Rosemount 2120 Level Switch: Vibrating Fork
Rosemount 2120 Level Switch: Vibrating Fork
00809-0100-4030, Rev FA
                                       December 2016
Contents
               1Section 1: Introduction
                      1.1 Using this manual . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
                      1.2 Models covered . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
                      1.3 Level switch overview . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
                            1.3.1 Features list . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 1
                            1.3.2 Measurement principle . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
                            1.3.3 Short fork technology . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 3
                            1.3.4 Special features. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
                      1.4 Product recycling and disposal . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 4
               2Section 2: Installation
                      2.1 Safety messages. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 5
                      2.2 Considerations before installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                            2.2.1 Safety considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                            2.2.2 Environmental considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 6
                            2.2.3 Application considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 7
                            2.2.4 Installation considerations . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 8
                            2.2.5 Installation recommendations. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 10
                            2.2.6 Installation examples . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 13
                      2.3 Installation procedures . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
                            2.3.1 Mechanical. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 14
                            2.3.2 Correct fork alignment . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 15
                            2.3.3 Tightening the threaded level switch . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
                            2.3.4 Insulation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 17
                      2.4 Setting the mode switch and switching time delay. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 18
                      2.5 LED indication. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 19
                      2.6 Electrical installation . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
                            2.6.1 Direct load switching electronics cassette. . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 20
                            2.6.2 PNP/PLC electronics cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 22
                            2.6.3 Relay output electronics cassette (standard version) . . . . . . . . . . . . . . . . . . . . . . . . . . . . 23
                            2.6.4 Relay output electronics cassette (12 Vdc nominal version) . . . . . . . . . . . . . . . . . . . . . . 24
                            2.6.5 NAMUR electronics cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 25
                            2.6.6 8/16 mA electronics cassette . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . . 26
Contents                                                                                                                                                                          i
Contents                                                                                                                                      Reference Manual
December 2016                                                                                                                           00809-0100-4030, Rev FA
ii                                                                                                                                                                      Contents
Reference Manual                                                                                            Title Page
00809-0100-4030, Rev FA                                                                                 December 2016
                 Read this manual before working with the product. For personal and system safety, and for optimum
                 product performance, make sure you thoroughly understand the contents before installing, using, or
                 maintaining this product.
                 For technical assistance, contacts are listed below:
                 Customer Central
                 Technical support, quoting, and order-related questions.
                 Americas 1 800 999 9307
                 Europe +41 (0) 41 768 6111
                 Middle east +971 4 811 8100
                 Asia +65 6777 8211
                 North American Response Center
                 Equipment service needs.
                 1-800-654-7768 (24 hours — includes Canada)
                 Outside of these areas, contact your local Emerson™ representative.
                 The products described in this document are NOT designed for nuclear-qualified applications.
                 Using non-nuclear qualified products in applications that require nuclear-qualified hardware or
                 products may cause inaccurate readings.
                 For information on Rosemount nuclear-qualified products, contact your local Emerson sales
                 representative.
Title Page                                                                                                            v
Title Page                                                                                          Reference Manual
December 2016                                                                                    00809-0100-4030, Rev FA
                Failure to follow these installation guidelines could result in death or serious injury.
                   Make sure only qualified personnel perform the installation.
                   Use the Rosemount 2120 Level Switch (“level switch”) only as specified in this manual.
                    Failure to do so may impair the protection provided by the equipment.
                   The weight of the level switch with a heavy flange and extended fork length may exceed
                    37 lb. (18 kg). A risk assessment is required to be done before carrying, lifting, and installing
                    the level switch.
                Explosions could result in death or serious injury.
                   Do not remove the level switch cover in explosive atmospheres when the circuit is alive.
                   The level switch cover must be fully engaged and locked to meet explosion-proof
                    requirements.
                   Verify the operating environment of the level switch is consistent with the appropriate
                    hazardous locations certifications.
                   Review Appendix B: Product Certifications for special conditions and safety instructions
                    associated with a hazardous location installation.
vi                                                                                                                     Title Page
Reference Manual                                                                                            Introduction
00809-0100-4030, Rev FA                                                                                     December 2016
Section 1 Introduction
                This combination of features makes the Rosemount 2120 Level Switch an ideal choice for a wide variety
                of challenging applications in the chemical, power generation, and oil and gas industries.
Introduction                                                                                                                 1
Introduction                                                                                                                                                                                                                                                                                           Reference Manual
December 2016                                                                                                                                                                                                                                                                                     00809-0100-4030, Rev FA
                                                                                                                                          Switch
                                                                                                                                                 www.rosemoun
                                                                                                                                                    Level
                                                                                                                                                            Fork
                                                                                                                                                                   Vibrating
                                                                                                                                                                               2120
                                                                                                                                                                                                                        Overfill protection
                                                                                                                                                                                                                        Spillage caused by overfilling can be hazardous to people and the environment, resulting in lost product
                                                                                                                                                                                                                        and potentially high clean up costs.
                                                   Fork
                                                          Level
                                                                  Switch
                                        t.com
                                                                                                                                                                                                                        temperature range of –40 to 302 °F (–40 to 150 °C) and operating pressures of up to 1450 psig (100
                                                                                                                                                                                                                        barg), making it perfect for use as a high or low level alarm. It is common practice to have an independent
                                 2120
                                                                                                                                                                                                                        high level alarm switch as a backup to an installed level device in case of primary failure.
                                     Vibrating
                  www.rosemoun
                                                   Fork
                                                          Level
                                                                  Switch
                                        t.com
                                                                                                                                                                                                                        standard user selectable time delay, from 0.3 to 30 seconds, virtually eliminates the risk of false switching
                                                                                                                                                                                                                        from splashing.
                                                   2120
                                                      Vibrating
                                    www.rosemoun
                                                                      Fork
                                                                           Level
                                                                                   Switch
                                                          t.com
                                                                                                                                                                                                                        Wireless applications
                                                                                                                                                                                                                        The advent of wireless communications allows process plant managers to save up to 90% on installation
                                                                                                                                                                                                                        cost compared with wired technologies. More data can be collected at central locations than has ever
                                                                                                                                                                                                                        been possible before. The Rosemount 2120 can be used with a Rosemount 702 Wireless Discrete
                                                                                                                                                                                                                        Transmitter to enable these benefits for your applications.
                                                                                                                                                                                                                        Hygienic applications
                                                                                                    www.rosemoun
                                                                                                                                                                                                                        With the highly polished forks option providing a surface finish (Ra) better than 0.4 μm, the Rosemount
                                                                                                                                 2120
                                                                                            t.com                    Vibrating
                                                                                                              Fork
                                                                                                      Level
                                                                                             Switch
                                                                                                                                                                                                                        2120 meets the most stringent hygienic requirements used in food and beverage, and pharmaceutical
                                                                                                                                                                                                                        applications. The Rosemount 2120 is robust enough to easily withstand CIP (Clean-In- Place) and SIP
                                                                                                                                                                                                                        (Steam-In-Place) cleaning routines.
2                                                                                                                                                                                                                                                                                                                       Introduction
Reference Manual                                                                                                             Introduction
00809-0100-4030, Rev FA                                                                                                      December 2016
J C
                 I
                                                                 E
H F
                A. Visible heartbeat LED                                  G. Wetted material in 316/316L SST, solid Alloy C and Alloy C-276,
                B. Mode switch, adjustable time delay                       or ECTFE-coated 316/316L SST
                C. Housings in glass-filled nylon, aluminum, or 316 SST   H. Short fork length or extensions up to 157.5-in. (4 m)
                D. Magnetic test point                                    I. Two cable/ conduit entries
                E. Threaded, flanged, or Tri Clamp connections            J. Direct Load, Relay DPCO, PLC/PNP, NAMUR,
                F. ‘Fast drip’ fork design                                  or 8/16 mA electronics
Introduction                                                                                                                                   3
Introduction                                                                                          Reference Manual
December 2016                                                                                     00809-0100-4030, Rev FA
                Fork design
                The “fast drip” fork design draws liquid away from the fork tips when mounted horizontally, and together
                with a short switching delay, allows the Rosemount 2120 Level Switch to react quickly and with greater
                sensitivity to density variations.
                Electrical hookup
                The terminal blocks extend above the housing and give easy terminal access. Electrical protections
                (see page 37) make electrical hook-up safe and easy.
4                                                                                                                   Introduction
Reference Manual                                                                                                                                             Installation
00809-0100-4030, Rev FA                                                                                                                                    December 2016
Section 2 Installation
                   Failure to follow these installation guidelines could result in death or serious injury.
                       Make sure only qualified personnel perform the installation.
                       Use the Rosemount 2120 Level Switch (“level switch”) only as specified in this manual.
                        Failure to do so may impair the protection provided by the equipment.
