Smartlyte Plus
Smartlyte Plus
The Diamond Diagnostics SMARTLYTE® PLUS analyzer is for In Vitro Diagnostic use.
The information in this manual was correct at the time of printing. However, Diamond Diagnostics
Inc continues to improve products and reserves the right to change the specifications, equipment, and
maintenance procedures at any time without notice.
If the system is used in a manner differently than specified by Diamond Diagnostics Inc., the
protection provided by the equipment may be impaired. See warning and hazard statements.
                                                          ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 2 of 58
                             Important Information!
This Service Manual contains important warnings and safety information to be observed by the user.
This instrument is only intended for one area of application which is described in the instructions. The
most important prerequisites for application, operation and safety, are explained to ensure smooth
operation. No warranty or liability claims will be covered if the instrument is applied in areas other than
those described or if the necessary prerequisites and safety measures are not observed.
The instrument is only to be operated by qualified personnel capable of serving these prerequisites.
Only accessories and supplies either delivered by or approved by DIAMOND are to be used with the
instrument.
Due to the operating principle of the instrument, analytical accuracy not only depends on correct
operation and function, but also upon a variety of external influences beyond the manufacturer’s control.
Therefore, the check results from this instrument must be carefully examined by an expert, before further
measures are taken based on the analytical results.
Instrument adjustment and maintenance with removed covers and connected power mains are only to be
performed by a qualified technician who is aware of the dangers involved.
Instrument repairs are only to be performed by the manufacturer or qualified service personnel.
For a detailed service training video for your SMARTLYTE PLUS Analyzer, please visit:
http://www.diamonddiagnostics.com/video/video_SMARTLYTE PLUS.htm
Symbol Explanation
                   Attention symbol - Refer to the Operator's Manual or Service Manual for further
                   instructions. This symbol is located on the inside of the instrument.
                   Type B instrument symbol - An instrument of the B type falls under safety categories
                   I, II, or III, or has an internal power supply providing the required insulation against
                   discharge current and reliable ground connections.
Important Information!
This device complies with Part 15 of the FCC Rules. Operation is subject to the following two conditions:
(1) this device may not cause harmful interference's, and (2) this device must accept any interference
received, including interference that may cause undesired operation.
Warning: Changes or modifications to this unit not expressly approved                            by    the    party
responsible for compliance could void the user's authority to operate the equipment.
Note: This equipment has been checked and found to comply with the limits for a Class B digital device,
pursuant to Part 15 of the FCC Rules. These limits are designed to provide reasonable protection against
harmful interference in a residential installation. This equipment generates, uses, and can radiate radio
frequency energy and, if not installed and used in accordance with the instructions, may cause harmful
interference to radio communication. However, there is no guarantee that interference will not occur in a
particular installation. If this equipment does not cause harmful interference to radio or television
reception, which can be determined by turning the equipment off and on, the user is encouraged to
try to correct the interference by one or more of the following measures:
Caution:
   • The instrument is designed as a conventional device (closed, not waterproof type).
   • Do not operate the instrument in an explosive environment or in the vicinity of explosive
       anesthetic mixtures containing oxygen or nitrous oxide.
   • This instrument is suitable for continuous operation.
   • The power plug is to be plugged into a ground socket only. When using an extension cord, make
       sure that it is of the proper size and is properly grounded.
   • Any breakage of the ground lead inside or outside the instrument or a loose ground
       connection can cause a hazardous condition when operating the instrument. Intentional
       disconnection of the grounding is not permitted.
   • When handling electronic components such as – CPU, RFID PCB, Interface-PC, and printer PCB
       – ESD (Electronic Sensitive Device) protection is recommended. ESD mats should be used when
       handling and working with these electronic components. ESD packaging is recommended for
       transportation.
