Model : SI 4410
Operation Manual
                                               -2-
6. Interface                                                         ---------------   68 page
  6-1. Serial Interface (RS-232C)                                    ---------------   68 page
  6-2. Current Loop Interface                                        ---------------   71 page
  6-3. Print Interface (Option 01 : Centronics Parallel Interface)   ---------------   72 page
  6-4. Analog Output Interface (Option 02 : 0~10V)                   ---------------   74 page
  6-5. Analog Output Interface (Option 03 : 4~20mA)                  ---------------   75 page
  6-6. Serial Interface (Option 04 : RS-232C / 422 / 485)            ---------------   76 page
  6-7. BCD INPUT (Option 05)                                         ---------------   77 page
  6-8. Command Mode                                                  ---------------   78 page
                                                    -3-
                 1. BEFORE INSTALLATION
1-1. Caution / Warning Marks
            Warning
                         This mark warns the possibility to arrive death or serious injury in case
                         of wrongly used.
            Caution
                          This mark cautions the possibility to arrive serious human body injury
                          or product lose in case of wrongly used.
 1-4. Inquiries
  If you have any kinds of inquiries for this model, please contact with your local agent or Head
  Office.
  Head Office : Sewhacnm Co.,Ltd.
  Website : http://www.sewhacnm.co.kr
  Email : sales@sewhacnm.co.kr
                                            -4-
                                        2. INTRODUCTION
         2-1. Introduction
            Thank you for purchase this “SI 4410” Industrial Digital Weighing Controller.
            This “SI 4410” model is advanced model of “SI 3200”, with powerful communication performance.
            With 2ports serial port interfaces and precise weighing control system, you can upgrade your
            weighing process.
            This “SI 4410” Weighing Controller has various kinds of “Weighing Mode”, like Limit, Packer, Loss-
            in Weighing(Minus Limit), so you can apply various kinds of weighing application.
            Enjoy your process with “SI 4410” Weighing Controller.
         2-2. Cautions
            1). Don’t drop on the ground or avoid serious external damage on item.
Cautions 3). Don’t install place where high voltage or heavy electric noise condition.
            4). When you connect with other devices, please turn off the power of item.
            5). Avoid from water damage.
            6). For the improvement of function or performance, we can change item specification without prior
            notice or permission.
            7). Item’s performance will be up-dated continuously base on previous version’s performance.
          2-3. Features
            1). All Modules and Option Cards are isolated to maximize accuracy and performance.
            2). Self diagnose function
            3). External input terminal inside.(4pcs:Can be set by F11 mode)
            4). By using “Photo-Coupler” on each module(Option, Analog board, In/Out), we improved “Impedance
                 problem”, “Isolation ability among inputs”, “Leading power problem”, and “Noise covering function”.
            5). Data back-up function, when the sudden power off.
            6). “Set value Error” check function added. – Check “Set values for each weighing mode”.
                 - If there is any wrong set value, “E” will be display and will not start weighing process
            7). Polycarbonate film panel, strong for dust and water.
            8). Weight Unit selection Function added. (“g”, “kg”, “t” selectable – F40)
            10). Variable options(Order in advance, Refer Chapter 6. Interface) – “RS-232C” Standard installed.
            11). Improved “Automatic Free Fall(In-flight) Compensation” function added. – Suitable for
                 “Liquid Filling” system (Can compensate “minus” weight)
            12) 2port Serial Interface available – various applications (monitoring, printing, and Command
                 mode) are available.
                                                          -5-
                             3. SPECIFICATION
3-1. Analog Input & A/D Conversion
  Input Sensitivity                                                     0.2㎶ / Digit
  Load Cell Excitation                                          DC 10V ( - 5V ~ + 5V )
  Max. Signal Input Voltage                                              Max.32mV
                                                                      [Zero] ±16PPM/℃
  Temperature Coefficient
                                                                  [Span] ±3.5PPM/℃
  Input Noise                                                            ±0.3㎶ P.P
  Input Impedance                                                        Over 10㏁
  A/D Conversion Method                                                 Sigma-Delta
  A/D Resolution(Internal)                                       520,000 Count(19bit)
  A/D Sampling Rate                                               Max. 500times / Sec
  Non-Linearity                                                          0.005% FS
  Display Resolution(External)                                            1/20,000
                      Number,       Function,
         Key                                    Number Key, Function, CAL. Lock key (14pcs)
                      CAL. Lock Key
                                                -6-
                 3-3. General Specification
                   Power Supply                                 AC110/220V(±10%), 50/60Hz, about 30VA
                   Operating Temperature Range                  -10℃ ~ 40℃
                   Operating Humidity Range                     Under 85% Rh (non-condensing)
                   External Dimension                           200mm(W)x105mm(H)x165mm(L)
                   Net Weight(kg)                               About 2.3kg
                   Gross Weight(kg)                             About 3.0kg
Weight Display
                                                                                                                    Final Set value
                                                                                                                    (weighing count)
  Status Lamp
                                                                                                                    PRE2 Set value
                                                                                                                    (Total Weight)
  Function key
                                                                                                                    PRE1 Set value
                                                                                                                    (Total Weight)
                 ※ Through the “Front display”, you can check various weighing information, like weight unit, each set
                 value, relay output, accumulated weight of each P/N or all P/N.
                                                                 -7-
3-5-1. Status Lamp (ANNUNCIATORS) : “▼” Lamp is “ON”.
                   When display Gross weight(Net weight + Tare Weight), “▼” Lamp is turn on.
      Gross
                   (Under F19-01 setting)
Print When print key input or Auto print, “▼” Lamp is turn on.
                   When indicator transfers or receives data from other devices, “▼” Lamp is turn
      Comm.        on. (If the “▼” is off although there is some data transference, please check
                   communication settings).
                                            -8-
1. To START or STOP weighing process.
   First input, SI 4410 Controller Starts weighing process, and Second input, SI
   4410 Controller stops weighing process.
   ※ This function will be activated under F21-02, 04, 05, 06, 07 setting, only.
                       -9-
                    Whenever pressing “0”key, digit will be change 1, 2, 5, 10, and 50.
                  - Decimal point position
                    Whenever pressing “0”key, decimal point will be change.
                 1. Modify the set value during setting process.
                 2. Calibration mode
                  - Move back to previous step.
                 F-function Mode.
                  - F-function Exit : Press “Clear” key, once.
                  - F-Test Mode Exit : Press “Clear” key, twice.
                 1. Save set value during setting process.
                 2. Calibration mode
                  - Save current setting and move to next step.
                 3. F-Function mode
                    - Save current F-function setting, and move to next F-function
                 1. “F-TEST” Mode Entrance : Press “F” key for 5sec.
                 2. Under “F-function Mode”, Move to next Function or move to certain function
                    No.(Press function No. and press “F” key)
                 3. Function key (Refer “Function keys”)
                 Enter/Exit to “Calibration” mode.
                                       - 10 -
                                     Print current P/N’s accumulated weighing count and weight.
                                     (Sub-Total Print)
※ After Pressing key, you have to input above function keys within 5sec. - After 5sec, the
※ If you set “F51-01” you can check the / key activation through Main display.
                                                     - 11 -
                 3-6. Rear Panel
AC Power Input
                     ① POWER AC IN
                       - Power switch : Power on/off switch.
                           - Fuse : AC250V / 0.5A , φ5.25 , 20mm.
                           - AC IN : Available Input AC 110V / 220V.
                 !     ※ The standard power supply is AC 220V(Fixed when ex-warehouse), if you want to have
          Warning          AC 110V, please inform in advance.