                       The weight of the level switch with a heavy flange and extended fork length may exceed
                        37 lb. (18 kg). A risk assessment is required to be done before carrying, lifting, and installing the
                        level switch.
                   Explosions could result in death or serious injury.
                       Do not remove the level switch cover in explosive atmospheres when the circuit is alive.
                       The level switch cover must be fully engaged and locked to meet explosion-proof requirements.
                       Verify the operating environment of the level switch is consistent with the appropriate hazardous
                        locations certifications.
                       Review Appendix B: Product Certifications for special conditions and safety instructions
                        associated with a hazardous location installation.
                   Electrical shock can result in death or serious injury.
                       If the level switch is installed in a high-voltage environment and a fault or installation error occurs,
                        high voltage may be present on the sensor leads and terminals.
                       Use extreme caution when making contact with the leads and terminals.
                       Make sure the main power to the level switch is off and the lines to any other external power
                        source are disconnected or not powered while wiring the level switch.
                       Ensure the wiring is suitable for the electrical current and the insulation is suitable for the voltage,
                        temperature, and environment.
Installation                                                                                                                                                                     5
Installation                                                                                      Reference Manual
December 2016                                                                                  00809-0100-4030, Rev FA
OK OK
6                                                                                                                Installation
Reference Manual                                                                                                  Installation
00809-0100-4030, Rev FA                                                                                       December 2016
                    Important
                There must be no risk of ‘bridging’ the forks (see Figure 2-2). Examples of media that can create
                ‘bridging’ of forks are dense paper slurries and bitumen.
OK
                Application guidelines
                   Ensure the process is operating within the level switch operating temperature and pressure ranges
                    (see “Specifications” on page 35).
                   Ensure the liquid viscosity is within the recommended range (see “Specifications” on page 35).
                   Check that the liquid density is higher than 37.5 lb/ft3 (600 kg/m3) (see “Specifications” on page 35).
                   Liquid density affects the switchpoint e.g. dry-to-wet (see Figure 2-3 on page 8).
                   Check for risk of coating build-up on the forks.
                    Avoid situations where drying and coating products may create excessive build-up. Problems may
                    occur if the product coats and then dries, causing caking.
                   Check the solids content in the liquid.
                    As a guideline, the maximum solid particle diameter in the liquid is 0.2 in. (5 mm).
                    Extra consideration is needed when dealing with particles bigger than 0.2 in. (5 mm).
                    Consult the factory for advice.
                   In almost all cases, the level switch is insensitive to foams (i.e. does not see the foam).
                    However in rare occasions, some very dense foams may be seen as liquid; known examples of this are
                    found in ice-cream and orange juice manufacturing.
Installation                                                                                                                  7
Installation                                                                                        Reference Manual
December 2016                                                                                   00809-0100-4030, Rev FA
                                                       0.1 in.
                                                     (2.5 mm)
SP HY
                            0.5 in.
                           (13 mm)
                                0.5 in.
                               (13 mm)
SP
                Note
                When mounted vertically, a low density media has a switching point closer to the process connection.
                A high density media has a switching point closer to fork tip.
                Device identification
                To identify the Rosemount 2120 Level Switch version, see the labels on the housing and on the
                electronics cassette inside the housing. See Appendix B: Product Certifications for approval information.
8                                                                                                                 Installation
Reference Manual                                                                                           Installation
00809-0100-4030, Rev FA                                                                                  December 2016
                Hygienic installation
                See “Hygienic installations” on page 62 for the hygienic approvals and compliance requirements.
                These requirements are also in the Rosemount 2120 Quick Start Guide (seeEmerson.com/Rosemount
                for other language versions).
Installation                                                                                                             9
Installation                                                                                       Reference Manual
December 2016                                                                                   00809-0100-4030, Rev FA
OK OK
10                                                                                                                Installation
Reference Manual                                                                Installation
00809-0100-4030, Rev FA                                                        December 2016
                      Maximum
                        3.28 ft.
                        (1.0 m)
3.28 ft.
                                   OK                                     OK
                                                   (1.0 m)
                                                   3.28 ft.
                                                   (1.0 m)
                                                              Maximum
                                                               3.28 ft.
                                                               (1.0 m)
                                        OK
                                        3.28 ft.       3.28 ft.
                                        (1.0 m)        (1.0 m)
OK
Installation                                                                              11
Installation                                                                               Reference Manual
December 2016                                                                            00809-0100-4030, Rev FA
                     Maximum
                        1.3 ft.
                       (0.4 m)
                                                              2.3 ft.
                                                             (0.7 m)
                                  OK                          2.3 ft.
                                                             (0.7 m)
                                                                                    OK
                                                              2.3 ft.
                                                             (0.7 m)
                                                             0.65 ft.
                                                             (0.2 m)
                                                                        Maximum
                                                                          1.3 ft.
                                                                         (0.4 m)
                                               OK
                                  0.65 ft.
                                  (0.2 m)
                                              2.3 ft.    2.3 ft.    2.3 ft.
                                             (0.7 m)
                OK
                                                        (0.7 m)    (0.7 m)
12                                                                                                      Installation
Reference Manual                                                                   Installation
00809-0100-4030, Rev FA                                                           December 2016
A B
A B
                A. Dry
                B. Wet
B B A
                A. Dry
                B. Wet
B B
A A
                A. Dry
                B. Wet
Installation                                                                                 13
Installation                                                                                                              Reference Manual
December 2016                                                                                                      00809-0100-4030, Rev FA
                      A
                                                                           C
B D
                A. PTFE
                B. NPT or BSPT (R) thread
                C. Gasket
                D. BSPP (G) thread
                E. Tri Clamp
                F. The Tri Clamp seal is supplied as an as accessory kit (see “Spare parts and accessories” on page 50)
14                                                                                                                                 Installation
Reference Manual                                                                                          Installation
00809-0100-4030, Rev FA                                                                                  December 2016
                                          B
                                                                                      C
                Pipe installation
                Figure 2-13. Pipe Installation
OK
                Note
                The glass-filled nylon housing of a Rosemount 2120 can be rotated to assist with cabling, but the metal
                housing cannot be rotated.
Installation                                                                                                         15
Installation                                                                                    Reference Manual
December 2016                                                                                00809-0100-4030, Rev FA
                Tank installation
                Figure 2-14. Tank Installation
                                                   OK                OK
                                     OK
                                                  OK                 OK
                                    OK
                Note
                The glass-filled nylon housing of a Rosemount 2120 can be rotated to assist with cabling, but the metal
                housing cannot be rotated.
16                                                                                                            Installation
Reference Manual                                                       Installation
00809-0100-4030, Rev FA                                               December 2016
OK
2.3.4           Insulation
                Figure 2-16. Insulation
A A
Installation                                                                     17
Installation                                                                                              Reference Manual
December 2016                                                                                           00809-0100-4030, Rev FA
                Note
                There is a five second delay when changing mode or time delay. The small cut-out in the rotating switch
                indicates time delay and mode.
                Recommended installation for high level is “Dry on” (Figure 2-18) and for low level it is “Wet on”
                (Figure 2-19). Do not install in the normally ‘off’ state.
A B
                A. LED
                B. Mode Switch/Time Delay
Figure 2-18. “Dry On” Mode, One Second Time Delay (Typical for High Level Applications)
Dry On Wet On
                                                         0.3                           0.3
                                                         1                             1
                                                         3                             3
                                                         10                            10
                                                         30                            30
                                                             Seconds Delay
Figure 2-19. “Wet On” Mode, One Second Time Delay (Typical for Low Level Applications)
Dry On Wet On
                         0.3                         0.3
                         1                           1
                 A       3                           3
                         10                          10
                         30                          30
                             Seconds Delay
18                                                                                                                     Installation
Reference Manual                                                                                        Installation
00809-0100-4030, Rev FA                                                                               December 2016
1 every 4 seconds Load fault; load current too high; load short circuit
Installation                                                                                                      19
Installation                                                                                                        Reference Manual
December 2016                                                                                                     00809-0100-4030, Rev FA
                Note
                When replacing a cassette, it is important to re-calibrate.
                Refer to “Replacement and calibration of electronic cassettes” on page 31.
                Note
                A DPST (Double Pole, Single Throw) on/off switch must be fitted for safe disconnection of the power
                supply. Fit the DPST switch as near as possible to the Rosemount 2120. Keep the DPST switch free of
                obstructions. Label the DPST switch to indicate it is the supply disconnection device for the
                Rosemount 2120.
