ML255 Electromagnetic Flow Meter Converter: Installation, Operation and Maintenance Manual
ML255 Electromagnetic Flow Meter Converter: Installation, Operation and Maintenance Manual
Electromagnetic
Flow Meter Converter
             www.mccrometer.com
Contents
SAFETY . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  1
1.0        CONVERTER OVERVIEW . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
2.0        INSTALLING THE CONVERTER AND CABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
           2.1        Verify Serial Numbers . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
           2.2        Mounting The Converter . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  2
           2.3        Installing Cables To Converter And Service Loop . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  3
           2.4        Pulling Sensor Cable Through Electrical Conduit . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  4
3.0        BATTERY INSTALLATION AND SYSTEM START-UP . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
           3.1        Battery Life . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
           3.2        Installing Batteries and Starting Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  5
           3.3        Setting the Time and Date  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
           3.4        Replacing Batteries . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  6
           3.5        Powering-Down and Powering-Up the System . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7
           3.6        Power Options . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  7
           3.7        Sampling Frequencies . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  8
4.0        SOLAR PANEL OPTION . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9
           4.1        Description . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .  9
5.0        ELECTRICAL CABLE CONNECTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
           5.1        Converter Electrical Cable Connections  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
           5.2        Terminal Board . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
           5.3        FPI Mag and Ultra Mag 4" through 48"  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 12
           5.4        Ultra Mag 2" & 3"  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
           5.5        SPI Mag 282L Cable Connection . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
           5.6        Converter Grounding . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .15
           5.7        Opto-Isolated Pulse Output Hook-Up  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
           5.8        Opto-Isolated Pulse Input . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 16
           5.9        Converter Power Hook-Up . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
6.0        CONVERTER START-UP - ALL SENSORS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . .18
           6.1        Menu Navigation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
           6.2        Front Panel Visualization Pages . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
           6.3        Factory Set Key Code  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
           6.4        Menu Structure . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
7.0        PROGRAMMING EXAMPLE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
8.0        MENU 0 - QUICK START . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
9.0        MAIN MENU DESCRIPTIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
           9.1        Menu 1 - Sensor  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 24
           9.2        Menu 2 - Scales . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
           9.3        Menu 3 - Measure  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
           9.4        Menu 4 - Alarms . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 29
           9.5        Menu 5 - Inputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
           9.6        Menu 6 - Outputs . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 30
           9.7        Menu 7 - Communication . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
           9.8        Menu 8 - Display . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 32
           9.9        Menu 9 - Data Logger . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 33
           9.10       Menu 10 - Diagnostic . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 34
           9.11       Menu 11 - Internal Data . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 35
10.0 ALARM MESSAGES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 36
11.0 SPECIFICATIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 40
12.0 DIMENSIONS . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 41
13.0 RETURNING A UNIT FOR REPAIR  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 42
14.0 TROUBLESHOOTING GUIDE  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 43
APPENDIX 1.0 UNITS OF MEASURE . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 44
APPENDIX 2.0 CONVERSION TABLES . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 45
WARRANTY STATEMENT  . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 46
This symbol indicates important safety information. Failure to follow the instructions can
              i       This symbol indicates important information. Failure to follow the instructions can result in
                      permanent damage to the meter or installation site.
Safety Warnings
When installing, operating, and maintaining McCrometer equipment where hazards may be present, you must protect
yourself by wearing Personal Protective Equipment (PPE) and be trained to enter confined spaces. Examples of confined
spaces are manholes, pumping stations, pipelines, pits, septic tanks, sewage digesters, vaults, degreasers, storage tanks,
boilers, and furnaces.
        WARNING!
   !    Incorrect installation or removal of FPI Mag meters can result in serious injury or death. Read the
        instructions in this manual on the proper procedures carefully.
         WARNING!
   !     Never enter a confined space without testing the air at the top, middle, and bottom of the space. The air
         may be toxic, oxygen deficient, or explosive. Do not trust your senses to determine if the air is safe. You cannot
 see or smell many toxic gases.
        WARNING!
   !    Never enter a confined space without the proper safety equipment. You may need a respirator, gas detec-
        tor, tripod, lifeline, and other safety equipment.
        WARNING!
   !    Never enter a confined space without standby/rescue personnel within earshot. Standby/rescue personnel
        must know what action to take in case of an emergency.
        WARNING!
   !    Pressurized pipes should only be hot tapped, cut, or drilled by qualified personnel. If possible,
        depressurize and drain the pipe before attempting any installation.
        WARNING!
   !    Carefully read all safety warning tags attached to the meter.
                  At the end of its lifetime, this product shall be disposed of in full compliance with the
                  environmental regulations of the state in which it is located.
   i
        IMPORTANT: Verify the Meter Serial Numbers on both the converter and sensor match. This will ensure a
        properly calibrated system. The Meter Serial Number is located on the side of the sensor, and the Converter
        Serial Number and the Meter Serial Number are located on a label on the side of the converter. Ensure the
 Meter Serial Number on the sensor and the converter tags match.
          WARNING: Do not connect any form of conduit directly to the converter enclosure. Doing so will allow moisture
   !      and potentially dangerous gasses to enter directly into the converter. Attaching any conduit to the enclosure, or
          altering the enclosure in any way will void the warranty.
     i
          CAUTION: Do not cut the cable cover off. Doing so may damage the sensor cable and adversely effect the
          calibration of the meter.
Cable Cover
                 CONTINUOUS SAMPLING
                  Maximum sample rate
                                                      Max S.R.                   Max S.R.                        Max S.R:
                 AVERAGE SAMPLING
                 Sample rate at 3 seconds
                                                        3 Sec.                         3 Sec.                  3 Sec.
S.R.=SAMPLE RATE
                  SMART SAMPLING
                  Automatic sampling switch
                  according to flow profile       VARIABLE S.R.                                  VARIABLE S.R.
Each sampling frequency mode has a different effect on battery life. Note that the battery life shown below is under optimal
conditions:
                  An internal algorithm allows the unit to automatically detect flow rate variations in the process.
• Continuous sampling: < 6 months
               This capability determines the automatic setting of the sampling frequency. No variations means
                lower sampling frequency, with less power consumption; high variations means higher frequency
• Average sampling: > 5 years
                to better follows the changes in the process.
• Max life sampling: > 7 years
• Smart sampling: 1-5 years depending on flow profile
Note: Continuous sampling is only available if energy saving is turned off. Energy saving can only be accessed through the
IF2 communication cable.
                                                                         193 mm
The solar panel provides power to the converter by converting
                                                                           7 ½”
sunlight into electrical energy to recharge the solar panels'
rechargeable battery. Its nominal power output is 5W. The solar
panel comes complete with all accessories, except for the fixing rod.
(Figure 6 and Figure 7)
                                                                                    295 mm
                                                                                     11 ½”
                                                                                                              193 mm
                                                                                                                7 ½”
                                                                                   50 mm
                                                                                     2“
! CAUTION - Always disconnect the power cord before attempting any electrical connections.
