5320A
Multifunction Electrical
Tester Calibrator
Extended Specifications
General Specifications
Warm-Up Time.........................................................30 minutes
Specifications Confidence Interval .....................99 %
Temperature Performance
Operating Temperature ........................................18 to 28 C
Calibration Temperature (tcal) ............................23 C
Temperature Coefficient.......................................Temperature coefficient for temperature outside of Tcal 5 C between +5 C to +40 C is
0.1 x /C
Storage Temperature............................................-20 to +70 C
Relative Humidity (operating) ..............................<70 % to 28 C
Altitude
Operating................................................................3,050 m (10,000 ft.)
Storage....................................................................12,200 m (40,000 ft.)
Dimensions ...............................................................450 mm X 480 mm X 170 mm (17.7 in. X 18.9 in. X 6.7 in.)
Weight ........................................................................18 kg (39.7 lbs.)
Power Line ................................................................115/230 V ac (50/60 Hz) 10 %
Power Consumption................................................150 VA Maximum
Safety Class...............................................................Class I, according to EN 61010-1
Electrostatic Discharge ..........................................This instrument meets class I for ESD requirements per EN 61326
W Fuse Protection
AC mains input ......................................................2 A, 250 V for 230 V, Time delay (T2L250V)
4 A, 250 V for 115 V, Fast (F4L250V)
RCD input ...............................................................3.15 A, 250 V, Fast (F3.15L250V)
Meter amps (A) input............................................20 A, 500 V, Fast (F20L500V)
Loop/Line impedance input.................................4 A, 250 V, Time delay (T4L250V)
Leakage current input ..........................................100 mA, 250 V, Fast (F100mAL250V)
Electrical Specifications
Low Resistance Source
Total Range...............................................................100 m to 10 k
Resolution .................................................................3 digits (continuously variable)
Uncertainty and Maximum Ratings
[2]
2-Wire Uncertainty
Maximum
AC[1]or DC
Range
Resolution
(tcal 5 C)
Current
100 m to 4.99 
0.1 m
400 mA
0.3 % + 25 m
5 to 29.9 
0.01 
250 mA
0.2 % + 25 m
30 to 199.9 
0.1 
100 mA
0.2 % + 25 m
200 to 499 
1
45 mA
0.2 %
500  to 1.999 k
1
25 mA
0.2 %
2 to 4.99 k
10 
10 mA
0.2 %
5 to 10 k
10 
5 mA
0.2 %
4-Wire Uncertainty
(tcal 5 C)
0.3 % + 10 m
0.2 % + 10 m
0.2 % + 10 m
0.2 %
0.2 %
0.2 %
0.2 %
Note: [1] Test current can exceed 120 % of maximum current for up to 3 seconds. Terminals automatically disconnect if test current
exceeds 120 % of specified maximum current.
[2] Uncertainty is valid to 200 mW. For higher power rating, add 0.1 % per each 300 mW above 200 mW.
Test Current Measurement
Range..........................................................................0 to 400 mA ac + dc rms
Resolution .................................................................1 mA
Uncertainty................................................................   20 
R = set resistance between 0.5  to 10 k.
 
 + 0.1 mA
 R 
Short Mode
Nominal resistance .................................................<50 m
Maximum current ....................................................400 mA ac + dc rms
Open Mode
Nominal resistance .................................................30 M 20 %
Maximum input voltage allowed .........................50 V ac + dc rms
Test voltage reading...............................................0 to 50 V ac + dc rms
Resolution .................................................................1 V
Uncertainty................................................................5 % + 2 V
5320A
Fluke Corporation
High Resistance Source
Range..........................................................................10 k to 10 G plus 100 G single value selection.