20                                                                                                                               Installation
Reference Manual                                                                                                                     Installation
00809-0100-4030, Rev FA                                                                                                             December 2016
                                 U                                                             U
        = Load off                                                <4mA                                                        <4mA
                                12V                                                            12V                             IL
                                                                   IL                     IL
                           IL                                            Fuse                        Fuse                             Fuse
                                      Fuse
                                      2A(T)                              2A(T)                       2A(T)                            2A(T)
        =Load on
                        DPST                                    DPST                   DPST                                 DPST
                            N          L                           N      L                N          L                        N       L
                            0V        +V                           0V    +V                0V        +V                        0V     +V
LED on continuously LED flashes every second LED on continuously LED flashes every second
Installation                                                                                                                                    21
Installation                                                                                                                                                                                          Reference Manual
December 2016                                                                                                                                                                                   00809-0100-4030, Rev FA
                                    U                                                                                                                      U
                                                                                                           <100 A                                                                                                <100 A
                                    <3V                                                                                                                     <3V
                                                     IL                                                         IL                                                IL                                                   IL
                                    U                                                                                                                      U
                                                                                                                                                                                                                       R
                                    <3V                                                                         R                                           <3V
                                                                                                                                                                                                                            IL
                                                                                                                     IL                                             R
                                                     R                                                                                          Fuse                                                  Fuse        < 100 A
                        Fuse                                                                Fuse           < 100 A                                                     IL                            1A(T)
                                                          IL                                1A(T)                                               1A(T)
                        1A(T)
                                                                                                                                                        +                                                     +
                                +                                                                      +
LED on continuously LED flashes every second LED on continuously LED flashes every second
22                                                                                                                                                                                                                                   Installation
Reference Manual                                                                                                                                                       Installation
00809-0100-4030, Rev FA                                                                                                                                              December 2016
                                                                                                                            RELAY
                                              3             3
                                             10
                                                            10 Dry
                                             30
                                                            30 Wet
                                                                      Wet On
                        1        2    3                                            4      5     6      7    8     9
                                                                                   NC     C     NO    NC    C    NO
                      PE                  Fuse 0.5 (T)                              Resistive Load         Inductive Load
                 (Ground)
                                                                                    cos φ = 1 ;            cos φ = 0.4 ;
                                                                                    L/R = 0 ms             L/R = 7 ms
                                             DPST                                                          IMAX = 3.5 A
                                                                                   IMAX = 5 A
                                                                                   ac:                     ac:
                                                  U = 20...264 V ~ (ac)
                              N           Live                                     UMAX = 250 V            UMAX = 250 V
                                                             (50/60 Hz)
                                                  I < 6 mA                         PMAX = 1250 VA          PMAX = 875 VA
                                                                                   dc:                     dc:
                                                  U = 20...60 V         (dc)       UMAX = 30 V             UMAX = 30 V
                              0V          +V      I < 6 mA                          PMAX = 240 W           PMAX = 170 W
                Note
                A Double Pole, Single Throw on/off switch must be fitted for safe disconnection of the power supply.
                Fit the DPST switch as near as possible to the Rosemount 2120 Level Switch. Keep the DPST switch free of
                obstructions. Label the DPST switch to indicate it is the supply disconnection device for the level switch.
NC C NO NC C NO NC C NO NC C NO
LED on continuously LED flashes every second LED on continuously LED flashes every second
Installation                                                                                                                                                                        23
Installation                                                                                                                                              Reference Manual
December 2016                                                                                                                                      00809-0100-4030, Rev FA
                                                                                                                      RELAY
                                       1              1
                                       3              3
                                      10
                                                      10 Dry
                                      30              30 Wet
                                                                Wet On
                       1   2    3                                            4      5     6      7     8     9
                                                                             NC     C    NO     NC     C    NO
                     PE             Fuse 0.5 (T)                             Resistive Load          Inductive Load
                (Ground)
                                                                             cos φ = 1 ;             cos φ = 0.4 ;
                                                                             L/R = 0 ms              L/R = 7 ms
                                      DPST                                   IMAX = 2 A                IMAX = 1 A
                                                                             ac:                     ac:
                                                                             UMAX = 125 V            UMAX = 125 V
                                                                             PMAX = 62.5 VA          PMAX = 37.5 VA
                                            U = 9...30 V         (dc)
                                                                             dc:                     dc:
                                            I < 4 mA
                                                                             UMAX = 30 V             UMAX = 30 V
                           0V       +V                                       PMAX = 60 W             PMAX = 30 W
                Note
                A Double Pole, Single Throw on/off switch must be fitted for safe disconnection of the power supply. Fit
                the DPST switch as near as possible to the Rosemount 2120. Keep the DPST switch free of obstructions.
                Label the DPST switch to indicate it is the supply disconnection device for the Rosemount 2120.
NC C NO NC C NO NC C NO NC C NO
LED on continuously LED flashes every second LED on continuously LED flashes every second
24                                                                                                                                                                     Installation
Reference Manual                                                                                                                                           Installation
00809-0100-4030, Rev FA                                                                                                                                   December 2016
                                                                        OPERATION MODE
                                                                Dry On Mode        Dry On   Wet On   EN 50227 / NAMUR
                                        -   +                     Dry
                                                                  Wet             0.3          0.3
                           8V                                                     1            1
                           dc
                                                                                  3            3
                                                                  Dry             10           10
                                                                  Wet             30           30
                                        1   2                   Wet On Mode       Seconds Delay
                                    -           +
                                                        A certified intrinsically safe
                  Ex                                    isolating amplifier to IEC 60947-5-6
                Note
                This electronics cassette is suitable for Intrinsically Safe applications and requires a certified isolating
                barrier. See “Product Certifications” on page 51 for Intrinsically Safe approvals.
                This electronics cassette is also suitable for non-hazardous (safe) area applications. It can only be
                interchanged with the 8/16 mA cassette.
                Do not exceed 8 Vdc.
- + - + - + - +
LED on continuously LED flashes every second LED on continuously LED flashes every second
Installation                                                                                                                                                         25
Installation                                                                                                                                        Reference Manual
December 2016                                                                                                                                00809-0100-4030, Rev FA
OPERATION MODE
                                                    I ON    = 15 ... 17 mA
                              PE                                                                U = 24 Vdc Nominal
                           (Ground)
                  Ex                                I OFF = 7.5 ... 8.5 mA
                                  -                 +
                  Ex                                       A certified intrinsically safe barrier
                                                           must be used to meet IS requirements
                                  -                 +
                           Drives 4-20 mA Analog Input
                Note
                This electronics cassette is suitable for Intrinsically Safe applications and requires a certified isolating
                barrier. See “Product Certifications” on page 51 for Intrinsically Safe approvals.
                This electronics cassette is also suitable for non-hazardous (safe) area applications.
                In this case, U = 11 - 36 V       (dc) and it can only be interchanged with a NAMUR cassette.
+ + + +
LED on continuously LED flashes every second LED on continuously LED flashes every second
26                                                                                                                                                               Installation
Reference Manual                                                                                                                  Service and Troubleshooting
00809-0100-4030, Rev FA                                                                                                                                           December 2016
                         Failure to follow these installation guidelines could result in death or serious injury.
                             Make sure only qualified personnel perform the installation.
                             Use the Rosemount 2120 Level Switch (“level switch”) only as specified in this manual.
                              Failure to do so may impair the protection provided by the equipment.
                             The weight of the level switch with a heavy flange and extended fork length may exceed
                              37 lb. (18 kg). A risk assessment is required to be done before carrying, lifting, and installing the
                              level switch.
                         Explosions could result in death or serious injury.
                             Do not remove the level switch cover in explosive atmospheres when the circuit is alive.
                             The level switch cover must be fully engaged and locked to meet explosion-proof requirements.
                             Verify the operating environment of the level switch is consistent with the appropriate hazardous
                              locations certifications.
                             Review Appendix B: Product Certifications for special conditions and safety instructions
                              associated with a hazardous location installation.
                         Process leaks could result in death or serious injury.
                             Install and tighten process connection before applying pressure.
                             Do not attempt to loosen the process connection while the level switch is in service.
                             Make sure that the level switch is handled carefully. If the process seal is damaged, gas might
                              escape from the tank.
                               41001029
                                 *41001029*
                                                    MAX WORKING PRESSURE AT 20°C: 0.25 bar
                                                      MAX PROCESS TEMPERATURE: 150°C
                                                         MADE IN UNITED KINGDOM
                                                                                              41001029
                                                                                                *41001029*
                                                                                                                      MAX WORKING PRESSURE AT 20°C: 0.25 bar
                                                                                                                        MAX PROCESS TEMPERATURE: 150°C
                                                                                                                           MADE IN UNITED KINGDOM
                                                            www.rosemount.com                                                 www.rosemount.com
                                                                                                                                                               MAGNETIC
                                                                                                                                                                 TEST
                              No magnet                                                          Magnet
                                                                                                                                                                POINT
                                                                                                                                                                          S
                                                                                                                                                                              N
                Output off                                                                                                                            Output on
                                                   TP                                                        TP
                                                                                                                         S                                     TP
                                                                                                                                 N
                                                                                                                                                                     S
                                                                                                                                                                          N
                                 No magnet                                                           Magnet
3.3                   Inspection
                         Visually examine the Rosemount 2120 Level Switch for damage. If it is damaged, do not use.
