Bilanciai: D400 Terminal
Bilanciai: D400 Terminal
BILANCIAI
D400 Terminal
Index
1. GENERAL                                                                    1-5
    1.1 Declaration of conformity                                             1-5
    1.2 Foreword                                                              1-6
    1.3 Symbols                                                               1-8
    1.4 Documentation                                                         1-9
    1.5 Description of the terminal                                           1-9
    1.6 Technical characteristics of the terminal                            1-10
    1.7 Dimensions and weight of the terminal                                1-11
    1.8 Obtaining technical assistance                                       1-12
    1.9 Replacement parts list                                               1-14
    1.10 Warranty                                                            1-14
2. SAFETY INSTRUCTIONS                                                       1-15
    2.1 Prohibited uses                                                      1-15
    2.2 Regulations                                                          1-15
    2.3 Prescriptions of use                                                 1-15
3. DELIVERY AND INSTALLATION                                                 1-17
    3.1 Connection of the terminal to the electrical supply line             1-18
    3.2 Connection of the terminal to the platform scale                     1-19
        3.2.1 Equipotentiality between the terminal and the platform scale   1-20
        3.2.2 Connection of analogue load cells                              1-20
        3.2.3 Connection of digital load cells                               1-21
    3.3 COM1 serial port connection                                          1-22
    3.4 COM2 serial port connection                                          1-23
        3.4.1 Connection of COM2 in RS232 configuration                      1-23
        3.4.2 Connection of COM2 in RS422 configuration                      1-24
        3.4.3 Connection of COM2 in RS485 configuration                      1-25
    3.5 Analogue output connection                                           1-26
    3.6 Input/Output connection                                              1-27
4. CONTROLS, DISPLAY, SWITCHING THE TERMINAL ON AND OFF                      1-29
    4.1 Display of weight and additional information                         1-29
        4.1.1 Selection display symbols                                      1-30
    4.2 Function keys                                                        1-31
    4.3 Switching the terminal on and off                                    1-34
5. USE OF THE TERMINAL                                                       1-37
    5.1 General                                                              1-37
        5.1.1 Using the keys to navigate the menus                           1-37
        5.1.2 Entering numeric data (Editor)                                 1-38
    5.2 User menu                                                            1-39
        5.2.1 Adjusting the contrast                                         1-39
        5.2.2 Changing the date and time                                     1-39
        5.2.3 Weight display                                                 1-39
        5.2.4 Diagnostics (for digital scale only)                           1-41
        5.2.5 Setting the outputs as Setpoints (if enabled)                  1-41
        5.2.6 Setting the outputs as ranges (if enabled)                     1-41
    5.3 Operating modes                                                      1-42
        5.3.1 Standard operation                                             1-42
                                         1-3
                                                            D400
       5.3.2 Sum weighing operation                                1-42
       5.3.3 Loading extraction operation                          1-42
       5.3.4 Unloading extraction operation                        1-43
   5.4 MPP operation                                               1-43
6. OPTIONS                                                         1-45
    6.1 4 I/O board                                                1-48
    6.2 BCD parallel 5V                                            1-49
    6.3 Calculator BCD                                             1-51
    6.4 Parallel 24V source current BCD (positive common)          1-53
    6.5 Serial port expansion board                                1-56
        6.5.1 Jumper RS485 - COMx2                                 1-58
    6.6 MPP memory expansion boards                                1-59
        6.6.1 Memory capacity                                      1-59
        6.6.2 Operation                                            1-59
        6.6.3 Disabling MPP                                        1-60
        6.6.4 Checking memorised weight data                       1-61
    6.7 12-24Vac-dc input power supplier                           1-62
7. MAINTENANCE                                                     1-65
    7.1 Battery                                                    1-65
    7.2 Changing the fuses                                         1-65
8. TROUBLESHOOTING                                                 1-67
    8.1 Malfunctions                                               1-67
    8.2 Error messages                                             1-68
                                       1-4
                                 D400
1. GENERAL
                           1-5
                                                     D400
1.2 Foreword
✔ The aim of this manual is to provide the operator, through the use of
  text and illustrations, with essential information regarding the
  installation, safe operation and maintenance of the weighing system.
