Users Manual

The document is a user manual for an intelligent clamp multimeter that measures AC and DC current, voltage, resistance, and other electrical parameters. It includes safety information, operational guidelines, and technical specifications for the device, emphasizing the importance of proper usage to avoid electric shock. Users are advised to read the manual thoroughly before operating the instrument and to follow safety regulations to ensure safe and effective use.

Uploaded by

Toño_jr
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
0% found this document useful (0 votes)
47 views19 pages

Users Manual

The document is a user manual for an intelligent clamp multimeter that measures AC and DC current, voltage, resistance, and other electrical parameters. It includes safety information, operational guidelines, and technical specifications for the device, emphasizing the importance of proper usage to avoid electric shock. Users are advised to read the manual thoroughly before operating the instrument and to follow safety regulations to ensure safe and effective use.

Uploaded by

Toño_jr
Copyright
© © All Rights Reserved
We take content rights seriously. If you suspect this is your content, claim it here.
Available Formats
Download as PDF, TXT or read online on Scribd
You are on page 1/ 19

USERSMANUAL

Intelligent clamp multimeter


digital
Smart Digital Ac& DC CIamp Meter
NCV

CAT A
I
600V MAX
400A

CATII
300V

U.0.0.0
HeS+MKSOMAVFC
T-RMS 6000Counts

SEL-HEd Lie A

REL
SEL

COM A INPUT

CATII
MAX 60OV

Designed 3nd Conforts to

IEC61010-1
CAT.III 600V

Before using the instrument, please


lease read this
manual
carefully, and saveit well for future usina,.
Table of content
1.Safety Information 1
1.1 Preparation...
1.2 Symbol..
2
1.3 Maintenance.
2. Description. 3
3.Specification

3.1Overview
3. 2Technical indexes
3.2.1 AC current

3.2.2 DC current

3.2.3 DC voltage 6
3.2.4 AC voltage 7
3.2.5 Line on-off test 7
3.2.6 Resistance.
7
3.2.7 DIODE 8
3.2.8 Capacitor..
3.2.9 Temperature.
3.2.10 Frequency
4. Operation guide. 10
5. Maintenance
17
6.Accessories.
17
1.Safety

EWarnings
Information

AWarning
-A
Pay special attention to use of this instrument, for
improper use may cause electric shock or damdge to
instrument. During use, observe usual safety regulations
and observe safety measures regulated in use manual.
Inorder to make full use of instrumentfunctions and
guarantee safe operation, please carefully read and
observe use methods in this manual.

Instrument complies with safety requirements


electronic instrument of
measuring EN-61010-1,
EN-61010-2-030 and EN-61010-2-032, level II pollution,

and over-voltage standard is CAT II 600V.


Please observesafety operation guide, and guarantee
to use instrument in a safe manner.

1.1 Preparation
1,1.1 When using this instrument, users must observe
standard safety rules:
General electric shock prevention.

Prevention of misuse of instrument.

1.1.2 After receiving the instrument, check whether it is


damaged during transportation.
1.1.3 After storing and shipping under adverse
conditions,
check whether the instrument has been
damaged.
1.14 Pens of the instrument must be in good condition.
Before use, check whether insulation of pens is

damaged, and whether metal wire is exposed.


1
1.2Symbol
io (refer to Use manual for important safety

information)
Able to
be |
used on dangerouselectrified

conductors.
Dual-insulation
protection (Category )
according to IEC-61010-1 standard
IOver-voltage
CAT III
level |II,pollution
(installation), degree 2 refers
level pulse
to protection of
withstand voltage
provided.

IcE Comply
Grounded
with EU standard.

1.3 Maintenance
1.3.1 Do not try to open bottom case to adjust or repair
instrument, for such operation could only be

conducted by technicians fully understanding the


instrument and electric shock danger.

1.3.2 Before opening instrument bottom case or battery


Cover, remove the pens from the wire being

measured.
1.3.3 In order to avoid electric shock possibly caused by
error reading, when symbol "N" displays on

instrument, replace batteryimmediately.


1.3.4 Use wet cloth and gentle detergent to clean the
instrument, and do no use any abrasive or solvent.

1.3.5 Power off when the instrument is not used, and rotate

range switch to OFF position.