                         Ensure the housing cover, cable glands, and blanking plugs are fitted securely.
                         Fit blanking a plug where required.
                         Ensure the LED flash rate is once every second or continually on.
                          If anything else is demonstrated, see “LED indication” on page 19.
OK
3.4                   Maintenance
                      Figure 3-4. Maintenance
OK
                      Note
                      Only use a soft brush for cleaning.
3.6             Troubleshooting
                If there is a malfunction, see Table 3-1 for information on possible causes.
                Table 3-1. Troubleshooting Chart
                                    • No LED; no power                  • Check the power supply; (check the load on direct
                                                                           load switching electronics).
                 Does not           • Fork is damaged                   • Replace the Rosemount 2120 Level Switch.
                 switch
                                    • Thick encrustation                • Clean the fork with care.
                                       on the fork
                 Incorrect          • “Dry = On” or “Wet = On”          • Set the correct mode on the electronics cassette.
                 switching             set incorrectly
                Individuals who handle products exposed to a hazardous substance can avoid injury if they are
                informed of, and understand, the hazard. If the product being returned was exposed to a hazardous
                substance as defined by OSHA, a copy of the required Material Safety Data Sheet (MSDS) for each
                hazardous substance identified must be included with the returned goods.
                      Note
                      Intrinsically Safe approved cassettes can only be replaced with the same type of IS cassette.
                      Non-I.S. cassette types can be interchanged with other non-I.S. cassettes, but a new label must be fitted
                      and the original part number transferred to the new label. Before starting the replacement and
                      calibration procedure, ensure that any controlled process will not be adversely affected.
                 0.3            0.3
                 1              1
                 3              3
                 10             10
                 30             30
                     Seconds Delay
                11.After a two second delay, the LED will go out (stop flashing).
                12.Within three seconds, rotate the mode switch two steps counter-clockwise.
                13.After a two second delay, the LED should flash twice per second.
                14.If the LED is flashing twice per second, the calibration has occurred correctly.
                   Remove the magnet from the test point. After a one second delay, the unit will return to normal
                   operation. Proceed to step 17.
                      15.If the LED is flashing once per second or it is on continuously, the calibration has failed. Remove the
                         magnet from the test-point, wait ten seconds, and repeat from step 2.
                      16.If the LED stays off after the two second delay of step 13, the sensor is not working correctly.
                         Check that the forks are clean and dry. Also, verify there is nothing jamming or touching the sensor.
                         If no fault is found with the sensor, the entire unit should be returned for repair
                         (see “Product recycling and disposal” on page 4).
                      17.Set the mode switch to the original setting noted in Figure 3-6 and wait five seconds.
                      18.Replace the cover and check that the system works.
A.1 Specifications                                                                         The maximum extended length is 157.5 in. (4000 mm) except for
                                                                                           ECTFE copolymer coating and mechanically-polished process
A.1.1 General                                                                              connection options which have a maximum length of 59.1 in.
                                                                                           (1500 mm) and 39.4 in. (1000 mm) respectively.
Product
                                                                                           Material selection
Rosemount 2120 Level Switch
                                                                                           Emerson provides a variety of Rosemount product with various
Measuring principle
                                                                                           product options and configurations including materials of
Vibrating fork technology                                                                  construction that can be expected to perform well in a wide range
                                                                                           of applications. The Rosemount product information presented is
Applications
                                                                                           intended as a guide for the purchaser to make an appropriate
Most liquids including coating liquids, aerated liquids, and slurries.                     selection for the application. It is the purchaser’s sole
                                                                                           responsibility to make a careful analysis of all process parameters
A.1.2 Mechanical
                                                                                           (such as all chemical components, temperature, pressure, flow
Housing / Enclosure                                                                        rate, abrasives, contaminants, etc.), when specifying product,
Table A-1. Housing/Enclosure Specifications                                                materials, options and components for the particular application.
Extended lengths
                                                                                           A.1.3 Functional
                                                                                                                                           32 (0)
Figure A-1. Process Pressure
                                                                                                                                           -40 (-40)
Up to 10000 cP (centiPoise).
Table A-3. Maximum Flange Pressure Rating                                                                    Solids content and coating
 Standard                                              Class/Rating                 SST flanges              Maximum recommended diameter of solid particles in the liquid is
 ASME B16.5                                           Class 150                     275 psig(1)              0.2 in. (5 mm).
 ASME B16.5                                           Class 300                     720 psig(1)              For a coating product, avoid bridging of forks.
 ASME B16.5                                           Class 600                     1440 psig(1)
                                                                                                             Switching delay
 EN1092-1                                             PN 10/16                      16 barg(2)
                                                                                                             User selectable 0.3, 1, 3, 10, 30 seconds delay for dry-to-wet and
 EN1092-1                                             PN 25/40                      40 barg(2)
                                                                                                             wet-to-dry switching.
 EN1092-1                                             PN 63                         63 barg(2)
                                                                                                             Clean-In-Place (CIP) cleaning
 EN1092-1                                             PN 100                        100 barg(2)
 JIS B2220                                            10K                           14 barg(3)               Withstands cleaning routines up to 160 °F (71 °C).
 JIS B2220                                            20K                           34 barg(3)               Steam-In-Place (SIP) cleaning
          1.                       At 100 °F (38 °C), the rating decreases with an increasing process
                                   temperature.                                                              Withstands cleaning routines up to 275 °F (135 °C).
          2.                       At 122 °F (50 °C), the rating decreases with an increasing process
                                   temperature.
                                                                                                             NACE
          3.                       At 248 °F (120 °C), the rating decreases with an increasing process
                                   temperature.
                                                                                                             NACE compliance to MR0175 / ISO 15156 or MR0103, depending
                                                                                                             on the option code selected for the model number.
Minimum and maximum operating temperatures                                                                   Safety integrity level
See Figure A-2 for operating temperatures.                                                                   The Rosemount 2120 FMEDA is certified for SIL2, and is SIL3
Clamp glands 02120-2000-0001 and 02120-2000-0002 (page 46)                                                   capable, for all electronics except the Direct Load option.
limit the maximum temperature to 257 °F (125 °C).                                                            Visit the Rosemount 2120 web page for additional information.
Hysteresis (water)                                                       The Rosemount 2120 must always be grounded either through the
                                                                         terminals or using the external ground connection provided.
±0.039-in. (±1 mm) nominal.
                                                                         Conduit plugs/cable gland
Switching point (water)
                                                                         Metal housing:
0.5 in. (13 mm) from tip (vertical) / from edge (horizontal) of fork
                                                                         Conduit entries for explosion-proof areas are shipped with one Exd
(this will vary with different liquid densities).
                                                                         plug (loose in bag) and two dust caps fitted. Use suitably rated
A.1.5 Electrical                                                         cable glands. Unused conduit entries must be sealed with a
                                                                         suitably rated blanking plug.
Switching mode
                                                                         Glass-filled nylon housing with direct load, PNP/PLC and IS
User selectable switching mode (“Dry = on” or “Wet = on”).               electronics are shipped with one PA66(1) cable gland and one
Protection                                                               blanking plug.
Polarity insensitive                                                     Glass-filled nylon housing with relay electronics are shipped with
– on Relay (except 12 Vdc version) and Direct Load electronics           two PA66(1) cable glands.