✔ This manual must be kept in a safe place where it is readily available
  for consultation. Always observe the instructions contained in the
  manual!
✔ The safe operation of the system is the responsibility of the operator,
  who must have a thorough knowledge of the system.
✔ The user is responsible for ensuring that the installation conforms to
  the applicable regulations.
✔ The equipment must be installed by specialised personnel who have
  read and understood this manual.
✔ "Specialised personnel" means any personnel who, by virtue of the
  training they have received and their professional experience, have
  been explicitly authorised by the "System safety supervisor" to install,
  operate and maintain the system.
✔ In the event of any problems, contact your nearest Service Centre.
✔ Any attempt on the part of unauthorised personnel to dismantle or
  modify the terminal is prohibited; any such attempt shall invalidate the
  warranty and release the manufacturer from all liability for any injury
  or damage.
✔ The alteration or removal of the data plates and seals is strictly
  prohibited; check that all plates and seals are present and legible, if
  not contact After-Sales Service.
✔ The manufacturer shall not be liable for any damages caused by
  incorrect handling of the terminal.
✔ The information and illustrations contained in this manual were up to
  date at the time of publication.
✔ The Manufacturer is committed to a policy of continuous product
  improvement and system components may therefore be subject to
  modification.
✔ All the technical information contained in this manual remains the
  exclusive property of the manufacturer and may not be divulged to
  third parties.
✔ No part of this document maybe reproduced or transmitted in any
                                   1-6
                                                     D400
  form, including publication in computerised form or on the World Wide
  Web, without the express written permission of the manufacturer.
✔ This manual may not be used for purposes other than those directly
  related to the installation, operation and maintenance of the terminal.
✔ In order to more clearly illustrate certain maintenance or adjustment
  operations, some of the illustrations in this manual show the weighing
  system with the safety guards removed. Under no circumstances may
  the system be operated in these conditions. Do not operate the
  system in these conditions under any circumstances whatsoever, but
  remove the safety guards for the time strictly required to carry out the
  required repairs or maintenance then fit them back in place.
                                   1-7
                                                      D400
1.3 Symbols
Below is a list of the symbols used in this manual to alert the reader to
the various hazards associated with the operation and maintenance of
the instrument.
                              DANGER
     Denotes an operation or procedure where failure to observe
     the instructions will result in death or serious injury.
                              CAUTION
     Denotes an operation or procedure where failure to observe
     the instructions could result in minor injury or damage to the
     instrument.
                             WARNING
     Information or instructions on how the system is to be
     operated correctly in order to maximise its service life or
     prevent loss or damage of programmed data or to optimise
     operation with regard to metrological standards.
Text and messages displayed on the terminal are printed in this manual
using special characters.
Messages:
Display messages appear like this.
Menu pathways:
2°F>MENU>Contrast.
The character > indicates the transition from one menu option to the
next.
                                    1-8
                                                       D400
1.4 Documentation
This manual is accompanied by a CD-ROM that primarily contains
information about the installation of the terminal.
Here you will find information on how to interface the terminal with a PC
and PLC.
                                    1-9
                                                 D400
                               1-10
                                                 D400
 Weight: 2 Kg
 Dimensions of panel space for rack mounting: 200 mm x 104 mm (base
 x height)
 The dimensions are given in mm.
Figure 1.1 - Dimensions and weight of rack-mounted version
(log0101.jpg)
                               1-11
                                                 D400
Weight: 2 Kg
The dimensions are given in mm.
Figure 1.2 - Dimensions and weight of table top version(log0102.jpg)
 Weight: 2.5 Kg
 The dimensions are given in mm.
Figure 1.3 - Dimensions and weight of version with tilted support
(optional) (log0124.jpg)
                               1-12
                                                     D400
1.8 Obtaining technical assistance
In the event of any operating faults requiring the intervention of
specialised technicians, contact the manufacturer or your nearest
Service Centre. To enable us to deal with your request swiftly, always
quote the serial number of your terminal, which can be found on the seal
label. Also provide information about the system in which the terminal is
installed.