1.3.6 If the instrument is not used for a long time, take out

the battery to avoid any damage to theinstrument.

2
2.
Description
2.1 Part 8
name

7 CAT

G00V

CATE

-BARB:
3 PS

AE
0.0.00
H6MkSQMAVFC
T-RMS 6000Counts

2
5 COM INPUT
4 MAYAAN

1
2 Data hold and
REL DC current
flashlight button
back to zero /NCV button
3 LCD display
4 Input socket
5 Power button /SEL function conversion button
6 Caution light
Current measuring clamp
8 Contactless induction zone
9 Eletric torch

3
3H AUTO )
A
2.2 LCD display

NCV
-BARA:
DC

AC

HE->+MKNMAVFC
T-RMS 6000Counts

AC & DC
Connection/Disconnectionindication

AUTO Automatic range mode


Automaticshutdown indication
Low battery
H Reading hold state

V,A Volt (voltage),


ampere (current)

2,k2, MQ Ohm kilohm and megohm (resistance)

Hz hertz
DIODE
nF uF mF icrofarad

°C°F Centigrade and Fahrenheit


NCV |Non-contactmeasuring voltage

Live Live line voltage identification

4
3. Specification
acycle, and
shall
shall
The instrument specifies one yearas
be re-calibrated under 18°C humidity
~ 28°C, with relative

less than 75%.


3.1 Overview
• Select measurement function and range automatically.
•Overload protection throughout the range.

voltage between measurement terminal and


Max.

ground: 600V DCor 600OV AC


Operating height: Max. 2000m

• Display: LCD
•Max. display value: 6000 digit.

Polarity indication: Automatic indication, and

indicates negative.

•Over range display: 'OL' or '-OL'

•Sampling time: About 3 times/second., bargraph 10


times/second
Unit display: Function and electricity quantity unit

display.

• Automatic shutdown time: 15 minutes


•Power supply: 1.5V AAA battery x 2
Battery under-voltage indication: LCD display

symbol
•Temperature coefficient: Less than 0.1xaccuracy/°C.
Operating temperature: 18°C ~ 28C.
• Storage temperature: -10°C ~ 50°C.
5
3.2 Technical indexes
3.2.1 ACCurrent
Resolution Accuracy
Range
60A 0.01A
+(2.5% reading + 30 digits)
400A 0.1A

Min. input current: AC Current in automatic mode;


1.0A

Manual mode 0.2A AC Current


Max, input current: 400A AC Current.
-Frequency range: 40 ~ 1000 Hz;

3.2.2 DCcurrent
Range Resolution Accuracy

60A 0.01A
+(2.5% reading +30 digits)
400A 0.1A

- Min. input current: Manual mode 0.2A DC current

- Max. input current: 400A DC current


Note: DC current accuracy refers to the above accuracy
when the return tozero mode is enabled.

3.2.3 DC voltage
3.2.3.1 Dc voltage V
Resolution Accuracy
Range
600V 0.1V + 0.5% reading + 5 digits)
-Min. inputvoltage 0.5V DC
- Max. input voltage: 600V DC
6
3.2.3 2Dc voltage mV
Range Resolution Accuracy

600 mV +3 digits)
0.1 mV ± 0.5% reading
- Min. input voltage 0.1m V Dc
-Max. input voltage: 600V DC

3.2.4 AC voltage
3-2.4-1 AC voltage V
Range Resolution Accuracy

600V 0.1V t(0.8% reading + 5 digits)


Min. input voltage: 0.5 V AC
-Max. input voltage: 600V AC(valid value)
-Frequency range: 45 ~ 1000Hz
3.2.4.2 AC voltagemV
Range Resolution Accuracy
600 mV 0.1 mV t(0.8% reading + 5 digits)
- Min.input voltage: 0.1 mV AC
Max. input voltage: 600V AC (valid value)