All electrical cables enter the converter through compression fittings located on the side of the converter. Ensure that all
compression glands are properly tightened and all unused fittings are plugged so the case remains sealed.
5.2     Terminal Board
All connections are made on the terminal board (Figure 8). To access the terminal board, loosen the four screws on the back
of the converter to remove the rear cover.
 NOTE: The terminal blocks unplug from the circuit board for easy connection.
         External power
        12-60VDC or 100-
        240VAC connector
       block battery socket
M1 connector block
                      IF2X socket
                                                                                  DIP switch for
                                                 SD memory card                  keyboard block
1 2 3 4 5 6 7 8 9 10
11 12 13 14 15 16 17
                                                                           SH        C1      C2
                                              SH        B1        B2                                E
                                                                                    OUT1 OUT2      OUT1 OUT2
                                                       COILS                            OPEN        COMMON
                                                                                    COLLECTOR
                                                               Black Ground
                                                               To Chassis Lug
1 2 3 4 11 12 13 Right Sensing
                                                                                                                                          8/16
                                                                                                                                                                   A        #1 (E1)           Blue
     ML255
                                                                                                                                                                                                            electrodes
                                                                             CHASSIS | BLACK
RED
                                                                                                                                 YELLOW
                    BLUE
PINK
BLACK
GREEN/YELLOW
                                                                                                          GREEN/YELLOW
                                                                                                                                                                                                           Left Sensing
                                                                                                                                          i
                                                                                                                                                                   *IMPORTANT: See Section 5.6.1, "Converter Grounding
                                                                                                                                                                   For FPI Mag Ultra Mag 4" - 48"", Figure 13 for instructions
                                                                                                                                                                   on attaching the chassis ground wire to the converter
                                                                                                                                                                   ground lug.
                                                                                                                                Shrink Tube
                                       CABLE A - ELECTRODES
                                                                                                                              Cable Diameters:
                                                                                        CABLE B - COILS
Cable Jacket
                                                                  Black Ground
                                                                  To Chassis Lug
1 2 3 4 11 12 13 Right Sensing
                                                                                                                                             8/16
                                                                                                                                                                    A        #1 (E1)          Blue
    ML255
                                                                                                                                                                                                             electrodes
                 BLUE
PINK
BLACK JUMPER
GREEN/YELLOW
CHASSIS | BLACK
GREEN/YELLOW
RED
                                                                                                                                    YELLOW
                                                                                                                                             p/n 17134 Rev. 1.0
                                                                                                                                                                                                            Left Sensing
                                                                                                                                                                    A        #2 (E2)          Pink
                                                                                                                                                                                                             Electrodes
                                                                                                                                                                                                              Reference
                                                                                                                                                                    B         #3 (C)         Black
                                                                                                                                                                                                               Ground
                                                                                                                                                                                                            Cable Shield
                                                                                                                                                                    A        #4 (SH)     Green/Yellow
                                                                                                                                                                                                              Ground
                                                                                                                                                                                                             Ground To
                                                                                                                                                                    B      Chassis Lug       Black
                                                                                                                                                                                                            Chassis Lug*
                                                                                                                                                                                                            Cable Shield
                                                                                                                                                                    B       #11 (SH)      Green Yellow
                                                                                                                                                                                                              Ground
                                                                                                                                                                    B       #12 (B1)          Red                 Coil
                                                                                                                                                                    B       #13 (B2)         Yellow               Coil
                                                                                                                                             i
                                                                                                                                                                  *IMPORTANT: See Section 5.6.2, "Converter Grounding For
                                                                                                                                                                  Ultra Mag 2" & 3"", Figure 14 for instructions on attaching
                                                                                                                                                                  the chassis ground wire to the converter ground lug.
                                                                                                                                 Shrink Tube
                            CABLE A - ELECTRODES
CABLE B - COILS
                                                                                                                             Cable Diameters:
                                                                                                                             Cable A (15035): 0.248”
                                                                                                                             Cable B (15036): 0.248”
Cable Jacket
                                                                                                     #1             Blue
                                                                                                                                   electrode
                           WHITE
                    BLUE
BLACK
BLACK
RED
                                                              YELLOW
                                                                                                                                    Sensing