1.  Introduction ................................................................................................................................................. 7
  1.1. General Description ............................................................................................................................ 8
    1.1.1. Measurement .................................................................................................................................. 8
    1.1.2. Operating Configurations .............................................................................................................. 8
  1.2. Software Operation ........................................................................................................................... 10
  1.3. Description of Modules.................................................................................................................... 11
    1.3.1. Mechanical Assemblies ............................................................................................................... 11
    1.3.1.1. Front Door Assembly ............................................................................................................... 11
    1.3.1.2. Rear Panel Subassembly ........................................................................................................ 14
    1.3.1.3. Power Supply Subassembly ................................................................................................... 15
    1.3.1.4. Power Inlet Module Subassembly ......................................................................................... 16
    1.3.1.5. CPU Board (CN-A4405) .......................................................................................................... 17
    1.3.1.6. Aspiration Assembly................................................................................................................. 21
    1.3.1.7. Electrode Tray Assembly ........................................................................................................ 26
    1.3.1.8. Peristaltic Pump Rotor (CN-A4014)....................................................................................... 29
    1.3.1.9. Peristaltic Pump Motor............................................................................................................. 30
    1.3.1.10. Solenoid Valve System ........................................................................................................ 31
    1.3.1.11. Printer Assembly (CN-A4060)............................................................................................. 34
    1.3.1.12. RFID Board (CN-A4054)...................................................................................................... 36
    1.3.1.13. Valve Board (CN-A4018) ..................................................................................................... 36
    1.3.1.14. Display Board (CN-A4406) .................................................................................................. 38
    1.3.1.15. Interface PC ........................................................................................................................... 39
  1.4. Fluidic Modules.................................................................................................................................. 40
    1.4.1. Standard A..................................................................................................................................... 40
    1.4.2. Standard B and C ......................................................................................................................... 40
    1.4.3. Reference Solution ...................................................................................................................... 40
    1.4.4. Waste Liquids ............................................................................................................................... 40
    1.4.5. Fluidic Flow Path .......................................................................................................................... 41
    1.4.6. Pinch Valve Tubing Set (AV-BP5193) ...................................................................................... 42
    1.4.7. Fluidic Diagram ............................................................................................................................. 45
  1.5. Electronics .......................................................................................................................................... 46
    1.5.1. Block Diagram of SmartLyte® Plus ........................................................................................... 46
    1.5.2. CPU Board .................................................................................................................................... 46
    1.5.3. Power Circuits ............................................................................................................................... 48
    1.5.4. Microcontroller .............................................................................................................................. 48
    1.5.5. EMF and Noise Shielding ........................................................................................................... 48
    1.5.6. Fluid Pack Sensor ........................................................................................................................ 48
    1.5.7. Valve Drivers ................................................................................................................................. 49
    1.5.8. Pump Motor Driver ....................................................................................................................... 49
    1.5.9. Input Amplifiers ............................................................................................................................. 49
    1.5.10. Temperature Sensor circuit..................................................................................................... 49
    1.5.11. Analog Channel and A/D Converter ...................................................................................... 49
    1.5.12. CPU board Connector ............................................................................................................. 49
    1.5.13. Valve Board ............................................................................................................................... 49
    1.5.14. Display Board ............................................................................................................................ 50
    1.5.15. Printer ......................................................................................................................................... 50
    1.5.16. Shielding and Grounding ......................................................................................................... 50
    1.5.17. Interface PC............................................................................................................................... 50
2. Scheduled Maintenance ......................................................................................................................... 51
                                                                                             ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 5 of 58
                                                                                                                                             Table of Contents
  2.1. Daily ...................................................................................................................................................... 51
  2.2. Weekly Maintenance ......................................................................................................................... 51
  2.3. Monthly Maintenance ....................................................................................................................... 51
  2.4. Six Month Maintenance ................................................................................................................... 51
  2.5. Annual Preventive Maintenance.................................................................................................... 51
3. Troubleshooting ....................................................................................................................................... 52
  3.1. Error Messages and Troubleshooting ......................................................................................... 52
4. Parts List .................................................................................................................................................... 58
  4.1. Consumables...................................................................................................................................... 58
  4.2. Analyzer Parts for Repair ................................................................................................................ 58
1. Introduction
                                              ®
    The Service Manual for the SmartLyte Plus Electrolyte Analyzer contains the technical information
    needed to ensure easy fault identification. This manual is intended to be complementary to the Operator's
    Manual where detailed instructions for operation, maintenance and troubleshooting are provided.
    As for all clinical instrumentation, a thorough understanding of the principles of operation is prerequisite to
    attempting service of this product. Training along with experience will enhance the use of this manual.
                                         ®
    Service and repair of the SmartLyte Plus analyzer should be performed only by qualified repair
    technicians. Care should be taken when removing the covers as hazardous voltages are exposed. Use
    only accepted electronic check procedures and static protection when replacing and handling all
    electronic parts.
    This manual is divided into 4 chapters to facilitate location of technical information. Chapter 1 includes all
    mechanical, fluidic and electronic assemblies are described. Chapter 2 outlines routine maintenance and
    troubleshooting procedures. Chapter 3 includes troubleshooting, including error messages. Finally,
    Chapter 4 includes part identification.
