GE Medical Systems Healthcare Institute
Seno DS
Seno DS Physical Installation
rev A 1
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Physical Installation Overview
The objective of this document is to give the GE Field Engineer a pictorial overview on how the
Seno DS is installed,
This document is for training only, and should be used in conjunction with the Senographe DS
Service Manual Manual. The Seno DS Service Manual should be used for exact procedures on
the physical installation.
rev A 2
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Seno DS arrives in three large
crates:
1. Control Station
2. Gantry Generator
3. Detector/Books/
Paddles/
Hardware/Cables/
etc.
rev A 3
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
!
To allow for clearance through
hospital doorways, the
gantry/generator crate will
probably need to be taken apart
either outside or at the loading
dock
The gantry and generator are
already connected together
rev A 4
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Gantry will need to be raised
high enough to allow the wood
blocks to be removed.
The lifting screws attach the base
plate to the pellet using the red flat
metal piece). You remove these
lifting screws from here…….
…….and screw it down here. Use
the red flat metal piece for the lifting
screw base
Wheels are not shown in this picture
rev A 5
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
When gantry has been raised
high enough, remove the
wooden blocks underneath the
gantry.
Wheels are not shown in this
picture
rev A 6
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Removal of lifting bolt
rev A 7
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Wheel Assembly
Comes shipped with system
rev A 8
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Remove bolts that stabilize
generator
rev A 9
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Lower the generator supports to
allow for the removal of the
blocks of wood
(there are 4)
rev A 10
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Ready to roll!
rev A 11
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
There are ramps to allow the
generator and gantry to roll off
the skid.
Remember the Generator and
the Gantry are cabled together
It is probably best to remove at least
this cover on the generator. The
reason for this is to allow more
clearance for the cables that exit
underneath the generator. They
won’t drag on the ground and get
beat up as much!
rev A 12
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Uncrating
Console (AWS)
The (AWS) now comes shipped
with a wheel assembly
rev A 13
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
When placing gantry at it’s
destination, use the lifting screws
along with the floor protectors to
lower the gantry to the floor
rev A 14
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Dolly's have been removed.
Lowering the gantry to the floor.
rev A 15
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
There are four screws that you
need to access to allow the
removal of the “omega” cover
rev A 16
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
With the “omega” cover
removed, continue removing
the nuts that attach the control
station to the skid
rev A 17
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Console cable
rev A 18
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Monitor support
rev A 19
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Console cable routing
rev A 20
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Detector, there is a lot to be
discussed here. Please use the
proper job card.
rev A 21
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Connect fiber optics, fluid lines,
and power connections to
detector (in that order) Fiber
optic lines are small and fragile.
Do not over tighten the power
cable connector screws! They
will break!
rev A 22
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Detector connections
rev A 23
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Detector
rev A 24
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
AWS (Control Station)
rev A 25
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Installation of glass shield
The new control stations have a
U bracket that supports the
glass shield while it is being
mounted.
rev A 26
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
Seno DS Installing
Finished!
rev A 27
Proprietary to GE Medical Systems 2003 B.Sidor rev 3
GE Medical Systems Healthcare Institute
Seno DS
End of Presentation
rev A 28
Proprietary to GE Medical Systems 2003 B.Sidor rev 3