                     ② Option Card 1
                     ③ Option Card 2
                      ※Option Card Connector installed for Optional Interface or Output.
                      (Printer I/F, Analog out, RS-422/485, or RS-232C(two port)
                     ④ LOAD CELL Connector (N16-05)
                     ⑤ SERIAL I/F
                      “RS-232C” or “CURRENT LOOP”(9Pin, D-Type Female) are built-in as standard
                     ⑥ External Input : External control input for wired remote control.
                      Refer to F-Function F11 to select desired function mode.
                      Input signal …………………………… Optical-Isolator
                                                               - 12 -
                                 4. INSTALLATION
4-1. External Dimension & Cutting Size
186
165
105 92
200 162.0
94
188
                                           - 13 -
4-2. Assembly
                - 14 -
4-3. Load Cell Installation
   4-3-1. Load Cell Connector Specification
                                                                                Red
                                                                                White
                                                                                Green
                                                                                Blue
                                                                                Shield
                                              - 15 -
4-3-3. Load Cell Wire Connection
   1). Please connect Indicator’s connector and Load cell cable basis on each color.
   2). It is possible to connect Max. 8pcs same capacity load cells with parallel. (350Ω)
   3). LOAD CELL Connector Standard : N16 - 05
   4). The load cell cable color can be different from each manufacturer, please refer following data
     sheet.
   5). Load Cell Wire Color Chart (Sorted by Manufacturer)
                                           - 16 -
          4-3-4. Formula to plan the precise weighing system
          This “SI 4410” weighing controller’s Max. input sensitivity is 0.2㎶ / Digit.
          And for precise weighing system, the following formula must be satisfied.
   !      Caution : “Input sensitivity” means Min. output voltage variation of weighing part to change 1digit.
                   So, please do not make large input voltage to make reliable weighing system.
Caution
                                                  ≤EⅹBⅹD
                                                                 A : Load cell capacity(kg)
           Single Load cell use         0.2㎶
                                                       A         B : Load cell Voltage(mV)
                                                                 D : Digit
                                                  ≤EⅹBⅹD
                                                                 E : affirmation Voltage of Load cell
           Plural Load cells use        0.2㎶                     N : Number of Load cell
                                                     AⅹN
          Example1.)
                     Number of Load cell : 1pcs
                     Load cell capacity : 500kg
                     Load cell Voltage : 2mV/V
                     Digit : 0.05kg
                     Affirmation Voltage of Load cell : 5,000mV
                     Max. Capacity of Weighing System : 300kg
          Then, estimation result for this weighing system with formula,
               5000ⅹ2ⅹ0.05                             The calculated value is larger than 0.2㎶, so this
                                      = 1 ≥ 0.2㎶
                       500                             system has no problem.
          Example2.)
                     Number of Load cell : 4pcs
                     Load cell capacity : 500kg
                     Load cell Voltage : 2mV/V
                     Digit : 0.10kg
                     Affirmation Voltage of Load cell : 5,000mV
                     Max. Capacity of Weighing System : 1,000kg
                                                   - 17 -
                                       5. SET-UP
5-1. Calibration
Adjust weight balance between “Real weight” on the load cell(Weight Part) and “Displayed weight of
Indicator”. When you replace LOAD CELL or Indicator, you have to do Calibration process once again
5-2. Test Weight Calibration Mode (Using Test weight)
Prepare At least 10% of Max. capacity of your weighing scale, and remove “CAL-BOLT” on the Front
panel and press “CAL - LOCK S/W” inside.
                                                      SeT-CAL
Then, you can enter the Calibration Mode with                            and start calibration mode with
pressing key.
※ For the save the each step, press key, for the cancel or move back, press key.
      Set Max. Capacity of Load cell                          Press “Enter” and move the next step.
Under this step, input Max. Capacity of you weighing scale.
Input Max. capacity of your weighing scale with No. keys.
The Max. capacity can not be exceed Max. capacity of load cell.
After finishing input, press           key and save and move to next step.
 Step No.2 - Setting “Decimal Point”
                                             - 18 -
 Decimal point position select with “0” key                   Press “Enter” and move the next step.
  Whenever pressing                key, Decimal point will be changed, like XXX. – XX.X – X.XX -
   .XXX
  The Max. Decimal point will be third place under Zero.
  On the Optimal position, stop pressing               key and press             key to save and move to
  next step.
Set Digit(Division) with “0”        Press “Enter” and move the next         Press “Enter” and move the
               key.                               step.                               next step.
Whenever pressing key, the Digit/Division value will be changed, like 01-02-05-10-50.
  On the optimal Digit/Division value, stop pressing               key and press               key to save
  and move to next step.
  ※ Caution
  (Max. capacity value / division value) can not be over 20,000.(as Indicator resolution is 1/200,00).
  If the value is over 20,000, Error message “ Err 01 “ will be displayed and move back “CAPA” mode
  again.
   Step No.4 - Measure the “DEAD Weight of Weighing Scale.
                                                - 19 -
Memorize “DEAD weight” of Weighing Machine                    Indicator will search “DEAD weight” during
                  with “Enter”.                                             5sec, automatically.
   When you press                 key, the indicator starts to find “Zero” condition of scale during the 5sec.
   After finding optimal “Zero” value, automatically move to next step.
   ※ Caution
   At this step, if there some force or Vibration on Weighing scale, and unstable condition will be
   continued, “ErrorA” will be display, and “DEAD value” will not be calculated.
   Under this condition, please remove force or vibration and process it again.
Step No.5 - Input Test Weight value and Calculate SPAN value.
   Under this step, input prepared “Test Weight’s value” and calculate “SPAN Value”.
   The “Test Weight’s value” must be at least 10% max. capacity of weighing scale.
                                                   - 20 -
                                                                           UP
Input prepared Test weight value and press               key, then                       message will be
displayed.
※ “At least 10%” means to guarantee precise weighing process you have to make standard with at least
    10% of weight of Max. capacity.
    We programmed the calibration will not be done, when you load less than 10% of max. capacity.
※ Caution
  In case of Max. load cell capacity value /Division value is under 5,000, please prepare Standard
  weight at least 10% of Max. load cell capacity value or max. display capacity value and set.
  In case of Max. load cell capacity value / Division value is over 5,000, please prepare Standard weight
  at least 20% of Max. load cell capacity value or Max. display capacity value   and set.
  This will be make more accuracy calibration setting.
  - If you set test weight value over max. load cell capacity, Error Message “Err 04” will be displayed.
  - If you set test weight value under 10% of max. load cell capacity value, Error Message “Err 05” will
    be displayed.
Calibration process completed. Push Calibration Lock S/W and tight Cal. Bol.
Press               key to save all calibration process. Now, Off the “CAL LCK S/W” and fasten the
Calibration Bolt.
※ Caution
To process “Simulation Calibration” process, All indicator has its’ own standard value of 2mV gap.
So, if you replaced analogue board, you have to input standard value of 2mv gap.
And you can check the this 2mV gap value on F96. (value is between 200,00 ~400,000)
                                              - 21 -
        5-3. Simulation Calibration Mode (Calibrate without Test weight)
        Through this “Simulation Calibration Mode” you can make simple calibration without Test weight.
        This calibration mode uses “Load cells’ max. capacity” and “Max. Output Rate(mV)”, the weight
        adjustment degree might be less than “Test weight Calibration”.
        The guaranteed resolution of this “Simulation Calibration” is 1/3,000.
Remove “CAL-BOLT” on the Front panel and press “CAL - LOCK S/W” inside.