                                  1-13
                                                       D400
 Code                                 Description
 460727                               Series D400/800 110/220 Vac
                                      power supplier
 460740                               Power supplier series D400/800
                                      12-24Vdc-ac
 290391                               D400 keypad
 403941                               CPU board
 460728                               Converter for analogue load cells
 403961                               Interface or digital load cells
 403981                               MPP FLASH memory expansion
                                      board
 403991                               Serial expansion board
 404001                               4 inputs/4 outputs board
 404521                               DIWIFI card
 404523                               DINET ethernet card 10/100
                                      Mbit/s, TCP/IP, UDP
 404231                               Ethernet card 10 Mbit/s, TCP/IP,
                                      UDP, FTP + RS232 modem
 404011                               BCD TTL or calculator board
 404012                               BCD 24V source board
 527313                               Battery
 460729                               Display
1.10 Warranty
The conditions of warranty are stipulated in the contract of sale.
                                   1-14
                                                      D400
2. SAFETY INSTRUCTIONS
2.2 Regulations
The operating conditions for the electronic terminal are subject to the
regulations in force in the country in which the terminal is used. All use
of the terminal in conditions which do not comply with these regulations
is prohibited.
                                   1-15
                                                    D400
✔ All maintenance and/or repairs must be carried out by authorised
  personnel only.
✔ Always disconnect the terminal from the electrical supply and wait a
  few minutes before accessing the internal components.
                                 1-16
                                                    D400
Key
1. 9-pin male connector (JBIL) for platform scale connection (15-pin
    male for digital load cells)
2. 9-pin female serial port connector (JCOM 1) for printer connection
3. 9-pin female serial port connector (JCOM 2) for connection of
    various devices
4. Data plate indicating voltage, frequency and fuse types
5. Analogue output connection terminals (JVI)
6. Input/Output connection terminals (JI/O)
7. Earth screw
8. Male 3-pin connector for power supply connection
9. Fuses
10. ON/OFF switch
11. Expansion slot 2
12. Expansion slot 1
Figure 3.1 - Rear of terminal (log0103.jpg)
                                 1-17
                                                       D400
                               DANGER
      Check that:
      ✔ The voltage and frequency of the electrical supply line
      corresponds to the indications on the warning plate on the
      rear of the terminal (see point 4 Figure 3.1 on page 1-17 );
      ✔ the mains outlet socket to which the terminal is connected
      is equipped with an earth;
      ✔ the warning and danger signs are present and legible;
      ✔ failing this, notify your maintenance personnel or contact
      our Assistance Service directly;
For the correct connection of the terminal to the electrical supply line,
proceed as follows:
✔ plug the 3-pin connector of the power lead into the connector on the
  rear if the terminal;
✔ insert the plug of the power lead into the correct mains outlet socket.
                                    1-18
                                                    D400
                            WARNING
     Do not route the terminal connection cables alongside power
     cables as these could cause disturbances that interfere with
     the correct operation of the terminal. Only use the connection
     cable supplied with the terminal. If the cable supplied is too
     short, do not attach an extension lead but contact the
     Manufacturer.
                            WARNING
     The cable screen should always be connected to the metal
     cap of the 9/15-pin connector. Do not route the scale
     connection cable alongside power cables.
                                 1-19
                                                     D400
Figure 3.2 - Pinout for the JBIL connector for connection to scales
with analogue load cells (log0001.gif)
                                  1-20
                                                     D400
                                  1-21
                                                    D400
                            CAUTION
     Operating limits stipulated by the standard RS232:
     Maximum transmission distance = 15 m
     Maximum line voltage = ± 12 Vdc
     For connection to external devices, use a screened cable
     and connect the screen to the metal cap of the 9-pin
     connector.
                                 1-22
                                                     D400
                             CAUTION
     The RS232 operating limits are indicated in par. 3.3 on page
     1-22.
                                  1-23
                                                   D400
                           CAUTION
     Operating limits stipulated by the standard RS422:
     Maximum transmission distance = 1200 m
     Maximum line voltage = +/- 7V
     For connection to external devices, use a screened twisted
     pair cable and connect the screen to the metal cap of the
     9-pin connector.