The LED flashes when the output is ‘off’ and is constantly lit when it       NAMUR (light blue) cassette
is ‘on’. The LED gives a constant indication that the Rosemount
2120 is functioning correctly (different flash rates are used to                                                                  OPERATION MODE
                                                                                                                              Dry On Mode     Dry On    Wet On    EN 50227 / NAMUR
indicate a product malfunction) and gives a local indication of the                                 -        +                  Dry
                                                                                                                                Wet          0.3           0.3
                                                                                      8V
process state.                                                                        dc
                                                                                                                                             1
                                                                                                                                             3
                                                                                                                                                           1
                                                                                                                                                           3
                                                                                                                                Dry          10            10
A magnetic test point is located on the side of the housing,                                                     ION = 2.2 ... 2.5 mA
                                                                                                                 IOFF = 0.8 ... 1.0 mA
allowing a functional test of the Rosemount 2120 and a system                 Ex                                 IFAULT < 1.0 mA
    8/16 mA (dark blue) cassette                                                                                          DPCO dual relay cassette (12 Vdc nominal version)
                                                              OPERATION MODE
                                                                                                                                     12 VDC NOM.                           OPERATION MODE                       Isolate Supply Before Removing
                                           +           Dry On Mode      Dry On        Wet On         8/16 mA                                                         Seconds Delay
                                                         Dry
                                                                                                                                                              +                       Dry On
                                                         Wet           0.3               0.3                                                                        Dry On Wet On Dry                      NC      C        NO    NC     C    NO
                                                                       1                 1                                                                          0.3           0.3 Wet
                                                                                                                                                                                                                                                      RELAY
                                                                       3                 3                                                                           1            1
                                                         Wet           10                10                                                                          3            3
                                                         Dry           30                30
                                                                                                                                                                    10
                                                       Wet On Mode     Seconds Delay                                                                                             10 Dry
                                                                                                                                                                    30           30 Wet
                         1       2         3
                                                                                                                                                                                            Wet On
                                                                                                                                     1      2                 3                                             4      5         6     7     8     9
                                               I ON   = 15 ... 17 mA                           U = 24 Vdc Nominal
                                               I OFF = 7.5 ... 8.5 mA                                                                                                                                      NC      C       NO     NC     C    NO
                     PE
                  (Ground)                     I FAULT < 3.7 mA                                                                     PE                            Fuse 0.5 (T)
      Ex                                                                                                                       (Ground)
                                                                                                                                                                                                            Resistive Load             Inductive Load
                                                                                                                                                                                                            cos φ = 1 ;                cos φ = 0.4 ;
                                                                                                                                                                                                            L/R = 0 ms                 L/R = 7 ms
                             -                 +                                                                                                                    DPST                                    IMAX = 2 A                   IMAX = 1 A
      Ex                                              A certified intrinsically safe barrier                                                                                                                ac:                        ac:
                                                      must be used to meet IS requirements                                                                                                                  UMAX = 125 V               UMAX = 125 V
                                                                                                                                                                                                            PMAX = 62.5 VA             PMAX = 37.5 VA
                                                                                                                                                                           U = 9...30 V      (dc)
                             -                 +                                                                                                                           I < 4 mA                         dc:                        dc:
                   Drives 4-20 mA Analog Input                                                                                                                                                              UMAX = 30 V                UMAX = 30 V
                                                                                                                                            0V                    +V                                        PMAX = 60 W                PMAX = 30 W
    DPCO dual relay cassette (standard version)                                                                           Solid state PNP output for direct interface to a PLC
                  Warning                  OPERATION MODE                      Isolate Supply Before Removing                                                                          OPERATION MODE                            PLC/PNP
                                      Seconds Delay    Dry On
                                     Dry On Wet On Dry                                                                                                                           Dry On Mode          Dry On      Wet On
                    N        L       0.3           0.3 Wet             NC         C      NO     NC    C    NO                                                                      Dry                                           Isolate Supply
                                                                                                                                            +          OUT             -
                                       1           1                                                                                                                               Wet                                           Before Removing
                                                                                                                RELAY
                                                                                                                                                                                                     0.3               0.3
                                       3           3                                                                                                                                                 1                 1
                                      10
                                                   10 Dry                                                                                                                                            3                 3
                                      30
                                                   30 Wet                                                                                                                             Dry            10                10
                                                             Wet On                                                                                                                   Wet            30                30
             1       2       3                                             4      5      6       7    8    9                         1      2                 3     4            Wet On Mode         Seconds Delay
                                                                                                                                                 Fuse 2A(T)
                                                                       NC         C      NO     NC    C   NO                                                                                                           IL (MAX) = 0 - 500 mA
                                                                                                                                                                                  U = 20 - 60 V            (dc)
            PE                   Fuse 0.5 (T)                               Resistive Load           Inductive Load                 PE                                            I < 4 mA + I L                       IPK = 5 A, 40 ms (inrush)
       (Ground)
                                                                            cos φ = 1 ;              cos φ = 0.4 ;               (Ground)
                                                                                                                                                                                                                       UOUT(ON) = U - 2.5 V
                                                                            L/R = 0 ms               L/R = 7 ms
                                     DPST                                                             IMAX = 3.5 A                                                                                                     IL (OFF) < 100 A
                                                                            IMAX = 5 A                                                    +V              O/P 0V
                                                                            ac                       ac
                                           U = 20...264 V ~ (ac)
                    N            Live                                       UMAX = 250 V             UMAX = 250 V
                                                     (50/60 Hz)
                                           I < 6 mA                         PMAX = 1250 VA           PMAX = 875 VA
                                                                            dc                       dc
                                           U = 20...60 V       (dc)         UMAX = 30 V              UMAX = 30 V
                    0V           +V        I < 6 mA                          PMAX = 240 W            PMAX = 170 W
Note
The external DPST switch that is shown in the wiring diagrams is an optional local disconnect (customer supplied).
                                                3.5 (89)
         Allow 1.2 (30)                                                              Allow 1.2 (30)
      to remove cover                 4 (102)                                     to remove cover            4.7 (120)
                   5                                                                             5.9
               (127)                                             A                             (151)                              B
                                                      C
                                                                                                                          C
                                                      D
                 2.7                                                                                                     D
                                                                                                 2.7
                 (69           1.7                                                              (69)
                                                  0.5 (13) switchpoint                                    1.7
                               (44)                                                                       (44)
                                                  (when mounted vertically)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                             C. 1.6 (40) A/F hexagon
B. Cable entry M20 x 1.5 or 3/4-in. ANPT                             D. 3/4- or 1-in. thread
Dimensions are in inches (millimeters). See the Rosemount 2120 web page for all 1-in. BSPP threaded dimensional drawings.
                5.9
              (151)                                                   A                                                                    B
                                                                                                  6.7
                                                                                                (171)
                                                       C
                                                                                                                                  C
                                                       D
                                                                                                                                 D
                                                  Ø1.1 (28) for 1-in. thread                           0.5 (13)                 0.5 (13) switchpoint
                                                  Ø0.9 (23) for 3/4-in. thread              switchpoint (when
                                                                                                                                (when mounted vertically)
                                                                                          mounted horizontally)
                                1.7                0.5 (13) switchpoint
                                (44)               (when mounted vertically)
                           0.5 (13)
                switchpoint (when
              mounted horizontally)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                                  C. 1.6 (40) A/F hexagon
B. Cable entry M20 x 1.5 or 3/4-in. ANPT                                  D. 3/4- or 1-in. thread
Dimensions are in inches (millimeters). See the Rosemount 2120 web page for all 1-in. BSPP threaded dimensional drawings.
Table A-4. Fork Length for 3/4- and 1-in. Threaded Rosemount 2120
              Process                              Standard length                           Minimum length                         Maximum length
            connection                            fork length code A                      fork length code E (M)                  fork length code E (M)
            3/4-in.   thread                        1.7 in. (44 mm)                          3.75 in. (95 mm)                         157.5 in. (4000 mm)
            1-in. thread                            1.7 in. (44 mm)                           3.74 in. (94 mm)                        157.5 in. (4000 mm)
Figure A-5. Tri Clamp Mounting (Standard Length, Surface Finish Codes 1 and 2)
                       Glass-filled nylon housing                                               Aluminum/stainless steel housing
                    (and not hygienically approved)                                              (and not hygienically approved)
                                       4 (102)                                                                     4.7 (120)
                                                                             Allow 1.2 (30)
       Allow 1.2 (30)                            3.5 (90)                 to remove cover
    to remove cover
A B
                 5.2                                                                     6.1
               (132)                                                                   (155)
                               C                            D                                               C
                                                                                                                                    D
                                                 Ø0.9 (23)
                 2.5                                                                                                       Ø0.9 (23)
                (64)           1.7                                                       2.5
                                                   0.5 (13) switchpoint                                     1.7
                               (44)                                                     (64)                                   0.5 (13) switchpoint
                                                   (when mounted vertically)                                (44)
                                                                                                                               (when mounted vertically)
              0.5 (13) switchpoint
                                                                                         0.5 (13) switchpoint
       (when mounted horizontally)
                                                                                  (when mounted horizontally)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                               C. 1.6 (40) A/F hexagon
B. Cable entry M20 x 1.5 or 3/4-in. ANPT                               D. 11/2-in. (38 mm) or 2-in. (51 mm) Tri Clamp, surface finish codes 1 and 2
Dimensions are in inches (millimeters).
Figure A-6. Tri Clamp Mounting (Standard Length, Surface Finish Codes 3, 4, 7, and 8)
                          Glass-filled nylon housing                                                Aluminum steel housing
                         (and hygienically approved)                                              (and hygienically approved)
                                                                                                                                          B
                                                                      A
                                                                                          6.1
              5.2                                                                       (155)
            (132)
                                                                                                                                    D
                        Ø1.65 (42)                          D                                      Ø1.65 (42)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                               C. 1.6 (40) A/F hexagon
B. Cable entry M20 x 1.5 or 3/4-in. ANPT                               D. 11/2-in. (38 mm) or 2-in. (51 mm) Tri Clamp, surface finish codes 3, 4, 7, and 8
Dimensions are in inches (millimeters).