                                                                                                     #2            White
                                                                                                                                   electrode
                                                                                                                                   Reference
                                                                                                     #3            Black
                                                                                                                                    ground
                                                                                                                                Magnet shield
                                                                                                     #11           Black        / overall cable
                                                                                                                                    shield
                                                                                                     #12            Red                Coil
                                                                                                     #13           Yellow              Coil
                                                    Cable Diameter:
                                                    Single Cable (36002): 0.250”
                                                                                    Black Ground
                                             Cable B                                Wire With Loop
                                                                                     From Cable B
                                                                                    Black Ground
                                              Cable B                               Wire With Loop
                                                                                     From Cable B
                                                                                     With Lead To
                                                                                      Terminal #3
     i
          IMPORTANT
          Outputs are not isolated from each other. All outputs MUST use the same power source.
                                                                     OUT 1
                                                                     15
                                                                     OUT 2
                                                                      16
                                                               COMMON EMITTER
                                                                    17
                                                                     Push button
                                                                       switch
                                                     10K      +
                                                              5                    3-40 VDC
                                                                                   24V
                                                              -
                                                              6
   i      WARNING!
          Hazardous supply voltage can shock, burn, or cause death.
The power supply line must be equipped with external surge protection for current overload (fuse or circuit breaker with
limiting capacity not greater than 10A). It must be easily accessible for the operator and clearly identified.
Power connection is made using the power terminal block on the upper right side of the terminal board. (Figure 17 and
Figure 18)
NOTE: The terminal block uplugs from the circuit board for easy connection. Connect earth ground to the protective
grounding terminal before making other connections. The power supply of this converter is 100 - 240 VAC, 44 - 66 Hz at
maximum 4 W or 12-60 VDC.
                                             Neutral Wire
                                             Typically A
                                             White Wire                                            Negative Wire
Figure 17. AC power supply terminal block Figure 18. DC power supply terminal block
Key:                Function:
                    UP/DOWN KEY (for moving cursor up or down)
                    SHORT PRESSING (< 1 SECOND):
                    Moves the cursor up to the previous subject on the menu
                    Increases the numeric figure of the parameter highlighted by the cursor
  Enter/Esc
    Key             LONG PRESSING (> 1 SECOND):
                    Confirms the selected function
                    Leaves the current menu
                                          Figure 19. Converter key conventions
"Visualization pages" are different display screens that present specific information.
If there is screen is in "screen saver" mode, long-press the Left/Right arrow key to wake it up. Short-press the Left/Right
arrow key to cycle through the different visualization pages.
i        ATTENTION!
         It is very important to record any customized code as it CANNOT be retrieved if it is lost! The converter will
         need to be returned to be reset to factory settings.
                                                    !2.UGPUQT!
                                                    PF>oo!       11467                                            1.1     Insert ND of sensor (0-300
                                                    MB>!      ,13/1111                                            1.2     Calibration data of sensor v
                                                                                                                  1.3     Sensors model: Enter the f
                                                    Kpu/rqukvkqp>! 1                                              1.4     Position for insertion senso
                                                    Mk>!        2/1111
                                                    Mr>!        2/1111                                            1.5     Factory parameter
                                                    M\>!       ,111111                                            1.6 Automatic setting accordin
                                                    ME>!        2/1111                                            1.7 Automatic setting accordin
                                                                                                                  1.8 Enables the empty pipe de
                                                    VE2>ou!        121                                            1.9 Value of empty pipe sensib
                                                    VE3>ou!        112                                            1.10 Perfom a SENSOR TEST ( s
                                                    G/R/fgvgev>!    QP                                            1.11*Enables the automatic zero
                                                    G/r/vjt>!      181                                            1.12 Reset the preceding value
                                                    \gtq!ecn/
                                                    \gtq!tgu/
!7.QWVRWVU
Qwv2>!         RNU,                 6.1* Output 1
                                    6.5* Output 2
Qwv3>!          QHH                 6.6 Power supply of pressure pr
!22.KPVGTPBN!FBVB
N3!eqfg>! ++++++                                  11.1 Level 2 access code enter
                                                  11.2 Load factory data pre-set
Nqcf!hcev/fcvc                                    11.3 Ks Coefficient
U0p>!      182341
MV>!      ,2/1111
MT>!      ,2/1111
MU>!      ,2/1111
Note: The function used in the example is the top menu                     You will start with a blank screen.
item, so you will not need to scroll through the Quick                     1. Long press on Enter/ESC.
Start menu. However, to do so, you need to have the
full menu item highlighted and press Left/Right or Up/
Down.
                                                                                This will wake up    
                                                                                  the front panel                                            
                                                                                                      .....  .....  ..... ..... ..... 
                                                                                                     .....
                                                              TER                                                                                               TER
                                                         EN                                                                                                EN
                                                                ESC                                                                                               ESC
                                                              TER                                                                                               TER
                                                         EN                                                                                                EN
                                                                ESC                                                                                               ESC
                                                              TER                                                                                               TER
                                                         EN                                                                                                EN
                                                                ESC                                                                                               ESC
NOTE: The Quick Start menu is configured at the factory. Some menus may not be available     in your configuration.
                                                                                                        If the
quick start menu is not present as the first menu option, then it has been disabled.              
                                                                                     (See Section 9.8 to enable the Quick
Start menu.)                                                                          
                                                                                       