1.1.1. Measurement
         Sample type                Whole blood, serum, plasma, urine, QC material, Aqueous samples
         Sample device              Syringe, sample cup, collection tube
         Sample size                95 uL
         Analysis time              28 seconds
         Sample rate                128 per hour
         Sodium (Na+) sensor        Ion-selective, flow-through, glass capillary electrode
         Potassium (K+) sensor      Ion-selective, flow-through, liquid membrane electrode
         Chloride (Cl-) sensor      Ion-selective, flow-through, liquid membrane electrode
         Calcium (Ca++) sensor      Ion-selective, flow-through, liquid membrane electrode
         Lithium (Li+) sensor       Ion-selective, flow-through, liquid membrane electrode
         Reference System           Open liquid junction calomel electrode
         Calibration                Fully automatic
                                    1 point with each sample
                                    2 point every 4 hrs (Lithium, 3 point)
         Warm-up time               2 minutes
         Temperature                Room temperature, 15 - 32° C, 60 - 90° F
         Humidity                   Maximum 85% relative humidity, non-condensing
                                                               ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 8 of 58
                                                                                                      1 Introduction
Data management             Quality Control memory storage, 3 levels, 500+ values each;
                            Calculation of mean, standard deviation, coefficient of variation (CV)
                            Samples, 10000+ measurements
Diagnostic Programs         Tests selected in Diagnostic Menu
Electronics                 Microprocessor-controlled and Interface-PC
Display                     Graphical touch screen, 800 x 400 pixels
Printer                     Integral thermal printer, 16 character width
Interface                   RS232C Serial Port
                            4 USB Ports
                            LAN/Ethernet
Electrical requirements     100 - 240 V, 50/60 Hz,
                            1.4 A max.
                            Self-adjusting
Nominal power
Consumption                 Minimum 10W/hour
                            Typical 25W/hour
                            Maximum 35W/hour
Dimensions
    Height                  335 mm; 13.2"
    Width                   315 mm, 12.4"
    Depth                   295 mm; 12.0"
    Weight                  approx. 6kg; 14 lbs;
Classifications
    Safety category         I
    Device Type             B (according to EN 60601-1, 6010-2-101)
    Mode of operation       Continuous
    Protection class        IP 20
    Explosion protection    The device is not designed for operation in explosive environments.
Data subject to change without notice. Technical information is supplied for general informational
purposes only.
Operations
Measurment
1. Front Door
2. Plastic Window
3. EMI Window
4. Door Magnets
                                                                                              Plastic Window
                                                                                              on Front
             Front Door
             Assembly
                                                      Copper Conductive
                                                      Tape
Door Magnets
1. Move the analyzer to the edge of a work surface so the analyzer door, when opened, will
   extend beyond and below the work surface.
2. Remove the red ground connector from the Right Angle Tab by grasping the red connector
   and pulling to the left exposing the Right Angle Tab. (Figure 1-5)
Ground Connector
3. Then remove the ground screw, external tooth washer, right angle tab, and hex nut. (Figure
   1-6)
Hex Nut
                                                                              External Tooth
                                                                              Washer
                                                                              Ground Screw
    Figure 1-6 Door Hinge Attachment Details
4. With one hand, hold the analyzer door near the right side hinge pin. (Figure 1-7)
6. With the other hand, gently apply pressure to the middle rear area of the door. This will flex
   the door slightly and allows the right hinge pin to clear the retaining hole in the main chassis.
7. Remove the right hinge pin then the left so the door can then be removed from the analyzer.
   (Figure 1-7).
Right Hinge
                                                                                         Apply Gentle
                                                                                         Pressure
The window is comprised of two layers, the EMI window and the plastic window. Both can be
replaced in the field.
1. From the rear side of the door, (Figure 1-4), gently press the window and snapping the EMI
   and plastic window out the front side of the door.
2. Verify that the copper conductive tape is still adhered to the both the front and back surface
   of the door.
4. Place the EMI Window first over the opening on the front side of the door.
5. Press along the frame of the window and ensure good contact of the EMI window with the
   copper tape on the frame of the window.
6. After removing the adhesive backing, apply the plastic window over the EMI window pressing
   along the frame again to ensure good contact.
   The Rear Panel Subassembly contains the Power Supply Subassembly, the Power Inlet Module
   and the External Communication Board. To separate the Rear Panel Subassembly, (Figure 1-8)
   from the Front Case perform the following steps,
                                                                               Interface PC with 4
                                                                               USB Ports & LAN
                                                                               connection
                                                                               Power Supply
                                                                               Module
4. Move the back case a few inches away from the chassis.
   5. Disconnect cables and wires so that the rear case can be separated from the front case.
      (Figure 1-8)
                                                                                   Interface PC
                                                                                   Ribbon Cable
                                                                                   Red Instrument
                                                                                   Ground Connector
   6. Disconnect the communication ribbon cable from the External Communication Board on the
      back panel leading to the Display Board.
   7. Disconnect the power supply cable attached to the CPU board by depressing the locking clip
      on the left side of the connector and gently pulling the power cable out.
8. On the left side of the chassis, pull the top red grounding wire connector off.
To access the Power Supply Subassembly, the Rear Panel must be removed first.