Resolution .................................................................4 Digit (continuously variable for 10 k to 10 G range)
Uncertainty and Maximum Ratings
[1]
Maximum Voltage
Uncertainty
Range
Resolution
(tcal 5 C)
(ac+dc) Peak
10.000 to 39.99 k
1
55 V
0.2 %
40.00 to 99.99 k
10 
300 V
0.2 %
100.00 to 199.99 k
10 
800 V
0.2 %
200.0 to 999.9 k
100 
1100 V
0.2 %
1.0000 to 9.999 M
100 
1100 V
0.3 %
[2]
10.000 to 99.99 M
1 k
1575 V
0.5 %
[2]
100.00 to 999.9 M
10 k
1575 V
1.0 %
[2]
1.0000 to 10.000 G
100 k
1575 V
3.0 %
[2]
[3]
100 G
NA
1575 V
5.0 %
Notes:
[1]
Uncertainty is valid to 500 volts. For test voltages above 500 V, add 0.1% for each 200 V above 500 V.
[2]
Maximum test voltage with the supplied banana leads is 1000 Vrms. For higher voltages, use leads rated at 1575 V or above.
[3]
Calibration value uncertainty is specified in the table. Nominal value uncertainty is 15 %.
Test Voltage Measurement
Range..........................................................................0 to 2000 V dc peak
Resolution .................................................................1 V
Uncertainty................................................................1 % + 5 V for R above 1 M
1 % + 2 V for R below 1 M
Settling Time ............................................................2 seconds for input deviations of <5 %
Short Mode
Nominal resistance .................................................<100 
Maximum input current allowed .........................50 mA ac + dc rms
Test current range ..................................................0 to 50 mA ac + dc rms
Resolution .................................................................0.1 mA
Uncertainty................................................................2 % + 0.5 mA
Resistance Multiplier Adapter (x1000 multiplier)
Resistance range .....................................................350 M to 10 T
Uncertainty................................................................Uncertainty of resistor to be multiplied, +3 % of multiplied value
Maximum test voltage ............................................5500 V dc
Ground Bond Resistance Source
Range..........................................................................25 m to 1.8 k
Resolution .................................................................16 discrete values
Minimum test voltage/current .............................10 V / 10 mA
Uncertainty and Maximum Ratings
Absolute Uncertainty
Maximum
Deviation from
of Characterized
Continuous Test
Test Current
Nominal Value Nominal Value
Resolution
Value
Current [1]
(tcal 5 C)
ACrms or DC
25 m
50 %
 10 m
30 A
100 mA
50 m
50 %
 10 m
28 A
100 mA
100 m
30 %
 15 m
25 A
100 mA
330 m
20 %
 15 m
18 A
100 mA
500 m
10%
 20 m
10 A
100 mA
1
10 %
 20 m
8A
10 mA
1.8 
10%
 30 m
6A
10 mA
5
10 %
 50 m
3.2 A
10 mA
10 
10 %
 80 m
2.0 A
10 mA
18 
10 %
 150 m
1.5 A
10 mA
50 
10 %
 300 m
0.8 A
1 mA
100 
10 %
 500 m
0.5 A
1 mA
180 
10 %
1
0.25 A
1 mA
500 
10 %
 2.5 
0.1 A
1 mA
1 k
10 %
 5
0.05 A
1 mA
1.8 k
10 %
 10 
0.025 A
1 mA
[1]
Test Current
Uncertainty
1.5 % + 0.7 A
1.5 % + 0.5 A
1.5 % + 0.35 A
1.5 % + 0.3 A
1.5 % + 0.2 A
1.5 % + 150 mA
1.5 % + 100 mA
1.5 % + 70 mA
1.5 % + 50 mA
1.5 % + 30 mA
1.5 % + 20 mA
1.5 % + 10 mA
1.5 % + 5 mA
1.5 % + 3 mA
1.5 % + 2 mA
1.5 % + 2 mA
Test currents up to 30 % of maximum continuous test current can be applied to the Calibrator with no time limitation. Test
current between 30 % and 100 % of the maximum continuous test current can be applied to the Calibrator for a limited
time. Minimum period of full current load is 45 seconds. The Calibrator calculates the allowed time period and when
exceeded, the output connectors are disconnected.
