1. Introduction
■   Introduction
This product as an indicator for weighing instruction and control is designed for the most
suitable use of any meters (packer, weigh separator, platform scale, various tensile and
compression test equipment) that can be applied in the widest way throughout industries.
This product ensures communications with external devices such as computers as it basically
installs an RS-232C port.
In addition, this product can be conveniently used for the weighing control as it basically has
three relay outputs necessary for control (applicable only if there is relay out).
Please utilize all the functions in this product sufficiently as you can use the device properly
with good knowledge of this manual before you use this product.
■   Features
· Measures to screen off external noises
· Built-in self-diagnosis and self-recovery functions from failure
· External resolution : 1/20,000
· Electrostatic data memory
· 2 built-in external relay input terminals (zero, tare/tare removal)
· Data memory
· 24V of input power supply can be used with the use of DC power supply regardless of
  polarity (500mA< Trans Type > or more is recommended)
· FAC (Full Auto Calibration) method for calibration
· Equivalent input method for Calibration
· Basic installation of RS-232C
· Option : RS-422/485
· Option : I/V-OUT
                                            4
                           All manuals and user guides at all-guides.com
■   Cautions
·   Do not drop or apply severe shocks to the product.
·   Do not install the product with high voltage or severe electric noises.
·   Do not install the produce with direct sunlight or severe vibrations.
·   Connect the product with external peripheral devices after turning off the power switch.
·   Keep the product from water or rain.
■   Accessories
· Product manual
· Option (separate option for any addition): 1 EA of adaptor (DC 24V / 500mA)
                                               5
                          All manuals and user guides at all-guides.com
2. Specifications
■   Analog Part & A/D Conversion
        Input sensitivity                                    0.3㎶ / D
    Adjustment range of Zero                           - 0.6mV ∼ + 1.5mV
Non-linearity 0.01% FS
■   Digital Part
                                                  7-Segment, 5 digit RED FND
            Display
                                                Character size : 12.7(H) ×7.3(W)mm
            Division                                 ×1, ×2, ×5, ×10, ×20, ×50
     Below Zero Indication                                “-“ Minus Sign
         Decimal Point                                  0, 0.0, 0.00, 0.000
          Status Lamp                       Stable, Zero, High, Low, OK, Hold, Comm
          Function Key                             ZERO, TARE, SET, ENTER
                                            6
                            All manuals and user guides at all-guides.com
■   General Specification
           Power              DC 24V / About 500mA (Trans Type)
    Operating Temperature
                              -5℃ ~ +40℃
            Range
     Operating Humidity
                              Under 85% Rh (Non-condensing)
            Range
        Product Size          (W) 100 X (H) 52 X (D) 125 (mm)
                                                      m
       Product Weight         About 450g
                                        .co
■   Options
          Option - 1          RS-422/485
                                               7
                         All manuals and user guides at all-guides.com
■   Display
· Weight: The current weight at the indicator is displayed.
- Whenever the set key is pressed, the value for high limit, low limit or differences is
             displayed.
■   Display of Status
·   STABLE: It is displayed when the weight is stable.
·   ZERO: It is displayed when the weight is “0”.
·   LOW: It is displayed when the weight is the low limit or less.
·   HIGH: It is displayed when the weight is the high limit or more.
·   OK: It is displayed when the weighing is completed.
·   HOLD: It is displayed while the weight is being held.
·   COMM(RTxD): It is displayed while the product is communicating with other devices.
                                              8
                     All manuals and user guides at all-guides.com
■ Key Operations
  Key                                        Descriptions
        It is used to set the weight display to „0‟ within a range of maximum weight
        (CAPA) defined by the user.
        (However, it is operated within the input range by the user to „Equipment Set‟ „08‟
        (Set for Zero Range).
        It is used to set the tare to „0‟ after a container is put on the weighing tray.