                                                                                                Power Inlet
                                                                                                Module Connector
                                                                                               Red Ground
                                                                                               Connectors
Figure 1-10 Rear Panel with Power Supply and Main Power Receptacle
4. Remove Ground Cables from Power Inlet Module and left side of Dual Tab.
   5. Finally, remove the four screws securing Blue Ring Connector and the Power Supply
      Mounting Plate to the Rear Panel. (Figure 1-10 screws circled in red)
                                                                                   Power Supply
                                                                                   Connect To
                                                                                   Mounting Plate
                                                                                Power Supply
                                                                                Connector to CPU
                                                                                Board
                                                                                      Connect to Right
                                                                                      side of Dual Tab
                                                                                      Connect to Left
                                                                                      side of Dual Tab
                                                                                      Connect to Power
                                                                                      Inlet Module
                   The Power Inlet Module can be found on the Rear Panel (Figure 1-13), to separate from the Rear
                   Panel.
                   1. First follow the instructions in Section 1.3.1.2 to separate the Rear Panel from the Front
                      Case.
                                                                                     Power Inlet
                                                                                     Module
                                                                                     Power Inlet
                                                                                     Module Connector
                                                                                     Ground Wire
                                                                                     Connector
4. Remove the two screws on the external face of the Rear Panel. (Figure 1-14)
   The CPU board can be replaced when there is a malfunction. ESD precautions should be
   exercised to prevent damaging the electronic components. To remove and replace the CPU;
   1. First remove the Rear Panel according to instructions in Section 1.3.1.2 to expose the CPU
      board.
2. Once the Rear Panel is removed, unplug the Display cable from the CPU board.
Display Cable
                                                                                        Interface PC
                                                                                        Ribbon Cable
Alignment Pin
                                                                                       Valve Board
                                                                                       Connector
CPU Screws, 6
                                                                                       Top Grounding
                                                                                       Screw
                                                                                       Power Cable
                                                                                       Connector
   Figure 1-16 Connectors and cables on CPU Board
3. On the CPU board, unclip the top and bottom tabs of the valve board connector. Holding the
   locking tabs open at the same time helps to push the ribbon cable connector out of the CPU
   connector.
5. Remove the 6 screws with their plastic washers that hold the CPU board to front case.
   (Figure 1-16 circled in red).
6. Carefully pull the CPU board off the alignment pin (Figure 1-16 circled in white). Ensure the
   RFID board and the printer is not dislodged.
7. Handle board carefully so as to not damage the Electrode Pins on the front of the CPU
   board.
8. Each Pin can be removed by pulling the Electrode Pin from its socket. (Figure 1-19)
                                                                   Display Cable
                                                                   Connector
Shield
                                                                   Valve Board
                                                                   Connector
                                                                   RS232 Serial
                                                                   Port
                                                                  Power Cable
                                                                  Connector
                                                                                 Printer
                                                                                 Connector
                                                                                 Electrode Pogo
                                                                                 Pin Socket
                                                                                 Electrode Pogo
                                                                                 Pin
Optical Sensor
                                                                                 RFID Board
                                                                                 Connector
To install the CPU board, certain steps should be taken to ensure proper placement.
1. Ensure RFID board is installed with its components and antenna facing the Electrode Pins.
2. Pull each Electrode Pin out slightly so that they extend beyond their normal positions. This
   will prevent Electrode Pins from being pushed too far into the socket.
4. Check that Electrode Pins are placed in front case without pushing pins into their sockets.
6. Secure CPU Board to the front case using the screws and plastic washers, previously
   removed. Leave screws slightly loose.
7. Check that each Electrode Pin is entirely visible through the Electrode mating holes. (Figure
   1-19)
                                                                       5 Electrode Pin
                                                                       Mating Holes
8. If pins are not all visible, adjust the position of the CPU board.
9. When all pins are visible, tighten all screws without over tightening.
   This module is comprised primarily of two subassemblies, the Aspiration Sub-Assembly and the
   Sample Probe.
Sample Door
                                                                                    White Sample
                                                                                    Probe Holder Block
                                                                                    Aspiration
                                                                                    Sub-assembly
Sample Probe
Fill Port
On the Aspiration Assembly, the sample probe and fill port are designed for easy replacement.
Intake Tubing
Probe
Sample Door
Fill Port
The sample probe assembly (Figure 1-22) is comprised of the probe and the intake tubing. The
assembly can be removed as one-piece from the Aspiration Assembly. To remove,
1. First, if the instrument is on, place in Hold in Maintenance mode to prevent reagents from
   cycling.
3. Disconnect the intake tubing from the sample sensor on the right side of the electrode tray.
   (Figure 1-25)
5. Grasp the probe near the white holder block, pull up to rotate and unsnap the sample probe
   from its position. (Figure 1-21)
It is recommended to replace the Sample Probe each year. To replace the Sample Probe
assembly;
1. Align the bottom of the sample probe in the white Sample Probe holder block and press the
   replacement assembly into place.
2. Press the top of the sample probe near the intake tubing to ensure the probe is completely
   snapped into the white Sample Probe Holder.