Fluke Corporation
5320A
Test Current Measurement
Range..........................................................................0 to 40 A ac+ dc rms
Resolution .................................................................1 mA to 100 mA depending on resistance output and test current
Open Mode
Nominal resistance .................................................>100 k
Maximum voltage ....................................................50 V ac+dc rms
Test voltage range ..................................................0 to 50 V ac+dc rms
Resolution .................................................................1 V
Uncertainty................................................................2 % + 2 V
Line/Loop Impedance Source
Range..........................................................................25 m to 1.8 k
Resolution .................................................................16 discrete values
Minimum test voltage/current .............................10 V/10 mA
Uncertainty and Maximum Ratings
Absolute
Maximum
Maximum
Nominal
Deviation from
Uncertainty of
Continuous Test
Short-term
Resistance
Nominal Value Characterized Value
Current [1]
Test Current[2]
Value
(tcal 5 C)
AC rms or DC
AC rms or DC
25 m
50 %
 10 m
30 A
40 A
50 m
50 %
 10 m
28 A
40 A
100 m
30 %
 15 m
25 A
40 A
330 m
20 %
 15 m
18 A
40 A
500 m
10%
 20 m
10 A
40 A
1
10 %
 20 m
8A
40 A
1.8 
10%
 30 m
6A
30 A
5
10 %
 50 m
3.2 A
21 A
10 
10 %
 80 m
2.0 A
15 A
18 
10 %
 150 m
1.5 A
10 A
50 
10 %
 300 m
0.8 A
5.0 A
100 
10 %
 500 m
0.5 A
3.0 A
180 
10 %
1
0.25 A
1.35 A
500 
10 %
 2.5 
0.1 A
0.6 A
1 k
10 %
 5
0.05 A
0.3 A
1.8 k
10 %
 10 
0.025 A
0.15 A
Test Current
Uncertainty
1.5 % + 0.7 A
1.5 % + 0.5 A
1.5 % + 0.35 A
1.5 % + 0.3 A
1.5 % + 0.2 A
1.5 % + 150 mA
1.5 % + 100 mA
1.5 % + 70 mA
1.5 % + 50 mA
1.5 % + 30 mA
1.5 % + 20 mA
1.5 % + 10 mA
1.5 % + 5 mA
1.5 % + 3 mA
1.5 % + 2 mA
1.5 % + 2 mA
Notes:
[1]
Test currents up to 30 % of maximum continuous test current can be applied to the Calibrator with no time limitation. Test
current between 30 % and 100 % of the maximum continuous test current can be applied to the Calibrator for a limited
time. Minimum period of full current load is 45 seconds. The Calibrator calculates the allowed time period and when
exceeded, the output connectors are disconnected.
[2]
Maximum short term test current is defined as the rms value of halfwave or fullwave test current flowing through the UUT.
Maximum time of test is 200 ms. A time interval of 200 ms represents 10 full waves of power line voltage at 50 Hz and 12
full waves at 60 Hz.
Test Current Measurement
Type of recognized test current...........................Positive impulse (halfwave), negative impulse (halfwave), symmetrical (fullwave).
Range..........................................................................0 to 40 A ac+dc rms
Resolution .................................................................1 to 100 mA depending on test current and resistance output
Prospective Fault Current
Range..........................................................................0 to 10 kA
Correction Manual Mode
Residual Impedance Range ..................................0 to 10 
Resolution .................................................................1 m
Uncertainty................................................................Uncertainty in manual (MAN) mode is the uncertainty of the setup resistor used. See table
above. Also, the uncertainty of the manually entered correction should be taken into
consideration.
Correction Scan Mode
Residual Impedance Range ..................................0 to 10 
Resolution .................................................................1 m
Uncertainty................................................................Resulting uncertainty in SCAN mode is the sum of uncertainty of the scanned residual
resistance and uncertainty of the set resistor per the above table, expressed in ohms.