                                1-24
                                                    D400
Key
NC = Reserved - do not connect
DATA +/- = Bidirectional data transmission line
TERMIN = Termination resistance to be connected to pin 6
Ri = Termination resistance inside terminal
NOTE: the free pins are reserved for RS232 connection
Figure 3.7 - Pinout for RS485 serial port connection (log0006.gif)
                                  1-25
                                                     D400
                             CAUTION
     Operating limits stipulated by the standard RS485:
     Maximum transmission distance = 1200 m
     Maximum line voltage = +/- 7V
     For connection to external devices, use a screened twisted
     pair cable and connect the screen to the metal cap of the
     9-pin connector.
     Maximum number of terminals that may be connected = 16
               Pin n°                              Signal
 1                                   Output 0-10 V
 2                                   Output 0(4)-20 mA
 3                                   Common
                                  1-26
                                                     D400
                             CAUTION
     Technical characteristics:
     Resolution= 10000 points
     Precision = 0.05 % FS
     Minimum voltage output load = 100 kohm
     Minimum current output load = 250 ohm
 Key
 Pi = Part inside terminal
 Pe = Part outside terminal
Figure 3.8 - JI/O terminal board for Input/Output connections
(log0008.gif)
                                  1-27
                                             D400
                      CAUTION
Technical characteristics:
Input:
Maximum input voltage = 24Vdc
Maximum input current = 5mA
The inputs can be controlled by a mechanical contact or by
an NPN-type transistor (negative common)
Output:
mechanical voltage-free contact
Maximum switchable voltage = 110 Vac/dc
Maximum switchable current = 200 mA
I/O refresh time = 1/10 s
                       DANGER
When the weighing system is installed in complex plants that
can represent a danger hazard for operators, enlist the
assistance of specialised personnel to perform several
manoeuvres without load in order to acquire the experience
necessary to work in safety.
                           1-28
                                                      D400
Key
1. Display of weight and additional information
2. Function keys
Figure 4.1 - Front of terminal (log0104.jpg)
The LCD (Liquid Crystal Display) (point 1 Figure 4.1 on page 1-29 ), in
addition to the universally recognised weighing symbols, also displays
information (in extended format) related to the operation of the terminal.
                                   1-29
                                                     D400
Listed below are the weighing symbols displayed and their meanings:
                      Weight stable symbol
                      Indicates that the weight value displayed is stable
                      and thus may be printed and/or transmitted.
                      Tare symbol
                      Indicates the presence in memory of an acquired
                      tare value.
Minimum weighment
                                  1-30
                                                      D400
                      Weight indication
                      Within the preset upper and lower limits.
                      Weight indication
                      Above the preset upper limit.
                      Enter/cancel tare
                      On pressing this key, the weight on the scale is
                      acquired as the tare value, provided the following
                      conditions are satisfied:
                      ✔ the weight must be stable;
                      ✔ the weight must have a positive value;
                      ✔ the weight must not exceed the maximum
                        capacity.
                                   1-31
                                D400
Enter/display tare
Pressing this key displays the preset tare.
With the option TARE a known tare value may be
entered.
Enter the numeric value as described in par. 5.1.2
on page 1-38.
On completion of the operation, the display will
show the net weight and tare values and illuminate
            1-32
                                             D400
               Customizable keys
               The functions associated with these keys are
               indicated on the display.
                      WARNING
The functions of all keys may vary, and will be signalled by
indications on the displays.
                           1-33
                                                   D400
Key
1. ON/OFF switch
Figure 4.2 - On/Off switch (log0105.jpg)
                            DANGER
     Before switching on the terminal, check that the safety
     guards are in place and in good working order.
     If the terminal is wall mounted on an electrical panel, there
     must be a main switch that disconnect powers to the entire
     panel.
     To identify this switch, contact your internal maintenance
     manager.
                                 1-34
                                                      D400
Press the switch on the rear of the terminal (see point 1 Figure 4.2 on
page 1-34 ) to position:
I                      to switch on the terminal;
O                      to switch it off
On power on, the display will show the Manufacturer's logo and the type
of operation for which the terminal is enabled (see par. 5.3 on page 1-42
).
Wait for the LOCK indication to appear (terminals subject to metrological
verification only).