        (However, it is operated within the input range by the user to „Equipment Set‟ „09‟
        (Set for Tare Range)
It is used for the input of low limit (SP1) and high limit (SP2).
Key Descriptions
                                             9
             All manuals and user guides at all-guides.com
                               10
                    All manuals and user guides at all-guides.com
5. Installation
■ External Dimension & Panel Cutting Size
(External Dimension X mm)
                                       11
                         All manuals and user guides at all-guides.com
 6. Calibration
 ■ What is calibration?
   Calibration refers to the adjustment to make the displayed value consistent with the
   actual weight on the display of weight.
                                                         m
  ☞ If the power is applied while any key is being pressed on the front, you may enter the test screen
   mode.
                                        .co
    - Press          Key to return to TEST1 after the sequential increase in TEST1~9.
    - Press         Key to enter the Test Mode as desired. Please refer to the following
      descriptions for each mode.
    - Press        Key again after setting or confirming each mode to return to TEST1.
      (TEST3 is exceptional).
    - Press           Key to enter ST.CAL.
                           es
       TEST 1             Zero weight value (A/D) can be confirmed.
                          Mode for the span calibration and equipment set under the SET, CAL mode.
       TEST 3
                          ☞ Press           Key to return to each mode selection from this status.
TEST 7 It is to select the analog amplification ratio. Press Key to return to TEST.
                                               12
                           All manuals and user guides at all-guides.com
■ Span Calibration
   The span calibration refers to the adjustment of linearity to set the actual weight consistent with
   the displayed value from “0” to the maximum weight as the standards in displaying the weight on
   the indicator.
                                                 13
                         All manuals and user guides at all-guides.com
I. Step 1
                                         It is a step to set the value of a division
                                         (minimum division of a scale to be displayed).
                                           In this, it is an abbreviation of division meaning “the
                                          value for a division of a scale (minimum division of a
                             scale to
                                         be displayed)”.
Ⅱ. Step 2
                                          It is a step to set the maximum display weight
                                        In this, “CAPA” is an abbreviation, which means
                                             the maximum display weight that can be weighted
                                             in the indicator.
                                         Enter the maximum display weight as the user
                                         desire instead of a random value that is currently
                                         displayed.
※ Do not set (a value for a division of a scale/maximum display weight) to more than (1/30,000). More than
(1/30,000) cannot be set. The product can be applied to at maximum 1/30,000.
                                              14
            All manuals and user guides at all-guides.com
Ⅲ. Step 3
                   It is a step to check zero status in the current
                    indicator.
                    A random value A/D is shown on the display,
                    which means zero for calibration.
Ⅳ. Step 4
                   It is a step to load the prepared standard
                   counterweights on the indicator.
                     In this, prepare standard counterweights for
                    10% or more of the maximum weight (CAPA)
                 Whenever             Key is pressed, the number increases in
                 0 -> 9
                Whenever              Key is pressed, it moves to the left
                Press            Key to check zero status and move
                Press       Key to store the currently displayed value for
                counterweights Move on to the next step.
                Proceed to Step 5 as it moves onto Step 3 of gage bar.
Ⅴ. Step 5
                   A span constant is shown on the display
                  It enters the weight mode as C.EnD. is
                  blinking and then reset.
                                 15
                    All manuals and user guides at all-guides.com
I. Step 1
                            It is a step to set the value of a division
                            (minimum division of a scale to be displayed).
                             In this, it is an abbreviation of division meaning “the
                             value for a division of a scale (minimum division of a scale to
                             be displayed)”.
Ⅱ. Step 2
                             It is a step to set the maximum display weight
                            written in the load cell
                            In this, “CAPA” is an abbreviation, which means
                            the maximum display weight that can be weighted
                            in the indicator.
                            Just enter the maximum weight displayed on the load cell
                            instead of the currently displayed random value.
                                       16
                               All manuals and user guides at all-guides.com
Ⅲ. Step 3
                                    It is a step to check zero status in the current
                                     indicator.