SeT-CAL
Then, you can enter the Calibration Mode with and press key to
CELCAL
enter “Simulation Calibration Mode” with and start calibration mode with
pressing key.
   Remove the CAL.Bolt.                         Press NO.1 key to enter               Check the “CELL-CAL”. Massage
    Push the CAL. S/W                         the “CELL Calibration” mode                          on display.
※ For the save the each step, press key, for the cancel or move back, press key.
                                         Check Max. Capacity of Load cell              Input the Max. Capacity of Load
Enter the Capacity input mode
                                               (Refer the Load cell Lable)                 cell(same value of Label
        Under this step, input Max. Capacity of Load cell. (Not weighing Scale)
        The Max. Capacity of load cell is stated on “Test report” or “Label”.
                                                       - 22 -
       Please input this max. capacity of load cell value with No. keys.
       After finishing input, press             key and save and move to next step.
       ※ Caution
       If the plural No. of load cells are installed, please make sum the all load cells capacity and input.
       Ex) There are 4pcs of load cells are installed, and each load cell’s Max. capacity is 1,000kg.
           Then, total Max. Capacity will be 4,000kg and you have to input 4,000kg.
    Enter the Decimal point input mode                       Input Decimal point and Press Enter to move next step
       Under this step, define the optimal position or Decimal point.
       Whenever pressing                 key, Decimal point will be changed, like XXX. – XX.X – X.XX - .XXX
       The Max. Decimal point will be third place under Zero.
On the Optimal position, stop pressing key and press key to save and move to next step.
Enter the division input mode              Select optimal division with pressing            Press Enter key and move to next
                                                         “o” key                                         step.
       Under this step, define the optimal Digit/Division value of weighing measurement.
Whenever pressing key, the Digit/Division value will be changed, like 01-02-05-10-50.
On the optimal Digit/Division value, stop pressing key and press key to save
                                                      - 23 -
           and move to next step.
           ※ Caution
           (Max. capacity value / division value) can not be over 20,000.(as Indicator resolution is 1/200,00).
           If the value is over 20,000, Error message “ Err 01 “ will be displayed and move back “CAPA” mode
           again. But, under “Simulation Calibration Mode”, please make (Max. capacity value / division
           value) setting is less than 3,000.
           When you press               key, the indicator start measurement and find optimal “Dead weight value
           of Scale” automatically.
           ※ Caution
           At this step, if there some force or Vibration on Weighing scale, and unstable condition will be
           continued, “ErrorA” will be display, and “DEAD value” will not be calculated.
           Under this condition, please remove force or vibration and process it again.
                                                         - 24 -
 Under this step, input Max. Output rate(mV) of load cell.
 The Output rate is stated on “Test report” or “Label” and please input this value with No. keys.
 Normally, the Output rate will be 4digits under Zero, then input 4digits and input “0” additionally and
 make full 6digits.
Calibration process completed.                            Push Calibration Lock S/W and tight Cal. Bol.
 Now, the Simulation Calibration is done.
   Press               key to save all calibration process and Off the “CAL LCK S/W” and fasten the
   Calibration Bolt.
                                               - 25 -
5-4. Set-up
Set-up means set the F-function and make SI 4410 weighing controller will perform more accuracy.
(Considering external / internal environmental condition)
5-4-1. Enter the Set-up Mode
      1). Method : Press           key for 4sec. Then you can enter “F-Test” mode. Under this mode,
                    press No.1 key and enter the “F-function” mode.
 5-4-2. F-Function Change
      Under F-function mode, Whenever press                 key, the Function No. will be increased one by
      one. Increase to F-90 and return to F-01
      If you move to certain function No., press f-function no. with number key and press            key.
      Ex.) If you want to call “F11-XX ” directly under “F-function mode”.
Under F-Function mode, input New set value with Number keys and press key to save.
      If you don’t press          key, the new set value will not be memorized.
      Ex.) If you want to change the “F01-01” to “F01-02”.
Under “F-function” mode, press key, you can move back to “F-Test” mode.
Under “F-Test” mode, press key once again, you can move back “Stand-by” mode.
                                             - 26 -
5-5. F-Function List
                                General Function Setting
    F01         Weighing Data Save Method selection       0, 1, 2
                                           - 27 -
                   Communication Mode setting
F30          Parity Bit selection Mode            0, 1, 2
                             Other setting
F80            Empty Range setting                XXXXXX
F89        Span calibration value check           X. X X X X X
F90            DATE check/change                  XX. XX. XX
                                 - 28 -
F91              TIME check/change                XX. XX. XX
                                  - 29 -
5-6. F-Function Detailed information
■ General Function Setting (● Factory default set value)
Finished Weighing data will be saved when               key input. And Print will be output simultaneously. So, the
saved data and printed data is same.
2. F01-00, F38-01 setting (Manual Save mode + Auto Print(When Finish relay output)mode)
When “Finish Relay” output, Print will be done automatically., But Finished weighing data will not be
saved automatically. If you want to save “Finished data”, you have to press            key separately.
※ Caution
Under this setting, there will be some difference between “Saved Data” and “Printed Data”.
3. F01-01, F38-00 setting (Automatic Save(When Finish relay output) + Manual Print mode)
When “Finish relay” output, the finished data will be saved automatically, but this data will not be
printed out. If you want to print out “Finished data”, you have to press          key separately.
※ Caution
Under this setting, there will be some difference between “Saved Data” and “Printed Data”.
4. F01-01, F38-01 setting (Automatic Save(When Finish relay output) + Automatic print(When
     Finish relay output) mode
When “Finish relay” output, finished data will be saved and printed automatically.
Under F21-01/03/04/07 mode(without Finish relay mode), indicator will apply same time period to
make finished data save.(After reaching to “Final” and will apply F22-XX time period)
                                                - 30 -
※ Caution
 Under this setting, “Saved data” and “Printed data” is same. So, this setting is recommended for
  Automatic weighing process.
5. F01-02, F38-00 setting (Combined Save mode(Auto + manual save mode) + manual print mode
When “Finish relay” output, finished data will be saved automatically, but same data will not be printed.
And also, if there is additional          key input, weighing data will be saved once again and printed
out.
※ Caution
After “Finish relay” output, if there is additional          key input, indicator will save two different
data for single process.
6. F01-02, F38-01 setting (Combined Save mode(Auto + manual save mode) + Auto print mode)
When “Finish relay” output, finished data will be saved and printed automatically.
But, if there is additional         key input, weighing data will be saved once again.
※ Caution
After “Finish relay” output, if there is additional          key input, indicator will save two different
data for single process.
                                       Weight-Back up selection
             ●          0      Normal Mode
  F02
                        1      Weight Back up Mode
※ Normal Mode
When the power is off , “Zero value” will be out and will display wrong weight value after power is
recovered.
Also, if there is material over 10% of max. capacity weight, the “Un-Pass” display will be appear and
you can not process weighing job.
If there is material less than 10% of max. capacity weight on the scale, indicator will initialize
with ”Pass” display, and display “0” whatever the real weight is. Under this case, remove the material
and press “Zero” key and input new “Zero” value. Then, new Zero value will be memorized and you
can process weighing job.
                                               - 31 -
               ※ Weight Back up Mode
               Under this mode, SI 4410 memory “Zero” value of weighing part. So, when the power is recovered
               SI4410 can display material weight.
               If you remove and turn on the power, please press “Zero” key and re-memorize Zero value.
               ※ This function is compensate vibration effect on weighing part with acceptable range setting.
               During the fixed time period(F06 setting), there is smaller weight variation than Motion band range, due
               to vibration, Indicator will display “Steady”, if there is larger weight variation than Motion band range,
               due to vibration, Indicator will its weight and “Steady” condition is broken and find new “Steady” point.