3.2.5 Line on-off test

Range Resolution Function

If the measured line


10 resistance is less than 50 Q.
•1)
the buzzer in the meter
will
sound
Overload protection: 250V DC or AC (valid value)
3.2.6 Resistance
Resolution
Accuracy
Range 0.12
6002 0.001kO
6k0
0.01kQ
60kQ
0.1kQ
+ 0.8% reading + 3
digits)
600k2
0.001MQ
6MQ
0.01MQ
60MQ
3.2.7 DIODE
Range Resolution Accuracy Testing environment
Testing
Diode test Open
current:aboutlmA;
circuit
1 V 0.001v about 2.8V.
voltage:

The display
showS an
approximation of the diode
forward pressure
drop

-Overload protection:
25OV DC orAC (valid
value)

3. 2.8Capacitor
Resolution
Range Accuracy
0.01nF 10% reading+40
60.00nF digits)
0.1nF
600.0nF
1nF
6.000uF

60.00uF 10nF
100nF 2.5%reading +20 digits)
600.0uF

6.000mF luF
60.00mF 10uF
-Overload protection: 250V DC or AC (valid value)

8
3.2.9
Temperature
Range Accuracy
Resolution
+3°C)
-40°C
1000°C 1°C ±1% reading

-40°F~ +3F)
2°F reading
1832°F t(1%
probe
The precision doesnot include the error of the
thermocouple
value)
- Overload protection: 250V DC or AC (valid

3.2.10Frequency
(Viagear A):
3.Z.10.1 Clamp frequency measurement
Range Resolution Accuracy
60.0Hz 0.1Hz
t(1% reading +5digits )
1000Hz 1Hz
- Measurement range:40Hz ~ 1000Hz

- Input signal range: >1/4 Ful- scale Value

3.2.10.2 Via gear V:


Range Resolution Accuracy
60.0Hz 0.1Hz
±(1% reading + 5 digits )
1000Hz 1Hz
- Measurement range: 40Hz ~ 1000Hz
value)
- Input signal range: >0.5~600V AC voltage (valid
4. Operation guide
4.1 Reading hold

During measurement, if it is required to hold readina.


touch button,value on display will be locked, touch
button again, to cancel reading hold.
4.2 Torch
1)Long press the flashlight button for 2 seconds, turn on
the flashliaht, and turn it off after about 1 minute.
4.3 Auto Power -Off
1) If no operation within 15 minutes after start,
there's

instrument will enter standby state, and shut down to


save energy.
2) After automatic shutdown, Press the "U" button, to
wake the instrument enter work state.
4.4To measure
1)Touch the electric source button to start the machine.
Ifthe battery voltage is insufficient,
the display willdisplay a lowvoltage symbol,
at which point the battery should be replaced.
Press the power button again to shut down the machine.
2)When the instrument is not measured, it will enter the
automatic scanning state. The instrument is shown as
follows:

Auto
10
4.5 NCV (Non
-contactvoltage detection) the
voltage down
Switch the hold
rotation switch to
any gear, close to
"NCV" key to display EF, sensor
and bring the NCV whether
the conductor detect
under test, instrument Could
co
the lead when
'g measured is sony AC Voltage.
instrument detects AC buzzer will

voltage, instrument
alarm and NCV alarm
light will flash.
L

EF
NGV

Note:
1) Even if there is no alarm indication, the voltage may
still be present. Do not rely on the non-contact voltage
todetermine whether a wire has a voltage.
Probe operation may be influenced by socket design,
type of insulation thickness, etc.

In NCV mode,the meter does not measure


detection
voltage, resistance,and current simultaneously.

11
current measurement
electric IACcurent frequency measurement
and,
4.6.1.ccurrent measurement:

|
Automatic
1) Hold
tested
the
linein

the
mode
trigger,

the
open the clamp head, and clamp awire of
clamp
measured signal >1A, the instrument main display
the

2)When the measured Current


value
displays
Screen
when the current value is >35A,the instrument
only
(Note: display itsfrequency value)
display
auxiliary

measurement:
Manual mode
to switch to AC the trigger, open
SEL key t
current, hold
1) Press
the clamp head,
wires of theline undertest in the clamp.
Clamp one ofthe
2) When the measured sigeO.2a, the main display screen
of the instrument will display the measured Current
valo

A~

o
12
4.6.2 Dccurrent measurement

CATS

CAT

rFL

A A
T-RSS 6000Counts S 8000Counts

REL
SEL NCV SEL
COM A NPUT COMA NPUT

Manual mode measurement:


1)Press SEL key to switch to DC current mode.
2) The front end of the lamp head is close to the measured
wire. Press the REL button to make its base close to
zero
Vice display rFL character.
3)Hold the trigger, open the clamp head, and clamp a wire of
the tested line in the clamp.
4) When the measured signal > 0.2a, the main
display screen
of the instrumentshows the measured current value

Note: To get an accurate reading, press the


to reset the meter to zero before "REL" key
measuring.