                                                                                  Intake Tubing,
                                                                                  Connected
Sample Sensor
The fill port can be removed and replaced as a complete assembly. It can also be disassembled
and cleaned if desired. To remove the fill port assembly;
1. Pull the fill port holder towards the front for easy access. (Figure 1-26)
Fill Port
2. Press the two plastic tabs on the fill port to allow removal of the fill port from the fill port
   holder. (Figure 1-26 and 1-27)
To remove the aspiration assembly Rear Panel (Figure 1-28) which contains the Power Supply,
Power Inlet Module, USB and Ethernet connectors must be removed first
Power Supply
4. Move the back case a few inches away from the chassis.
5. Disconnect cables and wires so that the rear case can be separated from the front case.
   (Figure 1-29)
       a. Disconnect the communication ribbon cable from the Communication Board on the
           back panel leading to the Display Board.
       b. Disconnect the power cable attached to the CPU board by depressing the locking clip
           on the left side of the connector and gently pulling the power cable out.
Once the CPU board has been removed, the Aspiration Sub-Assembly can be removed.
1. The cable on the far left side of the valve board is the aspiration assembly connector. Press
   the connector tab to disconnect. (Figure 1-29)
Valve Board
                                                                                        Aspiration Sub-
                                                                                        Assembly
                                                                                        Connector
                                                                                         Rear Hex
                                                                                         Screw
6. Remove the two rear screws using a hex key (Figure 1-29)
7. Then, remove the front securing screw using the same hex key. (Figure 1-30)
                                                        Front Hex
                                                        Screw
                                                      Reagent Supply
                                                      Tubing (tagged
                                                      red)
10. Check that the sample intake tubing is disconnected from the sample detector. (Figure 1-25)
11. Disconnect the reagent supply tubing (red tag) from beneath the fill port. (Figure 1-30)
12. The aspiration assembly can now be removed from the front of the analyzer for replacement.
   This assembly is comprised of the electrode tray and the sample sensor, AV-BP5036D. (Figure
   1-31)
Sample Sensor
                                                                                     Electrode Tray
                                                                                     Base
   1. First, if the instrument is on, place in Hold in Maintenance mode to prevent reagents from
      cycling. (Main Operations Menu>>>Maintenance>>>Hold in Maintenance)
2. Disconnect the sample sensor cable from the analyzer. (Figure 1-31)
   3. Remove the sample intake tubing on the right side of the assembly where the sample sensor
      is located. (Figure 1-31)
                                                                                                       Sample Sensor
                                                                                                       Connector
                                                                                                       Sample Intake
                                                                                                       Tubing
                                                                                                       Waste Tube
                                                                                                       (tagged Green)
                                                                                                       Reference Tubing
                                                                                                       Connector
       Figure 1-32 Electrode Tray with electrodes installed
5. Disconnect the waste tube (green tag) from the left side of the electrode tray.
6. Use one hand to press the Electrode Tray Plastic Tabs toward the center.
7. At same time, place fingers over the electrodes and thumb below the electrode tray, pull the
   tray towards. (Figure 1-33) The Electrode Tray should slide out easily and electrode pins will
   be out of the mating holes.
8. Use each hand to press the plastic tabs on the right and left side of the electrode tray
   assembly towards the center of the tray.
9. Slide the Electrode Tray towards you to completely release the tray and remove the
   assembly from the analyzer. (Figure 1-34)
The sample sensor (on the right) and the electrode locking mechanism (on the left) can be
removed and replaced by removing the screws on the underside of the tray. (Figure 1-35)
   The Peristaltic Pump Rotor works with the Peristaltic Pump Motor to manage all fluid flow through
   the analyzer. (Figure 1-37)
                                                                        Peristaltic
                                                                        Pump Rotor
1. Unwind front and back Peri-Pump tubing from the Motor shaft. (Figure 1-38)
Back Tubing
Front Tubing
2. Grasp the Rotor Assembly firmly and pull away from the instrument.
1. Align the rotor flat interface with the flat on the motor shaft.(Figure 1-39)
Flat Surface
3. There will be a small gap between the roller and the surface of the instrument.
   The Peristaltic Pump Motor manages all fluid flow in the analyzer through the Peristaltic Pump
   Rotor. These are the steps for removing the motor.
1. Remove the Peristaltic Pump Rotor according to the procedure above, Section 1.3.1.8.
2. Remove the rear panel and CPU Board as described in Section 1.3.1.2.
3. Locate the Pump Motor electrical connector plugged into the valve board. (Figure 1-40)
Valve Board
                                                                                   Pump Motor
                                                                                   Connector
                                                                                   Pump Motor
       Figure 1-40 Connectors and cables on CPU board, Pump Motor
5. On the front of the analyzer remove the four screws holding the motor in place.
6. The motor can now be removed from the back side of the instrument.
    The Solenoid Valve System is comprised of the solenoid valve subassembly, CN-A4015, the
    crimp pin, CN-4155, and the indicator cap, CN-4018. The Crimp Pin applies pressure to the
    tubing to shut off fluid flow. The Indicator Cap holds the Crimp Pin in place.