               If there are much vibration effect, please set with large set value.
                                                                                                                       Motion Band
           Zero
                                                                                                                            Range
          ※ In this case, if you increase “Motion Band Range”(F03-03), the “Steady Break point” will be in the
          steady range, and indicator will display “Steady” condition.
                                                                                                                            F03-03
Motion Band Range
                                                                                                                            F03-01
                                                                                                                       Motion Band
        Zero
                                                                                                                            Range
※ Each set value “1” is fixed 50% of one Digit, and the Digit is calculated after “Calibration Process”.
                                                                  - 32 -
                                            Zero Tracking Compensation Range setting
                                         Due to external causes(Temperature, wind, and dust), there are small weight
                                         difference, indicator will ignore the weight difference and display Zero.
                                  0
                                         For this compensation function, indicator will estimate the weight
              F04       2         ∫
                                         difference is over the set range during fixed time period.
                                  9
                                         If there is large weight difference over set range within fixed time period,
                                         the “Zero” is breaking and will find new zero point.
                    Weight
                                                                                        Zero breaks
                                                                                                               Display Current weight
    0.075kg
  (1.5ⅹdigit)
                                                                                                             Display Zero
                                                                                                            time
                             about 5msec.
         ※ Using this function, you can get the Zero value without pressing “Zero” key, when there is remained
                material on the scale within Auto Zero Range.
         Example) Max. Capacity : 10kg, Digit : 0.02kg, F005-30 setting,
         Under this setting, Indicator will display “Zero” automatically, when the weight is within ± 0.30kg(Set
         value + weight unit) and Steady.
              -0.30kg
Motion Band Range                                                                                                           F03-01
                                                            - 33 -
                                         Digital Filter setting
                                                                             If “B” set value is fixed,
                              A : Frequency Filter setting value (0~3)
                                                                             “A” set value is large, the
  F06         23     AB       (0 : about 200Hz/sec, 3 : about 500Hz/sec)
                                                                             indicator will response
                              B : Buffer Filter setting value (1~9)
                                                                             more sensitive.
※ Key TARE Function : You can set TARE value with No. keys, without weight.
※ To input Key TARE value(F10-01 setting),
  Press “TARE” key, and input “target TARE value” with No. keys and press “ENTER”.
                                               - 34 -
                                      External Input Selection
             Set Value          Input 1           Input 2             Input 3               Input 4
                                                   TARE/
                      1       RUN/STOP                                ZERO                  PRINT
                                               TARE RESET
   F11                                                                                    SUB-TOTAL
            ●         2            RUN             STOP               PRINT
                                                                                            PRINT
                                                                                         NET WEIGHT/
                      3            ZERO            TARE           TARE RESET
                                                                                         GROSS WEIGHT
 ※ Within “Motion Band Range” and “Zero Tracking Compensation Range”, weight variation is not
     over/under the range during “STEADY” condition check time period(F12), indicator will take
     “STEADY” and “STEADY” lamp is “ON”.
                                              - 35 -
                                   Display Up-date rate selection
                         0    About 10ms
                         1    About 30ms
               ●         2    About 50ms
  F13
                         ∫         ∫
                         8    About 170ms
                         9    About 190ms
※ Under F14-01 setting, after set this value and you have to input new set values for each set point
  (FINAL, PRE1, PRE2, FREE FALL). These new set values will be applied to all P/N.
※ The Sub/Grand Total data will be displayed on Sub-display windows, when                and
  key input.
※ Max. Display value of Accumulated weighing Count is “999,999”, and over this value, count will be
  return to “000,000”.
※ Max. Display value of Accumulated weight is “2,147,483,647”, and over this value, accumulated
    weight will be return to “0,000,000,000”.
                                                - 36 -
                           “NEAR ZERO” relay output mode selection
                             Display weight is Zero(Including “TARE” Zero)Æ Near Zero relay
            ●         0
  F17                        output
※ Under F17-01 setting, even if the TARE is set and display weight is Zero, there is no NEAR Zero
    relay output. – Only when the Net weight is Zero, there is Near Zero relay output.
            ●         0      Not Display
  F19
                      1      Display on PRE2 display window
※ Gross weight will be display on PRE1 and PRE2 display window, and PRE1 and PRE2 set values
    will be display, only when PRE1, PRE2 key input.
                                             - 37 -
■ Relay Output Mode Setting
※ If the set values (FINAL, PRE1, PRE2, FREE FALL) are changed, or new calibration process is
    done, or power is off, “F-function” set value is changed, this “Automatic Free Fall Compensation”
    function will be initialized.
※ This “Automatic Free Fall Compensation” can compensate minus(-) Free Fall, and can be displayed
    Max. “-999”.
※ To make more accuracy of “Automatic Free Fall Compensation” function, please set “t1(F22)”
    value with large set value, or “0(on Steady Mode)”.
    If the “t1” set value is too small, there might be some error on making accuracy compensation.
                                              - 38 -
                                     Weighing Mode selection
                                             - 39 -
             ◆ Weighing Mode 1. Limit Mode (F21-01 setting) - No Finish Relay output.
   Weight
                                                    FINAL             ※     This   Flow   chart   was   made   based   on
                                      PRE 2                                 “LOW/HIGH” set value is “0”
                     PRE 1
Near Zero
TARE input
PRE 1 Output1
PRE 2 Output2
FINAL Output3
LOW Output4
HIGH Output5
                                                             - 40 -
             ◆ Weighing Mode 2. Packer Mode (F21-02 setting) - Finish / Error Relay output
Near Zero
TARE input
START
PRE 1                                                                                                                Output1
PRE 2                                                                                                                Output2
FINAL                                                                                                                Output3
FINISH                                                              t1         t2
                                                                                                                     Output4
                                                                               t4
ERROR                                                                                                                Output5
Near Zero                                                                                                            Output6
                                                            - 41 -
             ◆ Weighing Mode 3. Loss-in Weight Mode 1. (F21-03 setting) - No Finish/Error Relay output
Near Zero
PRE 1
                                                                                                         FALL
                                                                  PRE2
                                                                                FINAL
                                                                                                     FINAL Set
TARE input
TARE Reset
PRE 1                                                                                                      Output1
PRE 2                                                                                                      Output2
FINAL                                                                                                      Output3
LOW                                                                                                        Output4
HIGH                                                                                                       Output5
Near Zero                                                                                                  Output6
                                                          - 42 -
             ◆ Weighing Mode 4. Loss-in Weight Mode 2 - No Finish/Error Relay output
Weight PRE 1
Near Zero
                                                                                                     FALL
                                                       PRE 2
                                                                  FINAL
                                                                                     FINAL Set
TARE input
TARE reset
START
 PRE 1                                                                                                     Output1
 PRE 2                                                                                                     Output2
FINAL                                                                                                      Output3
LOW                                                                                                        Output4
HIGH                                                                                                       Output5
Near Zero                                                                                                  Output6
                                                         - 43 -
             ◆ Weighing Mode 5.. Loss-in Weight Mode 2)    - Finish / Error Relay output
   Weight
                                 PRE 1
Near Zero
                                                                                               FALL              HIGH
                                                     PRE 2                                                       PASS
                                                              FINAL        FINAL Set                             LOW
TARE input
TARE reset
START
 PRE 1                                                                                                      Output1
 PRE 2
                                                                                                            Output2
FINAL
                                                                                                            Output3
FINISH                                                                  t1        t2                        Output4
ERROR                                                                              t4                       Output5
Near Zero                                                                                                   Output6
                                                       - 44 -
               ◆ Weighing Mode 6. Loss-in Weight 3. - Finish / Error Relay output
                                       PRE 2
    Weight
                          PRE 1
Near Zero
                                           LOW
                                                                                                        FALL
                                           PASS
                                           HIGH                    FINAL
                                                                                    FINAL Set
TARE Input
TARE Reset
 START
   PRE 1                                                                                                       Output 1
PRE 2 Output 2
FINAL Output 3
FINISH t1 t2 Output 4
 ERROR                                                                                     t4                  Output 5
                                                                                                               Output 6
 Near Zero
                                                          - 45 -
                ◆ Weighing Mode 7. Loss-in Weight 4. (2step Feeding, 1step discharge, F21-07 setting)
                    - No Finish/Error Relay output
       Weight                              PRE 2
                                                                ※    This   Flow   chart   was   made   based   on
                              PRE 1
                                                                “LOW/HIGH” set value is “0
Near Zero
                                                                                                           FALL
                                                                    FINAL
                                                                                            FINAL Set
TARE input
   START
TARE Reset
   PRE 1                                                                                                          Output 1
PRE 2 Output 2
FINAL Output 3
LOW Output 4
HIGH Output 5
                                                           - 46 -
            “FINISH Relay” delay time(t1) setting (Under F01, 05, 06 setting)
                      After current weight is reached to FINAL, you can set some delay time
                      of “FINISH relay ON time.