Figure A-7. Tri Clamp Mounting (Extended Length, Surface Finish Codes 1 and 2)
                                       4 (102)
                                                                                          Allow 1.2 (30)
         Allow 1.2 (30)
                                                                                       to remove cover             4.7 (120)
      to remove cover                            3.5 (90)
A B
               5.2
             (132)                                                                            6.4
                                                                                            (162)
                                                            C
                          Ø1.6 (41)
                                                                                                      Ø1.6 (41)                     C
                  E(M)                                                                           E(M)
                                                                                                                            Ø1.1 (28)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                                C. 11/2-in. (38 mm) or 2-in. (51 mm) Tri Clamp, surface finish codes 1 and 2
B. Cable entry M20 x 1.5 or 3/4-in. ANPT
Table A-5. Fork Length for Tri Clamp Rosemount 2120 (Not Hygienically Approved)
              Process                               Standard length                      Minimum length                          Maximum length
            connection                             fork length code A                 fork length code E (M)                   fork length code E (M)
             Tri Clamp
                                                     1.7 in. (44 mm)                     4.13 in. (105 mm)                      157.5 in. (4000 mm)
       (surface finish code 1)
             Tri Clamp
                                                     1.7 in. (44 mm)                     4.13 in. (105 mm)                      157.5 in. (1000 mm)
       (surface finish code 2)
Figure A-8. Tri Clamp Mounting (Extended Length, Surface Finish Codes 3, 4, 7, and 8)
                                    4 (102)
                                                                                      Allow 1.2 (30)
     Allow 1.2 (30)                                                                to remove cover             4.7 (120)
                                              3.5 (90)
  to remove cover
                                                                      A
                                                                                                                                      B
               5.2
             (132)                                                                       6.1
                                                                                       (155)
                      Ø1.65 (42)                         C
                                                                                                                                C
                                                                                                Ø1.65 (42)
A. Cable entry M20 x 1.5 or 1/2-in. ANPT                               C. 11/2-in. (38 mm) or 2-in. (51 mm) Tri Clamp, surface finish codes 3, 4, 7, and 8
B. Cable entry M20 x 1.5 or 3/4-in. ANPT
Table A-6. Fork Length for Tri Clamp Rosemount 2120 (Hygienically Approved)
                Process                              Standard length                       Minimum length                        Maximum length
              connection                            fork length code A                  fork length code E (M)                 fork length code E (M)
                Tri Clamp
                                                         1.7 in. (44 mm)                   4.13 in. (105 mm)                    157.5 in. (1000 mm)
  (surface finish codes 3, 4, 7, and 8)
                                       3.5 (90)
     Allow 1.2 (30)                                                                  Allow 1.2 (30)
  to remove cover            4 (102)                                              to remove cover              4.7 (120)
        4.8
      (121)                                              A                                                                             B
                                                                                               6
                                                                                           (154)
                                        3.5 (89)
     Allow 1.2 (30)
  to remove cover
                              4 (102)
                                                                            Allow 1.2 (30) to
                                                                               remove cover             4.7 (120)
      4.8
    (121)
                                                          A                             6
                                                                                    (154)
                                                                                                                                B
G PNP/PLC (3-wire) 20 to 60 Vdc NA, E1, E5, E6, E7, EM, G5, G6 ★
V Relay DPCO, 20 to 264 Vac 50/60Hz, 20 to 60 Vdc NA, E1, E5, E6, E7, EM, G5, G6 ★
K NAMUR All ★
H 8/16 mA All ★
                                                                                                             Available              Available
 Surface finish
                                                                                                            connections             housings
1           Standard surface finish                                                                               All                     All    ★
2(3) Hand polished (Ra < 0.4 μm) Tri Clamp only All ★
7(4) Mechanically-polished to Ra < 0.1 μm, hygienically approved Tri Clamp only A, D, X, Y ★
8(4) Electro-polished to Ra < 0.38 μm, hygienically approved Tri Clamp only A, D, X, Y ★
G6(7) CSA Ordinary Locations (unclassified, safe area) All except option E Y, T ★
D Glass Filled Nylon, 1/2-in. ANPT conduits/cable threads NA, I1, I5, I6, and I7 ★
X Aluminum Alloy, M20 conduits/cable threads All except G5, G6, E5, E6 ★
Y Aluminum Alloy, 3/4-in. ANPT conduits/cable threads All except E1, E7, and EM ★
S Stainless Steel, M20 conduits/cable threads All except G5, G6, E5, E6 ★
T Stainless Steel 3/4-in. ANPT conduits/cable threads All except E1, E7, and EM ★
H(8) Standard length flange 4.0 in. (102 mm) All flanged options ★
E(9) Extended, customer specified length in tenths of inches All except 1-in. BSPP O-ring (1P) ★
M(9) Extended, customer specified length in millimeters All except 1-in. BSPP O-ring (1P) ★
XXXX(9) Specific customer specified length in tenths of inches, or millimeters (XXX.X inches or XXXX mm) ★
 Material certification(8)(10)
Q15           NACE® MR0175 / ISO 15156                                                                                                                                            ★
 Safety certification(11)
QS            Prior-use certificate of FMEDA Data                                                                                                                                 ★
 Special procedures(12)
P1            Hydrostatic testing with certificate                                                                                                                                ★
 Hygienic certifications(13)
QA            3-A certificate                                                                                                                                                     ★
QE EHEDG certificate ★
 ASME-BPE statement(13)
QB            ASME-BPE statement                                                                                                                                                  ★
  7.   See “Product Certifications” on page 51. E6 includes G6 requirements. G6 is for use in unclassified, safe area locations only.
  8.   Not available for hand polished wet side.
  9.   Minimum length available for 3/4-in. threaded connection is 3.8 in. (95 mm); for 1-in. threaded, it is 3.7 in. (94 mm); for flanged, it is 3.5 in. (89 mm);
       and for Tri Clamp, it is 4.1 in. (105 mm). Maximum length is 157.5 in. (4000 mm), except for ECTFE co-polymer coating and polished process where the maximum length
       is 59.1 in. (1500 mm) and 39.4 in. (1000 mm) respectively. Examples: Code E1181 is 118.1 inches. Code M3000 is 3000 millimeters.
  10. Only available for process-wetted parts.
13. Available only for a Rosemount 2120 with a Tri Clamp fitting, Product Certification code NA, G*, or I*, and Surface Finish code 3, 4, 7, or 8.
                                Seal for 1-in. BSPP (G1A).                                                              Seal for 3/4-in. BSPP (G3/4A)
                                Material: Non-asbestos BS7531 grade X carbon                                            Material: Non-asbestos BS7531 grade X carbon
                                fiber with rubber binder                                                                fiber with rubber binder.
                                      1.54                                                                                     1.26
                                      (39)                                                                                     (32)
02100-1000-0001                                                                              02100-1040-0001
                                      1.30                                                                                     1.04
                                      (34)                                                                                    (26.5)
                                Adaptor boss 1-in. BSPP to                                                              Adjustable clamp gland for 1-in. extended
                                11/2-in. (38mm) Tri Clamp.                                                              lengths. It is not explosion proof.
                                Material: 316 SS fitting. FPM/FKM ‘O’ ring                                              Material: 316 St. steel, (Si) Silicone rubber seal
                                              1.20                                                                              1-in.
                                             (30.5)                                                                                              Pmax = 1.3 barg
                                   0.18                                                                                                          Tmax = 257°F (125°C)
                                   (4.5)                    1-in.
02100-1010-0001                                             BSPP                             02120-2000-0001
                                                                                                                                               11/2-in.
                                                                                                                                                BSPP
                                  2.36                    1.54
                                  (60)                    (39)
                                2-in. (51 mm) Tri Clamp kit including vessel                                            Adjustable clamp gland for 1-in. extended
                                fitting, clamp ring, seal.                                                              lengths. It is not explosion proof.
                                Material: 316 St. steel, NBR Nitrile                                                    Material: 316 St. steel, (Si) Silicone rubber seal
                                                                                                                                1-in.            Pmax = 1.3 barg
02100-1020-0001                                                                                                                                  Tmax = 257°F (125°C)
(1)                                                                                          02120-2000-0002
                                                                                                                                               11/2-in.
                                                                                                                                                NPT
 2.      The Quick Release kit is a set of accessories requiring a Rosemount 2120 with the 2-in. Tri Clamp option and an existing 2-in. NPT process connection on the vessel.
         For additional information, see Rosemount 2120 Quick Release kit – Quick Start Guide.
 3.      Check the Electronic Type and Product Certification sections in Table A-8 on page 46 for availability conditions.
