Specifications & more

HumaStar 200
> System Specifications
> Scope of Supply
> Obligatory Items
> Optional Item

System Overview REF 16895


Analyzer type Automated clinical chemistry analyzer
Reagent system HUMAN multipurpose reagents
Throughput (max. and typical) 200 tests/hr maximum
120-130 tests/hr typical
(applies to most parameters)
Operation modes Random access / STAT / Batch
Barcode reader Internal sample barcode reader
Reaction / reading System Reaction cuvettes with multicycle washing station
Languages available English, French, Spanish
Additional languages can be translated by distributor
User interface External PC, touch screen recommended

Key specification
HumaStar 200
Random Access Clinical Chemistry Analyzer

Tests Measuring modes Photometry (colorimetry, UV-tests, turbidimetry)


Analysis modes Endpoint (bichromatic) with sample/reagent blank
Differential endpoint with sample/reagent blank
Fixed time
Kinetic (bichromatic)
Calculation modes Factor 1-point and multi-point
Calculation algorithms Linear (factor, linear, linear regression, average),
Non linear (cubic spline, poly-linear, four parameters (sigmoidal),
five parameters (sigmoidal), multi-parameters (logit-log)
# of programmable profiles Unlimited
# of standards per test Up to 8

Sample / Control /
Calibrator Sample type Serum, plasma, urine, CSF, whole blood, others
Sample vessels Primary tubes (12.5 x 100 mm), optional tray (16 x 100 mm)
Sample cups (10 mm), optional tray (Hitachi 3.5 ml)
Sample positions Total of 60 positions for samples, optional tray 20 + 20 positions
Sampling volume 2 to 300 µl/test
Loading Continuous
Pre-Dilution  Automatic (In-needle or reaction cuvette)
Post-Dilution Automatic (abnormal levels, excessive substrate consumption
and/or lack of linearity)

Reagent Reagent positions 30 reagent positions (20 ml, 50 ml, adapter for tubes and cups, removeable tray)
Reagent cooling Approx. 9 °C below room temp. at bottom of reagent, independent switch
Reagent 1-2 volume range 5 to 350 µl

Reaction unit Reaction wells /


positions for incubation  80 washable Bionex® cuvettes
Reaction volume (minimum) 210 µl
Reaction volume (maximum) 350 µl
Cuvette washing system 8-step wash station, 8 needles
2 bottles: systemic- and special wash solution
Incubation temperature 37°C ± 0.2°C heat transfer by air
Incubation time 792 sec. max. incubation + reading time

Liquid Handling Liquid transportation Pipetting arm


Liquid level sensors Capacitive
Collision detection for probe Yes (vertical)
Water consumption < 2l/h (8ml/test)
Water quality < 10 μS filtered, distilled or de-ionized

Key specification

2 of 6
Reading Optical system Interference filters
Readings Mono- or bichromatic
Light source Halogen lamp (Phillips, 6 V, 10 W)
Spectral range 340 to 900 nm
Wavelengths pre-installed 340, 405, 505, 546, 578, 600, 650, 700 nm
Max. # of wavelengths installed 9
Wavelength error (accuracy) ± 2 nm on peak wavelength
Wavelength bandwidth (precision) Half bandwidth 10 mm
Detector Silicon photodiode
Absorbance range (linearity) 0 to 2.5 OD
Resolution 0.0001 OD

Data Processing Memory for Sample results, calibration, patient data, QC data, error-log, absorbance curves
Memory capacity Unlimited (HDD)
Reports for Patient, single test, complete sample, work sheet, method and
QC´s, calibration curves, kinetics, continuous printing
Quality control module Levey-Jennings, Westgard multirules, SD, CV%
Max. number of control levels Up to 3 levels per test
Test statistics Number executed, SD, CV%, Mean
Warnings Analytical limits, reagent and reaction integrity check
(blank, linearity, substrate depletion, reaction OD etc.),
Printer Printer connected to external PC
LIS ASTM Bi-directional , ethernet of external PC, polling mode on LAN

General PC min. requirements Intel Core i3, 100 GB HDD, 4 GB RAM , 1280 x 768 dots resolution,
keyboard and mouse, 3 USB 2.0 ports (not including mouse and keyboard),
Windows ™10, NET framework (version which is automatically in­stalled
together with the software), All updates need to be installed.
Only English, German, French and Spanish operating systems have been tested.
All other languages need to be tested locally. No complex Chinese operating
system. English (USA) keyboard and language drivers must be installed in any case.
Recommended PC requirements Intel Core i5 CPU or higher, 8 GB RAM, touch screen monitor, FULL HD resolution,
SSD (Solid-State-Drive) memory instead of HDD highly recommended,
no Atom CPU or use dedicated graphic GPU, 6 USB 2.0 ports
(instrument, barcode scanner, mouse, keyboard, printer, touch screen)
LAN port for LIS host communication, External printer, A4
Physical dimensions (W x D x H) Instrument without any components: 69 x 76 x 52 cm
Space required for routine use: 150 x 90 x 100 cm
Packaging: 92 x 86 x 79 cm + 58 x 38 x 58 cm
Weight: Gross: 90.0 kg, net: 51.0 kg
Mains fuse T2 A / 250V, 5 x 20 mm
Electrical requirements 110...120, 220...240 VAC; 50/60 Hz; < 200 VA
Environmental Operating: temperature 16...30 °C, humidity < 80 % rel. non condensing
Transport: temperature 0...50°C, humidity < 85% rel. non condensing
Wash / waste tank 20 l systemic, 2 l special wash, 20 l waste, level sensors

3 of 6
HumaStar 200
Random Access Clinical Chemistry Analyzer

Scope of Supply Unit/Size REF.