                                                                 Steady point
                 00
F22   10          ∫
                          FINISH Relay
                                                                                   Com-Out4
                 99     “00” setting : At Steady point, FINISH relay output
                        “20” setting : After 2.0sec from Steady point, FINISH relay output
                        “99” setting : After 9.9sec from Steady point, FINISH relay output
                                                                 Steady point
                 01
F23   10          ∫                                              t2
                          FINISH Relay
                 99                                                                Com-Out4
                        “01” setting : FINISH relay will be “ON during 0.1sec.
                        “20” setting : FINISH relay will be “ON” during 2.0sec.
F27   10          ∫
                                                       t4
                 99                                                             Com-Out5
                        “01” setting : ERROR relay will be “ON during 0.1sec.
                        “20” setting : ERROR relay will be “ON” during 2.0sec.
                                         - 47 -
                  Manual Discharge selection (Under F21-04, 05, 06, 07 setting)
※ Under Loss-in weight Mode2, 3, 4, Feeding start time will be determined by SP1 relay.
  Example 1.) FINAL set value is 1,000kg, F29-03 setting,
  When there is only 1,000kg in the feeding hopper, the PRE 1 relay will not be OFF, and you can start
  additional weighing process.
  But, at the same condition, if there is only 995kg in the feeding hopper, the SP1 relay will be OFF,
  and you can not start additional weighing process.
  If you want to start additional weighing, please change the F29 setting as large set value than 03.
  After finishing all the process, you want to discharge remained material, please use “Manual
  discharge function”.(“F” key + RUN/STOP key)
                                               - 48 -
■ Communication Mode setting (Serial Port 1. - Standard installed port)
2 Even Parity
※ When you set the “Serial Communication Speed”, you have to estimate your P.C’s spec. If you use
  normal P.C, you had better to set with 0, 1, or 2 set value.
※ Under Command Mode setting(F32-01), you can not use serial port 1. to use Data transference or
    printing purpose. So the following “F-functions, F35, F36, F37, F38” will not be activated.
    So, ignore the F35, 36, 37, 38 settings.
    Under Command Mode setting(F32-01), F33, and F34 function will be activated.
※ If you want to fix “Serial Port 1.” for Command mode, and you also want to data transference or
    print purpose, you have to add optional serial port and set up for data transference or print through
    optional serial port, Please refer F65, 66, 67, 68.
                                                - 49 -
                            Print port selection (Under F32-01 setting, only)
            ●           0      Same port as using for Command Mode.
  F33
                        1      The other port.
※ Through this function, you can selection print port, which will be used for print, when indicator takes
    a “Print Command” from P.C. If you choose F33-00, the print data will be transfer to P.C with print
    format. If you choose the other port, the print data will be transfer through the other port – option
    Port(Com. Port 2)
    If you set F33-01, but you just connect with another P.C, you can take print data with print format
    though the another P.C In this case, the transferred data format is same as print format, not a data
    format(please refer “Print Interface section”)
                                                 - 50 -
                     Print Mode selection (Under F32-00, F35-01 setting, only)
            ●         0       Manual Print : Whenever “Print” key input.
  F38
                      1       Auto Print : When Finish relay output, automatically print.
■ Print Mode Setting (These settings will be apply to Serial and Parallel print)
※ Selected “Weight Unit” will be apply to “LED” display and Print format.
                                                 - 51 -
                       SUB/GRAND Total Data Delete selection
                       Manual Delete Mode
      ●         0      SUN Total Delete : “Clear” key + “P/N” key
F44                    GRAND Total Delete : “Clear” key + “S/N” key
                       Automatic Delete Mode
                1
                       After SUB/GRAND Total Print, Automatically Deleted.
                0
F45   3         ∫      Whenever set value increased, 1line will be added.
                9
                0
F46   3         ∫      Whenever set value increased, 1line will be added.
                9
                                      - 52 -
                                       Parallel Print Port selection
            ●            0      Parallel Port is not installed.
  F50                    1      Share Standard Serial Port.
                         2      Share Optional Serial Port.
※ If you install “Print interface option(Centronics Parallel), you have to choose, which port you will
    use for print data transference to “Print Interface”.
    If you installed “Print interface”, but you didn’t choose the port, you can not print out through
    Parallel print.
※ If there is an mal-function on Parallel print device, “PrinT-Er” will be displayed.
    Then, please check the Parallel print device condition.
                             Function / Clear key Activation display selection
            ●            0      Activation display not use
  F51
                         1      Activation display use
※ Under F51-01 setting, whenever you press “Function” / “Clear” key, “F” or “C” display will be
appear on main display window, and let you know, “Function” or “Clear” key is activated.
■ Other Setting
※ Under “Other setting mode”, you can not move to other function directly.
Press key and move to F01 and move to other function No. directly.
                                                 - 53 -
                                 SPAN Calibration Value Check
                             Span Calibration Value Check
                             “       ”.
  F89      X.X.X.X.X.X.
※ Under this mode, use           key to choose the each data and input new data with No. keys.
                                      TIME check / Change
  F91     Check Current TIME data or you can Change to new date
※ Under this mode, use key to choose each data and input new data with No. keys.
                                             - 54 -
■ Communication Mode setting (Serial Port 2. - Optional Serial port)
This setting will be activated only when “Optional Serial Port” is installed.
2 Even Parity
※ When you set the “Serial Communication Speed”, you have to estimate your P.C’s spec. If you use
normal P.C, you had better to set with 0, 1, or 2 set value.
※ Under Command Mode setting(F62-01), you can not use serial port 1. to use Data transference or
    printing purpose. So the following “F-functions, F65, F66, F67, F68” will not be activated.
    So, ignore the F65, 66, 67, 68 settings.
    Under Command Mode setting(F62-01), F63, and F64 function will be activated.
※ If you want to fix “Serial Port 2.” for Command mode, and you also want to data transference or
    print purpose, you have to set the F32-00, and F35-01 or 02.
                                                - 55 -
                           Print port selection (Under F62-01 setting, only)
            ●          0       Same port as using for Command Mode.
