When syringes are filled:
               • Master pump starts infusing
               • Second pump primes (if enabled), then pauses
      - When the infusing pump reaches the overlap position, if enabled, or changes pumping direction,
        the paused pump will start to infuse
      - Sequence repeats:
               • Refilling pump fills, primes, pausing and waits for the other pump.
               • Infusing pump infuses to the overlap position or syringe empty, and then signals the
                  other pump to start infusing. When empty, changes direction and starts refilling
      While Pumping
      The pumping rate can be changed. The new pumping rate will be transmitted to the other pump,
      into the corresponding pumping sequence.
      Pressing the Stop key on either pump, will Pause the sequence on both pumps. Pressing the Start
      key, on the Master pump, will continue the pumping sequence.
      Power Failure Mode – Auto-Synchronization
      After a power failure restart, or to start the infusion with partially filled syringes, limit switches
      can be attached to the pump at the refill position of the syringes. Attach the limit switch wires
      between the Trigger Input (pin 2) and Ground (pin 9) on the pump’s 9 pin TTL connector.
      Then enable Power Failure Mode in the Master pump’s setup menu.
      When the pumps start, they will begin by filling the syringes until either the syringe volume
      target is reached or the limit switch is triggered. Power Failure Mode overrides pre-filled syringe
      mode settings.
                                              RUN.O
                                                       Pump 1
                                                       Pump 2
                                                                                       RUN.1      …
                                                      Time
  Maintains a constant total flow rate between 2 syringe pumps. A gradient (linear function)
  dispensing function is programmed into the Master pump. The secondary pump will be
  automatically programmed with the inverse of the Master pump’s gradient function. The sum of
  the two pump’s flow rates will be a constant. Both pumps will increase and decrease pumping
  rates in tandem.
  2 Modes of Operation
  - Single cycle: Single ramp up or ramp down, then stop.
          Set TTL setup setting RUN.0
  - Continuous cycle: Ramp up, then ramp down, and repeat continuously.
         Set TTL setup setting RUN.1
  Requirement:
      ϭͿ ϮWƵŵƉƐĨƌŽŵƚŚĞEͲϭϬϬϬƐLJƌŝŶŐĞƉƵŵƉƐĞƌŝĞƐǁŝƚŚ͚yĨŝƌŵǁĂƌĞƵƉŐƌĂĚĞƐ͘
      ϮͿ >Ͳh>ͲϯĚƵĂůƉƵŵƉĐŽŵŵƵŶŝĐĂƚŝŽŶƐĐĂďůĞ͘
  Quick Setup:
  On the Master pump only – DO NOT change default settings on the second pump.
      ϭͿ ƚƚĂĐŚƚŚĞĐŽŵŵƵŶŝĐĂƚŝŽŶƐĐĂďůĞ͕>Ͳh>Ͳϯ͕ƚŽƚŚĞ͞dŽŽŵƉƵƚĞƌ͟ƉŽƌƚŽŶƚŚĞďĂĐŬŽĨďŽƚŚ
         ƉƵŵƉƐ͘
      ϮͿ ^ĞůĞĐƚƚŚĞĚĞĨĂƵůƚ'ƌĂĚŝĞŶƚDŽĚĞƉƵŵƉŝŶŐƉƌŽŐƌĂŵ͗
                - Turn on power to the pump WHILE pressing the “Volume / Program Function” key.
                - Use any up-arrow key to select Gradient Mode, displayed as:
                            Volume
                - Press the          key to select, or wait for the time out.
      ϯͿ ^ĞƚƚŚĞƐLJƌŝŶŐĞĚŝĂŵĞƚĞƌŽŶƚŚĞDĂƐƚĞƌƉƵŵƉ͘ŽƚŚƉƵŵƉƐǁŝůůƵƐĞƚŚĞƐĂŵĞĚŝĂŵĞƚĞƌ͘
         ;ĞĨĂƵůƚƉƌŽŐƌĂŵĂƐƐƵŵĞƐĂƚůĞĂƐƚĂϭϰ͘ϰϯŵŵĚŝĂŵĞƚĞƌͿ͘
                        Start
      ϰͿ WƌĞƐƐ                 ŽŶƚŚĞDĂƐƚĞƌƉƵŵƉƚŽƐƚĂƌƚƚŚĞĚƵĂůƉƵŵƉ'ƌĂĚŝĞŶƚƉƵŵƉŝŶŐƉƌŽŐƌĂŵ͘
      Setup:    On the Master pump only – DO NOT change default settings on the second pump.
         ϭͿ ƚƚĂĐŚƚŚĞĚƵĂůƉƵŵƉĐĂďůĞ͕>Ͳh>Ͳϯ͕ƚŽƚŚĞ͞dŽŽŵƉƵƚĞƌ͟ƉŽƌƚŽŶďŽƚŚƉƵŵƉƐ͘
         ϮͿ ^ĞůĞĐƚ'ƌĂĚŝĞŶƚĐŽŵŵƵŶŝĐĂƚŝŽŶƐŵŽĚĞ͗
                   From the Setup menu (Press and hold “Diameter / Setup” key), select Gradient
                   communications Mode (displayed as Grad). When Ad:00 (or other communications
                   mode) is displayed, use the left-most arrow key to select.
         ϯͿ ^ĞƚƚŚĞƐLJƌŝŶŐĞĚŝĂŵĞƚĞƌ͘
         ϰͿ ŶƚĞƌĂ>ŝŶĞĂƌĨƵŶĐƚŝŽŶƉƵŵƉŝŶŐƉƌŽŐƌĂŵĂƐĂϮWŚĂƐĞWƵŵƉŝŶŐWƌŽŐƌĂŵ͗
                  Phase 1: Linear function starting rate, and [hours:minutes] portion of pumping time.
                  Phase 2: Linear function ending rate, and [seconds:tenths] portion of pumping time.
         5) Change RUN.0 (default value), if needed to change from single cycle mode.
         Note: The secondary pump is set to the default communications settings of Address 0 and
         19,200 baud rate.
      Operation
      When the Master pump is started, the Linear function will be expanded to a 3 or 5 Phase
      Pumping program, depending on mode, representing the forward and inverse of the linear
      function, then repeat. The second pump will then be programmed with the syringe diameter and
      the inverse of the linear pumping program. The Second pump will start pumping when the
      Master pump starts pumping.
      With the starting and ending pumping rates reversed on the secondary pump, the sum of the flow
      rates of the two pumps will be virtually constant.
      Note: The linear function updates the pumping rate every 100 ms, resulting in a synchronization
      error of up to 100 ms between the two pumps. This will cause a slight difference in the total
      volumes dispensed between the two pumps.
                                     ADDENDUM NE-1010/NE-510/NE-511
                                                  69
                                                 www.k echnology.it
NE-1010/510/511 High Pressure                                                                        New Era Pump Systems™
         1. Specifications
            Model            Style         Stall Detection   Number of Syringes         Maximum Syringe Size
           NE-1010        Stand-Alone            Yes                 1               60 mL; 140 mL partially filled
           NE-510            OEM                 No                  1               60 mL; 140 mL partially filled
           NE-511            OEM                 Yes                 1               60 mL; 140 mL partially filled
         1.2 Operational
         Maximum force:                               100 lbs. at minimum speed, 18 lbs. at maximum speed
                                          ADDENDUM NE-1010/NE-510/NE-511
                                                       70
                                                    www.k echnology.it
NE-1010/510/511 High Pressure                                                                 New Era Pump Systems™
    then press and hold the ‘Start/Stop’ key. The display will indicate            and the pusher block will move at
    top speed in the set direction. Release the ‘Start/Stop’ key to stop the pump.
   Maintenance
       Periodic lubrication is required for proper operation. Failure of the nut block can occur if not properly
        lubricated.
       Lead-screw: Grease
       Guide Rods: Oil
       Remove accumulation of dirt or debris
   WARNING
   Use extreme caution. The NE-8000 can deliver enough force to break syringes or any objects that get caught in
   the mechanism. Broken syringes and associated plumbing can create dangerous projectiles that can cause bodily
   harm. Keep fingers and loose clothing away from mechanism.
                                                ADDENDUM
                                                       71 NE-8000
                                                  www.k echnology.it
NE-1010/510/511 High Pressure                                                                  New Era Pump Systems™
                                              ADDENDUM
                                                     72 NE-8000
                                                www.k echnology.it
NE-1010/510/511 High Pressure                                                                           New Era Pump Systems™
           Specifications
           Mechanical & Electrical
             0RGHO         6W\OH        6WDOO'HWHFWLRQ   1XPEHURI6\ULQJHV       0D[LPXP6\ULQJH6L]H
                                                                                    P/
            1(    6WDQG$ORQH            1R                  
                                                                                    P/3DUWLDOO\ILOOHGP/
           