HumaStar 200 1 16895
Packlist
UPS 1 18961
Sample Tray 60 positions (installed on instrument) 1 16890/10
Accessories Kit HumaStar 200 REF. 16895AI contains the following items:
Reagent Bottles Tray 30 positions 1 16890/13
Reagent Bottle 50ml 30 pcs. 1 16890/35
Reagent Bottle 20ml 30 pcs. 1 16890/34
Cap for reagent bottles 30 pcs. 2 16890/36
Bottle Adapter 20 ml 15 pcs. 1 16890/14
Reagent Bottles Tray (30 pos.) Closing Lid for empty reagent bottle positions 15 pcs. 1 16890/41
REF. 16890/13 Reagent Adapter for Cup or Tube 1 16890/15

Reaction Cuvettes Starter Kit 20 pcs. 1 16890/40


Sample Tubes 12 mm, 5 ml 50 pcs. 1 16890/30
Sample Cups 1 ml Adapter (for 16890/10) 10 pcs. 1 16890/12
Sample Cups 1 ml (for 16890/10) 50 pcs. 1 16890/31
Waste Tank 1 16890/55
Tank for Systemic Solution 20 l 1 16890/56
Tank for Special Wash Solution 2 l 1 16890/57

Reagent Bottles 20 ml Tanks Tubing Group 1 16890/58

Reagent Bottles 50 ml Power Cord 1 16890/146

REFs. 16890/34, 16890/35 USB Cable 3 m 1 16890/231


Halogen lamp 1 16890/51
Fuses Kit 1 16890/50
Software and Settings USB Card for HS200 1 16895/24
User Manual 1 16890/1

Obligatory Items Service Kits Unit/Size REF


Need to be ordered with the first instrument:
Starter Spare Part Kit for HS 100|200 (Sufficient for up to 5 instruments) 1 16890/253

Need to be ordered with each instrument:


Maintenance Kit 12 month for HS 100|200 (Sufficient for 1 year) 1 16890/250-1
Maintenance Kit 24 month for HS 100|200 (to be used 1 year after yearly maintenance) 1 16890/250-2
HC-Acute 50ml (4 pieces sufficient per maintenance kit) DG 1 17400/56

DG = this item is classified as dangerous goods


4 of 6
Obligatory Items Consumables REF
Wash Additive 4 x 25 ml 18971
Special Wash Solution 12 x 30 ml DG 18974
Cuvette Clean (necessary for turbidimetry) 4 x 100 ml DG 18973

Accessories
Personal Computer with monitor (printer) is necessary for the operation of this instrument.
HUMAN offers them as optional items below.

Optional Items Service Kit REF


Maintenance Kit 12 month for HS 100|200 (Sufficient for 1 year) 16890/250-1
Maintenance Kit 24 month for HS 100|200 (to be used 1 year after yearly maintenance) 16890/250-2
HC-Acute 50ml (4 pieces sufficient per maintenance kit) DG 17400/56

Accessories REF
Personal Computer with Windows incl. Keyboard and Mouse 18992P
Monitor LCD 19 inch 17901M
Touch Screen Monitor 18995MT
HP Laser Printer (USB + Parallel) 18993L
Desktop Barcode Reader 16890/19
Sample Tray (60 pos.) Sample Tray 60 Positions (12 x 75 mm – 12.5 x 100 mm or 1 ml (10 mm) cups with adapter) 16890/10
REF. 16890/10 Sample Tray 20 Positions (12 – 16 x 100 mm + 20 positions of 2 ml cups) 16890/11
Sample Tray 60 Positions (12 – 12.65 x 75 mm or 1 ml (10 mm) cups with adapter) 16890/26
Matched Interference Filter 420 nm 16890/300
Matched Interference Filter 492 nm 16890/301
Matched Interference Filter 520 nm 16890/302
Matched Interference Filter 620 nm 16890/303
Matched Interference Filter 630 nm 16890/304
Matched Interference Filter 670 nm 16890/305
Matched Interference Filter 880 nm 16890/306

Consumables REF
Sample Tubes 5 ml (12 x 85 mm) 1000 pcs. 16890/30
Sample Cups 1 ml (10 mm) (for Sample Tray 16890/10 with 16890/12) 1000 pcs. 16890/31
Sample Cups 2 ml (for Sample Tray 16890/11) 500 pcs. 17470/59
Reaction Cuvettes 200 pcs. (up to 75.000 tests) 16890/33
Reagent Bottle 20 ml 30 pcs. 16890/34

Sample Tubes 5 ml Reagent Bottle 50 ml 30 pcs. 16890/35

Sample Cups 1 ml Cap for Reagent Bottles 30 pcs. 16890/36

REFs. 16890/30, 16890/31

5 of 6
16895/2020-12 © 2020 HUMAN

Legal statement
While Human is making every effort to include accurate and up-to-date information, we make = parts of this item are classified as dangerous goods but are classified as

„Limited and Excepted quantities (EQ) of dangerous goods“ and therefore


no representations or warranties, express or implied, as to the accuracy or completeness of the
do not fall under restrictions that aplly to dangerous goods.
information provided in this document and disclaim any liability for the use of it.

HUMAN Gesellschaft für Biochemica und Diagnostica mbH


Max-Planck-Ring 21 · 65205 Wiesbaden · Germany
Tel. +49 6122-9988-0 · Fax +49 6122-9988-100 · e-mail: human@human.de · www.human.de