  F63
                       1       The other port.
※ Through this function, you can selection print port, where indicator takes a “Print Command” from
    P.C. If you choose F63-00, the print data will be transfer to P.C with print format. If you choose the
    other port, the print data will be transfer through the other port.
    If you set F63-01, but you just connect with another P.C, you can take print data with print format
    though the another P.C
※ The transferred data format is same as print format, not a data format(please refer “Print Interface
    section”)
                                                 - 56 -
          Print Mode selection (Under F62-00, F65-01 setting, only)
      ●    0      Manual Print : Whenever “Print” key input.
F68
           1      Auto Print : When Finish relay output, automatically print.
                                  - 57 -
■ Serial Port Setting Examples
※ Other set values, except example, are not influence the performance.
                                                    - 58 -
D. Serial Print installed
                                      Standard Serial Port Setting
 F32-00 : Simplex Mode setting
 F35-01 : Print Mode setting
 F38-00/01 : 00 setting : Manual Print Mode(             key input, 1time Print out)
             01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
※ Even if F38-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode).
                                                - 59 -
CASE 2.
             Standard Serial Port Setting – Data Transference Mode(P.C Monitoring)
 F32-00 : Simplex Mode setting
 F35-00 : Data Transference Mode
 F36-00/01/02 : 00(Steam Mode)
                01(When “Finish Relay” Output, 1time transfer)
                                                - 60 -
CASE 2.
                       Standard Serial Port Setting – External Display Mode
 F32-00 : Simplex Mode Setting
 F35-00 : Data Transference Mode
 F36-00 : Stream Mode Fix
 F37-00/01/02 : Transferred Data format setting.
 F40-00 : Under “External Display” mode, F40-00 Fix
                           Optional Serial Port Setting – Command Mode
 F32-01 : Command Mode
 F33-00(Nominate Printing port for “PRINT COMMAND”)
  – There is not installed print, So Print data will be transfer to P.C directly.
 F34-00/01(“Check-Sum” selection)
  - Not use : 00 setting,, Use : 01 setting
 F68-00/01 : 00 setting : Manual Print Mode(       key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
※ Even if F68-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode).
                                                - 61 -
CASE 2.
                          Standard Serial Port Setting – Serial Print Mode
 F32-00 : Simplex Mode setting
 F35-01 : Print Mode setting
 F38-00/01 : 00 setting : Manual Print Mode(             key input, 1time Print out)
             01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
                           Optional Serial Port Setting – Command Mode
 F62-01 : Command Mode
 F63-00(Nominate Printing port for “PRINT COMMAND”)
  – There is not installed print, So Print data will be transfer to P.C directly.
 F64-00/01(“Check-Sum” selection)
  - Not use : 00 setting,, Use : 01 setting
 F40-00/01 : Transferred Data Byte Setting
 – Apply only for Weight related data.
※ Even if F38-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
   Save Mode).
                                                - 62 -
CASE 2.
                      Standard Serial Port Setting – External Display Mode
 F32-00 : Simplex Mode Setting
 F35-00 : Data Transference Mode
 F36-00 : Stream Mode Fix
 F37-00/01/02 : Transferred Data format setting.
 F40-00 : Under “External Display” mode, F40-00 Fix
                     Optional Serial Port Setting – Data Transference Mode
 F62-00 : Simplex Mode setting
 F65-00 : Data Transference Mode
 F66-00/01/02 : 00(Steam Mode)
               01(When “Finish Relay” Output, 1time transfer)
 F68-00/01 : 00 setting : Manual Print Mode(       key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
※ Even if F68-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode).
                                             - 63 -
CASE 2.
                         Standard Serial Port Setting – Serial Print Mode
 F32-00 : Simplex Mode setting
 F35-01 : Print Mode setting
 F38-00/01 : 00 setting : Manual Print Mode(            key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
                     Optional Serial Port Setting – Data Transference Mode
 F62-00 : Simplex Mode setting
 F65-00 : Data Transference Mode
 F66-00/01/02 : 00(Steam Mode)
               01(When “Finish Relay” Output, 1time transfer)
※ Even if F38-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode).
 F68-00/01 : 00 setting : Manual Print Mode(       key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
※ Even if F68-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode).
                                               - 64 -
CASE 2.
                          Standard Serial Port Setting – Serial Print Mode
 F32-00 : Simplex Mode setting
 F35-01 : Print Mode setting
 F38-00/01 : 00 setting : Manual Print Mode(             key input, 1time Print out)
             01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
                        Optional Serial Port Setting – External Display Mode
 F62-00 : Simplex Mode Setting
 F65-00 : Data Transference Mode
 F66-00 : Stream Mode Fix
 F67-00/01/02 : Transferred Data format setting.
 F40-00 : Under “External Display” mode, F40-00 Fix
※ Even if F38-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
    Save Mode)
                                                - 65 -
H. Standard Serial Port & Optional Serial Port – Data Transference Mode(P.C Monitoring)
                     Standard Serial Port setting – Data Transference Mode
 F32-00 : Simplex Mode setting
 F35-00 : Data Transference Mode
 F36-00/01/02 : 00(Steam Mode)
               01(When “Finish Relay” Output, 1time transfer)
I. Standard Serial Port & Optional Serial Port – External Display Mode(SE6005/SE6105)
                      Standard Serial Port setting – External Display Mode
 F32-00 : Simplex Mode Setting
 F35-00 : Data Transference Mode
 F36-00 : Stream Mode Fix
 F37-00/01/02 : Transferred Data format setting.
 F40-00 : Under “External Display” mode, F40-00 Fix
                      Optional Serial Port Setting – External Display Mode
 F32-00 : Simplex Mode Setting
 F35-00 : Data Transference Mode
 F36-00 : Stream Mode Fix
 F37-00/01/02 : Transferred Data format setting.
                                             - 66 -
J. Standard Serial Port & Optional Serial Port – Serial Print Mode
                        Standard Serial Port Setting – Serial Print Mode
 F32-00 : Simplex Mode setting
 F35-01 : Print Mode setting
 F38-00/01 : 00 setting : Manual Print Mode(            key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
                         Optional Serial Port Setting – Serial Print Mode
 F62-00 : Simplex Mode setting
 F65-01 : Print Mode setting
 F68-00/01 : 00 setting : Manual Print Mode(            key input, 1time Print out)
            01 setting : Auto Print Mode (When “Finish Relay” output, 1time auto Print out)
※ Even if F38-01(Auto print mode), finished data will not be automatically save when F01-00(Manual
                                               Save Mode)
                                               - 67 -
                                              6. INTERFACE
     6-1. Serial Interface (RS-232C)
     RS-232C Serial Interface is sensitive/weak for electric Noise.
     So, please isolate with AC power cable and use shield cable to reduce the electric noise effect.
       6-1-1. Signal Format
            ①. Type : EIA-RS-232C
            ②. Communication Method : Half-Duplex, Full Duplex, Asynchronous
            ③. Serial Baud Rate : Selectable on “F-function31”
            ④. Data Bit : 8(No Parity mode, only)Bit – Refer “F30”.
            ⑤. Stop Bit : 1
            ⑥. Parity Bit : Non, Even, Odd (Selectable on “F-function 30”) - Refer “F30”
            ⑦. Code : ASCII
                   STX 02H
                   ETX 03H
                   CR 0DH
                   LF 0AH
            ⑧. Check-Sum (Error Detecting, “F-Function 36”)
                                LSB                                   MSB
 1     +9V                                1    2     3     4    5
                                  0                                       6
0 -9V
                                                                              Parity    Stop
                    Start Bit             Data Byte – Refer “F40”
                                                                                Bit      Bit
6-1-2. Data Format(1) : ID Number will not be transferred. (Refer “F-function 35”)
, , k g CR LF
                                                      - 68 -
           ②. Header 2. : NT : Net-Weight
                               GS : Net-Weight, under TARE.