                                     A random value A/D is shown on the display,
                                     which means zero for calibration.
                                                                m
                                    Press        Key to move to the setting for the maximum
                                    display weight
                                    Press         Key to store the currently displayed
                                               .co
                                    zeroing value.
                                    Move on to the next step.
Ⅳ. Step 4
                                    It is a step to enter mV/V displayed on the loadcell.
                                    Ex) Enter 2.0000 if 2mV/V is displayed on the load cell.
                                 es
                                     Whenever               Key is pressed, the number increases in
                                     0 -> 9
                                     Whenever               Key is pressed, it moves to the left
                                     Press            Key to check zero status and move
                                     Press       Key to store the currently displayed value for
                uid
Ⅴ. Step 5
                                      A span constant is shown on the display
                                     It enters the weight mode as C.EnD. is
     -g
                                                      17
                  All manuals and user guides at all-guides.com
                                     18
                              All manuals and user guides at all-guides.com
7. Set-up
■ Set-up
- Overview
It means the setting works to ensure the indicator to be operated under the optimal conditions by
setting F-FUNCTION properly to the operating devices and surrounding environment.
                                                    19
                        All manuals and user guides at all-guides.com
※ Caution: Be aware that the value for the Function is stored when Key is pressed.
  ※ The value can be stored in the internal memory when you press
       Key after the value is changed to your desired set value.
                                            20
                               All manuals and user guides at all-guides.com
■ F-FUNCTION LIST
  Function
                                                    21
                            All manuals and user guides at all-guides.com
                                                             m
 ※F80~F89 can be entered after FUNCTION is designated.
                                            .co
     81         0       ZERO DISPLAY RANGE SETTING                             xxxxxx
     82        xxxxxx   ZERO MINUS VALUE SETTING                               xxxxxx
     83         0       MAXIMUM ANALOG OUTPUT VALUE SETTING                    xxxxxx
     85        xxxxxx   INPUT OF EQUIVALENT CIRCUIT VALUE                      xxxxxx
     89          -      CHECK FOR CALIBRATION SPAN CONSTANT                    x. x x x x x
                              es
      Input                IN1                         IN2
       Use                 Zero               Tare / Tare Removal
     Output               OUT1                        OUT2                OUT3
            uid
                                                   22
                              All manuals and user guides at all-guides.com
 * It is a function to stabilize the state when the range of variations for weight in the set time does not
 go over the A/D Count set range.
 ※ There are subtle vibrations in general when a weighing object is placed on the indicator.
 Hence, the state indicator “Stable” is turned on the display once the vibration is stabilized. “Motion
 Band” sets the time range to stabilize the vibration in this way.
                                                     23
                            All manuals and user guides at all-guides.com
 Ex) F04 is set to “3” when the maximum display division is set to 120.00Kg and a division to 0.05Kg;
                                                     If there is weight change within 250g within 0.75 SEC,
                           4.5digit                  the indicator automatically displays
                                                                                       t 0.
                           (0.25Kg)
※ This function prevents the indicator from any impact from the accumulated dusts or dirt where you might have
a lot of dusts or dirt in the environment during the use.
 * The automatic zeroing can be done (before the weighing is done again) if there is any remaining
 under the set value with no use of „ZERO‟ Key once this function is set.
 * Ex) If the maximum display weight is set to 120.00Kg, a division to 0.02Kg and F05 to 30 in the
 indicator, the display value becomes “0.00Kg” as the zeroing is immediately operated once
 (STEADY LAMP ON) is carried while the remaining amount exists up to ±(0.02~0.30Kg).
 * This function should be used after the adjustment for the set value depending on the environment
 (surrounding vibration).
 * Decrease the set value for the quicker response speed in the indicator
                                                        24
                           All manuals and user guides at all-guides.com
* As the indicator set 10% more for its allowable range, the maximum weight in reality becomes
110kg on the indicator if the maximum weight is set to 100kg.