 4.      Intrinsically Safe approved cassettes can only be replaced with the same type of IS cassette. Non-I.S. cassette types can be interchanged with other non-I.S. cassettes,
         but the new label must be fitted and the original part number transferred to the new label.
5. This replacement cassette is for versions of the Rosemount 2120 shipped since June 2013.
TÜV-tested and approved for overfill protection according to the      B.8 Canadian Registration Number
German DIBt/WHG regulations. Certified under safety devices for
                                                                      Certificate Number CRN 0F04227.2C
tanks and piping related to water pollution control.
                                                                      The requirements of CRN are met when a Rosemount 2120 CSA
                                                                      IS-approved (G6, E6, or I6 option codes) vibrating fork level switch
B.4 Marine approvals                                                  model is configured with 316/316L stainless steel (1.4401/1.4404)
ABS – American Bureau of Shipping                                     wetted parts and either NPT threaded or 2 to 8-in. ASME B16.5
GL – Germanischer Lloyd                                               flanged process connections.
SRS – Russian Maritime Registered Shipping (RMRS)
                                                                      B.9 Safety Integrity Level (SIL)
B.5 Drinking water approval                                           The Rosemount 2120 has been independently certified to IEC
                                                                      61508 as required by IEC 61511. Certification was conducted by
Rosemount Measurement Ltd. (United Kingdom) confirms that the
                                                                      Exida. The Rosemount 2120 is SIL2-certified and SIL3-capable.
wetted parts of the Rosemount 2120 vibrating fork level switch are
suitable and approved for drinking water usage. The wetted parts
of the vibrating fork level switches executed in stainless steel
(option code D) and Alloy C / Alloy C-276 (option code C) with
flanged, NPT thread, BSPT(R) thread, or Tri-clamp process
connections, are in accordance with the requirements of DVGW*-
Worksheet W270.The materials used are classified as
toxicologically and microbiologically safe.
Product Certifications                                                                                                                  51
Reference Manual                                                                                            Product Certifications
00809-0100-4030, Rev FA                                                                                                   December 2016
B.10 Hazardous locations certifications                                  7. If the equipment is likely to come into contact with
                                                                            aggressive substances, it is the responsibility of the user to
B.10.1 American and Canadian approvals
                                                                            take suitable precautions that prevent it from being
Factory Mutual (FM) explosion-proof approval                                adversely affected, thus ensuring that the type of protection
                                                                            is not compromised.
E5   Project ID: 3012658
                                                                            Aggressive substances – e.g. acidic liquids or gases that may
     Explosion-proof for Class I, Div. 1, Groups A, B, C, and D             attack metals or solvents that may affect polymeric
     Temperature Class: T6                                                  materials.
     Enclosure: Type 4X                                                     Suitable precautions – e.g. regular checks as part of routine
                                                                            inspections or establishing from the material’s data sheet
Canadian Standards Association (CSA)                                        that it is resistant to specific chemicals.
explosion-proof approval                                                    The metallic alloy used for the enclosure material may be at
E6   Project ID: 1786345                                                    the accessible surface of this equipment; in the event of rare
     Explosion-proof for Class I, Div. 1, Groups A, B, C, and D             accidents, ignition sources due to impact and friction sparks
                                                                            could occur. This shall be considered when the Rosemount
     Temperature Class: T6
                                                                            2120 is installed in locations that specifically require Class 1,
     Enclosure: Type 4X                                                     Div 1 equipment.
     Single seal
                                                                         8. It is the responsibility of the user to ensure:
Instructions for hazardous area installations (E5 and E6)                       a. The voltage and current limits for this equipment are
                                                                                   not exceeded.
Model numbers covered:                                                          b. That the joint requirements between the switch and
2120*****E5Y**, 2120*****E5T**, 2120*****E6Y**,                                    the vessel tank are compatible with the process
2120*****E6T** (“*” indicates options in construction, function                    media.
and materials – see Table A-8 on page A-46).                                    c. That the joint tightness is correct for the joint material
                                                                                   used.
The following instructions apply to equipment covered by CSA and                d. That only suitable certified cable entry devices will be
FM explosion-proof approvals:                                                      used when connecting equipment.
                                                                                e. That any unused entries are sealed with suitably
  1. The equipment may be used with flammable gases and                            certified stopping plugs.
     vapors with apparatus Class 1, Div 1, Groups A, B, C and D.
                                                                         9. The switch fork is subjected to small vibration stresses as part
  2. CSA and FM explosion-proof approved versions of the                    of its normal function. As this provides a partition wall, it is
     Rosemount 2120 are certified for use in ambient                        recommended that the fork should be inspected every
     temperatures of –40 °F to 176 °F (–40 °C to 80 °C), and with a         two years for signs of defects.
     maximum process temperature of 302 °F (150 °C).
                                                                       10. Technical data:
  3. Installation of this equipment shall be carried out by suitably
                                                                                a. Coding: Class 1, Div 1, Groups A, B, C, and D
     trained personnel, in accordance with the applicable code of
     practice.                                                                  b. Temperatures: See Table B-1.
  4. Inspection and maintenance of this equipment shall be             Table B-1. Temperatures (E5 and E6)
     carried out by suitably trained personnel, in accordance with
     the applicable code of practice.                                  2120*****E5Y**, 2120*****E5T**, 2120*****E6Y**,
                                                                       2120*****E6T**:
  5. The user should not repair this equipment.
                                                                        Temperature              Maximum ambient        Maximum process
  6. The certification of this equipment relies upon the following                              air temperature (Ta)    temperature (Tp)
                                                                        classes
     materials used in its construction:
     Body:                                                             T6, T5, T4, T3, T2, T1          75 °C                    75 °C
     Aluminum alloy (ASTM B85 360.0) or 316 stainless steel
                                                                       T5, T4, T3, T2, T1              70 °C                    95 °C
     Cover:
     Aluminum alloy (ASTM B85 360.0) or 316 stainless steel            T4, T3, T2, T1                  65 °C                   125 °C
     Probe:                                                            T3, T2, T1                      50 °C                   150 °C
     316 stainless steel, or alloy C276 (UNS N10276) and
     alloy C (UNS N10002)                                              Minimum ambient air temperature (Ta) = –40 °C
     Probe filling: perlite                                            Minimum process temperature (Tp) = –40 °C
     Cover seal: silicone
52                                                                                                                       Product Certifications
Product Certifications                                                                                                 Reference Manual
December 2016                                                                                                      00809-0100-4030, Rev FA
I6   Certificate Number: 06 CSA 1786345                                        5. Suitably trained personnel shall carry out installation in
     Intrinsically Safe for Class I, Div. 1, Groups A, B, C, and D                accordance with the applicable code of practice.
     Class I, Zone 0, Ex ia IIC                                                6. The user should not repair this equipment.
     Non-incendive for Class I, Div. 2, Groups A, B, C, and D
     Temperature Code: T5 (See Control Drawings)
     Single Seal
Product Certifications                                                                                                                         53
Reference Manual                                                       Product Certifications
00809-0100-4030, Rev FA                                                         December 2016
54                                                                             Product Certifications
                                                                                                                                                    10      MOB-02714                41
Product Certifications
                                                                                                                        T5, T4, T3, T2, T1               80°C                60°C          T5, T4, T3, T2, T1            80°C                   80°C
                         CLASS I, ZONES 0,1 & 2, GROUPS IIA, IIB, IIC.                                                  T4, T3, T2, T1                   60°C                115°C         T4, T3, T2, T1                77°C                   115°C
                                                                                                                        T3, T2, T1                       50°C                150°C         T3, T2, T1                    71°C                   185°C
                                                                                                                                                                                           T2, T1                        65°C                   260°C
                                                                                                                                  Minimum Ambient Air Temperature (Ta) = -40°C
                                                                                                                                   Minimum Process Temperature (Tp) = -40°C                          Minimum Ambient Air Temperature (Ta) = -50°C
                                                                                                                                                                                                      Minimum Process Temperature (Tp) = -70°C
                                                                                                                                                                                                                                                                                                                                                                       Product Certifications
                           3. THE ENTITY CONCEPT ALLOWS INTERCONNECTION OF INTRINSICALLY SAFE                                                                       NONINCENDIVE FIELD WIRING PARAMETERS
                             APPARATUS NOT SPECIFICALLY EXAMINED IN COMBINATION AS A SYSTEM WHEN THE
                             APPROVED VALUES OF Uo (Voc ) AND Io (Ioc ) OF THE ASSOCIATED APPARATUS                                                      DESCRIPTION          VALUE
                             ARE LESS THAN OR EQUAL TO Ui (Vmax ) AND Ii (Imax ) OF THE INTRINSICALLY
                             SAFE APPARATUS AND THE APPROVED VALUES OF Co AND Lo OF THE ASSOCIATED                                                       MAXIMUM INPUT VOLTAGE (Vmax)                     15V
                             APPARATUS, ARE GREATER THAN THE TOTAL VALUES Ci AND Li OF ALL THE                                                           INTERNAL CAPACITANCE (Ci)                       211nF
                                                                                                                                                                                                                                                                     Figure B-1. FM Intrinsically Safe Control Drawing (NAMUR Electronics)
                             INTRINSICALLY SAFE APPARATUS INCLUDING ALL THE CABLE.                                                                       INTERNAL INDUCTANCE (Li)                      0.06mH
                                                                                                                                                                                                                                     PRELIMINARY
                           4. FOR FURTHER INFORMATION REFER TO MANUAL:
                              2120 : 00809-0100-4030                                                                                                                                                                               A
                              2130 : 00809-0100-4130
55
                                                                                                                                                                                                                                                                                                                                             00809-0100-4030, Rev FA
56
                                                                                                                                                            3       MOB-02714                   41
                          HAZARDOUS LOCATION.