           ③. Data Bit(Number) 2B(H) : “+” Plus
                                      2D(H) : “-“ Minus
                                      2D(H) : “ “ Space
                                      2E(H) : “.” Decimal Point
           ④. Unit : kg, g, t
, , , CR LF
6-1-4. Data Format(3) : CAS “CI5101A” Data Transference) – CAS 22byte Format
, , k g CR LF
                                                                                               Weight
                                        ID      Lamp
Header 1            Header 2                                      Data(8)              Space    Unit
                                       No.     Display
                                                       - 69 -
         ①. Header 1. : OL : Over Load, Under Load
                         ST : Display “Steady”
                         US : Display “Un-Steady”
         ②. Header 2. : NT : Net-Weight
                         GS : Net-Weight, under TARE.
         ③. Lamp Display : Current Lamp Condition (ON/Off Data)
Bit 7          Bit 6          Bit 5           Bit 4            Bit 3       Bit 2         Bit 1        Bit 0
                                                                           Gross
 1            Steady           1              Hold             Print                     Tare         Zero
                                                                           Weight
2 : RxD TxD : 3
                                    3 : TxD                    RxD : 2
        Personal Computer
                                    5 : SG                      SG : 5                      SI 4410
         (9pins Standard)
2 2 : RxD
3 3 : TxD
5 5 : SG
                                                      - 70 -
    6-2. Current Loop Interface
    “Current Loop” Interface is stronger for Electric Noise than “RS-232C” interface.
    So, it can be used for long distance communication.(About 100m long distance).
3 : RxD TxD : 8
4 : RxD TxD : 9
                                                                                        SI 4410
              Remote Display
            (External Display)
                                                 - 71 -
6-3. Print Interface (Option 01 : Centronics Parallel Interface)
This Print Interface Option is based on “Centronics Parallel Interface”, so this print interface can be
connected other printers using this communication method.
But, the print format is programmed based on our “SE7300”, and “SE7320” Industrial Printers, so you
had better to use these printer for convenience.
 6-3-2. Specification
     ①. Data Transference Speed : 1,000 to 6,000 characters/sec.
     ②. Synchronization : STROBE Pulses
     ③. Handshaking : ACK and BUSY Signals
     ④. Logic Level : Compatible with TTL level
DATA
STROBE t t t
                                                - 72 -
6-3-4. Typical Interface Circuit
                       BUSY, ACK
      Input
                      (From Printer)
                      Data0 ~ Data7
                        (To Printer)
     Output
                         STROBE
                        (To Printer)
SUB-TOTAL GRD-TOTAL
                                            - 73 -   T-WEIGHT :                2.000kg
6-4. Analog Output Interface (Option 02 : 0~10V Output)
This Option card converts weight value to Analog Voltage output(0~10V) and transfers to external
devices(Recorder, P.L.C), controlled by voltage output.
 6-4-1. Specification
     ①. Output Voltage : 0~10V DC output
     ②. Accuracy : More than 1/1,000
6-4-2. Circuit
1 : HI(+), 5 : (-)
※ For 0~5VDC or 1~5VDC analog output, please inform when you inquiry.
                                             - 74 -
6-5. Analog Output Interface (Option 03 : 4~20mA Output)
This Option card converts weight value to Analog Electric Current output(4~20mA) and transfers to
external devices(Recorder, P.L.C), controlled by electric current output.
 6-5-1. Specification
     ①. Output Current : 4~20mA (Output Range : 2~22mA)
     ②. Accuracy : More than 1/1,000
     ③. Temperature Co-efficiency : 0.01%℃
     ④. Max. Loaded Impedance : Max. 500Ω
      ※ When Weight display is “Zero”, 4mA current will be output, when Weight display is “Full
          Capacity”, 20mA current will be output.
6-5-2. Circuit
1 : HI(+), 5 : (-)
                                              - 75 -
   6-6. Serial Interface (option 04 : RS-232C/422/485)
   RS-422/485 serial interface is more stable for electric noise effect compare with other communication
   method, using electric current difference.
   But, install isolated place from Power cable or other electric cables and wires, and please use shielded cable
   for better performance.
   Recommendable communication distance is about 1.2km.
If you install additional RS-232C interface, please refer “6-1. Serial Interface” section.
                                            LSB                                        MSB
               1                                      1      2     3        4   5
                                                0                                        6
               0
   1.5V Electric
Potential Difference                                                                     Parity Bit   Stop 1 Bit
                                  Start 1 Bit                    Data Bit
6-6-3. RS-422 Circuit (In case of RS-485, Use No6 and 7 pin - No.8 and 9 will be use)
                                                    - 76 -
6-7. BCD Input Interface( Option 05) – Input for Part No. selection.
This “BCD interface” option card can be applied on PLC (Programmable Logic Controller), or Score
Board applications.
Each Input circuit is isolated with “Photo-Coupler”, from external devices electrically.
                                                                                           100X1
                                                                                           100X2
                                                                                           100X4
                                                                                           100X8
                                                                                           101X1
                                                                                           101X2
                                                                                           101X4
                                                                                           101X8
                                                                                           100X1
                                                                                           100X2
                                                                                           100X4
                                                                                           100X8
                                                                                           101X1
                                                                                           101X2
                                                                                           101X4
                                                                                           101X8
                                               - 77 -
6-8. Command Mode
Under “Command Mode”, Indicator will recognize the receipt of Order based on 02h(Header) and
03h(END) signal, and transfers ACK(06)/ NAK(15).
 6-8-1. Read Command (Standard Serial Port and Optional Port is same.)
 ※ Caution : Please use Capital Letter to input Command.
 ※ Each “READ” Command’s interval must be guaranteed at least 100ms. If you command
      another one within 100ms, the indicator will not response. (Under “Check-Sum”, the
      interval will be 150ms).
                                               - 78 -
                                                  Finished Weight Data Transfer
  STX ID NO. RFIN ETX         Finished Weight
                                                  -STX ID NO. RFIN Finished Weight(7/8byte) ETX
※ In case of Weight, FINAL, PRE1, PRE2, SUB-TOTAL, GRAND-TOTAL, Finished Weight, LOW,
    HIGH set value can be transfer 7 or 8byte data.(F40-00 : 7byte, F40-01 : 8byte)
※ Under “RWRS Command”, External Input Data and Relay Output Data will be transferred only when the
    key input time and Command receipt time is same.
※ Under RCWT Command, “ST/US” means “STEADY/UNSTEADY” and “NT/GS” means
    “NET/GROSS weight”.
                                             - 79 -
6-8-2. Write Command
※ Each “WRITE” Command’s interval must be guaranteed at least 150ms. If you command
    another one within 150ms, the indicator will not response. (Under “Check-Sum”, the
    interval will be 200ms).
STX ID NO. WZER ETX To make Current Weight as Zero ACK or NAK
                                     - 80 -
6-8-3. Command Mode Example - “Check-sum” not use (F34-00 setting).