                                                                                       
                                                                                         
                                                                                       
                                                                                          
                                                                                                  
9.0     MAIN MENU DESCRIPTIONS                                                        
                                                                                      
Press the key Enter/Esc key to go to the Main menu directly when the “Quick start menu” is disabled. When
                                                                                                  it is not disabled
                                                                                                        
you can select the Main menu from the “Quick start menu”. The functions in the Quick 
                                                                                      Start menu and the
                                                                                                            Main menu are
                                                                                     
explained below. Please note that some functions are only displayed if other functions         or with the insertion
                                                                                          are enabled
of additional modules.
                                                                                            
                                                                                             
                                                                                         
9.1       Menu 1 - Sensor                                                                   
                                                                                            
                                                                                            
                                                                                           
                                                                                                
                                                                                                     
ND Inside Pipe Diameter                                                                           
                                                                                         
The inner diameter of the pipe entered in millimeters. NOTE: The flow accuracy will             
only be as good as the accuracy of the actual pipe ID entered into the converter. The   
                                                                                                       
                                                                                                       
“Nominal Pipe Size” often referred to can be different from the actual pipe ID and if 
                                                                                                 
                                                                                                      
                                                                                             
                                                                                           
used cause a substantial error in flow reading. Setting the ND value to zero will cause   
                                                                                                          
                                                                                           
the converter to display the velocity instead of flow rate. Totalizer will then increment   
                                                                                            
                                                                                        
in feet or meters.                                                                         
                                                                                              
                                                                                           
                                                                                                   
                                                                                            
KA                                                                                          
Factory calibrated gain for the forward flow. Do not change the value.
Ins. Position
This is for a single point sensor. Position for insertion sensors: 0=1/8DN, 1=1/2DN, 2=7/8DN
Ki
Automatic setting according to ID (insertion meter only).
Kp
Automatic setting according to ID (insertion meter only).
KZ
This is the Zero Calibration constant. The KZ updates every time the Zero Calibration is performed.
KC
This is a coil current adjustment factor for use between sensor families.
TC2
This is the time needed to stabilize coil current when switching coil polarity.
E.P.detect
Enables the empty pipe detection function.
E.P.thr.
Sets the empty pipe detection threshold.
Zero cal.
Zero point calibration function for the forward flow. Enables the automatic zero calibration system.
       i
                                                                                    
                  IMPORTANT - The water must be perfectly still before starting
                                                                                the Zero Cal process
                                                                                                    or an offset
                  will be introduced into the flow report.                                  
                                                                                                   
                                                                                            
                                                                                          
To perform the zero point calibration:                                                 
                                                                                      
1. Select the Zero cal. menu and press                      will ask if you want
                                         the Enter/Esc key. It                to execute.             
                                                                                             
                                                                                         
                                                                                   
                                                                  
                                                                 
                         2.
                                            Long press the Enter/ESC      button and release when the message
                                   "Measuring. . . " appears.
                                                              The converter counts up from zero to 700,
                                     
                                  after which the  zero point   is set.
                                