If on completion of the operation the display shows a value other than
                                   1-35
       D400
1-36
                                                       D400
5.1 General
pressing       to select the option 2°F. Now use the other keys to select
the next option. The arrow keys ↑ ,↓ ,→,← enable you to move up,
down, left and right through the menus.
By pressing SELECT you can select the required option or enter a
submenu.
The keys # and \\ allow you to change the way the menu pathway is
displayed : # extended pathway, \\ pathway expressed numerically.
Selecting ESC returns you to the previous menu level. By pressing ESC
repeatedly you can therefore exit through the menus to return to the
normal weight display mode. Form here on in this manual, the
instructions will only indicate the pathway without referring to the specific
keys to be pressed. For example, the instruction on how to change the
contrast will be given as follows:
2°F>MENU>Contrast>+ o ->SAVE>ESC
                                    1-37
                                                     D400
                                  1-38
                                                      D400
2°F>MENU>Contrast
By pressing + or - you can adjust the contrast. To save the new setting
in memory press SAVE .
2°F>MENU>Display
2°F>VISUAL
✔ 15mm digits
  The display shows the gross weight (or net) and the weighing
  symbols.
                                   1-39
                                                     D400
✔ 30mm digits
  The display shows the weight in 30 mm high digits and the weighing
  symbols.
✔ Selection
  In addition to the weight and the usual weighing symbols, the display
                              WARNING
        The display range described here is independent of the
        range associated with the outputs (see par. 5.2.5 on page
        1-41 and par. 5.2.6 on page 1-41 .
✔ Extraction
  This display mode is only possible if loading or unloading extraction
  operating mode is selected (see par. 5.3.3 on page 1-42 and par.
  5.3.4 on page 1-43 ).
  The display shows the gross weight, the extracted weight and the
  weighing symbols.
                                  1-40
                                                      D400
2°F>MENU>Diagnostics
Enter the required weight values as described in par. 5.1.2 on page 1-38
.
2°F>MENU>Data management>Range
                                   1-41
                                                          D400
                                     1-42
                                                     D400
to slow extraction mode.
On pressing the key in correspondence with the indication START , the
extracted weight value is zero-set and scale loading is enabled. The
display shows the gross weight (the weight on the scale) and the
extracted weight (the weight gradually loaded onto the scale).
The display format used in this operating mode is Extraction (see
par. 5.2.3 on page 1-39 ).
The extraction operation can be concluded by pressing STOP.
                                  1-43
       D400
1-44
                                                     D400
6. OPTIONS
Optional expansion boards (with the exception of the MPP memory
expansion board) are installed in the slots provided on the rear of the
terminal (see points 11,12 Figure 3.1 on page 1-17 ).
For the positions and numbering of the outputs (I/O, COM, BCD), refer
to the label on the bottom of the terminal.
                                  1-45
                                        D400
Inputs
          Slot              N°input       designation
CPU board (slot0)   IN 1              INPUT 01
                    IN 2              INPUT 02
Slot1               IN 1              INPUT 11
                    IN 2              INPUT 12
                    IN 3              INPUT 13
                    IN 4              INPUT 14
Slot2               IN 1              INPUT 21
                    IN 2              INPUT 22
                    IN 3              INPUT 23
                    IN 4              INPUT 24
Outputs
          Slot             N°output       designation
CPU board (slot0)   OUT 1             OUTPUT 01
                    OUT 2             OUTPUT 02
Slot1               OUT 1             OUTPUT 11
                    OUT 2             OUTPUT 12
                    OUT 3             OUTPUT 13
                    OUT 4             OUTPUT 14
Slot2               OUT 1             OUTPUT 21
                    OUT 2             OUTPUT 22
                    OUT 3             OUTPUT 23
                    OUT 4             OUTPUT 24
                             1-46
                                            D400
Serial ports
         Slot                  N°com          designation
 CPU board (slot0)     COM 1             COM 01
                       COM 2             COM 02
 Slot1                 COM 1             COM 11
                       COM 2             COM 12
 Slot2                 COM 1             COM 21
                       COM 2             COM 22
BCD
                Slot                     designation
 Slot1                           BCD 1
 Slot2                           BCD 2
                               1-47
                                                     D400
Key
Pi = Part inside terminal
Pe = Part outside terminal
Figure 6.2 - Connection diagram for optional 4 INPUT - 4 OUTPUT
board (log0009.gif)
The figure shows the connection terminals of the 4 I/O board.