Correction COMP Mode (Active Loop Compensation) (5320A/VLC only)
Residual Impedance Range ..................................0 to 2 
Maximum Test Current...........................................<25/N A pk, where N equals number of UUT generated test current periods.
5320A
Fluke Corporation
Uncertainty of compensation ...............................15 m. Resulting uncertainty in COMP mode is the sum of the uncertainty of the
compensated residual resistance and the uncertainty of the setup resistor used per the
above table, expressed in ohms. Uncertainty is valid at the point in time when the COMP
function is initiated.
Leakage Current Source
Range..........................................................................0.1 to 30 mA
Resolution:
Passive Mode .........................................................10 A setting, 1 A measurement
Differential Mode...................................................10 A setting, 1 A measurement
Substitute Mode.....................................................10 A
Active Mode (5320A/VLC only)...........................10 A
Test Voltage:
Passive Mode .........................................................60 to 250 V ac+dc rms
Differential Mode...................................................60 to 250 V ac+dc rms
Substitute Mode.....................................................10 to 250 V ac+dc rms
Active Mode (5320A/VLC only)...........................50 to 100 V ac+dc rms
Uncertainty:
Passive Mode .........................................................0.3 % + 2 A ac+dc rms
Differential Mode...................................................0.3 % + 2 A ac+dc rms
Test uncertainty can be influenced by power line voltage instability
Substitute Mode.....................................................0.3 % + 2 A ac+dc rms
Active Mode (5320A/VLC only)...........................0.3 % + 1 A ac+dc rms
RCD (Residual Current Device)
Trip Current Range:
0.5 X I and 1 X I mode: ........................................3 to 3000 mA in 1 mA steps
1.4 X I and 2 X I Mode .........................................3 to 1500 mA in 1 mA steps
5 X I Mode ..............................................................3 to 600 mA in 1 mA steps
Trip Current Measurement Resolution...............1 A on 30 mA range
10 A on 300 mA range
100 A on 3A range
Uncertainty:
0.5 X I and 1 X I mode: ........................................1 % rms
1.4 X I and 2 X I Mode .........................................2 % rms
5 X I Mode ..............................................................5 % rms
Trip Time Range ......................................................10 to 5000 ms
Trip Time Uncertainty ............................................0.25 ms
Series Resistance ....................................................0.025 , 0.05 , 0.1 , 0.33 , 0.5 , 1 , 1.8 , 5 , 10 , 18 , 50 , 100 , 180
, 500 , 1000 , 1800 
Line/Touch Voltage Range....................................250 V
Line/Touch Voltage Uncertainty..........................5 % + 3 V
AC/DC Voltage Calibrator (5320A/VLC only)
Range..........................................................................3 to 600 V, ac or dc
Resolution .................................................................4 digits
Internal Ranges:
AC Mode .................................................................30, 100, 300, and 600 V (Autoranging only)
DC Mode .................................................................30, 150, and 600 V (Autoranging only)
Frequency:
Range......................................................................40 to 400 Hz
Resolution ..............................................................3 digits
Settling Time ............................................................300 ms to 3 s, depending on output value
AC Voltage
Range
3  29.99 V
30  99.99 V
100  299.9 V
300  600 V
Fluke Corporation
Uncertainty and Maximum Burden Current
Uncertainty
Maximum Burden Current
(% of Reading + mV)
0.001 V
0.1 % + 9
500 mA
0.01 V
0.1 % + 30
300 mA
0.1 V
0.1 % + 90
150 mA
0.1 V
0.1 % + 180
50 mA
Resolution
5320A
DC Voltage
Range
3  29.99 V
30  149.9 V
150  600 V
Uncertainty and Maximum Burden Current
Uncertainty
Resolution
Maximum Burden Current
(% of Reading + mV)
0.001 V
0.1 % + 9
2 mA
0.01 V
0.1 % + 45
3 mA
0.1 V
0.1 % + 180
5 mA
AC Output Signal Distortion ..................................0.2 % 10 mV (harmonic distortion and non-harmonic noise from 20 Hz to 500 kHz), for
output power lower than 10 VA on each range.