Ex) If the maximum weight (CAPA) is set to 100kg and F08 to “T”, then “ZERO” Key can be
operated within 1kg (10%).
Ex) “Tare” Key can be operated within 50kg when the maximum weight (CAPA) is set to 100kg and
F09 to “2”.
                                               25
                        All manuals and user guides at all-guides.com
                  Division
                                                      Tare/Tare                           Hold/Hold
                         0           Zero                                   Set
                                                      Removal                             Removal
    F12                  1           Zero               Hold                Set         Hold Removal
                         2           Zero               Tare                Set         Tare Removal
                 ●       3           Zero               Start               Set             Stop
                                                                                          Hold/Hold
                         4           Zero             Start/Stop            Set
                                                                                          Removal
                                              26
                    All manuals and user guides at all-guides.com
                                              m
                                .co
                      es
          uid
   -g
all
                                       27
                 All manuals and user guides at all-guides.com
                                   28
                   All manuals and user guides at all-guides.com
                                      29
                All manuals and user guides at all-guides.com
                                  30
                     All manuals and user guides at all-guides.com
                 00
F22     10       ∫
                 99
                          Delay time to the relay operation after SP2 relay operations can be
                          set.
                          * Reference
                          00 : Relay output during the stable weight
                          01 : Relay output after 0.1 second
                          99 : Relay output after 9.9 seconds
                 01
F24     10       ∫
                                                                                  123
                 99
                          The delay time to the operation of relay for judgment after the
                          completion of weighing can be set.
                          * Reference
                          01 : Relay output after 0.1 second.
                          99 : Relay output after 9.9 seconds
                                          31
             All manuals and user guides at all-guides.com
                                          m
               00
 F25    10      ∫
               99                                                             123
                         .co
                      ON time for relay output for the completion of weighing can be set.
                      * Reference
                      01 : Relay ON for 0.1 second
                      99 : Relay ON for 9.9 seconds
              es
       uid
 -g
all
                                32
                       All manuals and user guides at all-guides.com
※ Communication Setting
                                            33
           All manuals and user guides at all-guides.com
                              34
                  All manuals and user guides at all-guides.com
                                           35
                       All manuals and user guides at all-guides.com
8. Interface
■ RS-232C Interface
        1                5
RS-232C Interface is sensitive to electric noises.
Therefore, it should be wired away from AC power cable or other electric wires.
The shielded cable should be used in any case.
Communication mode: It can be set from F-Function(F30~F35).
-    Signal Format
     ①Type : EIA-RS-232C
     ②Method : Half-Duplex, full-duplex, asynchronous method
     ③Baud-rate : Optional for 2400,4800,9600, 19200, 38400,
                        57600,76800,115200
     ④Data bit : 7 or 8(No, Parity)
     ⑤Stop bit : 1
     ⑥Parity bit : Optional for Even, Odd, No, Parity
     ⑦Code : ASCII
                                                                                  MSB
          1   +9V                    LSB0   1    2          3         4       5
                                                                                   6
0 -9V
      ⑧ Printer Format
Data format1(Basic Format)
              ,                       ,                                                    k           g    CR        LF
                                                36
                            All manuals and user guides at all-guides.com
                                                               m
       2D(H) : “-” MINUS
       2O(H) : “ ” SPACE
       2E(H) : “ ” Decimal point
 - UNIT (Common for data format 2,3)
                                         .co
      - Kg
       -t
       -g
Data format(2)
           ,                ,                 ,          Data(8)             K     g       CR      LF
                                es
Header1                               Lamp Status                    Blank
                  Header2       Equipment ID                                        unit
                uid
weight
37
PERSONAL
        COMPUTER
                            All manuals and user guides at all-guides.com
Indicator PC
8 TXD(+) RxD( +)
9 TXD(-) RxD(-)
6 RXD(+) TXD( +)
7 RXD(-) TXD(-)
                                               38
                           All manuals and user guides at all-guides.com
※ Reference
Although any of the signal level conversion methods as stated above is used, there are