                                                                                           
                                                                            
                 
                  
                                                                     
                                                                     
                                                          
                                          
                   
Zero
  res. 
This resets the Zero Calibration.
               
 
      Once the desired value is entered quick push the Enter/Esc key to save the changes and then long push Enter/
      Esc to exit back to the display.
Tot1MU                                                                                    
                                                                                              
The totalizer unit/multiplier as well as the visible decimal place is set here.           
To change the totalizer unit/multiplier:
                                                                                         
                                                                                               
                                                                                          
                                                                                              
                                                                                         
                                                                                             
1.   Highlight the menu and press the Enter/Esc key.            2. Press the Up/Down key.   
                                                                                             
                                                                                          
   This will highlight                                          This will scroll 
                                                                                          
                                                                                                      
  the unit/multiplier.                            through the different       
                                                                                              
 See Appendix 1.0 for                                   available units.            
                          
                                                                            
                                                                                               
    available units of     
                                                    NOTE: The totalizer 
             measure.                                                       
                                                                                                      
                                                                          
                                                                                          
                                                                     multiplier is built     
                           
                                                                            
                                                                                        
                                                                                             
                                                        into the unit of          
                                                                                          
                                       TER
                                                                      measure, so for 
                                                                                               
                                                                                                       T ER
                           
                                           EN C                                   
                                                                gallons multiplied by 
                                                                                                 EN C
                                                                                                     
                                          ES                                        ES
                                                     1000, select Kgal. 
                                                                                         
                                                                                            
                                                                                                      
                                                                                                                
                          
                                                                            
                                                                                               
                          
                                                                    
                                                                                          
                                                                                                      
                          
                                                                    
                                                                                          
                                                                                             
                           
                                                                          
                                                                                         
                                                                                             
                                                                         
                                                                                               
                                                                          
                                                                                         
                                                                                               
                                                                                                      
                           
                                                                                
                                                                             
                                                                                              
3. Press the Right/Left     key twice.                          4. Press the Up/Down       key.
                                                                                        
                                                                                             
                           
     This will highlight 
                                                     This will change the      
                                                                                         
                                                                                               
the numeric value to 
                                                                     
                                                                   decimal resolution         
                                                                           
                                                                                             
               the right.                                                  
                                                                     displayed for the       
                           
                                                                          
                                                                                               
                                                  totalizers. Changing 
                                                                                               
                          
                                                             the decimal 
                                                                                         
                                                                                              
                                                                                                       
                            
                                                                                   
                                                                        resolution will 
                                                                                              
                           R                                             
                                                                                         
                                                                       not change the             
                          
                                             TE
                                                                                        
                                                                                                           N TER
                                                   EN C                                   
                                                                       multiplier. The           E
                          
                                   ES                                               ESC
                          
                                             available selections 
                                                                                         
                                                                                          
                                                                                                      
                                                                                                      
                                                                                                   
                              
                                                                            
                                                                     are 00001, 001.0, 
                                                                                             
                                                      01.00, and 1.000. 
                                                                                               
                          
                                                                          
                                                                                               
                                                                                                      
                          
                                                                      
                                                                                           
                                                                               
                                                                          
                           
                          
                                
                          
                          
  If the desired unit of measure               
                                 is not in the current  
                                                        list, press the Right/Left key    
                              
                                       
  and scroll to the blank space between the unit/multiplier and the numeric                 
                                                                           
  decimal resolution selection   and press the
                                         Up/Down key to switch between            
  U.S. Standard and Metric   units.
                                                                                      
                                                                                            
                                                                                          
Tpls1
                                                             
Duration of the pulse output 1 expressed in milliseconds.      
This option is only available when "out1" in Menu 6 -        
Outputs is set to PLS+, PLS-, or PLS. The pulse duration     
                                                             