For specific information on connection refer to par. 3.6 on page 1-27 .
                                  1-48
                                                      D400
                              CAUTION
     Operating limits
     V out Max = +5V
     I out Max = +/- 10 mA
                                   1-49
                                                    D400
                                       Key
                                       U1,U2,U4,U8 = BCD units
                                       (Binary Coded Decimal)
                                       DE1,DE2,DE4,DE8 = BCD Tens
                                       C1,C2,C4,C8 = BCD Hundreds
                                       M1,M2,M4,M8 = BCD
                                       Thousands
                                       DM1,DM2,DM4,DM8 = BCD
                                       Tens of thousands
                                       CM1 = Hundreds of thousands
                                       DVALID = (OUTPUT) can
                                       assume the following values:
                                       0: the data will be valid after 10
                                       ms
                                       1: the data will remain valid for
                                       at least another 5 ms
                                       NEG = (OUTPUT). Assumes the
                                       value "0" (zero) when the weight
                                       is negative
                                       NC = Reserved - do not connect
                                       DGND = Signal reference
                                       common
                                1-50
                                                         D400
                             CAUTION
     The operating limits for this type of output are:
     V out Max = +5V
     I out Max = +/- 10 mA
                                   1-51
                                                   D400
                                      Key
                                      CU1,2,4,8 = BCD units (Binary
                                      Coded Decimal)
                                      CDEC1,2,4,8 = BCD Tens
                                      CC1,2,4,8 = BCD Hundreds
                                      CAM1,2,4 = BCD Thousands
                                      NEG = (OUTPUT) negative that
                                      assumes the value "0" (zero) if
                                      the weight is negative
                                      CALC0,1,2,3,4,5 = Address pins
                                      for scale numbers 1 to 64
                                      CDVALID = (OUTPUT) can
                                      assume the following values:
                                      0: The data will be valid after 10
                                      ms
                                      1: The data will remain valid for
                                      at least 10 ms
                                      NC = Reserved - do not connect
                                      DGND = Signal reference
                                      common
                               1-52
                                                       D400
                              CAUTION
      The operating limits for this output type are as follows:
      V out Max = +24V
      I out Max = + 10 mA
                                   1-53
                                                   D400
                                      Key
                                      U1,U2,U4,U8 = BCD Units
                                      (Binary Coded Decimal)
                                      DE1,DE2,DE4,DE8 = BCD Tens
                                      C1,C2,C4,C8 = BCD Hundreds
                                      M1,M2,M4,M8 = BCD
                                      Thousands
                                      DM1,DM2,DM4,DM8 = BCD
                                      Tens of thousands
                                      CM1 = Hundreds of thousands
                                      DVALID = (OUTPUT) can
                                      assume the following values:
                                      0: the data will be valid after 10
                                      ms
                                      1: the data will remain valid for
                                      at least another 5 ms
                                      NEG = (OUTPUT). Assumes the
                                      value "0" (zero) when the weight
                                      is negative
                                      +24V = External voltage
                                      DGND = Signal reference
                                      common
                               1-54
                                                 D400
Key
L = Generic external load
T = Output transistor inside terminal
+24V = External voltage
Figure 6.6 - Example of 24V source current connection (log0063.gif)
                               1-55
                                                   D400
Key
1. Comx1
2. Comx2
Figure 6.7 - Serial port designation (log0129.jpg)
The operating limits stipulated by the RS232/422/485 standards are
indicated in par. 3.3 on page 1-22 and in par. 3.4 on page 1-23.