           'ULYHEORFNW\SH                            6ROLG0XVWXVH3XUJHIXQFWLRQWRPRYHSXVKHUEORFN
           
           0RWRUW\SH                                  6WHSPRWRU
           0RWRUVWHSVSHUUHYROXWLRQ                  
           0RWRUWRGULYHVFUHZUDWLR                  
           'ULYHVFUHZSLWFK                           UHYROXWLRQV´
           
           3RZHUVXSSO\W\SH                           ([WHUQDOFRXQWU\DQGSRZHUVRXUFHVSHFLILF
           3RZHUVXSSO\RXWSXWUDWLQJ                  9'&#P$
           3RZHUFRQQHFWRU                             PPFHQWHUSRVLWLYH'&
           9ROWDJHDWSRZHUFRQQHFWRU                  9'&DWIXOOORDG
           $PSHUDJH                                    P$DWIXOOORDG
           
           'LPHQVLRQV                                  ´[´[´/[:[+
                                                         FP[FP[FP
           :HLJKW                                      OEVNJ
           Operational
           0D[LPXPIRUFH                               OEV
           
           0LFURVWHSSLQJ                             WRGHSHQGLQJRQPRWRUVSHHG
           $GYDQFHSHUVWHS                            ȝPWRȝPGHSHQGLQJRQPRWRUVSHHG
           
           0D[LPXPVSHHG                               FPPLQ
           0LQLPXPVSHHG                               FPKU
           0D[LPXPSXPSLQJUDWH                        P/PLQZLWKDP/V\ULQJH
           0LQLPXPSXPSLQJUDWH                        P/KUZLWKD%'P/V\ULQJH
           6\ULQJHLQVLGHGLDPHWHUUDQJH               WRPP
           
           1XPEHURI3URJUDP3KDVHV                    
           
           56SXPSQHWZRUN                         SXPSVPD[LPXP
           56VHOHFWDEOHEDXGUDWHV                
           
           
                                                   ADDENDUM
                                                          73 NE-8000
                                                      www.k echnology.it
             KF Technology Srl - Via Amedeo Bocchi, 84/8 - 00125 Roma - Italy
Ph: +39 06.454.34.179 - Mobile: +39 339.533.03.22 - Fax +39 06.972.53.131 - Skype: kft2002
                        info@kftechnology.it - www.kftechnology.it