can be set from 8.0 to 7999.9.                                 
IMPORTANT
i    The converter cannot detect the type of device it is connected to so it is up to the user to verify the setting
     is compatible with the external device receiving the pulse. Incorrect settings can damage the receiving
     device. See section 10.6 Menu 6 - Outputs for output specifications.
Prof
Measurement Profile
         0 = Continuous Power
         1 = Smart
         2 = Average
         3 = Max Life
Continuous Power uses more battery power and is not recommended. Max Life uses the least battery power and is the
default.
Damping
Measure Filter Damping Profile Enable is only valid in the Smart Profile setting. This applies a fixed filter to the reading.
Al.max-
Maximum Reverse flow rate exceeded. Can be 0-125% of Full Scale
Al.min-
Minimum Reverse flow rate exceeded. Can be 0-125% of Full Scale.
Hyst.
Alarm Hysteresis. Set 0-25%. This sets the lag in response based on a percentage of the full scale. Example if the alarm
triggers at 100% and the hysteresis is set to 2% then once triggered the current rate must change beyond 2% to exit
out of the current alarm state. This setting applies to all alarms.
Count lock
Totalizer Count Lock Input Enable, Set by turning on or off. This allows for the totalizers to be locked (frozen) when the
input is active.
Alarm
Allows the input to be wired to an external alarm output such as a water intrusion alarm.
Wake-up
Wake up the display with an input.
Out 2
Transistor output channel 2. See tables on the next page for available settings.
 OFF                DISABLED
 PLS                PULSE FOR POSITIVE AND NEGATIVE FLOW VOLUME
 PLS-               PULSE FOR NEGATIVE FLOW VOLUME
 PLS+               PULSE FOR POSITIVE FLOW VOLUME
 F.SIGN             FLOW DIRECTION OUTPUT (ENERGISED = -)
 DIRECT. DR.        DIRECT OUTPUT DRIVE
 ALL ALARMS         ALL ALARMS
 OVR.RANGE.         OUT OF RANGE ALARM OUTPUT (ENERGISED = FLOW RATE OK)
 HARDW.AL.          CUMULATIVE ALARM OUT interrupt coils, empty pipe, meas. error (ENERG. = NO ALARMS)
 P.EMPTY            EMPTY PIPE ALARM OUTPUT (ENERGISED = FULL PIPE)
 MX+MN ALL          MAX AND MIN ALARM OUTPUT (ENERGISED=AL. OFF)
 *MX+MN TEMP        MAX AND MIN TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
 *MIN TEMP          MIN TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
 *MAX TEMP          MAX TEMPERATURE OUTPUT (ENERGISED= AL. OFF)
 *MX+MN P12         MAX AND MIN PRESSURE 1 AND 2 OUTPUT (ENERGISED= AL. OFF)
 *MX+MN DP          MAX AND MIN DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
 *MIN DP            MIN DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
 *MAX DP            MAX DIFFERENTIAL PRESSURE OUTPUT (ENERGISED= AL. OFF)
Comp.mode
Format the Data Logging Data to the previous version of this converter. Default=Off
int.1
Data Logging Interval. HH:MM:SS
Log T+
Include the T Positive Totalizer for Data Logging.
Log P+
Include the P Positive Totalizer for Data Logging.
Lot T-
Include the T Negative Totalizer for Data Logging.
Log P-
Include the P Negative Totalizer for Data Logging.
Log NT
Include the Net T Totalizer for Data Logging.
Log NP
Include the Net P Totalizer for Data Logging.
Log Q
Include the Flow Rate for Data Logging.
Log STAT
Include Statistics like the board temperature and cycles/h for Data Logging.
M.units
Include the Units of Measure for Data Logging.
% values
Include the Flow Rate % of full scale and optional analog inputs for Data Logging.
   i       CAUTION - if the passcode is changed from the default value and is lost, it CANNOT be recovered. In
           the event the passcode is lost, the converter can be returned to the factory to be reset.
Load fact.data
This will configure the converter to the configuration when it left McCrometer.
S/n
Displays the board serial number.
KT
Used to adjust gain errors. This factor is generated by the converter at power up.
KS
Field adjustment coefficient. This value is a direct multiplier that is used as a field adjustment / correction coefficient.
                                                                                                            ALARM
                                                                                                              VIA
 CODE            MESSAGE                       CAUSE                             ACTION                      SMS *
  000     NO ALARMS                -----                            -----                                      NO
  001     SYSTEM STARTUP           The system was started           -----                                     YES
  003     PRESS.1>FS               The Pressure at input 1 is       Check the maximum full scale set          YES
                                   higher than the full scale set   and the process conditions
  004     FLOW>FS                  The flow rate is higher than     Check the maximum full scale set          YES
                                   the full scale set               and the process conditions
  005     PULS.1>F.MAX             The pulse frequency at           Reduce the pulse duration if the          YES
                                   output 1 is too high             connected counter allow this or
                                                                    increase the pulse unit value
  006     B3 LOW                   The battery B3 powering the      Replace the battery                       YES
                                   MODEM is exhausted
  007     B2 LOW                   The battery B2 powering the      Replace the battery                       YES
                                   main board is exhausted
  008     B1 LOW                   The battery B1 powering the      Replace the battery                       YES
                                   main board is exhausted
  077     SYSTEM RESTART           The system was restarted         If this message appears after an          YES
                                   after a reset signal             AUTO-TEST command, it is not an
                                                                    error indication, otherwise check
                                                                    all the connections, the batteries
                                                                    and the grounding
  094     ALARM INPUT ACT.         The digital input detected an    Check the process condition               YES
                                   alarm signal condition
  103     B.TEMP.OUT R.            The measured board               Ensure that the instrument is             YES
                                   temperature is out of the        operating within the specified
                                   allowed range                    temperature conditions
  113     FIRMW.FILE ERR           The file containing the          Re-send the firmware file                 YES
                                   firmware was received with
                                   errors
  115     S.OUT OVERLOAD           An overload condition is         Check the connections and                 YES
                                   present in the power source      the external sensors (pressure,
                                   that feeds the external          temperatures, input circuitry)
                                   sensors and inputs
  117     ERR.SENS.TEMP            The external temperature         Check the external temperature            YES
                                   sensor is not working            sensor and its connections