13
4.7 DC voltage measurements
4.7.1 DC voltage measurements (V)
Connect the test pen to the signal being tested. When the
measured DC signal is greater than 0.5V, the display voltage
value When the measured DC signal is <0.5V, the instrument
uses the default resistance value and displays the resistance
value of the measured signal.

4.7.2 Dc voltage measurement(mV)


1)Press SEL button to dc voltage mV,
the instrument shows mV
2)Connect the test pen to the test signal.
When the test signal is <600mV,
the instrument will display the current
dcvoltage value under test;
(v
when the test signal is >600mV,
the instrument will display "OL".

A8 ACvoltage measurement
4.8.1 voltage measurement (V)
A C

Connect the test pen to the signal


being tested.
When the measured AC signal is
greater than 0.5V, the display
displays the voltage and
frequency values.
When the measured V

ACsignal <0.5V,the instrument


default resistance value, display
5000
the measured signal internal
resistance value.

14
4.8.2 Acvoltage measurement(mV)
1)Press SEL display mV
to enablethe mV
function. Instrument test.
2)Connect the meter and under
meter pen to the signal
When the
measured is less than 600mV,
the meter displays thesignal
dc measured current
voltage value.
When the measurement
signal
is > 60OmV, the instrument
displays "OL".

4.9 resistance measurement


1)Connect the test pen to the resistance to be measured.
The main display of the instrument will display the measured
resistance value.
When the measurement resistance
is less than 50
the instrument Buzzer
sound alarm, indicator
light on.

20

15
4.10Single watch pen wire test
1.Press " " button to switch
NCV
to LIVE mode, and the
meter shows LIVE,
2. Insert the meter stylus
into the power socket.If the
instrumentdetectsacvoltage, it will display
the
voltage signal strength based on the detected
signal
strength. The instrument induction LCD
segment code
will display the voltage signal
strength.The red
backlight is lit up and the buzzer emits an
alarm of
varying intensity.
4. 11Capacitance
measure
1. Press "SEL" key to switch to capacitance
2.Measure the capacitance with both ends of the test
pen and read the measurement from the LCD.
Notice:
OWhen measuring large capacitors, it takes
time to stabilizethe reading
2When measuring the capacitance with
polarity, pay attention to the corresponding
polarity to avoid damaging the instrument.

4. 12 Temperature
measurement
Press ""
SEL
key to switch TEMP, and the red plug of
the thermocouple is inserted into the C/°F end.The
black plug can be inserted into the COM jack. When
the reading is stable, the temperature value can be
read directly from the display screen.
Notice:
D The maximum temperature of the randomly
packaged K-type thermocouple is 250°C, and
the instantaneous measurement can be up to
300°C.

16
5.
Maintenance
5.1 Change
battery
EWarnings
Before
opening
Warning
battery cOver
of the
-A
instrunment,remove the
pens from the circuit being
measured, to avoid electric shock.

1)If symbol" A" appears it indicates to change the


battery.
2) Unfasten bolts on battery cOver of instrument and
remove the cOver.
3) Change the old battery.
4) Place the battery cover.
E Note:] Do not reverse battery polarity.
5.2 Change pens
EWarnings
Warning -A
When changing the pen, it is required to

replace with an identical pen or a pen of the same


level. The pen must be in good condition,and level
of the pens is: 1000V 10A.

Ifinsulation layer of the pen is damaged, or metal wire


of the lead exposes,it is required to change the pen.
6. Accessories
1) Pens Level: 1000V 10A One pair
2) Use Manual One copy

3) Battery 1.5V AAA battery 2PCS


Optionalfunction accessories 1PCS
A) K-Type