Motor Shaft
Crimp Pin
                                               Indicator Cap/
                                               Opening
    There are 5 solenoid valves, one each for Standard A (A), Standard B (B), Standard C (C),
    Reference Solution (R), and Air (V). The removal of each solenoid valve is identical. (Figure 1-
    45)
                                                                        V Valve Solenoid
                                                                        Screws
Solenoid Valve assembly removal requires the removal of the indicator cap and crimp pin first.
Proceed as follows;
1. If the instrument is on, place in Hold in Maintenance mode to prevent reagents from cycling
   and valves from activating.
   (Main Operations Menu>>>Maintenance>>>Hold in Maintenance)
2. Locate the opening of the indicator cap where the solenoid shaft is visible. Orient opening
   upwards.
                                                   Indicator Cap
                                                   Opening
3. Push down on Indicator cap in the direction of the arrow to slide it off the solenoid valve shaft.
To install the crimp pin and cap, the instrument must be on.
3. Press the button for the valve that needs the Crimp Pin and Indicator Cap installed.
(Check Valves>>>Check Valve A (B, C, R or V)
4. This energizes the solenoid and the shaft is extended to the outer most position
6. Orient the opening of the Indicator Cap on the shaft and snap on by sliding in the direction of
   the arrow.
7. If the solenoid shaft retracts before the Crimp Pin and Indicator Cap are in place, press ON
   button to re-energize and move the shaft.
1. First follow the steps above for the removal of the Indicator Cap and Crimp Pin.
2. Remove the two screws on each side of the solenoid shaft. (Figure 1.44)
3. Remove the rear panel and the CPU Board as described in Section 1.3.1.2.
4. Each solenoid has an electrical connection to the valve board which must be unplugged prior
   to removal of the solenoid. (Figure 1-47)
Retaining Ring
Valve Board
Connectors
    The printer is a graphics printer mounted to a board which is connected directly to the CPU
    board. It uses thermal paper, AV-BP5025D.
    It is important to follow the instructions below to avoid damage to the printer or CPU board.
       Printer Board
       Guides
                                                             Printer Paper
                                                             Feed Button
Printer Paper
                                                             Printer Paper
                                                             Tray
1. Ensure the power to the instrument is off before removing the printer.
2. Open the Main Door so that the printer compartment is fully exposed.
    4. Tear the paper roll from the printer and completely remove the paper roll and the paper tray.
       Do not pull the paper from the printer as this can damage the printer.
    6. Slide fingers under the printer green board along the side to the back until an opening is
       detected.
    7. Holding by the opening, pull the printer towards you along the printer board guides on the
       analyzer until the printer is disengaged from the CPU board
Removal of the printer should be performed for replacement and for removal of a paper jam. To
replace the printer;
2. Locate the printer board guiding tracks and slide a printer onto the analyzer.
3. Press the printer firmly into place to ensure proper connection to the CPU board.
4. Verify the printer is working by running a printer test under the Diagnostic Menu.
(Main Operations Menu>>>Diagnostics>>>Check Printer)
Note: Never attempt to dislodge paper from the printer with a paper clip or similar object
to avoid damage to the print head or printer platen.
The RFID board is attached to the CPU board at JP10. (Figure 1-29 CPU Board, Front)
    The RFID board can be accessed after removal of the CPU board as described in Section
    1.3.1.5.
    This board must be connected all the time to the CPU board for proper functioning of the
    instrument.
    DO NOT unplug the RFID board while the CPU board is powered ON. Irreparable damage may
    occur.
                                                                               RFID
                                                                               Antenna
                                                                           RFID
                                                                           Components
                                                                                 Connector to
                                                                                 CPU Board
    Valve board has multiple connections to devices on a single platform. The Aspiration Assembly,
    Solenoid Valves, Peri-Pump Motor, all plug into the valve board. The lamp that lights the
    measurement pathway is also located on this board. To access the board;
1. The Rear Panel and CPU with RFID boards, described in Section 1.3.1.2 must be removed.
    2.   Remove one screw with washer on the right side which holds the board in place. (Figure
         4.36 below)
    3. Then remove the seven electrical connectors, aspiration, solenoid valves, pump motor, to
       slide this board out.
                                                                                 Peri-Pump
                                                                                 Motor Plug
                                                                                 Solenoid Valve
                                                                                 Plugs
                                                                                 Aspiration
   Figure1-52 Valve Board                                                        Assembly Plug
The top side of the Valve Board has three lamps that illuminate the measurement pathway
when fluids flow.
Lamps
The Bottom side of the Valve Board seven connectors for the Aspiration Assembly, Solenoid
Valves and Pump Motor.