                                   properly
  118     MAIL S.FAILED            The last email sending           Check the transmitting                    YES
                                   process has failed for some      conditions (antenna signals, GPRS
                                   reasons                          parameters, server status and
                                                                    configurations)
  119     SMS NOT AUTH             A SMS from a not authorized      Check the events log for the              YES
                                   number was received              sender number and do the
                                                                    appropriate actions
    *Not available at this time.
                                                                                                         ALARM
                                                                                                           VIA
CODE          MESSAGE                       CAUSE                             ACTION                      SMS *
135    V.MODEM LOW              The internal voltage of the    Check the battery B3 and its                YES
                                MODEM is not sufficient for to connections, verify if any visible
                                guarantee a proper operation damage has occurred to the
                                                               board or the MODEM
144    NO CMD RECEIVED          The command list sent to the      Check the command list and re-           YES
                                instrument does not contain       send it again
                                any executable commands
150    MAX Q+                   The Positive flow rate is         Check the maximum threshold              YES
                                higher than the maximum           positive flow rate alarm value and
                                threshold value set               the process conditions
151    MIN Q+                   The Positive flow rate is lower   Check the minimum threshold              YES
                                than the minimum threshold        positive flow rate alarm value and
                                value set                         the process conditions
153    MAX Q-                   The Negative flow rate is         Check the maximum threshold              YES
                                higher than the maximum           negative flow rate alarm value
                                threshold value set               and the process conditions
154    MIN Q-                   The Negative flow rate is         Check the minimum threshold              YES
                                lower than the minimum            negative flow rate alarm value
                                threshold value set               and the process conditions
157    MAX P1                   The pressure at input 1 is        Check the maximum threshold              YES
                                higher than the maximum           pressure 1 alarm value and the
                                threshold value set               process conditions
158    MIN P1                   The pressure at input 1 is        Check the minimum threshold              YES
                                lower than the minimum            pressure 1 alarm value and the
                                threshold value set               process conditions
160    MAX P2                   The pressure at input 2 is        Check the maximum threshold              YES
                                higher than the maximum           pressure 2 alarm value and the
                                threshold value set               process conditions
161    MIN P2                   The pressure at input 2 is        Check the minimum threshold              YES
                                lower than the minimum            pressure 2 alarm value and the
                                threshold value set               process conditions
163    MAX DP                   The pressure difference           Check the maximum threshold              YES
                                (delta) P1-P2 is higher than      pressure difference (delta) alarm
                                the maximum threshold set         value and the process conditions
164    MIN DP                   The pressure difference         Check the minimum threshold                YES
                                (delta) P1-P2 is lower than the pressure difference (delta) alarm
                                minimum threshold set           value and the process conditions
167    MAX TEMP                 The external temperature is       Check the maximum threshold              YES
                                higher than the maximum           temperature alarm value and the
                                threshold value set               process conditions
168    MIN TEMP                 The external temperature          Check the minimum threshold              YES
                                is lower than the minimum         temperature alarm value and the
                                threshold value set               process conditions
171    P.EMPTY                  The system detected an            Check the empty pipe settings            YES
                                empty pipe alarm condition        and the process conditions
 *Not available at this time.
                                                                                                           ALARM
                                                                                                             VIA
CODE         MESSAGE                        CAUSE                              ACTION                       SMS *
193    CONFIG.ENTERED           The system detects an             Check the events log and do the            YES
                                access to the configuration       appropriate action depending on
                                parameters                        your data protection policy
194    CLOCK S.FAIL             The clock synchronization         Check the server configuration,            YES
                                with the remote server failed     the GPRS network conditions and
                                                                  antenna signal
195    POWER FAILURE            The system experienced a          Check the status of the batteries          YES
                                power failure (loss of power)     and all their connections. THIS
                                                                  MESSAGE MAY APPEAR WHEN
                                                                  THE SYSTEM IS SWITCHED OFF
                                                                  WITHOUT PRIOR PUTTING IT IN
                                                                  STAND-BY MODE
206    F.SENSOR ERROR           The system detected one or     Check the error code in the                   YES
                                more errors in the flow sensor events log
207    TEMP.>FS                 The external temperature is       Check the maximum full scale set           YES
                                higher than the full scale set    and the process conditions
208    PULS.2>F.MAX             The pulse frequency at            Reduce the pulse duration if the           YES