                                   1-56
                                                D400
                                    Key
                                    RX422 +/- = Data reception
                                    RX232 = Data reception
                                    TX422 +/- = Data transmission
                                    TX232 = Data transmission
                                    TERMIN = Termination
                                    resistance to be connected to
                                    pin 6
                                    NC = Reserved - do not connect
                                    GND = Signal ground
                                    Ri = Termination resistance
                                    inside terminal
Figure 6.8 -Supplementary serial port connector (9-pin female
D-type) (log0064.gif)
                               1-57
                                                   D400
                                       Key
                                       GND = Signal ground
                                       RX232 = Data reception
                                       TX232 = Data transmission
                                       TERMIN = Termination
                                       resistance to be connected to
                                       pin 8
                                       TX422 +/- = Data transmission
                                       RX422 +/- = Data reception
                                       Ri = Termination resistance
                                       inside terminal
                                1-58
                                                      D400
                              CAUTION
     When maximum memory capacity is reached, the oldest data
     are deleted and replaced by the most recent.
6.6.2 Operation
The type of MPP operation is defined during installation.
There are two main types of operating mode (for further details see the
advanced user manual):
✔ in the first type, the user operates from the terminal by pressing
  to save and transfer the weight data;
✔ in the second type, the user works from the PC keyboard in the way
  determined by the specific software.
                                   1-59
                                                    D400
In both operating modes, on completion of the saving operation, the user
can display the progressive number associated with the weighing
operation (see par. 6.6.4 on page 1-61 ).
Code MPP will only appear on the display when the memorising
operation is enabled with the key.
Both the weighing terminal and the PC may signal errors caused by
failure to transfer or save the data correctly. In this event, follow the
on-screen instructions.
                            WARNING
     Disabling MPP may compromise the operation of equipment
     connected to the weighing system in that data transfer to
     such equipment will be inhibited.
                                  1-60
                                                      D400
                                   1-61
                                                     D400
Key
1. ON/OFF switch
2. T4A250V fuses
3. Input connector on panel
Figure 6.11 Connection to the power source with very low safety
voltage rating (citi0201.jpg)
                                  1-62
                                       D400
                   CAUTION
Peak use conditions:
DC (direct current): Vmin=11Vdc Vmax=28Vdc
AC (alternate current): Vmin=12Vac Vmax=24ac
Maximum power draw=50W
Power cable (not supplied): 2 conductors with 1-2mm²
section
                        1-63
       D400
1-64
                                                        D400
7. MAINTENANCE
7.1 Battery
On switching on the terminal, it performs a check on the condition of the
internal lithium battery. If the battery is discharged, the terminal displays
the message CHANGE THE BATTERY . To change the battery, contact
Assistance Service.
Key
1. Fuse holder
2. Removal tab
3. ON/OFF switch
Figure 7.1 - Fuse (log0126.jpg)
                               DANGER
      Disconnect the INPUT/OUTPUT cables and the mains lead.
                                    1-65
                                                   D400
✔ Refer to Figure 7.1 on page 1-65
✔ Turn the ON/OFF switch to O (off).
✔ Press the tab and withdraw the fuse holder.
✔ Replace blown fuses with new ones of the same rating (T630mA250V
  time delay).
✔ If the fuse blows again on switching on the terminal, do not make any
  further attempts to replace it but contact Assistance Service.
                                 1-66
                                                    D400
8. TROUBLESHOOTING
If the suggested remedy does not solve the problem, contact Assistance
Service.
8.1 Malfunctions
 Problem           Cause             Remedy
 The terminal      No power          Check that power arrives at
 does not switch                     the mains outlet socket.
 on                                  Check that the mains lead is
                                     correctly plugged in.
                                     Check that the ON/OFF switch
                                     is in the position I (ON).
                                     Check the fuses.
 The display       Adjust contrast Switch on the terminal while
 backlighting is   (set to minimum)
 on, but no data                    holding pressed        .
 are displayed                      After a few seconds, the
                                    display will show the "CB" logo
                                    with the contrast set to
                                    maximum, after which the
                                    contrast will gradually be
                                    reduced.
                                  1-67
                                                   D400
                                 1-68
                                          D400
                        1-69
                                                        D400
 excep n.nn
 in task xx.x
The Diagnostics item of the user menu will allow you to establish the
cause of errors concerning the digital cells. Consult par. 5.2.4 on page
1-41.
                                      1-70
                      SOCIETÀ COOPERATIVA
BILANCIAI