    The Rear Panel and CPU board, referenced in Section 1.3.1.2 must be removed to gain
    complete access to the display board. To remove the display board;
1. Follow the steps for removing the Rear Panel in Section 1.3.1.2
Display Spacers
Display Screws
4. Then remove three screws and each of the 50 degree spacers to remove the display board.
Display Screen
                                                                                    Sample Sensor
                                                                                    Connector
        Figure 1-56 Valve Board, Front
                                                                                                     Display Cable
                                                                                                     Connector
                                                                                                     Data Module
                                                                                                     Connector
1.3.1.15. Interface PC
                                                                               Interface PC
                                                                               Module
SD Card
1.4.1. Standard A
         Standard A, STD A, is drawn to the electrode module by a vacuum provided by the peristaltic
         pump. When STD A is to be aspirated into the electrode module, solenoid valve A is opened and
         solenoid valves B, C, V and R are closed.
         STD A is drawn from the Fluid Pack, to the fill port, through the sample probe up to the sample
         sensor. When the sample sensor detects fluid, solenoid valve V is opened and solenoid valve A is
         closed. The peristaltic pump continues to pump STD A into the electrode module. At the trailing
         edge of the STD A is air.
         When the sample sensor detects air, the peristaltic pump stops. Standard A should fill the
         electrode module. During the time STD A is aspirated into the electrode module, the reference
         housing is pressurized due to the peristaltic pump providing pressure to the reference solution
         line and solenoid valve R being closed. A small amount of reference solution is forced out through
         the reference junction to provide contact to the sample.
         Once measurement is complete, the peristaltic pump moves the fluid back to the Fluid Pack into
         the Waste Container.
         The reference solution is circulated back into the Reference Solution container in the Fluid Pack.
         The reference housing has an in and an out port. It is filled automatically using the second
         winding of the peristaltic pump and solenoid valve R. The reference solution connector allows for
         the reference housing tubing to be plugged into the reference solution circuit. As the peristaltic
         pump aspirates reference solution, solenoid valve R is opened to allow excess solution to be
         pumped into the reference return line of the Fluid Pack.
         Calibration standards and sample waste are all pumped out of the left side of the electrode tray
         (green-banded tubing) through the peristaltic pump to the waste line and into the waste bag of the
         Fluid Pack. The Waste Container has an internal one-way valve is incorporated to prevent waste
         material from leaking out of the Fluid Pack.
      The pinch valve tubing set contains all tubing required for the analyzer. The tubing set has a pre-
      cut tubing, a harness with fluidic interconnections preassembled. The Peristaltic Pump Tubing,
      AV-BP5027D, is also included in the set. (Figure 1-60)
 Waste Out
                                                                     ↑
                                                                     Reference
                                                                     In
                                           Reference Out
< --Waste In
Replacement of the tube set should be performed annually. Follow the procedure below;
      1. Place the instrument in Hold in Maintenance mode to prevent reagents from cycling.
      (Main Operations Menu>>>Maintenance>>>Hold in Maintenance)
      3. Next, follow the procedure outlined in Section 1.3.1.10 to remove the indicator cap and crimp
         pin from each solenoid valve. This facilitates the removal of tubing from each solenoid.
4. Pull the two pump tubes of the peristaltic pump tube set off the pump rotor.
      5. Unplug the reference from the connector. Use a towel to absorb any reference solution that
         leaks from the housing
7. Gasp the front of the white TPR block and slide it out of the analyzer towards you.
      8. Disconnect the green tagged tube from the electrode tray and the red tagged tube from the
         bottom of the fill port.
      10. Remove parts from the tubing kit which include the tubing harness, green tagged waste tube
          and Peristaltic Pump Tubing separately.
11. Connect the green tagged tube to the electrode tray as shown in Figure 1-62.
12. Connect the waste tube from the TPR block to the Peristaltic Pump.
14. Connect the reference tube from the reference solution connector to the peristaltic pump.
    Ensure the reference tubing’s are connected correctly. Misplacement will cause malfunction
    of the reference electrode.
16. Pull tubes out below the TPR block on left side.
17. Install the reference tube and reference solution tube connector.
18. Place tubing into Left Guide Track, with shortest first (Waste In) to the longest, Standard A.
19. Place all Standard A, B, C, Reference solution and Air tubings into their tracks followed by
    their respective crimp pin then the indicator cap.
21. Connect the red tagged tube to the bottom of the fill port.
23. Install the Fluid Pack and run Prime All to fill reagent lines.
(Main Operations Menu>>>Maintenance>>>Prime All).
                                                         Peristaltic
          TPR          Tubing                            Pump Tubing
                       Cradle
Green
Tube                                                                                                    Blue
                                                                                                        Tube
                                                                                                 Red
                                                                                                 Tube
    R
                                     V
1.5. Electronics
            •      Pump
            •      Valve
            •      Printer
            •      RS232 LIS
            •      Lamp
            •      Temperature Sensor
            •      Sample Sensor
            •      Signal processing
            •      RFID
            •      Fluid pack sensor
            •      Door
            •      Buzzer
            •      Power distribution
1.5.4. Microcontroller
         The STM32F407 microcontroller, represented as U12, can be located on the CPU board
         (slight above & on right of the buzzer). This processor is preloaded with customized software
         to control the peripherals – valve, pump, printer, RFID, signal processing, temperature
         sensor, sample sensor and buzzer.