Sensing Ammeter Current Range........................500 mA
Resolution .................................................................1 mA
Uncertainty................................................................5 mA
Multimeter
Voltage
Range..........................................................................0 to 1100 V ac rms or dc
Resolution .................................................................4 digits
Internal Ranges........................................................10, 100, and 1100 V (Autoranging only)
Frequency Range.....................................................DC, 20 Hz to 2 kHz
Input Resistance ......................................................10 M 1 %
Time Constant ..........................................................1.5 s
Readings/Second.....................................................2
Safety class ...............................................................1000V CAT I, 300 V CAT II
AC/DC Voltage Uncertainty
Range
Resolution
10 V
100 V
1100 V
0.001 V
0.01 V
0.1 V
Uncertainty
(% of Reading + mV)
0.15 % + 5
0.20 % + 50
0.20 % + 550
Current
Range..........................................................................0 to 20 A continuous, 30 A for up to 30 minutes, ac rms or dc
Resolution .................................................................4 digits
Internal Ranges........................................................300 mA, 3 and 30 A (Autoranging only)
Frequency Range.....................................................DC, 20 to 400 Hz
Time Constant ..........................................................1.5 s
Readings/Second.....................................................2
AC/DC Current Uncertainty
Uncertainty
Range
Resolution
(% of Reading + mA)
300 mA
0.1 mA
0.15 % + 0.15
3A
1 mA
0.15 % + 1.5
30 A
10 mA
0.30 % + 15
Phantom Power
Range..........................................................................0 to 33 kVA
Resolution .................................................................3 digits
Uncertainty................................................................ (V )2 + (I )2 where Vunc is specified uncertainty of measured voltage and Iunc is specified
unc
unc
uncertainty of measured current.
10 kV Adapter (1000:1 voltage divider)
Range..........................................................................0 to 10 kV ac rms/dc
Resolution .................................................................4 digits
Uncertainty................................................................0.3 % of value + 5 V dc
0.5 % of value + 5 V ac at 50 or 60 Hz
80K-40 High Voltage Probe
Range..........................................................................0 to 40 kV ac rms/dc
Resolution .................................................................4 digits
Uncertainty................................................................0.5 % of value + 10 V dc
0.5 % of value + 10 V ac at 50 or 60 Hz
5320A
Fluke Corporation
Ordering information
Models	
5320A	
5320A/40 	
5320A/VLC	
	
5320A/VLC/40	
Description	
Multifunction Electrical Tester Calibrator
Calibrator with 40 kV Probe
Calibrator with 600 V Source
and Active Loop Compensator
5320A/VLC Calibrator with 40 kV Probe
Note: All models include the 10 kV divider/resistance multiplier adapter as standard
Accessories
5320CASE	
Y5320	
Rugged Transit Case
Rack Mount Kit (Slides)
Fluke.	Keeping your world
	up and running.
Fluke Corporation	
PO Box 9090, Everett, WA USA 98206
Fluke Europe B.V.	
PO Box 1186, 5602 BD 	
Eindhoven, The Netherlands
For more information call:	
In the U.S.A. (800) 443-5853 or 	
Fax (425) 446-5116	
In Europe/M-East/Africa +31 (0) 40 2675 200 or 	
Fax +31 (0) 40 2675 222	
In Canada (800)-36-FLUKE or 	
Fax (905) 890-6866	
From other countries +1 (425) 446-5500 or 	
Fax +1 (425) 446-5116	
Web access: http://www.fluke.com
2006 Fluke Corporation. All rights reserved. 	
Printed in U.S.A. 8/2006 2727996 D-EN-N Rev A	
Pub_ID: 11144-eng Rev 01