Parameters: ECG, RESP, NIBP, SpO 2, TEMP
MEC-1200
      Patient Monitor
      Technical Specifications
      Safety                                                                                                    Monitor mode: ≥ 105 dB
      Meet requirement of IEC60601-1,CE Marking according to MDD93/42/EEC                                       Surgery mode: ≥ 105 dB
                                                                                           Calibration Signal:  1mV 5%
      Dimension and weight                                                                 Bandwidth:           Diagnostic mode: 1~15 Hz
      Dimension:         258(W) x 128(D) x 244(H) mm                                                            Monitor mode: 0.5~35 Hz
      Weight:            4.0 kg (max)                                                                           Surgery mode: 0.05~100 Hz
                                                                                           Alarm Range:         Adult: 15-300 BPM
      Power Supply                                                                                              Pediatric/Neonatal: 15-350 BPM
      Power Source:       AC mains power or Internal battery power                         S-T detection:       YES
      AC Mains Power Source                                                                                     Measurement range: -2.0mV~2.0mV
      Power Requirements: AC 100~240V (± 10%)                                                                   Alarm range: -2.0mV~2.0mV
      Line Frequency:     50/60 Hz (± 3Hz)                                                 Arrhythmia analysis: YES
      Consumption:        80W                                                              Pacemaking analysis: YES
      Battery Power(optional)                                                              Alarm:               YES, audible and visual alarm, alarm events recallable
      The maximum number of installed battery:  1
      Battery Capacity:   2.3amp-hour                                                      Respiration
      Operating time:     120 minutes under the normal use and full charge                 Method:            RA-LL impedance
                                                                                           Measurement Range: Adult: 0~120 BrPM
      Storage Environment                                                                                     Pediatric/Neonatal: 0~150BrPM
      Temperature:        -20℃ to +60℃                                                     Resolution:        1 BrPM
      Humidity:           10% to 95%, non-condensing                                       Accuracy:          ± 2% or± 2 BrPM, whichever is greater
                                                                                           Alarm Range:       Adult: 0~120 BrPM
      Operation Environment                                                                                   Pediatric/Neonatal: 0~150 BrPM
      Temperature:       0℃ to +40℃                                                        Apnea Alarm:       YES, and apnea delay: 10-40 seconds
      Humidity:          15% to 95%, non-condensing
                                                                                           NIBP
      Patient Range                                                                        Method:               Oscillometric
      Adult, pediatric and neonatal                                                        Operation Modes:      Manual/Automatic/Continuous
                                                                                           Measurement Unit:     mmHg/kPa selectable
      Performance Specification                                                            Measurement Types:    Systolic pressure, Diastolic pressure and Mean pressure
      Display:            8.4" color active matrix TFT                                     Measurement Range:    Adult:       10-270 mmHg
      Resolution:         800x600                                                                                Pediatric: 10-200 mmHg
      Trace:              4 waveforms                                                                            Neonatal: 10-135 mmHg
      Sweep Speed:        6.25, 12.5, 25 and 50mm/s                                        Over-pressure Protection:          YES
      Indicator:          Alarm indicator light                                            Resolution:           1 mmHg
                          Power indicator light                                            Accuracy:             Maximum Mean error: 5mmHg
                          Audio indicators for QRS beep and alarm sound                                          Maximum Standard deviation: 8mmHg
      Interface:          Networking                                                       Alarm:                Systolic, Diastolic and Mean
      Battery:            Rechargeable
      Trend time:         1~72 hours                                                       Temperature
      Alarm:              3-level audible and visual alarm                                 Scale:                0℃ and oF selectable
      Recorder(optional): Built-in, thermal array, 2 channels                              Measurement Range:    0~50 ℃ or 32-122 oF
      Record width:       48mm, (record paper 50mm)                                        Resolution:           0.1 ℃ or 0.1 oF
      Record speed:       25mm/s, 50mm/s                                                   Accuracy:             ± 0.1 C (0℃ to 50℃), exclusive of probe errors
                                                                                                                 ± 0.2 F (32 o F to 122 o F), exclusive of probe errors
      ECG                                                                                  Channel:              1 channel
      Lead Type:              3-lead                                                       Alarm Range:          0~50 ℃
      Input:                  3-lead: RA; LA; LL or R; L; F
      Lead Selection:         3-lead: I; II; III                                           SpO 2
      ECG Waveform:           1 channel                                                    SpO 2
      Gain Selection:         x0.25, x0.5; x1; x2; auto                                    Range:                0~100%
      Sweep Speed:            12.5mm/s, 25mm/s, 50mm/s                                     Accuracy:             70% ~ 100%: 2%
      Heart Rate Range:       Adult: 15~300BPM                                                                   0% ~ 69%: unspecified
                              Pediatric/Neonatal: 15-350BPM                                Pulse Rate
      Accuracy:               1BPM or ±1%, whichever is greater                            Range                 20-254 BPM
      Resolution:             1BPM                                                         Accuracy:             ±2BPM or 2%, whichever is greater
      Protection:             Withstand 4000VAC/50Hz voltage in isolation.
                              Anti-electrosurgical interference and defibrillation
      CMRR:                   Diagnostic mode: ≥ 90 dB
      Standard Configuration                 Options
      ECG, RESP, NIBP, TEMP, SpO 2           Thermal Recorder, Lead Acid Battery
P/N:ENG-MEC1200-210285-20070128