                          CLASS I, DIVISION 1 & 2, GROUPS A, B, C, D.
                                                                                                                                                         2120***H*I***                                                    2130M**E***********I5**
                          CLASS I, ZONES 0,1 & 2, GROUPS IIA, IIB, IIC.
                                                                                                                                Temperature Classes Maximum Ambient Air       Maximum Process        Temperature Classes Maximum Ambient Air        Maximum Process
                                                                                                                                                     Temperature (Ta)         Temperature (Tp)                            Temperature (Ta)          Temperature (Tp)
                                                                                                                                T5, T4, T3, T2, T1           80°C                  60°C              T5, T4, T3, T2, T1           80°C                     80°C
                                                                                                                                T4, T3, T2, T1               60°C                  115°C             T4, T3, T2, T1               77°C                    115°C
                                                                                                                                                                                                                                                                                                                                                                                     Reference Manual
                                                                                                                                                                DESCRIPTION          VALUE
                         NOTES:                                                                                                                                 MAXIMUM INPUT VOLTAGE (Ui)                         30V
                                                                                                                                                                MAXIMUM INPUT CURRENT (Ii)                         93mA
                         1.   INSTALLATION OF EQUIPMENT TO BE IN ACCORDANCE WITH NEC ARTICLES 504
                                                                                                                                                                MAXIMUM INPUT POWER (Pi)                           0.65W
                         AND 505, AND ISA RP12.6 RECOMMENDED PRACTICE FOR INSTALLATION OF
                         INTRINSICALLY SAFE CIRCUITS.                                                                                                           INTERNAL CAPACITANCE (Ci)                           12nF
                                                                                                                                                                INTERNAL INDUCTANCE (Li)                          0.035mH
                         2.   THE ELECTRICAL CIRCUIT IN THE HAZARDOUS LOCATION MUST BE CAPABLE OF
                         WITHSTANDING AN A.C. TEST VOLTAGE OF 500V RMS TO GROUND OR THE FRAME                                                                            NONINCENDIVE FIELD WIRING PARAMETERS
                         OF THE APPARATUS FOR 1 MINUTE.
Product Certifications
Product Certifications                                                     Reference Manual
December 2016                                                            00809-0100-4030, Rev FA
Product Certifications                                                                        57
Reference Manual                                                           Product Certifications
00809-0100-4030, Rev FA                                                             December 2016
58                                                                                 Product Certifications
Product Certifications                                                                                           Reference Manual
December 2016                                                                                                00809-0100-4030, Rev FA
  3. The equipment is suitable for installation across the             12. Technical data
     boundary between an area that specifically requires                        a. Coding
     Equipment Protection Level Ga (Zone 0) and an area that                       ATEX:
     specifically requires Equipment Protection Level Gb or Db                     II 1/2 G D, Ex db IIC T6...T2 Ga/Gb,
     (Zone 1 or 21). The probe forks (and extension tube) only to                  Ex tb IIIC T85°C...T265°C Db
     be installed in Zone 0.                                                       IECEx:
                                                                                   Ex db IIC T6...T2 Ga/Gb,
  4. The equipment has not been assessed as a safety related
                                                                                   Ex tb IIIC T85°C...T265°C Db
     device (as referred to by directive 94/9/EC (2014/34/EU)
     Annex II, clause 1.5).                                                     b. Temperature: See Table B-3.
  5. Installation of this equipment shall be carried out by suitably   Table B-3. Temperatures (E1 and E7)
     trained personnel, in accordance with the applicable code of
     practice.                                                         2120*****E1X**, 2120*****E1S**, 2120*****E7X**,
                                                                       2120*****E7S**
  6. Inspection and maintenance of this equipment shall be
     carried out by suitably trained personnel, in accordance with                                          Maximum           Maximum
                                                                        Temperature classes and
     the applicable code of practice.                                                                      ambient air         process
                                                                        maximum surface
                                                                                                          temperature        temperature
  7. The user should not repair this equipment. Repair or               temperature (T)                       (Ta)               (Tp)
     modification of flame-paths is not permitted.
                                                                       T6, T5, T4, T3, T2, T1   T85 °C        75 °C              75 °C
  8. The certification of this equipment relies upon the following
     materials used in its construction:                               T5, T4, T3, T2, T1       T100 °C       70 °C              90 °C
     Housing and cover:                                                T4, T3, T2, T1           T135 °C       65 °C             125 °C
     Aluminum alloy (ASTM B85 360.0) or stainless steel 316C12
                                                                       T3, T2, T1               T160 °C       50 °C             150 °C
     Probe (partition wall): Stainless steel 316L or 316/316L, or
     Alloy C276 (UNS N10276) and Alloy C (UNS N10002 or
                                                                       Minimum ambient air temperature (Ta) = –40 °C
     N30002)
     Probe filling: perlite                                            Minimum process temperature (Tp) = –40 °C
     Cover seal: silicone
Product Certifications                                                                                                                     59
Reference Manual                                                                                             Product Certifications
00809-0100-4030, Rev FA                                                                                                     December 2016
60                                                                                                                         Product Certifications
Product Certifications                                                                                           Reference Manual
December 2016                                                                                                00809-0100-4030, Rev FA
Product Certifications                                                                                                                 61
Reference Manual                                                                                                  Product Certifications
00809-0100-4030, Rev FA                                                                                                        December 2016
Note
A certified isolating amplifier to IEC 60947-5-6 is required for
intrinsic safety if the NAMUR electronics is used in a hazardous area
installation (see below).
A certified intrinsically safe barrier is required for intrinsic safety if
the 8/16 mA electronics is used in a hazardous area installation
(see below).                                                                                        OK
Technical Regulation Customs Union (EAC)
flameproof approval
EM Certificate: TC RU C-GB.BH02.B.00175
   (Housing codes X and S only)
   Markings:
   1Exd IIC T6 X (-40 °C  Ta  +75 °C);
   1Exd IIC T5 X (-40 °C  Ta  +70 °C);
   1Exd IIC T4 X (-40 °C  Ta  +65 °C);
   1Exd IIC T3 X (-40 °C  Ta +50 °C);
   See certificate for special conditions for safe use (X)
                                                                               2. Installation of the level switch shall be carried out by suitably
Technical Regulation Customs Union (EAC)                                          trained personnel, in accordance with the applicable
intrinsically safe approvals                                                      standards and code of practice.
IM   Certificate: TC RU C-GB.BH02.B.00175                                      3. Inspection and maintenance of the level switch shall be
     (NAMUR and 8/16 mA electronics only)                                         carried out by suitably trained personnel, in accordance with
     Markings:                                                                    the applicable standards and code of practice.
     0Exia IIC T5 X (-40 °C  Ta  +80 °C);
                                                                               4. If the level switch is installed in a stub then, to ensure
     0Exia IIC T4 X (-40 °C  Ta  +60 °C);                                       clean-ability, the length (L) must not exceed the diameter (D)
     0Exia IIC T3 X (-40 °C  Ta  +50 °C)                                        with a minimum diameter of 46 mm.
     See certificate for special conditions for safe use (X)                      If the stub diameter is < 46 mm then L < 2 * (D - 23).
62                                                                                                                            Product Certifications
Product Certifications                                                     Reference Manual
December 2016                                                            00809-0100-4030, Rev FA
Product Certifications                                                                        63
Reference Manual          Product Certifications
00809-0100-4030, Rev FA            December 2016
64                                Product Certifications
                                                                                                       Reference Manual
                                                                                                  00809-0100-4030, Rev FA
                                                                                                          December 2016
Global Headquarters
Emerson Automation Solutions
6021 Innovation Blvd.
Shakopee, MN 55379, USA
   +1 800 999 9307 or +1 952 906 8888
   +1 952 949 7001
   RFQ.RMD-RCC@Emerson.com