                                output 2 is too high              connected counter allow this or
                                                                  increase the pulse unit value
209    PRESS.2>FS               The Pressure at input 2 is        Check the maximum full scale set           YES
                                higher than the full scale set    and the process conditions
210    EL.SIG.ERROR             The system detected an error      Check the flow sensor cables, the          YES
                                condition on the measuring        electrodes surface, the grounding
                                electrodes                        and the process condition
211    EXCIT.ERROR              The system detected an error      Check the flow sensor cables, the          YES
                                in the coils excitation circuit   coils insulation and resistance
212    SD MEMORY FULL           The SD card memory is full        Change the SD card memory or               YES
                                (no more space for new data)      erase all the unwanted data
215    POWER SUPPLY OFF         The power coming from the         Check the power supply source              YES
                                universal power supply went       and the relative cables
                                off
216    SD MEMORY ERROR          The SD card installed is          Replace the SD memory, format it           YES
                                missing, not valid, not           or check if it is properly inserted
                                formatted or incompatible
                                with the system
ELECTRICAL CHARACTERISTICS:
        Power supply version             Power supply            Power supply frequency               Max power
                                        12V - 60V        *                    //
                  AU                                                                                       4W
                                         100V - 240V~*                     44-66Hz
       LITHIUM BATTERY NOT
                                           3.6v-19aH                          //                         30mW
       RECHARGEABLE
      * Voltage must not exceed ±10% of the above limits.
                                              Mod
                                                   el
                                              Seria #: 88
                                                          00
                                              Met l #: E12- 03xxx
                                                  er #:    34
                                                        MI1 567
                                                           2-34
                                                                56
                                              3255
                                                    W.
                                                PH (95 STET
                                                      1) 652 SON
                                                            -6811 AVE,
                                                                  • htt HEME
                                                                       p://ww T, CA
                                                                             w.m     92
                                                                                 ccrom 545
                                                                                      eter.coUSA
                                                                                             m
                                                                                                                     R
                                                                                                                  E       C
                                                                                                                 T       S
                                                                                                              N       E
                                                                                                             E
              U.S. Standard
  in3                Cubic Inches
  Gal                U.S. Gallons
  IGL              Imperial Gallons
   ft3                Cubic Feet
  bbl              Standard Barrels
  BBL                 Oil Barrels
  KGL                Kilo Gallons
  IKG            Imperial Kilo Gallons
  Aft                  Acre Feet
  MGL               Mega Gallons
  IMG           Imperial Mega Gallons
   hf   3
                Hundreds of cubic feet
   kf   3
               Thousands of cubic feet
  ttG        Tens of thousands of gallons
                   Metric
   ml                  Milliliters
  cm3             Cubic Centimeters
    l                    Liters
  dm3              Cubic Decimeter
  dal                  Decaliter
   hl                 Hectoliter
  m3                 Cubic Meter
   Ml                 Megaliter
                    Time
   s                   Seconds
   m                   Minutes
   h                    Hours
   d                     Days
                                  Fraction                     Decimal
                                    1/8                          .125
                                    1/4                          .25
                                    3/8                          .375
                                    1/2                           .5
                                    5/8                          .625
                                    3/4                          .75
                                    7/8                          .875
Table of Conversions
              Multiply                                 By                     To Get
             Centimeters                            0.3937                    Inches
             Centimeters                            0.03281                    Feet
                Inches                                25.4                  Millimeters
                 Feet                                30.48                 Centimeters
             Square Feet                             144.0                 Square Inches
           Square Inches                           0.006944                 Square Feet
            Cubic Inches                          0.0005787                 Cubic Feet
             Cubic Feet                              7.481                    Gallons
             Cubic Feet                             1728.0                 Cubic Inches
             Cubic Feet                             0.02832                Cubic Meters
             Cubic Feet                              28.32                     Liters
            Cubic Meters                             35.31                  Cubic Feet
            Cubic Meters                             264.2                    Gallons
             US Gallons                              3.785                     Liters
             US Gallons                             0.1337                  Cubic Feet
             US Gallons                            0.003785                Cubic Meters
             US Gallons                            .8326748               Imperial Gallons
                Liters                              0.2642                    Gallons
         ˚F = ( ˚C x 9/5 ) + 32                ˚C ( ˚F - 32 ) x 5/9
Propeller Flowmeters
Magnetic Flowmeters
DURA MAG ™
FlowConnect ™