          IC U24 – U26 Provide high to low input impendence matching for each electrode (Na, K, Cl,
          Ca, & Li) input. These amplifiers provide a gain of 10 for all incoming signals.
          The analog inputs from the amplifiers are connected to IC U23. This IC converts analog
          signal to digital readings that is then used by software for instrument operation.
          In order to fulfill all responsibilities of the CPU board and the program installed on it, is
          equipped with a number of connectors to interface it with all peripherals. The description of
          the JP connectors is as below:
          The valve board contains the LED array used to illuminate the measuring chamber and the
          interconnectors used for all solenoids, the peristaltic pump and the door sensor. A ribbon
          cable from the CPU Board supplies the electrical connection for each signal through
          connector JP5.
                                                           ECO 9099 SOP05-5090F Rev.00 Effective Date: 12/20/18 Page 49 of 58
                                                                                                             1 Introduction
          The display board is comprised of 2 stacked boards. The 1st board has the 5” touch screen
          display mounted on it. This display is then connected to display adaptor board that gets
          power from CPU board and HDMI signals from interface PC. Touch signals are sent back to
          the interface PC via flex ribbon cable.
          Additionally, the display adaptor board has one RJ11female connector mounted on it. Sample
          sensor male RJ11 connector from electrode tray plugs to RJ11 female connector.
1.5.15. Printer
          Beside the shield for eliminating the internal noise from the power supply, the entire
                     ®
          SmartLyte Plus is equipped with a protective shield to prevent the influence of magnetic
          fields, and other noise sources in the laboratory. The internal protective shield creates a
          “Faraday Cage” that protects instrument measurements from outside noise. This shield is
          also grounded with a grounding wire in the power receptacle in the power switch input.
1.5.17.   Interface PC
                     ®
          SmartLyte Plus has an interface PC which has pre-installed Linux OS installed on it. This Interface-
          PC is responsible for following peripherals/features:
2. Scheduled Maintenance
                                                     ®
     Maintenance procedures for the SmartLyte Plus Electrolyte Analyzer require minimal time by the
     operator to perform. The procedures outlined below should be performed according to the schedule
     indicated. Detailed instructions for these procedures can be found in the Operator's Manual describing the
     correct method to perform routine analyzer maintenance.
2.1. Daily
3. Troubleshooting
                       ®
    The SmartLyte Plus Electrolyte Analyzer is designed to provide long, trouble-free service. However, any
    scientific measuring device may occasionally malfunction, requiring the operator to identify the cause of the
    problem.
    The messages in the table below provides the likely cause and recommends steps that should help you return
                  ®
    your SmartLyte Plus Electrolyte Analyzer to operation.
                           ®
    If the SmartLyte Plus Electrolyte Analyzer does not perform correctly after conducting the basic steps outlined in
    this chapter, you should contact Technical Service at Diamond Diagnostics Inc. Support staff can assist you over
    phone or email. Online-Remote Support feature under Network > Remote Support menu is also available for
    remote support/training.
ERR                  When displayed in place of sample        •   Ensure the electrode securely plugged in place.
                     results, the analyzer is unable to       •   Check for proper filling of the reference
                     obtain valid voltage reading from the        electrode.
                     electrode.
                                                              •   Ensure sufficient sample size and proper filling
                                                                  of measurement chamber by running a sample
                                                                  and examining flow when backlight is on.
Expired Fluid        Inserted Fluid Pack has passed           •   Insert new pack
Pack                 expiry date..
Inspect electrodes   Stable reading of Standard A could       •   Ensure that electrodes are properly plugged in
                     not be obtained within 6 aspirations         and aligned correctly.
                     for all electrodes.                      •   Check the locking mechanism on electrode tray
                                                                  to ensure properly engaged.
                                                              •   Check Electrodes with STD-A. The mV should
                                                                  fall between -100 and 100. The stability is
                                                                  expressed as SD in parenthesis. Expected
                                                                  value is 0 to 10. If stability of electrodes is out of
                                                                  range, replace the reference electrode.
                                                              •   Perform daily maintenance.
                                                              •    If necessary, clean the reference electrode
                                                                  housing or replace the reference electrode.
                                                              •   Replace the Fluid Pack.
                                                              •   Check shielding connection to ground for short.
Interface error      Only displayed during ACTIVE             •   Check if pins 2 and 3 of the serial port were
                     INTERFACE. The interface test (see           connected properly.
                     chapter 8:”Testing the interface”) was   •   Check the serial cable.
                     not successful.
                                                              •   If the error persists, contact Technical Support.
4. Parts List
  4.1. Consumables