Example) PRE1 Setting Command, ID No : 01, New PRE1 Set value : 0.600kg
P.C Command Format (F40-00, 7byte Data Transfer setting)
(STX ID NO. WPR1 000.600 ETX)
   02h       30h 31h        57h 50h 52h 31h         30h 30h 30h 2Eh 36h 30h 30h             03h
  Header      ID No.              WPR1                REQUEST DATA(7byte)                   End
P.C Command Format (F40-01, 8byte Data Transfer setting)
(STX ID NO. WPR1 0000.600 ETX)
   02h       30h 31h     57h 50h 52h 31h       30h 30h 30h 30h 2Eh 36h 30h 30h              03h
  Header     ID No.             WPR1                REQUEST DATA(8byte)                     End
                                           - 81 -
Indicator Response Format (Normal Operation)
   02h       30h 31h                                  06h                               03h
  Header      ID No.                                 ACK                                End
Indicator Response Format (Error)
   02h       30h 31h                                  15h                               03h
  Header      ID No.                                 NAK                                End
                                          - 82 -
WRITE COMMAND Format
 STX(1)      ID NO.(2)     COMMAND(4)              REQUEST DATA               BCC(2)      ETX(1)
※ Under “Check-Sum” using mode(F-34-01 setting), summing all the data from “02” to “03”
    and convert to ASCII code and transfer to P.C. If the calculated value is “02”, the
    transferred data will be “30h 32h”.
Example) PRE1 Setting Command, ID No : 01, New PRE1 Set value : 0.600kg
(Under “BCC Check” setting, F34-01)
P.C Command Format (F40-00, 7byte Data Transfer setting)
(STX ID NO. WPR1 000.600 ETX)
   02h       30h 31h      57h 50h 52h 31h      30h 30h 30h 2Eh 36h 30h 30h      45h 34h     03h
 Header      ID No.            WPR1              REQUEST DATA(7byte)             BCC        End
P.C Command Format (F40-01, 8byte Data Transfer setting)
(STX ID NO. WPR1 0000.600 ETX)
   02h      30h 31h    57h 50h 52h 31h      30h 30h 30h 30h 2Eh 36h 30h 30h     31h 34h     03h
 Header     ID No.         WPR1                REQUEST DATA(8byte)               BCC        End
Indicator Response Format (Normal Operation)
   02h       30h 31h                             06h                            36h 43h     03h
 Header      ID No.                              ACK                             BCC        End
Indicator Response Format (Error)
   02h       30h 31h                             15h                            37h 42h     03h
 Header      ID No.                              NAK                             BCC        End
                                          - 83 -
                           7. Error & Treatment
7-1. Load Cell Installation
         Error                      Cause                        Treatment                  Remark
                                                                                      1).Input Resistance of
                            1). Load cell broken
                                                        1). Measure input/output      “EX+” and “EX-“ is
                            2). Load cell isolation
                                                        resistance of Load cell.      about 350Ω~450Ω.
                            resistance error
                            3). Weighing part
   Weight Value is                                      2). Measure Load cell         2). Output Resistance
                            touches other devices
       unstable                                         isolation resistance          of “EX-“ and “EX+”
                            or some weight is on
                                                                                      is about 350Ω.
                            the weighing part
                            4). Summing Board
                                                                                      3). Isolate Resistance
                            Error
                                                                                      is more than 100Ω
   Weight Value is                                      1).    Check    Load   cell
                           1). Load cell Error
   increased regular                                    connection
                           2). Load cell connection
 rate, but not return to                                2). Measure Load cell
                           Error
        “Zero”                                          Resistance
                                                 - 84 -
7-2. Calibration Process
   Error                   Cause                                          Treatment
              Standard weight value is over      Re-input Standard weight value with Number keys,
   Err 04
              than Max. Capacity                 under Max. Capacity
              Standard weight value is less      Re-input Standard weight value with Number keys,
   Err 05
              than 10% of Max. Capacity          more than 10% of Max. Capacity
              1. Amp. Gain is too big            Check standard weight’s weight with set value.
              2. Sig+ and Sig- wire              If there is difference between set value and real
   Err 06
                 connection error                weight, please re-input the value
              3. Test weight is not loaded       (set value is too small)
              1. Amp. Gain is too small          Check standard weight’s weight with set value.
              2. Sig+ and Sig- wire              If there is difference between set value and real
   Err 07
                 connection error                weight, please re-input the value
              3. Test weight is not loaded       (set value is too big)
                                             - 85 -
7-3. Digital Weighing Indicator
 Error No.      Display               Cause                              Treatment
※ Under “CELL-Er”, Relay will not be Output, and Analogue Output(4~20mA/0~10V), either.
                                            - 86 -
7-4. Indicator Test mode
 Through this “Test Mode”, you can check basic conditions of Indicator.
 This Test consists with total 7 tests.
                            Under “TEST” display, press No.1 key and Enter “TEST1” mode.
           Test 1.          Under this mode, you can check the A/D value.
         Analogue           If the A/D value is close to 520,000, or there is no change although
         Value Test         pressing or loading some force on/in weighing part, please check load cell,
                            load cell, cable, connecter, A/D board.
           Test 2.          Under “TEST” display, press No.2 key and Enter “TEST2” mode.
          Key test          Press each key, and check the pressed key is operated.
                            Under “TEST” display, press No.3 key and Enter “TEST3” mode.
                            This Test will be operated automatically from Relay1 to Relay6.
           Test 3.
    Output Relay Test           ※ This test will operate automatically, so please remove all materials
                                in/on weighing parts.
                                If you can not remove materials, please remove relay terminals.
                                                - 87 -
                 Under “TEST” display, press No.4 key and Enter “TEST4” mode.
   Test 4.
External Input
                 If you press External input S/W, the External S/W No. will be displayed.
    Test
                 If the S/W No. is not displayed, please check connecting condition.
                 Under “TEST” display, press No.5 key and Enter “TEST5” mode.
   Test 5.       After connecting No.2 and 3 pin of 9pin connector, you can test
Communication    communication condition, like TXD or RXD/TXD.
    Test         If there is an error in communication, “232-Err” will be displayed with
(Com. Port 1)    3times buzzer sound. The communication is working properly, “232Pass”
                 will be displayed with one time buzzer sound.
                 Under “TEST” display, press No.6 key and Enter “TEST6” mode.
   Test 6.       After connecting No.2 and 3 pin of 9pin connector, you can test
Communication    communication condition, like TXD or RXD/TXD.
    Test         If there is an error in communication, “232-Err” will be displayed with
(Com. Port 2)    3times buzzer sound. The communication is working properly, “232Pass”
                 will be displayed with one time buzzer sound.
                                     - 88 -
                                       WARRANTEE CETIFICATION
WARRANTEE CLAUSE
1. The Warrantee period, we can guarantee, is one(1) year from your purchasing date
2. Warrantee Exception Clause
      - Warrantee period is expired.
      - Any kinds of Mal-function or defection caused by Modification or Repair without Sewhacnm’s
        permission.
      - Any kinds of Mal-function, Defection, or External damage, caused by operator
      - Any kinds of Mal-function, Defection, caused by using spare part from Non-Authorized
        Distributor or Agent.
      - Any kinds of Mal-function, Defection, caused by not following Warnings or Cautions mentioned
        on this manual.
      - Any kinds of Mal-function, Defection caused by “Force Majeur”, like Fire, Flood.
      - Without presentation of this “Warrantee Certification”.
3. Other
      - Any kinds of “Warrantee Certification” without authorized Stamp is out of validity
.
Manufacturer                                                Product        Digital Weighing Indicator
SEWHACNM Co.,Ltd.                                             Model                 SI 4410
310, Kolon Techno Valley, 60-4, Kasan-dong,
                                                            Serial No.
Keumcheon-gu, Seoul, Korea
Website : http://www.sewhacnm.co.kr
                                                              AUTHORIZED
Made in KOREA
                                                                 STAMP
- 89 -