many cases with problems occurring from the fact that the power supply (the grounding line
to be exact) for the transmission side is connected from a remote distance connection
through the communication line. Noises flown in through this grounding line might lead the
transmitting or receiving system to the instability. For such reasons, it is desirable to insulate
and separate the power supply for the system from the power supply for communication
line if it involves a long-distant transmission or it is used under the surrounding environment
                                               39
              All manuals and user guides at all-guides.com
▶ COMMAND MODE
PC→ (ASCII )
response (HEX )
PC→ (ASCII )
response (HEX)
PC→ (ASCII)
response ( HEX )
PC→                     (ASCII)
                                                                            Command to transmit
Indicator                                               ( HEX )
                                                                            the weight value
Indicator                               (ASCII)
                                                                            of “KEY Tare”
response                                                          ( HEX )
                                   40
                            All manuals and user guides at all-guides.com
PC→ (ASCII)
PC→                         (ASCII)
                                                                               Command to transmit
Indicator                                               ( HEX )
                                                                               the Lowest limit DATA
Indicator                              (ASCII)
response ( HEX )
PC→ (ASCII)
response ( HEX )
                                                 41
                   All manuals and user guides at all-guides.com
▶ COMMAND MODE
( WRITE COMMAND)
                                               m
 PC→                          (ASCII)
                               .co
 Indicator                                                                  the “Tare”
                               (ASCII)
 response                                                     ( HEX )
                    es
 PC→                           (ASCII)
response ( HEX )
PC→ (ASCII)
 response                                                     ( HEX )
all
                                        42
                         All manuals and user guides at all-guides.com
PC→ (ASCII)
response ( HEX )
PC→                             (ASCII)
                                                                          Command to change
Indicator                                                       ( HEX )
                                                                          the “Part Number”
Remark      STX(1) ID(2) Command(4) P/N(2) ETX(1)
                                                                          to the DATA that is now
Indicator                     (ASCII)                                     being transmitted
response                                                   ( HEX )
PC→ (ASCII)
response ( HEX )
                                               43
                All manuals and user guides at all-guides.com
                         (ASCII)
PC→ Indicator
                                                    ( HEX )      Command to set
response ( HEX )
                         (ASCII)
PC→ Indicator                                                    Command to “Start”
                                                    ( HEX )
                                                                 (F21 – 2)
Indicator                  (ASCII)
                                                                 (PACK MODE)
response                                               ( HEX )
                         (ASCII)
PC→ Indicator                                                    Comand to “stop”
                                                    ( HEX )
                                                                 (F21 – 2)
Indicator                  (ASCII)
                                                                 (PACK MODE)
response                                               ( HEX )
                                     44
                         All manuals and user guides at all-guides.com
PC→                                     (ASCII)
                                                                         Command to
Indicator                                                   ( HEX )
                                                                         change the Lowest
Remark      STX(1) ID(2) Command(4) LOW(6) ETX(1)
                                                                          limit (SP1)
Indicator                     (ASCII)
response ( HEX )
PC→                                     (ASCII)
                                                                         Command to
Indicator                                                      ( HEX )
                                                                         change the upper
Remark      STX(1) ID(2) Command(4) HIGH(6) ETX(1)
                                                                         limit (SP2)
Indicator                     (ASCII)
response ( HEX )
                                             45
                   All manuals and user guides at all-guides.com
- While Key Test Mode in TEST2 in relation with the analog setting;
                                       46
      All manuals and user guides at all-guides.com
MEMO
                                m
                  .co
       es
  uid
 -g
all
                         47
All manuals and user guides at all-guides.com
MEMO
                  48
All manuals and user guides at all-guides.com
MEMO
                   49
All manuals and user guides at all-guides.com
MEMO
                  50
All manuals and user guides at all-guides.com
51