ENIGMA LAB
Shreenath nagar B/H ST Bus stand, Junagadh, Gujarat
Email-service.enigmalab@gmail.com
CALIBRATION CERTIFICATE
Certificate No.- EN/CAL/KR/01
CUSTOMER NAME : Date of Receipt- 23/02/2023
IMOST ACADEMY (INDIA) PVT. LTD. Date of Issue- 23/02/2033
Njarackal, Kochi, Kerala- 682503 Cal. Date- 23/02/2023
Sug. due date - 23/02/2024
Receipt Cond. - As found
Type of Cal. - Full calibration
Location of Cal. - In house
DETAILS OF DUC/DUT :
Name - AED
Make - M & B ELECTRONICS Calibration SOP No.- SOP-DEF.29
Model - A7000 Ambient Temp. - 28 Degree Centigrade
Serial No. - AE D2208079 Relative Humidity- 23 %
MASTER EQUIPMENT DETAILS :
Name Serial No Validity Traceability from
IMPUSE 7000D DEFIBRILLATOR ANALYZER 5388071 30–FEB 23 Tektronix, Mumbai
THERMOHYGROMETER 9900399 29-MAR-23 Prism.AHM
CALIBRATION REPORT
PHYSICAL INSPECTION :
Condition- As found Alarm - Checked Battery- Normal Temp.
CALIBRATION DATA :
Parameter Standard Reading UUC Reading Error Unit Expanded
Uncertainty (±)
50 51.2 1.2 Joule 1.20
70 71.3 1.3 Joule 1.32
Discharging Energy 90 90.3 0.3 Joule 0.58
100 100.4 0.4 Joule 0.61
150 151.2 1.2 Joule 1.25
200 201.3 1.3 Joule 1.34
Parameter Energy Selected Standard charge UUC Error Expanded
(Joule) time (Sec) Reading (sec) Uncertainty (±)
(Sec)
50 2.6 2.7 0.1 0.36
Charge time 70 3.7 3.8 0.1 0.36
90 4.8 4.9 0.1 0.36
100 5.3 5.4 0.1 0.36
150 8 8.2 0.2 0.39
200 10 10.3 0.3 0.52
Nominal voltage 11.9 V Checked as per manuf. guideline
Nominal current 2.9 Ah Checked as per manuf. Guideline
Battery type Non rechargeable
Battery testing parameters Shelf Life 5 Year on standby – Decaying current checked as per IEC
60601-5 Guideline.
5 year on storage- Decaying current checked as per IEC-
60601-4 Guideline.
Impedance Value UUC Behaviour Result
10 Ohm Not connected to patient Checked & Pass
15 Ohm Not connected to patient Checked & Pass
20 Ohm Patient Detected Checked & pass
60 Ohm Patient Detected Checked & pass
Output impedance test 100 Ohm Patient Detected Checked & pass
150 Ohm Patient Detected Checked & pass
200 Ohm Patient Detected Checked & pass
210 Ohm Not connected to patient Checked & Pass
220 Ohm Not connected to patient Checked & Pass
Entries in the Uncertainty column show the expanded measurement uncertainties evaluated and reported in
accordance with the lSO Guide to the Expression of Uncertainty in Measurement (GUM) The coverage factor (k) is 2
with coverage probability of approximately 95.45%.
The result reported is valid at the time of under the stated conditions of measurement.
Any error in this report should be brought to our knowledge within 15 days from the date of this report.
This certificate shall not be reproduced except in full or part without the written approval of the calibration facility.
ENIGMA LAB certifies that the performance of the above instrument has been verified using test equipment of known
accuracy and had been calibrated using measurement standards traceable to the International System of Units (Sl)
through the National and international Standards.
________________ ___________________
CALIBRATION ENGINEER AUTHORIZED SIGNATORY